分散法とエッジ法による ディジタルマンモグラフィの解像特性の比較
松林 由紀子濱野 美月石井 里枝北川 薫石井 美枝眞田 泰三吉田 彰
著者情報
ジャーナル フリー

2021 年 77 巻 2 号 p. 182-190

詳細
抄録

We studied the image resolution properties in digital mammography using the variance and edge response function methods. The associated measurements were made using a polymethylmethacrylate phantom of 0–40 mm thickness while maintaining the incident dose to the detector constant, as well as with and without using the phantom but by varying the level of exposure to the detector. The results obtained using the two methods were compared on the object plane of the mammography unit with compensation for the magnification and rejection of the scattered radiation. In this unit, the imaging process of resolution enhancement may be performed according to the phantom thickness, shape of test device, exposure level, and so on. The modulation transfer function values of the variance method were slightly higher than those of the edge method, approximately 6% at 2 cycles/mm and 8% at 4 cycles/mm without resolution enhancement processing. We concluded that the variance method is more suitable than the edge method for the constancy test or quality control because of its lower coefficient of variation.

著者関連情報
© 2021 公益社団法人 日本放射線技術学会