Dental Materials Journal
Online ISSN : 1881-1361
Print ISSN : 0287-4547
ISSN-L : 0287-4547
Original Paper
In vitro microleakage of six different dental materials as intraorifice barriers in endodontically treated teeth
Ki Sun LEEJong Sook KIMDong Yul LEERyan Jin Young KIMJoo Hee SHIN
Author information
JOURNAL FREE ACCESS

2015 Volume 34 Issue 4 Pages 425-431

Details
Abstract

The aim of this study was to compare the coronal sealing ability of six different dental materials: Three MTA-based cements and three established restorative materials by in vitro dye penetration method. For in vitro infiltration experiments, seventy extracted single-rooted human teeth were used. After crowns of teeth were reduced, root canals were prepared, and filled with gutta-percha cone. Teeth were randomly divided into 6 groups with 10 teeth per group. The orifice of each tooth was prepared to 3 mm depth and filled with the following materials: (I) ProRoot WMTA; (II) EndoCem Zr; (III) Angelus White; (IV) LuxaCore; (V) Fuji II LC; and (VI) Elite. After 5,000 cycles of thermocycling between 5°C and 55°C, dye penetration of each specimen was measured. The order of less dye infiltration of coronal filling materials was: ProRoot WMTA<LuxaCore, Angelus White<EndoCem Zr<Fuji II LC<Elite (p<0.05).

Content from these authors
© 2015 The Japanese Society for Dental Materials and Devices
Previous article Next article
feedback
Top