Dental Materials Journal
Online ISSN : 1881-1361
Print ISSN : 0287-4547
ISSN-L : 0287-4547
Original Paper
Dental color matching: A comparison between visual and instrumental methods
Christopher IGIELMichael WEYHRAUCHStefan WENTASCHEKHerbert SCHELLERKarl Martin LEHMANN
Author information
JOURNAL FREE ACCESS

2016 Volume 35 Issue 1 Pages 63-69

Details
Abstract

The aim of this study was to compare the agreement rate (%) and color difference (ΔE*ab) of three dental color-measuring devices, with the visual shade identification. The tooth color were determined by two operators, which were advised to select a VITA classic shade tab in each other’s agreement. The Shadepilot (SP), CrystalEye (CE) and ShadeVision (SV) were used to measure tooth color. Statistically analyses include agreement rate (%), color difference (ΔE*ab), McNemar test (p=0.05), Student’s t-test (p=0.05) and Bland Altman scatterplots. The SP had an agreement of 56.3% with the visual shade determination, the CE 49.0% and SV 51.3%. ΔE*ab of the visually and instrumentally selected shade tabs and natural teeth were frequently above the threshold for acceptability. Comparing both methods, for SP ΔE*ab values differ in a range of clinical acceptability.

Content from these authors
© 2016 The Japanese Society for Dental Materials and Devices
Previous article Next article
feedback
Top