Three-dimensional quantification of magnetic resonance imaging artifacts associated with shape factors
Iku KAWABATAHaruki IMAIZuisei KANNOAkemi TETSUMURAYusuke TSUTSUMIHisashi DOIMaki ASHIDATohru KURABAYASHITakao HANAWAToru YAMAMOTOTakashi ONO
Author information
JOURNAL FREE ACCESS

2019 Volume 38 Issue 4 Pages 638-645

Details
Abstract

Differences in the volumes of artifacts caused by variously shaped titanium objects on magnetic resonance imaging (MRI) were evaluated. Spherical-, square cubic-, and regular tetrahedron-shaped isotropic, and elongated spherical-, elongated cubic-, and elongated tetrahedron-shaped anisotropic objects, with identical volumes, were prepared. Samples were placed on a nickel-doped agarose gel phantom and covered with nickel-nitrate hexahydrate solution. Three-Tesla MR images were obtained using turbo spin echo and gradient echo sequences. Areas with ±30% of the signal intensity of the standard background value were considered artifacts. Sample volumes were deducted from these volumes to calculate the total artifact volumes. Isotropic samples had similar artifact volumes. For anisotropic samples, the artifact volume increased in proportion with the normalized projection area. MRI artifact size can be reduced by high anisotropic designs, and by positioning the long axis of the metal device as parallel as possible to the magnetic field axis.

Content from these authors
© 2019 The Japanese Society for Dental Materials and Devices