ノート
CT ガイド下IVR における術者頭頸部被ばくの測定
飯塚 崇文松本 一真萩原 芳明名定 敏也池内 陽子菊池 圭介光家 千恵美琴浦 規子山門 亨一郎小林 薫高木 治行加古 泰一
著者情報
ジャーナル フリー

2019 年 75 巻 7 号 p. 625-630

詳細
抄録

In April 2011, the International Radiological Protection Committee recommended that “The equivalent dose of the crystalline lens should not exceed 20 mSv/year, averaged over defined periods of 5 years, with no single year exceeding 50 mSv”. Based on this recommendation, it is predicted that the equivalent dose limit of our crystalline lens can be lowered in the near future. Therefore, it is important to grasp the current situation of radiation exposure. The purpose of this study is to measure the crystalline lens of surgeons by focusing on the CT-fluoroscopy guided interventional radiology’s (IVRs). We also examined whether the exposure dose of the crystalline lens can be correctly evaluated by measuring the unequal exposure dose of the neck, which is usually used for the unequal exposure measurement. Results of the analysis of 200 CT-fluoroscopy guided IVR procedures showed that the unequal exposure dose of the neck was significantly correlated with the exposure dose of the crystalline lens which was measured near the left eye ball (R=0.83). However, the exposure dose of the crystalline was 33% lower than those of the neck. Therefore, although the individual dosimeter worn on the neck can be used as the useful index of the exposure dose of the crystalline lens, the results can be overestimated.

著者関連情報
© 2019 公益社団法人 日本放射線技術学会