Annals of Vascular Diseases
Online ISSN : 1881-6428
Print ISSN : 1881-641X
ISSN-L : 1881-641X
CASE REPORTS
Spontaneous Collapse of Balloon Expandable Stent in Heavily Calcified Aorta
Taehwan KimShigeo Ichihashi Amane KozukiDaichi FujimotoSatoru NagatomiFrancesco BolstadJunya ShiteKimihiko Kichikawa
Author information
JOURNAL OPEN ACCESS

2020 Volume 13 Issue 2 Pages 187-190

Details
Abstract

Balloon expandable stents (BES), which are made of stainless steel, are vulnerable to external compression, leading to deformation or collapse/fracture of the stents. In this report, two cases of complete collapse of BES are presented. In both cases, BES were placed in a heavily calcified aorta and subsequently collapsed without any evidence of external compression. Repeated pulsation of heavily calcified aorta was presumed to be the cause of the stent collapse.

Content from these authors
© 2020 The Editorial Committee of Annals of Vascular Diseases. This article is distributed under the terms of the Creative Commons Attribution License, which permits use, distribution, and reproduction in any medium, provided the credit of the original work, a link to the license, and indication of any change are properly given, and the original work is not used for commercial purposes. Remixed or transformed contributions must be distributed under the same license as the original.

この記事はクリエイティブ・コモンズ [表示 - 非営利 - 継承 4.0 国際]ライセンスの下に提供されています。
https://creativecommons.org/licenses/by-nc-sa/4.0/deed.ja
Previous article Next article
feedback
Top