日本放射線技術学会雑誌
Online ISSN : 1881-4883
Print ISSN : 0369-4305
ISSN-L : 0369-4305
原著
Web camera を用いた高線量率小線源治療装置の品質保証システム
廣瀬 麻子上田 悦弘大平 新吾五十野 優辻井 克友乾 翔輝正岡 祥谷口 真宮崎 正義手島 昭樹
著者情報
ジャーナル フリー

2016 年 72 巻 3 号 p. 227-233

詳細
抄録

Purpose: The quality assurance (QA) system that simultaneously quantifies the position and duration of an 192Ir source (dwell position and time) was developed and the performance of this system was evaluated in high-dose-rate brachytherapy. Methods: This QA system has two functions to verify and quantify dwell position and time by using a web camera. The web camera records 30 images per second in a range from 1,425 mm to 1,505 mm. A user verifies the source position from the web camera at real time. The source position and duration were quantified with the movie using in-house software which was applied with a template-matching technique. Results: This QA system allowed verification of the absolute position in real time and quantification of dwell position and time simultaneously. It was evident from the verification of the system that the mean of step size errors was 0.31±0.1 mm and that of dwell time errors 0.1±0.0 s. Absolute position errors can be determined with an accuracy of 1.0 mm at all dwell points in three step sizes and dwell time errors with an accuracy of 0.1% in more than 10.0 s of the planned time. Conclusion: This system is to provide quick verification and quantification of the dwell position and time with high accuracy at various dwell positions without depending on the step size.

著者関連情報
© 2016 公益社団法人 日本放射線技術学会
前の記事 次の記事