Research Article - Interventional Cardiology (2019) Volume 11, Issue 1

A systematic literature review and meta-analysis of impella devices used in cardiogenic shock and high risk percutaneous coronary interventions

Corresponding Author:
Tim Spelman Syn
ergus AB, Danderyd, Sweden
E-mail:
tim.spelman@synergusrwe.com

Received date: January 21, 2019; Accepted date: February 06, 2019; Published date: February 10, 2019

Abstract

Background: To perform a meta-analysis on pooled survival and complications rates of Impella® heart pumps (Abiomed Inc., Danvers, USA) use in cardiogenic shock and high-risk coronary percutaneous coronary intervention (PCI). Methods and Findings: Articles were searched in Medline, Medline In-Process, EMBASE and the CENTRAL bibliographic databases on the 30th April, 2017. Prospective and retrospective studies with ≥10 patients supported with Impella. Survival and complication rates were pooled. The literature review identified 33 articles. Data on patient characteristics, indication of support, type of Impella device and outcomes were extracted. A random effect was used to pool the various outcomes. Low heterogeneity (I2 ≤0.25) results are presented. A total of 2,827 patients (40.5% cardiogenic shock, 59.5% high-risk PCI) were included (mean age 64.9±11.4, male 74.6%). In the cardiogenic shock indication, survival rate to 90 and 180 days was 62.6% and 58.3%, respectively; the rates of hemolysis and device malfunction were 8.8% and 2.5%, respectively. In the high-risk PCI indication, the 30 day major adverse cardiac and cerebrovascular event (MACCE) rate was 15.3% with a 30 day survival, stroke, myocardial infarction and repeat revascularization rates of 92.2%, 0.3%, 13.5% and 1.6%, respectively. Hemolysis and device malfunction affected respectively 0.7% and 1.4% of the high-risk PCI patients. Conclusions: This is the largest meta-analysis summarizing the literature outcome data of Impella heart pumps. The present study demonstrates very encouraging survival in cardiogenic shock patients and very good 30 day outcomes in patients undergoing prophylactic support for high-risk PCI.

Keywords

Heart failure; Cardiac disease; Intervention; Surgery; Transplantation; Quality and outcomes; Vascular disease; Acute coronary syndromes; Coronary artery disease

Introduction

Short-term mechanical circulatory support (MCS) is an established treatment option across a diverse range of clinical indications including cardiogenic shock and high risk percutaneous coronary interventions (HRPCI) [1-4]. The Impella platform of left ventricular assist devices (Abiomed, Inc., Danvers, Massachusetts) are minimally invasive, catheter-based, axial flow pumps which directly unload the left ventricle by driving blood from the left ventricle into the ascending aorta. Impella devices were introduced onto the market in 2003 in Europe and 2008 in the United States. Since the advent of the first generation of Impella pumps a number of revised devices have since become available, including the Impella 2.5, Impella CP (cardiac power), Impella 5.0 and Impella LD (Left direct) (Table 1) [5].

Impella 2.5® Impella CP® Impella 5.0®
Flow < 2.5 L/min < 4.0 L/min < 5 L/min
Catheter Size 9F 9F 9F
Pump Insertion Size 12F 14F 21F
Approved Duration 4 days (US)
5 days (EU)
4 days (US)
5 days (EU)
6 days (US)
10 days (EU)
FDA Approved Indications High Risk PCI
AMICS/PCCS
High Risk PCI
AMICS/PCCS
AMICS/PCCS
Insertion Sheath 13cm
Peel-Away
(femoral artery)
13 cm/25cm
Peel-Away
(femoral artery)
6 cm
Peel-Away
(axillary/femoral graft)
Valve Interaction Smooth Cannula Smooth Cannula Smooth Cannula

Table 1: Types of Impella devices, included into the study

Whilst the current evidence base supporting the survival benefits and documenting complications associated with Impella support in cardiogenic shock and HRPCI is growing, these studies tend to be limited by under-powering and/or design, with the majority of the available data being sourced from small clinical trials or case series [6,7]. Whilst Impella has been available since 2003, no contemporary systematic overview of the combined safety and effectiveness profile of Impella is currently available. The objective of this study was to present a pooled and up-to-date review of the survival and safety profile associated with the use of Impella devices in the cardiogenic shock (CS) and high-risk percutaneous coronary intervention (HRPCI) indications.

Methods

A systematic literature search was performed in Medline, Medline In-Process, EMBASE and the CENTRAL bibliographic databases on the 30th April, 2017. Inclusion criteria are detailed in (Table 2). Only full-text peer-reviewed articles with 10 or more Impella patients supported for CS or HRPCI were included. The full search strategy is detailed in the Supplemental Material. Summary data for each included study is described in Supplemental Material. Benefit and safety outcomes analyzed and stratification groups used in the analysis are described in Table 3.

Inclusion criteria Exclusion criteria
Indication Study type Outcomes reportedc
Cardiogenic shock (CS) Multiple-patient observational and experimental studies of Impella device with ≥10 Impellacasesb
Patients implanted with Impella devices in 2004 and later
Complications and safety outcomes including stroke/TIA, MACE, bleeding, hematoma, hemolysis, renal dysfunction, limb ischemia, device malfunction and revascularization Reviews
Conference abstracts
Device name was not reported in study
Study population of less than 10 patients
Studies of percutaneous ventricular assist devices other than Impella
No clinical outcomes or no clinical outcomes of interest reported
Mixed devices (results of Impella reported combined with results of other devices)
Mixed indications (results of patients with CS or HRPCI reported combined with results of patients treated with other devices)
Other indication than cardiogenic shock and HRPCI
Health economic studies
Prediatric population
Right ventricular support
Concomittant use of Impella and ECMO
Support during Balloon Aortic Valvuloplasty (BAV) procedure
Support during Electrophysiology (EP) procedure
Prophylactic use in HRPCIa Multiple-patient observational and experimental studies of Impella device with ≥10 Impellacasesb
Patients implanted with Impella devices in 2004 and later

Table 2: Inclusion criteria

Stratification levels Benefit & safety events
Indication
use of Impella as prophylactic circulatory support in patients undergoing a non-emergent HRPCI
emergent circulatory support in patient with cardiogenic shock following an acute myocardial infarction (AMI), open heart surgery (post-cardiotomy cardiogenic shock (PCCS)) or an acute decompensated heart failure (ADHF)
Study type
randomized controlled trials (RCT)
non-randomized studies including prospective cohort studies, retrospective cohorts, case series and chart reviews.
Benefits
survival to next therapy
post-discharge survival (at 1, 3, 6 and 12 months where available)
Safety1
bleeding (including all forms of bleeding as reported, e.g. minor/major bleeding, bleeding required surgery or transfusion etc.)
hematoma (incliding all forms of hematoma)
hemolysis during hospitalization
leg/limb ischemia during hospitalization
stroke and/or TIA (during hospitalization, at 1 and 3 month)
MACE
device malfunction (where explicitly reported as either a “device malfunction” or “device related technical failure”).

Table 3: Stratification groups and study outcomes

Statistical analyses

Categorical variables were summarized using frequency and percentage. Continuous variables were summarized using mean and standard deviation (SD) or median and inter-quartile range (IQR) as appropriate. A random effects meta-analysis using DerSimonian– Laird method was used to pool the various benefits and safety outcomes across included studies and expressed as proportions with 95% confidence intervals. Arcsine transformation of proportions was made for analysis of data from single-arm studies [8,9]. Inter-study heterogeneity was analyzed by Cochran's Q and I2 statistics. A significant p-vale of Q (p<0.05) indicated that there might be significant heterogeneity between studies. Heterogeneity measured by I2 was quantified as low, moderate, and high (low: 0-25%; moderate 26-50%; high 51-100%) [9]. The primary results were limited to those outcomes associated with an I2 of less than 50%. For all analyses, p-value <0.05 was considered significant. All analyses were conducted in R (R Foundation for Statistical Computing, Vienna, Austria) and validated in Stata version 15 (StataCorp, College Station, Texas) [10].

Results

Patient characteristics

A total of 33 publications [11-42] reported clinical outcomes for 2,827 patients with Impella support were included in the analysis (Tables 2 and 3). The mean (SD) age of patients in included studies was 64.9 (11.4) years. Males constituted 74.6% of patients. The median (IQR) number of patients treated with Impella devices in included studies was 36 (18-119). A total of 1,144 (40.5%) patients were supported for CS, of which 890 (78%) for AMICS (Acute Myocardial Infarction complicated by Cardiogenic Shock). A smaller group of 93 (8%) patients had CS secondary to acute decompensated heart failure (ADHF) whilst 48 (4%) patients had post-cardiotomy cardiogenic shock (PCCS). The median (IQR) duration of circulatory support was 43 hours (25-53 hours) for AMICS, 295 hours (231-358 days) for ADHF. The median (IQR) duration of support was 1.9 hours (1.5-2.1 hours) in the 1,715 patients supported prophylactically with Impella for HRPCI.

Meta-analysis

The pooled proportions of selected survival and complications (survival, stroke/TIA, MI, revascularization, MACE, bleeding, hemolysis, renal dysfunction, limb ischemia and device malfunction) across Impella device types within the cardiogenic shock indications (CS of any etiology and AMICS, ADHF-CS, PCCS) and HRPCI, are summarized in Tables 4, 5 and Supplemental Figures.

Group Outcome Patients (N) Studies pooled (N) Pooled proportion (95%CI) I2 Degree of I2 Q p-Value for Q
CS – prospective studies Survival
Survival to next therapy 26 2 0.887 (0.597-1) 0.74 High 3.79 0.052
Survival at 30-days 35 3 0.697 (0.383-0.933) 0.82 High 10.82 0.004
Survival at 6 months 25 2 0.658 (0.334-0.916) 0.77 High 4.38 0.036
Complications
Bleeding* 15 3 0.247 (0.042-0.547) 0.82 High 9.89 0.007
Hematoma** 3 2 0.075 (0.015-0.175) 0 Low 0.06 0.803
Device malfunction 2 3 0.054 (0.006-0.144) 0.27 Moderate 2.68 0.261
Hemolysis 3 3 0.078 (0.023-0.162) 0 Low 0.66 0.72
Limb ischemia 1 2 0.059 (0.005-0.167) 0 Low 0.8 0.372
CS – observational retrospective studies Survival
Survival to next therapy 178 8 0.717 (0.565-0.847) 0.83 High 39.16 0
Survival to discharge 352 10 0.63 (0.539-0.716) 0.75 High 37.27 0
Survival at 30-days 263 9 0.58 (0.472-0.684) 0.82 High 45.74 0
Survival at 90-days 77 4 0.626 (0.539-0.709) 0 Low 0.48 0.924
Survival at 6 months 28 3 0.583 (0.442-0.718) 0 Low 0.01 0.993
Survival at 1 year 109 5 0.463 (0.321-0.609) 0.8 High 22.06 0
Complications
Bleeding* 106 10 0.157 (0.089-0.239) 0.8 High 38.48 0
Hematoma** 11 3 0.048 (0.024-0.08) 0 Low 0.72 0.699
Device malfunction 17 5 0.051 (0.019-0.096) 0.59 High 8.68 0.07
Hemolysis 60 7 0.143 (0.042-0.291) 0.93 High 58.45 0
Limb ischemia 11 6 0.042 (0.023-0.067) 0 Low 3.85 0.571
Stroke/TIA in-hospital 10 6 0.035 (0.018-0.057) 0 Low 2.97 0.705
AMICS – observational retrospective studies Survival
Survival to next therapy 97 3 0.704 (0.532-0.851) 0.64 High 5.87 0.053
Survival to discharge 278 6 0.563 (0.464-0.659) 0.71 High 16.4 0.006
Survival at 30-days 158 5 0.472 (0.361-0.584) 0.73 High 11.41 0.022
Survival at 6 months 20 2 0.588 (0.421-0.746) 0 Low 0 0.966
Survival at 1 year 71 3 0.372 (0.247-0.506) 0.69 High 5.84 0.054
Complications
Bleeding* 88 5 0.214 (0.159-0.276) 0.42 Moderate 8.31 0.081
Hematoma** 10 2 0.049 (0.023-0.083) 0 Low 0.69 0.406
Device malfunction 8 3 0.025 (0.011-0.045) 0 Low 0.92 0.63
Hemolysis 30 3 0.081 (0.056-0.111) 0 Low 0.26 0.879
Limb ischemia 10 4 0.036 (0.017-0.063) 0 Low 3.2 0.362
Renal Dysfunction 98 3 0.459 (0.147-0.79) 0.97 High 15.95 0
Stroke/TIA in-hospital 9 4 0.037 (0.018-0.062) 0 Low 1.59 0.662
ADHF – observational retrospective studies Survival
Survival to next therapy 59 3 0.624 (0.27-0.915) 0.92 High 26.41 0
Survival to discharge 63 3 0.678 (0.58-0.769) 0 Low 0.59 0.743
Survival at 30-days 43 2 0.672 (0.553-0.781) 0 Low 0 0.945
Complications
Stroke/TIA in-hospital 1 2 0.027 (0.001-0.085) 0.14 Low 1.16 0.281
* any type of bleeding
** any type of hematoma

Table 4: Meta-analysis of survival and complications outcomes in cardiogenic shock patients

Group Outcome Patients (N) Studies pooled (N) Pooled proportion (95%CI) I2 Degree of I2 Q p-Value for Q
HRPCI- Prospective studies Survival
Survival at 30-days 235 3 0.922 (0.886-0.951) 0 Low 0.2 0.906
Complications
Device malfunction 0 2 0.004 (0.002-0.029) 0.15 Low 1.18 0.277
Hemolysis 2 2 0.09 (0.017-0.212) 0 Low 0.55 0.46
MACE at 30-days 39 3 0.153 (0.112-0.2) 0 Low 0.56 0.754
MI at 30-days 33 2 0.135 (0.095-0.18) 0 Low 0.25 0.616
Revascularization at 30-days 3 2 0.016 (0.004-0.036) 0 Low 0.07 0.792
Stroke/TIA at 30-days 0 2 0.003 (0-0.014) 0 Low 0.89 0.344
HRPCI – observational retrospective studies Survival
Survival to next therapy 587 9 0.99 (0.981-0.997) 0 Low 2.75 0.949
Survival to discharge 938 6 0.979 (0.964-0.99) 0.24 Low 5.39 0.37
Survival at 30-days 398 6 0.961 (0.94-0.977) 0 Low 3 0.7
Bleeding* 35 7 0.074 (0.035-0.126) 0.72 High 18.93 0.004
Hematoma** 21 6 0.075 (0.036-0.127) 0.5 High 10.21 0.069
Device malfunction 1 5 0.007 (0.001-0.017) 0 Low 3.25 0.516
Hemolysis 12 7 0.014 (0.008-0.021) 0 Low 3.69 0.719
Limb ischemia 4 3 0.046 (0.002-0.139) 0.62 High 5.64 0.06
MACE at 30-days 17 3 0.051 (0.03-0.077) 0 Low 0.76 0.685
MI at 30-days 1 4 0.008 (0.001-0.02) 0 Low 1.42 0.701
Renal Dysfunction 45 6 0.033 (0.015-0.059) 0.61 High 14.28 0.014
Revascularization in-hospital 5 3 0.009 (0.003-0.017) 0 Low 0.18 0.912
Revascularization at 30-days 3 4 0.019 (0.004-0.044) 0.15 Low 3.54 0.316
Stroke/TIA peri-procedural 0 2 0.006 (0-0.022) 0 Low 0.31 0.578
Stroke/TIA in-hospital 0 5 0.002 (0-0.008) 0.1 Low 3.42 0.49
Stroke/TIA at 30-days 1 4 0.008 (0.001-0.02) 0 Low 1.15 0.764
* any type of bleeding
** any type of hematoma

Table 5: Meta-analysis of survival and complications outcomes in HRPCI studies

Cardiogenic shock

The pooled proportions of survival and complications across Impella device types for cardiogenic shock indications are summarized in Table 4.

Survival

In the two RCTs identified in the systematic search, survival at 30-days was reported to be 54% in both (29, 33). Additionally, in the IMPRESS trial (n=24) survival at 6-months was 50% (29). Survival at 90-days in the CS of any etiology in retrospective cohort was 62.6% (95%CI: 53.9%-70.9%) and survival at 6 months was 58.3% (95%CI: 44.2-71.8%). Other survival outcomes including survival to next therapy and survival to discharge were associated with unacceptably high heterogeneity and were thus considered unreliable. Similarly, in prospective studies of CS of any etiology, the outcomes on survival were similarly limited by excessive heterogeneity. Survival at 6 months in the AMICS subgroup was 58.8% (95%CI: 42.1%- 74.6%). In the subgroup of ADHF patient's survival to discharge was 67.8% (95%CI: 58%-76.9%), whilst 30-day survival was 67.2% (95%CI: 55.3%-78.1%).

Complications

In the IMPRESS trial, bleeding incidence among Impella patients was 33.3% and hemolysis 8.3% (29), while in the ISAR-SHOCK trial (n=26) no bleeding was observed (33). Across prospective CS studies, the pooled rates of hemolysis (7.8%; 95%CI: 2/3%-16.2%) and limb ischemia (5.9%; 95%CI: 0.5%-16.7%) were low. Similarly, in the retrospective CS studies the low rates of in-hospital stroke (3.5%; 95%CI: 1.8%-5.7%) and limb ischemia (4.2%; 95%CI: 2.3%-6.7%) were observed. When retrospective studies were pooled for the subgroup of AMICS patients, event rates within the observational retrospective studies were generally low (device malfunction 2.5% (1.1%-4.5%); in-hospital stroke 3.7% (95%CI: 1.8%-6.2%); limb ischemia 3.6% (95%CI:1.7%-6.3%), hematoma 4.9% (95%CI:2.3%- 8.3%), hemolysis 8.1% (95%CI: 5.6%-11.1%). The exception was bleeding, observed in 21.4% of patients (95%CI: 15.9%-27.6%). In the ADHF group inhospital stroke rate was 2.7% (95%CI: 0.01%-8.5%).

High-risk percutaneous coronary intervention

The pooled proportions of survival and complications across Impella device types for high risk PCI are summarized in Table 5.

Survival

A single RCT (PROTECT II) comparing HRPCI patients on Impella 2.5 (n=225) to patient on intraaortic balloon pump (n=223) reported 92.4% of patients randomized to Impella 2.5 had survived to 30-days post insertion, decreasing marginally to 87.9% at 90-days (28). For the HRPCI prospective group, 30-day survival was 92.2% (95%CI: 88.6%-95.1%). Survival was very high within retrospective studies of patients supported prophylactically with Impella 2.5 for HRPCI. Survival to next therapy was 99% (95%CI: 98.1%-99.7%), 97.9% at discharge was (95%CI: 96.4%-99%) and 96.1% at 30-days (95%CI: 94%- 97.7%).

Complications

The incidence of stroke/TIA in the PROTECT II intention-to-treat (ITT) Impella arm was low – from 0.0% at 30-days to 0.9% at 90-days. Acute renal dysfunction was associated with 4.0% of insertions at both 30 and 90-days (28). The use of Impella 2.5 in HRPCI patients (prospective study cohort) was associated with low rates of strokes/TIA at 30-days (0.3%; 95%CI: 0%-1.4%), device malfunction (0.4%; 95%CI: 0.02%-2.9%), revascularization at 30-days (1.6%; 95%CI: 0.4%-3.6%) and hemolysis (9%; 95%CI: 1.7%-21.2%). The rate of MACE (Major Adverse Cardiac Events) at 30-days was 15.3% (95%CI: 11.2%-20%) and MI (myocardial infarction) at 30-days was 13.5% (95%CI: 9.5%-18%). In the retrospective studies cohort, the rate of strokes/TIA at 30-days was 0.8% (95%CI: 0.1%-2%) device malfunction (0.7%; 95%CI 0.01%-1.7%), revascularization at 30-days (1.9%; 95%CI 0.4%-4.4%) and hemolysis (1.4%; 95%CI 0.8%-2.1%). The rate of MACE at 30-days was 5.1% (95%CI: 3%-7.7%) and MI at 30-days was 0.8% (95%CI: 0.1%-2%).

Device type

The pooled proportions of survival and complications disaggregated by Impella device types within the cardiogenic shock indication are summarized in Supplemental Material.

Survival

Across all indications, 67.8% (95%CI: 58%-76.9%) of patients supported with Impella 5.0 survived to discharge. This fell marginally to 67.2% survival at 30- days (95%CI: 55.3%-78.1%). All results of survival outcomes from the Impella 2.5 group had unacceptably high heterogeneity and were thus considered unreliable.

Complications

In-hospital stroke/TIA was again low, associated with only 2.9% (95%CI: 1.1%-5.7%) of Impella 2.5/CP insertions and 2.7% (95%CI: 0.1%-8.5%) of Impella 5.0 supports. Consistent with the other indications and study types analyzed, in-hospital bleeding was associated with 23.1% (95%CI: 16.7%- 30.3%) of Impella 2.5/CP insertions. Similarly, device malfunction was again low associated with just 2.3% of Impella 2.5/CP insertions (95%CI: 0.9%-4.3%). For Impella 2.5/CP, hemolysis was reported in 8.6% (95%CI: 5.7%-12%) of insertions while for Impella 5.0 it was 6.9% (95%CI: 1.8%-15%). The rates of limb ischemia in Impella 2.5/CP was 4.7% (95%CI: 2.1%- 8.2%) and 3.6% (95%CI: 0.3%-10%) in Impella 5.0.

Discussion

Impella devices were associated with good survival and generally low rates of complications and safety outcomes across all combinations of indication and study types analyzed. In the absence of sufficiently powered randomized clinical trials covering relevant indications and patient cohorts, the presented metaanalysis provides the best evidence to date and confirms observations from individual studies that the use of Impella in CS is likely to be safe and effective. It further extends the existing evidence base by demonstrating that these low event rates and favorable survival outcomes are generally consistent across both the indication and the study design used to study Impella outcome data.

Survival in CS secondary to either ADHF or AMI supported with Impella was particularly encouraging, with pooled 90-day survival across both indications at 62.6%. When the analysis was limited to the ADHF cohort, an indication characterized by the use of the more powerful Impella 5.0 device, survival to discharge was 67.8% (with an upper limit high of 83.0%) whilst survival at 30-days was 67.2%. In the context of ADHF, the 5.0 device is employed to reverse tissue hypoxemia, end organ dysfunction, and cardiorenal syndrome facilitating bridge to recovery, durable LVAD insertion of heart transplantation [25].

The relatively high rates of survival consistently observed across these often severely decompensated patients supports the effectiveness of the Impella 5.0 as a "bridge to decision" in ADHF [25]. Survival in the AMICS patients at 6 months was 58.8% (when the meta-analysis was limited to case series). Elsewhere, prophylactic Impella support for patients undergoing HRPCI was associated consistently high survival rates at 30-days (92.2% in prospective studies and 96.1% in retrospective studies), in line with expectations.

Notably, the rate of stroke/TIA was particularly low – regardless of indication, device type or study design. The maximum inpatient stroke rate observed amongst those pooled analyses associated with acceptable heterogeneity was 2.7% in both the ADHF case series and in the pooled CS retrospective observational studies it was 3.5%. In-hospital stroke rate was 3.7% in the AMICS and just 0.02% of the HRPCI retrospective studies group. This is broadly consistent with the low rates of stroke/TIA reported in the Impella arm of the pivotal PROTOCOL II study; which observed 0.0% and 0.9% stroke or TIA rates at 30 and 90-days respectively.

Limb ischemia is a significant risk for CS patients managed with a combination of mechanical support and catecholamine therapy. However, our meta-analysis suggests that limb ischemia is a relatively infrequent event at 5.9% of patients in the pooled CS prospective studies group (upper limit 16.7%) and 4.2% in CS retrospective group (upper limit 6.7%). In the AMICS subgroup limb ischemia rate was 4.4%. Bleeding events were reported in 23.5% of the AMICS subgroup, Hemolysis rates were consistently low across indication/ study type groups, ranging from a high of 8.8% in the AMICS PCI case series cohort to just 1.4% of the HRPCI retrospective subgroup. In the retrospective studies device malfunction was low in both the AMICS PCI s (2.5%) and the HRPCI subgroups (0.7%).

Whilst this meta-analysis provides the largest pooling of survival and complications to date in CS and HRPCI patients on Impella support, it does have a number of limitations. Firstly, it is not a comparative analysis and thus should not be used to make inferences around the exact benefit attributable to Impella relative to any other left ventricular support device in these specific clinical scenarios, with regards to either efficacy or in terms of harm avoided. Secondly, a large proportion of eligible studies included in the meta-analyses were lowquality case series. This led, in part, to the unacceptably high levels of heterogeneity for several key outcomes. However, by stratifying the meta-analysis by the level of evidence (i.e. study type) we were able to statistically demonstrate for the first time that those favorable survival and low incidence of adverse event signals previously only observed in individual studies were broadly consistent across study types (randomized trial, prospective cohort, retrospective, case series) – suggesting these signals are genuine. An appropriately powered clinical trial and/or larger prospective cohort study, preferably of longer follow-up duration than the studies included here, would be required to better characterize the benefit of Impella in these patient cohort, particularly in relation to competitor support devices.

Funding

This study was supported by Abiomed Inc. The sponsor had no influence or editorial control over the content of the study.

Conflicting Interests

Natalia M Stelmaszuk-Zadykowicz, Sun Sun and Tim Spelman are employees of Synergus AB – MedTech consulting company and received consulting fees from Abiomed Inc. Jonathan Hill, Bernhard Schieffer, Andreas Schäfer have received honoraria for speaking and chairing at symposia. The other authors report no conflicts.

Acknowledgements

We would like to thank Vladica Veličković for his contribution to this project.

References

Awards Nomination 20+ Million Readerbase

Select your language of interest to view the total content in your interested language

Journal Metrics:

Impact Factor 1.34
Scimago Journal Rank (SJR) 123
SJR Total Cites 15
Source Normalized Impact per Paper (SNIP) 0.144
h-index (2023) 12
PubMed NLM ID:  10148499
Google Scholar h5 index: 6
Iindex Copernicus Value: 105.52


Google Scholar citation report
Citations : 1297

Interventional Cardiology received 1297 citations as per Google Scholar report


Interventional Cardiology peer review process verified at publons

Indexed In

flyer