Dosimetry and Calorimetry Performance of a Scientific CMOS Camera for Environmental Monitoring

by
Alexis Aguilar-Arevalo
1,
Xavier Bertou
2,
Carles Canet
3,
Miguel Angel Cruz-Pérez
4,
Alexander Deisting
5,*,
Adriana Dias
5,
Juan Carlos D’Olivo
1,
Francisco Favela-Pérez
1,2,
Estela A. Garcés
6,
Adiv González Muñoz
6,
Jaime Octavio Guerra-Pulido
1,
Javier Mancera-Alejandrez
7,
Daniel José Marín-Lámbarri
6,
Mauricio Martinez Montero
1,
Jocelyn Monroe
5,
Sean Paling
8,
Simon J. M. Peeters
9,
Paul Scovell
8,
Cenk Türkoğlu
10,
Eric Vázquez-Jáuregui
6 and
Joseph Walding
5
add Show full author list remove Hide full author list
1
Instituto de Ciencias Nucleares, Universidad Nacional Autónoma de Mexico, A. P. 70-543, Mexico 04510, Mexico
2
Centro Atómico Bariloche, CNEA/CONICET/IB, Bariloche 8400, Argentina
3
Centro de Ciencias de la Atmósfera, Universidad Nacional Autónoma de México, Mexico 04110, Mexico
4
Programa de Posgrado en Ciencias de la Tierra, Universidad Nacional Autónoma de Mexico, Ciudad Universitaria, Coyoacán 04510, Mexico
5
Royal Holloway, University of London, Egham Hill TW20 0EX, UK
6
Instituto de Física, Universidad Nacional Autónoma de México, A. P. 20-364, Mexico 01000, Mexico
7
Facultad de Ingeniería, Universidad Nacional Autónoma de Mexico, Mexico 04510, Mexico
8
Boulby Underground Laboratory, Boulby Mine, Saltburn-by-the-Sea TS13 4UZ, UK
9
Department of Physics and Astronomy, University of Sussex, Brighton BN1 9QH, UK
10
SiPM Systems for Astroparticle Physics and Medical Physics Group, Particle Astrophysics Science and Technology Centre, 00-614 Warsaw, Poland
*
Author to whom correspondence should be addressed.
Sensors 2020, 20(20), 5746; https://doi.org/10.3390/s20205746
Submission received: 16 July 2020 / Revised: 16 September 2020 / Accepted: 23 September 2020 / Published: 10 October 2020
(This article belongs to the Special Issue Optical Technologies for Environmental Monitoring)

Abstract

:
This paper explores the prospect of CMOS devices to assay lead in drinking water, using calorimetry. Lead occurs together with traces of radioisotopes, e.g., 210 Pb , producing γ -emissions with energies ranging from 10 k e V to several 100 k e V when they decay; this range is detectable in silicon sensors. In this paper we test a CMOS camera (Oxford Instruments Neo 5.5) for its general performance as a detector of X-rays and low energy γ -rays and assess its sensitivity relative to the World Health Organization upper limit on lead in drinking water. Energies from 6 k e V to 60 k e V are examined. The CMOS camera has a linear energy response over this range and its energy resolution is for the most part slightly better than 2%. The Neo sCMOS is not sensitive to X-rays with energies below 10   k e V . The smallest detectable rate is 40 ± 3 m Hz , corresponding to an incident activity on the chip of 7 ± 4 Bq . The estimation of the incident activity sensitivity from the detected activity relies on geometric acceptance and the measured efficiency vs. energy. We report the efficiency measurement, which is 0.08(2)% (0.0011(2)%) at 26.3 k e V ( 59.5 k e V ). Taking calorimetric information into account we measure a minimal detectable rate of 4 ± 1 m Hz ( 1.5 ± 1 m Hz ) for 26.3 k e V ( 59.5 k e V ) γ -rays, which corresponds to an incident activity of 1.0 ± 6 Bq ( 57 ± 33 Bq ). Toy Monte Carlo and Geant4 simulations agree with these results. These results show this CMOS sensor is well-suited as a γ - and X-ray detector with sensitivity at the few to 100 ppb level for 210 Pb in a sample.

1. Introduction

Ingesting lead can have acute and chronic health effects and it is especially harmful to infants and children. There is no safe threshold for the onset of lead’s negative effects on the human condition and damages are permanent—for example, a child loses three IQ points on average when it consumes as much lead as 25 μ g k g 1 body weight per week over a longer period [1]. It is estimated that globally 26 million people in low- and middle-income countries are at risk of lead exposure [2] as for example, people living in rural areas in Mexico [3]. The main source of lead pollution is from improper recycling of lead-acid batteries. The tonnage of lead in production is rapidly increasing and has grown by more than an order of magnitude in the past decade [2]. The trend towards clean technologies, such as electric cars, will likely increase the demand on lead-acid battery recycling, therefore the corresponding pollution can be expected to increase.
Lead screening tests in the UK, for example, are usually carried out in both public and private water supplies [4]. In public water supplies these are done at water treatment facilities, service reservoirs, water supply points and customer taps in water supply zones. In private water supplies, samples are collected at the point of use. The laboratories that carry out the analysis of these samples are accredited by the United Kingdom Accreditation Service (UKAS) and the Drinking Water Testing Specification (DWTS). In low- and middle-income countries (LMICs), access to such testing is more limited and may be prohibitively expensive. Therefore, a low-cost sensor of lead in drinking water would allow a wider range of people access to on-demand assay methods. A broader programme of measurements enabled by a low-cost technology could have important impacts on mitigating lead intake through contaminated water.
Most people carry a CMOS sensor in their pocket – the silicon chip in their mobile phones’ cameras. Even lower cost CMOS sensors are available off-the-shelf. These silicon chips are in principle capable of measuring radiation as X-rays, γ -rays, and radiation from β - and α -decays. Radioactive isotopes are found in trace amounts together with stable isotopes of lead [5]. Detecting the decay radiation of these isotopes can potentially enable lead detection in food and drinking water. Thus, radio assay methods based on a cheap and already common sensor such as a CMOS chip can be of great help to mitigate lead ingestion, particularly in LMICs. A major challenge is to detect the small signal from trace contamination. This paper reports on a first step towards developing lead radio assay in CMOS by exploring the potential of a scientific CMOS, operated to minimize noise. In this paper we qualify how a scientific CMOS, built for optical light detection, performs as a radiation detector. Previous studies have shown that it is possible to use CMOS sensors to distinguish between α -decays and other types of radiation, as well as counting events from different radiation dosages [6]. It has also been shown that these sensors can provide good spatial resolution, since they allow for a geometrical confinement of the received signal, for either X-rays or for charged particles [7]. Spatial information can be used to distinguish electronic noise and background radiation from the actual signal of interest, although it is by it self not a pre-requisite for dosimetry applications.

1.1. Lead-210 Measurement Applications

One of the radio-isotopes occurring with stable lead is 210 Pb , which decays by β decay (Q-value of 63.5 ± 0.5 k e V ) to an excited state of 210 Bi , which de-excites practically immediately while emitting a γ -ray with 46.6 k e V energy. This lead isotope can occur in trace amounts with Pb , because it is a precursor of the stable lead (i.e., 206 Pb in this case) in the 238 U (or 222 Rn ) decay series. Measurements in for example, Reference [5] show a range of 3.9 × 10 −8 ppb to 2.4 × 10 −5 ppb for certain Pb samples. Assaying 210 Pb at trace levels is required for sediment layer dating in geology [8], to assess pollution levels in environmental monitoring [9], and to identify trace radioactivity in materials for particle physics experiments (e.g., DM searches) [10].
The necessary radio-isotope assays for such studies are done with (high purity) germanium detectors reaching concentrations as low as 1 × 10 −6 ppb. Germanium detectors are also used for, for example, 210 Pb assays of materials for DM experiments [11,12] and many more applications in science and technology. To measure even lower concentrations of 210 Pb in a sample, ashing of that sample has been applied with success (plants: [9], acrylic: [13]). Heavy (radio-)isotopes are retained in these processes, so the decay-rate per volume is enhanced after ashing. The lowest concentration measured reaches limits of down to 1.1 × 10 −20 g 210 Pb per g of acrylic or 1 × 10 −11 ppb [11].
The measurement sensitivity goal for assay of lead in drinking water is the World Health Organization (WHO) upper limit. The WHO quotes a value of 10 ppb [1] as a guideline value for lead in water, although it states explicitly that there is no safe threshold for lead ingestion. In instances of drinking water contamination, levels from 100 to 1000 times this value have been measured [14].

1.2. Outline of This Paper

For this paper we study a CMOS sensor, which has been designed for imaging optical wavelengths. The paper is structured as follows: in Section 2 we give the technical details of the scientific CMOS sensor—Oxford instruments Neo 5.5 scientific CMOS camera [15] (Neo sCMOS); in Section 3 we outline the experimental set-up and our measurement procedures; Section 4 describes the analysis procedure—the steps taken to go from camera exposure frames to the reconstruction of the energy deposited in the CMOS chip. Results from measurements with radioactive sources and a X-ray tube are presented in Section 5. We examine the general performance of the Neo sCMOS when exposed to X-rays and γ -rays. Since the camera has a glass window before the sensor, we can not measure α particles emitted by the sources tested in Section 5. The performance analysis includes a study of the calorimetric capabilities of the camera, that is, its capabilities to measure the energy of incident photons and its energy resolution as function of the incident energy (Section 5.1). Furthermore we measure the background rate without any source present, the minimal detectable rate with an 241 Am source and the camera’s detection efficiency for X-rays and γ -rays (Section 5.2). We use the measured efficiencies to make a rough estimate on the sensor thickness in Section 5.2.4. To check our measured results for consistency we construct a simulation of different layers of the Neo sCMOS in Geant4 [16], which is described in Section 5.3 and we compare our measured spectra with these simulations in order to further understand the effects of the thickness of different layers in the chip. The paper concludes with a discussion of the results (Section 6) in which we estimate the sensitivity of the CMOS approach to measure lead concentration in water down to the WHO of 10 ppb [1].

2. Cmos Camera Specifications

An ideal silicon sensor for the measurement of X-rays and low energy γ -rays would have a thick conversion region to enhance the probability that photons are absorbed in the silicon, low noise, and ideally allows the photons to pass unhindered by for example, an entry windows to the chip. The DAMIC CCDs [17] are a good example for scientific sensors designed for this purpose, as they search for DM by measuring energy deposits 10 ± k e V in their silicon. In this work we explore the potential of commercial CMOS cameras for dosimetry applications, with the aim of understanding the prospects for CMOS-based sensors in the field. The Oxford instruments Neo sCMOS camera [15] was chosen because it provides readout noise comparable with CCDs with relatively more widespread CMOS technology.
The Neo sCMOS features a chip with 2560 × 2160 active pixels, each with a height and width of 6.5 μ m . The active size of the sensor is thus 16.6 × 14.0   m m 2 (height × width). The spatial granularity of the pixels is relevant for the image background characterisation and event classification as discussed in Section 4.1, Section 4.2.3 and Section 4.2.4. Each pixel in the Neo sCMOS’s chip has a typical well depth of 30 × 10 3   e (electron) and is equipped with its own micro lens. This micro lens array ensures that light arriving at the chip’s surface is focused into the active region of the pixels. The Neo sCMOS has two acquisition modes: global shutter, where a full image frame is acquired and rolling shutter, where image data is acquired one row at a time. The detection limit for 210 Pb we are aiming to reach is small, and thus we chose operating conditions to minimize the noise. For rolling shutter with a readout frequency of 200 M Hz , the Neo sCMOS read noise specification is 1   e ( 1.5   e ) mean (RMS) at 30   C . This temperature is enabled by the Neo sCMOS’s cooling system, which allows for cooling the chip to 30   C ( 40   C ) in a room temperature environment (with additional water cooling). For comparison, when a global shutter is used the mean read noise is 2.3   e . Groups of hardware pixels ( 2 × 2 , 3 × 3 , 4 × 4 , 8 × 8 ) can be binned together, prior to readout, in order to reduce the overall contribution of readout noise and increase readout speed. These are then read together as one readout pixel. Operating parameters were chosen to minimize the read noise, to optimize the SNR for the expected small signal. Table 1 lists the operating parameters for the CMOS during the measurements presented in this paper.
No information on the cross section of the camera chip, that is, its different layers and the thickness of the active silicon, is provided by the supplier. Information on the quantum efficiency is only available for radiation in the wavelength range from 300 n m to 1000 n m and not for short wavelengths (X-ray and γ -ray energies). Furthermore, it is not known how layers on top of the silicon (i.e., the micro lenses) affect a measurement of X-rays/ γ -rays. To understand the possible effect of these layers on our results, we performed Geant4 simulations for the different radiation sources, cf. Section 5.3. We furthermore attempt to assess the thickness of the sensor by comparing our measurements to toy Monte Carlo simulations treating the camera as only one silicon layer (Section 5.2.4).
Many of the Neo sCMOS features are not present for cheap, commercial CMOS sensors, especially the various features reducing the noise as the chip cooling or the general noise figures. On the other hand, the Neo sCMOS is a sensor optimised for optical wavelengths and the thickness of its conversion layer can be expected to be on the order of a few μ m to 10   μ m . The micro lense array on top of the actual pixels is an addtional layer as is for example, a Bayer filter on top of the pixel matrix of a different camera chip. Furthermore incident radiation needs to passs through a window before it can reach the chip as is the case for commercial sensors enclosed in a housing, like the camera chip of a mobile phone. The camera has all these properties in common with cheap CMOS cameras.

3. Experimental Description

CMOS data for radioactive source measurements are acquired inside a dark box, with dimensions of 244   c m × 122   c m × 122   c m ( L × W × H ). The large size of the box allows the distance between the radioactive source and the camera to be increased up to 2.5   m . This allows for a measurement of the minimal detectable source activity, as described in Section 5.2.2. The camera is positioned inside the box as shown in Figure 1a and all data is taken whilst the box is closed. Several calibration sources are used to understand the camera’s behaviour and to obtain different spectra for calorimetry. In particular, we take background data without any sources and also use 241 Am , 55 Fe , and 210 Pb sources. Data taking with the X-ray tube is also performed under dark-box conditions, in a different enclosure, shown in Figure 1b and described in Section X-ray Data Taking.
Prior to a data acquisition run the camera is cooled to 30   C and the respective source is positioned in front of the camera, with the source supported such that it is aligned on the chip centre, as in Figure 1a. The NEO sCMOS is controlled via a cable, which is fanned out of the dark-box and connects to a custom PCIe card hosted in the data acquisition computer. We used the Andor SOLIS for Imaging software package for the data acquisition as well as to set camera’s parameters. The operation settings in Table 1 are chosen for several reasons: a readout binning of 4 × 4 is preferred over 1 × 1 due to limitations of the data transfer rate and also to optimise data processing; the number of exposures and exposure time are chosen to ensure enough energy deposits to result in a clear peak in the energy spectra; all the other settings are chosen to ensure a low readout noise. The analysis procedure developed to identify energy deposits in the camera chip relies on a low occupancy. For this reason, an even larger readout binning than 4 × 4 is not chosen, although a larger binning would be even more beneficial for data transfer and data processing. As an exposure time of 10 s is used, the readout time of the camera is not an issue.

X-ray Data Taking

A LD Didactic X-ray apparatus (554 800) [18] is used for the X-ray data taking. Figure 1b shows the camera inside the apparatus. The door to the compartment with the camera is closed before the data taking and the compartment is sealed light-tight. The X-ray tube and camera develop substantial heat, therefore the camera’s water cooling is used to ensure stable operation at 30   C . The rate of X-rays of the apparatus is larger than the rate of any radioactive source we use in this paper, allowing much shorter exposure times than stated in Table 1: 0.004 s and 0.025 s .
The anode in the X-ray tube is made of Mo, with its characteristic K α and K β lines at 17.41 k e V and 19.61 k e V , respectively. Data is also acquired with a Cu or Zr filter between the X-ray tube and the camera. The observation of absorption edges adds more energy measurements in addition to the two X-ray lines, which makes these tests valuable for the energy calibration of the sensor. In Section 5.1.1 our results with the X-ray source are discussed.

4. Data Analysis

The camera control software produces files in the FITS (Flexible Image Transport System) [19] format. Each FITS file can contain several frames, that is, 2D arrays with one Analogue-to-Digital Unit (ADU) measurement for each camera pixel. For example, Figure 2 shows such a frame. After data taking, frames are processed by python code and CERN ROOT [20] routines. During normal data acquisition conditions (Table 1) we take several runs of N f = 100 frames. The data analysis described below identifies clusters due to radiation in each frame. Ultimately all clusters found in the frames of several runs are combined into one set and further data analysis is done on this set (Section 5).

4.1. Frame Background Calculation

The image processing analysis corrects for the pixel pedestal in two steps, described in Section 4.1.1 and Section 4.1.2, then finds clusters of signal pixels, described in Section 4.2, and measures the energy within each cluster. These clusters are identified by their difference to the remaining pedestal value, after the corrections.

4.1.1. Column Correction

The first step of the analysis is to correct for the raw image pedestal, which is defined as the background ADU measurement in each pixel in the absence of a source. We observe that the ADU values of each pixel in a given column are correlated with each other, giving rise to the distinct columns in Figure 2. This correlation pattern is no fixed-pattern-noise as it changes form frame to frame. Therefore we employ the following approach to correct for it on a single frame basis. First, the mean column value C col x , n f and its standard deviation σ C col x , n f for each column is calculated:
C col x , n f = 1 N y y = 0 N y C ( x , y , n f )
σ C col x , n f = 1 N y 1 y = 0 N y ( C ( x , y , n f ) C col x , n f ) 2 , n f = j , x = k ,
where C ( x , y , n f ) is the charge (in ADU ) measured by a pixel at a given x , y position in frame n f . The column coordinate and the frame number are fixed ( x = k , n f = j ) while the sum runs over the row coordinate ( y = 0 N y ). After a first calculation of the column mean and standard deviation using Equation (1) and Equation (2), all pixel values C ( k , y , j ) C col k , j ± 5 · σ C col k , j are excluded and C col k , j and σ C col k , j are calculated again until σ C col k , j changes less than 0.5% between two iterations. This iterative approach is necessary in order to exclude pixels with a high charge value as for example, hot pixels – transient high values in a certain x , y position – or pixels with a higher charge value due to a signal induced by incident radiation. After the final mean and standard deviation is found, that mean is subtracted from each pixel value:
C col sub ( x , y , n f ) = C ( x , y , n f ) C col x k , n f ,
x k in C indicates that the column mean is the same for all the C ( x , y , n f ) along a column with x = k , that is, in y direction.
The result of this column-pedestal correction procedure is shown in Figure 3a, for the raw data of Figure 2. A raw frame recorded during data taking with no source and with 241 Am has a mean of 129 ± 20 ADU and 130 ± 73 ADU, respectively, where the uncertainty is chosen to be one standard deviation. After the column correction the mean moves to 0 ± 18 ADU and 0 ± 73 ADU, respectively.

4.1.2. Time-Series Analysis

At this stage it is possible that there is still fixed-pattern-noise in the recorded frames, for example, pixels which have, in every frame, a C value elevated over the neighbouring pixel’s values. Such pixels may be hot pixels or pixels with charge values of only a few 100 ADU (Figure 4). In order to correct for these we adopt a time-series approach: all charge values C col sub ( x , y , j ) in the n f = j frame in a run, with N f frames in total, are subtracted from their corresponding values in the n f = j + 1 frame, C col sub ( x , y , j + 1 ) . The result are N f 1 frames with pixel intensities T ( x , y , n f ) given by
T ( x , y , n f ) = C col sub ( x , y , n f ) C col sub ( x , y , n f + 1 ) n f = 0 N f 2 .
Note that in Equation (4) the n f starts at zero – hence the last index is N f 2 for N f 1 frames. The time series pair-wise subtraction removes effects that are persistent in time, hence the name. Transient features, for example, radiation from a source, may create negative entries during this procedure.This can be tolerated as long as the source rate is not too high such that transient features occur at the same x , y coordinate in two subsequent frames. Figure 3b shows the effect on the 1D charge distributions. The tails of the distributions change due to the subtraction of transient pixels with a high charge value. This leads to a mean of 0 ± 21 ADU and 0 ± 112 ADU for the time-series corrected frame with no source and 241 Am , respectively. The standard deviation increases, since there are now more negative pixel values in the distribution, from the pairwise subtraction of transient features.

4.2. Clustering

The column corrected and time-series subtracted data, following Equation (3) and Equation (4) are then searched for clusters. A cluster is defined as one or more spatially adjacent pixels which have a charge value larger than the remaining pedestal value.

4.2.1. Threshold Calculation

The threshold value is constructed by a data driven method: We check all N f 1 individual values a pixel at coordinates x = m , y = i measures over the course of a run. In the notation introduced before, these values correspond to all T ( x , y , n f ) , where x and y are held constant and n f runs from 0 to N f 1 . From these charge values a run-averaged pixel pedestal value p ( x , y ) and its standard deviation σ p ( x , y ) are calculated for all pixels at coordinates x , y .
p x , y = T x , y = 1 N f 1 n f = 0 N f 1 T ( x , y , n f )
σ p x , y = 1 N f 2 n f = 0 N f 1 ( T ( x , y , n f ) p x , y ) 2 , x = m , y = i .
This is again an iterative procedure, similar to what is done to calculate the column mean. From subsequent iterations all pixel values T ( x , y , n f ) p x , y ± 5 · σ p x , y are rejected when using Equations (5) and (6) to (re)calculate p ( x , y ) and σ p x , y . Both values are regarded as final when σ p x , y changes less than 0.5% between two iterations. While p ( x , y ) is by construction close to zero for column subtracted and time-series subtracted data, σ p x , y has a minimal value slightly above 6 ADU and most probable value between 14 ADU and 15 ADU, skewed towards higher values. Setting aside time-series subtraction and column correction, the pedestal values and their standard deviation for every pixel should allow to discriminate between background and a charge signal, provided the fluctuations of the background are randomly distributed. However, the column mean of a specific column changes from exposure to exposure, motivating the approach described in this section. The camera’s manual states a read noise RMS value of 1.5   e and a dark current of 0.015   e / pixel / s . Combining these values in quadrature, whilst taking into account the exposure time (10 s ) and the readout binning ( 4 × 4 ), and converting to ADU , we get 4.22   ADU , using the conversion factor of 0.67   e / ADU as specified by the supplier. The modification of this RMS value by the before described column correction and the pair-wise subtraction have to be taken into account before comparing the RMS to σ p x , y . While the corrections described in Section 4.1.1 lead to a negligible reduction of the RMS, the pairwise subtraction (Section 4.1.2) increases the resulting RMS by a factor of 2 . The smallest measured σ p x , y value of 6   ADU fits this expectation of 5.97 ADU, indicating that the corrections applied here do remove most other noise contributions than the read- and dark-noise.

4.2.2. Cluster Finding

The cluster finding algorithm employs two values, a seed and a skirt pixel threshold intensity. The seed is a higher threshold value designed to quickly find the cluster’s largest charge values. The skirt is a lower threshold value designed to find potentially dimmer adjacent pixels to the seed pixel associated with the cluster. The following threshold condition is used to discriminate whether a pixel value T ( x , y , n f ) is part of the background or part of the charge deposit of a signal for example, by radiation incident on the chip:
T ( x , y , n f ) > p x , y + k · σ p x , y k = k seed k skirt .
We distinguish two cases ( k seed or k skirt ) for the multiplier k: First, the factor to find the seed pixel for a cluster ( k seed ). After the seed pixel has been found we check in its vicinity for pixels fulfilling Equation (7) with k skirt , where k skirt k seed . All contiguous pixels with charge values larger than the skirt threshold, as well as the seed pixel, constitute one cluster. For each cluster we store its defining properties such as size, charge, x , y position, frame number, pedestal value and an identification number (cf. Section 5). Figure 5c shows the clusters identified in the previously shown zoomed image of a frame taken with the 241 Am source. The difference in the bare cluster counts and the shape of the cluster charge spectra in Figure 6 obtained during measurements with a radioactive source versus measurements without the presence of a source shows that the clusters selected in Figure 5c are not just noise, but due to the source radiation. The parameter k seed and k skirt are optimised using data obtained without the presence of a radioactive source (background data), while aiming for a low cluster count by recorded frame and a small cluster size. This optimisation is done on all the clusters found in our set of background run. Clusters in the background data will be due to cosmic radiation passing through the chip, due to noise fluctuations and created by radiation from natural radio-isotopes. For these sources we expect a cluster size to be small—especially since we use 4 × 4 readout binning—since γ - and X-ray photons should depose their energy localised and it is not likely that cosmic muons or β particles pass exactly parallel through the chip. With k seed = 10 less than 0.5 clusters per frame are found while a further increase to for example, k seed = 20 does not result in a further reduction. The cluster size decreases exponentially with k skirt and approaches a mean of 3   clusters for all tested k seed values. The change is no longer significant for k skirt 3 . Therefore, we use k seed = 10 and k skirt = 3 during our analysis. These values are to some extent arbitrary. However, using k seed = 10 and k skirt = 3 reduces the amount of background clusters found as stated before. When combined with additional cuts on cluster properties, as discussed in Section 5, these cluster finding settings allow to create almost background free samples.

4.2.3. Cluster Parameters

For each cluster we store the following properties:
  • the frame number
  • the cluster number, which is a counter for all clusters in one frame
  • the x / y position, that is, the coordinates of the seed pixel
  • the cluster size, that is, how many pixels make up a cluster
  • the cluster charge, that is, the integral over all T ( x , y , n f ) in a cluster subtracted by the cluster pedestal (In case of the time-series approach p x , y     0 . Without the time-series approach subtracting the cluster pedestal is essential since it is different from zero.)
  • the cluster pedestal, that is, the integral over all the clusters pixels’ p x , y
  • the charge of the pixel with the highest charge in the cluster (maximal charge).
Analysis based on clusters are done on the full set of clusters found in specific set of data taking runs. Figure 6a,b show the cluster charge as well as the maximal charge without additional cuts on cluster properties. In some data taking runs the camera took a short time to reach a stable state. Therefore, as a precaution, we do not use clusters of the first five frames of each run. Occasional runs with no-stable camera conditions have in general been rejected.
Cluster charge and maximal charge spectra of the background data peak at a few 100 ADU (Figure 6b) and have a tail towards higher values. Their most probable cluster size is 2 pixel (Figure 6d). The shape of the spectra as well as the cluster size are compatible with the expectations that the background counts are created by noise fluctuations and cosmic radiation passing through the chip.
A clear peak-structure is observed in the case of the 241 Am data in Figure 6a. The charge of a cluster should be proportional to the energy deposited by the incident radiation. From 241 Am -decay energy spectra in the literature there should be 5 prominent lines [21,22] at energies stated in Table 2. The spectra presented here show four prominent peaks (Figure 6, first column), which will be discussed in detail in Section 5.1.

4.2.4. Sizes of Identified Clusters

The peaks visible in Figure 6a,e sit on a floor which is itself related to the decay radiation of the source. Examining the cluster charge spectrum as a function of the cluster size (Figure 6c) shows that this floor is mainly due to clusters with a size of 1 and 2 pixels. The peaks are furthermore significantly less prominent for these cluster sizes than for example, for cluster sizes of 3 pixels and 4 pixels. For cluster sizes larger than four pixels the peak heights decrease again. It is interesting to note that there is a correlation between cluster size and cluster charge, that is, energy deposited in the sensor. For increasing cluster size the ratio of clusters with a large charge value to such with a low charge value increases. Gamma radiation and X-rays are expected to interact in the CMOS sensor and to release their energy locally. Therefore, the observed cluster sizes are larger than expected, even more so, given the readout binning of 4 × 4 , resulting side length per readout pixel of 4 × 6.5   μ m = 26   μ m each. As stated in Section 3 the exact layout of the actual CMOS is not known – its different layers may lead to a spread of the charge which reaches a few 10 μ m . Incident radiation can for example, be absorbed in a non-active layer of the chip and then diffuse towards the collection zones. Another possible explanation is that a substantial fraction of the incoming γ energy gets transferred to a few δ -electrons which can then travel more than a pixel length in the sensor, while they produce further ionisation. It can be excluded that the cluster size gets inflated by pixels accidentally assigned to the respective cluster. Comparing the spectrum where only the most energetic pixel per cluster is shown (maximal charge) with the cluster charge spectrum shows that the information from the lower energy pixels in a cluster is needed to measure a spectrum with distinguishable peaks (Figure 6a,e).
For the analysis of the Neo sCMOS’ calorimetric response we require the cluster size to be larger than 2 pixels in order to improve the quality of the peak spectrum. Figure 6e shows the data displayed in Figure 6a but with a size > 2 pixels cut. The floor below the peaks is substantially reduced. Approximately half of the entries in the background spectrum are removed when this cut is applied to the background data (Figure 6f vs Figure 6b).

5. Performance of the Cmos as Radiation Detector

5.1. Calorimetric Capabilities of the Neo Scmos

The camera response to 55 Fe , 210 Pb and 241 Am source radiation is examined in order to determine the Neo sCMOS’ calorimetric measurement capabilities. Background data obtained with no source present is used as well for this analysis. Figure 7 shows cluster charge spectra measured for these sources. To create these, the analysis procedures detailed in Section 4 are applied to the raw frames and a cluster size larger than two pixels is required for all entries in the plots.
The contribution of the source radiation to the spectra has to be disentangled from the contribution of the background radiation. To this end, spectra obtained with radioactive sources and the background spectrum are normalised to the same live-time and then the background spectrum is subtracted from the source spectra. The results are shown in Figure 7b–d.
241 Am :
The cleanest spectrum is obtained with the 241 Am source, which has an activity of 344 ± 17 k Bq as of at the time the measurement. The uncertainty on the initial source activity is not known, therefore a 5% error is assumed. Americium-241 decays via an α -decay to 237 Np . There are many possible α -decays with different Q values from 5000 k e V to 5500 k e V [23], where the most probable (85%) decay has an energy of 5485 k e V . These α -decays occur together with γ -ray emission and X-ray emission by the 237 Np atom [23]. Table 2 lists the two γ energies with the largest yield per decay as well as X-ray lines measured in 241 Am spectra elsewhere. The CMOS chip of the Neo sCMOS camera is housed behind a glass window, with an assumed thickness of 1 m m —therefore the α -particles will not reach the sensor, since the range of α s of this energy is less than 100 μ m [24]. The energy deposits measured with the 241 Am source are thus for the most part due to γ - and X-rays. Attenuation lengths for different γ - and X-ray energies are given in Table 3.
55 Fe :
Iron-55 decays via electron capture to 55 Mn [23]. After the decay, the electron shell re-arranges to match the levels of 55 Mn and to fill the hole from the electron capture. By doing so, Auger-Meitner electrons with an energy of up to 6 k e V are released as well as X-rays of 5.9 k e V and 6.5 k e V . For these X-ray energies the yield per decay is 16.6% and 7%, respectively. Although the source used has a rate of 100   k Bq , the background subtracted spectrum in Figure 7c is compatible with zero. For low charge values, that is, low energy deposits, the spectrum is more erratic—however, no clear peak can be identified. For photon energies 10   k e V we estimate a lower limit for the photon absorption in glass with the data from Reference [25], assuming 10 k e V photon energy and a glass density of 2.23 g / c m 3 . For a window of 1 m m and 2 m m , at least 97% and 99.04% of the X-rays are absorbed in the glass before they reach the chip, respectively. Therefore, the non observation of any clear peak is most likely due to the X-ray absorption in the Neo sCMOS window.
210 Pb :
Lead-210 decays via β decay to 210 Bi as mentioned in Section 1.1. The most probable β decay (84%) results in an excited state of 210 Bi , whilst emitting an electron with an average decay energy of 4.16 k e V . The nucleus de-excites by emitting a γ of 46.5 k e V with a 4% yield per decay. The de-excitation is accompanied by the emission of X-rays from approximately 9 k e V to 16 k e V with a yield of 22% per decay. The second most probable decay (16%) is a β decay with an electron mean energy of 16.2 k e V to 210 Bi in the ground state [23]. The 210 Pb source used has a rate of 185   k Bq . It holds the lead diluted in nitric acid in a small glass vial. It is not likely that any of the low energy β -radiation is detected by the Neo sCMOS, given that the decay electrons have to traverse the liquid, the glass of the vial and of the Neo sCMOS before it can be detected by the CMOS chip. Therefore, similarly to the 241 Am source, only the X-rays and γ -rays are measured.
The 210 Pb spectrum in Figure 7d contains fewer counts than the 241 Am spectrum (Figure 7b). There are several factors contributing to this: First, the activity of the 210 Pb source is a factor of 1.85 lower than the activity of the 241 Am source. The latter source has also a significantly smaller extent—compared to the CMOS sensor it can be considered as a point source, while the lead source extends over a vial of more than 1 c m length and 0.5 c m diameter. Next, the γ yield for the two sources differs greatly—comparing 4 % to 36 % for the 210 Pb 46.5 k e V   γ -ray and the 241 Am 59.5 k e V   γ -ray. In order to establish whether the 241 Am and 210 Pb spectra are consistent with each other, we first need to establish the overall energy scale and compare peaks at a known energy directly.

5.1.1. Energy Response Calibration

All the spectra presented so far are shown with analogue-to-digital units as unit of the deposited energy in the detector. The Neo sCMOS’s manuals consulted during this work do not state a conversion factor from ADU to energy in e V . However, the report [27] specifies a gain of either 0.59 e / ADU or 0.67 e / ADU according to the supplier. These gain values translate to a conversion factor of either 2.154 e V /ADU or 2.446 e V /ADU, respectively, accounting for the W factor in Si of 3.65 e V to create an electron-hole-pair [28]. In order to establish the exact energy scale, the known energies of radioactive sources from literature are matched to the ADU values at which peaks are observed. All large peaks in the cluster charge spectra are fitted with a Gaußian curve and their mean energy, ε peak , and σ is extracted. Table 4 lists all peaks used for this analysis and the result of the fits. The Gaußians are furthermore plotted in Figure 7b,d and Figure 8a. The fits are done locally—in the ranges from ε min to ε max as specified in Table 4—and where necessary a polynomial of order one is added to the Gaußian curve to account for the floor due to other radiation. For the peaks at the high energy end of the 241 Am and 210 Pb spectrum an error-function is used instead of a polynomial.
The second, independent, dataset to obtain the energy scale calibration uses measurements where the CMOS is irradiated by an X-ray tube, described in Section 3. For the spectra obtained with the X-ray tube using only Gaußian fits with a local background is not sufficient (Figure 8a): The two characteristic peaks of the molybdenum X-ray tube are expected to be located on top of the bremsstrahlung spectrum of the tube. For low X-ray energies the camera has negligible calorimetric capabilities as seen in the measurements with the 55 Fe source (previous section, Figure 7c). Hence, the Neo sCMOS should become efficient for X-rays of the molybdenum X-ray tube somewhere after 6   k e V – from that point onwards there should be an increasing number of counts due to bremsstrahlung and eventually the molybdenum K α and K β peaks at 17.4 k e V and 19.6 k e V , respectively. Figure 8a shows the spectrum, the fit to the spectrum, and the fit’s components. An onset of counts is observed at 5   k ADU however, no clear double peak structure is observed. As there is no clear expected functional shape for the bremsstrahlung contribution, we model it as the minimal functional addition ( B r e m s ( ε ) ) needed so B r e m s ( ε ) + G a u ( ε ) K α + G a u ( ε ) K β fits the data well.
B r e m s ε = p 0 · exp 1 2 ε p 1 σ 2 + p 2 · 1 erf ( p 3 · ε p 4 ) + p 5 · ε + p 6
G a u ( ε ) K j = p 0 j · exp 1 2 ε ε peak j σ j 2 j = α β .
In these equations ε is the cluster charge (or energy deposited in the chip) in ADU . The parametrisation (8) for the bremsstrahlung contribution yields the lowest χ 2 / N dof of 4.62 for the total fit of B r e m s ( ε ) + G a u ( ε ) K α + G a u ( ε ) K β to the data, whilst all fit parameters are free. The extracted parameters of the two K lines ( ε peak , σ peak ) are listed in Table 4.
Using a Cu or a Zr foil to filter the molybdenum X-rays results in the spectra shown in Figure 8b. The absorption edges of those two elements for energies higher than 6   k e V are at 8.98 k e V (Cu) and at 18 k e V (Zr). The shape of the spectrum recorded with the Cu filter does not feature a drop which can be identified with an absorption edge—the edge is thus placed in the energy range where the Neo sCMOS is not sensitive to allow X-ray calorimetry. There is a larger reduction of counts for energies 7   k ADU , relative to the not filtered spectrum and the one with the Zr filter. After, the number of counts increases again. Starting from low energies, the shape of spectrum with the zirconium filter matches the un-filtered spectrum, until the edge at 7.5   k ADU . This drop is identified with the Zr absorption edge. The ADU value at which the amplitude reaches the 50% value between maximal peak height and the floor in the spectrum is taken as its energy position (Table 4). The uncertainty on the edge’s position is taken to be half of the ADU range between the edge’s 10% and 90% value.
Figure 9a displays the measured charge values plotted against their expected energies for all peaks and edges in Table 4. Fitting a linear function without an axis intercept to these points yields the conversion factor from e V to ADU (and vice versa) to be 0.405 ± 0.001 ADU/ e V ( 2.467 ± 0.007 e V /ADU). For this fit χ 2 / N dof is 1.64 while using a function with an intercept results in a χ 2 / N dof of 0.56, an intercept of − 0.36 ± 0.09 k e V , and a slope of 2.434 ± 0.009 e V /ADU ( 0.41 ± 0.01 ADU/ e V ). These values are compatible with the conversion factor mentioned before, although slightly different from an intercept of zero. In the next sections the conversion factor without an intercept is favoured over the conversion with an intercept, since the low χ 2 / N dof in the latter case indicates over-fitting.
The measured conversion factors matches well with the higher of the two gain values discussed before, that is, 2.446 ADU/ e V which is located between the two different fit values. This agreement is taken as another reason to use in the following the conversion factor determined without an intercept, given the supplier does not specify an offset.

5.1.2. Energy Resolution

In Figure 9b the energy resolution is shown as σ peak divided by the peak positions ε peak (Table 4). As the uncertainty on σ the uncertainty of the fit is used while σ peak itself is used as the uncertainty of the peak position ε peak . The uncertainty on the energy resolution, Δ σ peak / ε peak includes both of these contributions. For the most part the resolution is better than 2%. Outliers from this trend are the two molybdenum X-ray lines and the Np, L γ 1 line (at 20   k e V ). The two X-ray lines are extracted from a more complicated fit with the worst χ 2 / N dof and the uncertainty on their σ peak values is likely to be underestimated.
The energy resolution is determined by several factors. The full containment of all electrons produced during the photon conversion and their subsequent readout will play an important role. Furthermore their can be pixel-to-pixel variations of each pixels’ amplifier gain. An increasing trend in cluster size with increasing energy has been observed, as stated in Section 4.2.4, but the good linearity of the e V to ADU relation suggests that all electrons produced by a photon interaction are read out. The pixel-to-pixel amplifier gain variations for the Neo sCMOS are not known. Typical values are in the few % range—for example, Reference [29] shows a variation of about 2 % [29]. Such a variation would be consistent with the energy resolution shown here. Note that the energy measurements in Figure 9b are the result of summing the energy measured in each pixel of a cluster. If the per-pixel amplifier variation would be the dominating factor for the energy resolution, the Neo sCMOS would have variations slightly larger than 2 % .

5.2. Radiation Detection Efficiency

In this section we quantify the minimum detectable radioactivity using the Neo sCMOS, and measure the efficiency of the sensor as a detector for γ - and X-rays. Both are done using the 241 Am source, since this source has a well suited activity and an energy spectrum with clear peaks.

5.2.1. Geometric Acceptance of the Experimental Set-Up

The fraction of the 241 Am activity detected by the sensor depends on the source-detector distance. The geometric acceptance ( ϵ G ) is calculated assuming the 241 Am emits radiation as a point source, as
ϵ G = A spherical cap A sphere A camera A camera plane = 0.0889   c m 2 r 2 ( r in c m ) .
The calculation of this expression exploits the sphere into which the source emits radiation and its intersection with the plane of the camera chip. Therefore, A sphere is surface area of that sphere, A spherical cap is the surface area of the base of the spherical cone covering the camera chip, A camera plane is the corresponding surface area of an otherwise similar cone with a flat base, A camera is the surface area of the camera and r is the camera to source distance. Figure 10 compares the analytical estimate of Equation (10) with a toy Monte Carlo simulation, using the actual source geometry, showing good consistency.

5.2.2. Minimum Detectable Radioactivity

Data is acquired for 3800 s each at distances from 1.8 c m up to 258.6 c m between source and sensor. From now on we omit “between camera chip and radioactive source” when referring to the distance between these two components. The detected spectra, the inferred incident activity using the geometric acceptance from Equation (10) and the known source activity are shown in Figure 11. At a distance of 258.6 ± 0.1   c m this activity is as low as 0.42 ± 24   Bq . It is also of interest to determine the incident activity considering only certain γ -ray energies emitted by the 241 Am source, as opposed to the full source activity. These are obtained by multiplying the incident activity for all 241 Am decays by the respective peak yields which are 2.40(4)% and 35.9(4)% for the 26.3 k e V and 59.5   k e V energy peaks, respectively [23]. These incident activities can be seen in Figure 11a. To obtain the uncertainty on the incident activities in this figure we propagate the distance uncertainty through the calculations in Equation (10), as well as the uncertainty of the initial activity of the 241 Am source and the uncertainty of the emission probabilities. In the case of the toy Monte Carlo simulation, the statistical uncertainty on the counts obtained is used.
In order to estimate the minimum detectable radioactivity and eventually the Neo sCMOS’ detection efficiency, the detected rates in Figure 11b are determined as
r a t e = s p e c t r u m d t i m e t i m e counts s .
The measured rate without any source present is calculated as well using the spectrum in Figure 6f. During 30400 s of data taking 619 clusters with a size > 2 pixels are recorded, yielding a background rate of 20.4 ± 0.88 m Hz . In windows corresponding to 5 σ of the integration limits around the 26.3 k e V and the 59.5 k e V γ -peak energies (Table 4) the background rate measured with no source present is 1.1 ± 0.2 m Hz ( 0.10 ± 0.06 m Hz ) for the 26.3 k e V ( 59.5 k e V ) peak region. The exposure time per frame and the number of considered frames are known with great certainty, hence the statistical uncertainty on the cluster count is the only contribution to the uncertainty of the quoted rates. Figure 11b shows the background rates added with five times their uncertainty as yellow shaded regions and as horizontal lines. The green shaded regions correspond to 90% Confidence Level (CL), that is, 1.28 σ , around the measured background rate, where we use the background uncertainty as standard deviation, that is, σ . Thus count rates higher than 21.41 m Hz for the full background rate, 1.4 m Hz for the background rate in the 26.3 k e V peak window and 0.17 m Hz for the background rate in the 59.5 k e V peak window, respectively, are larger than the background at 90% CL. Overall, our results are for the most case the same when when using 5 σ or a 90% CL criterion to differentiate between the measured background rate and the source rate. Table 5 shows all result for both cases, in the following we refer to the results obtained with the 5 σ criterion. The detected rate measured for a distance of 128.5 c m is the first to be compatible with the background rate added with five standard deviations ( 27 ± 3 m Hz ), while the rate measured at 64.5 c m distance ( 40 ± 3 m Hz ) is significantly larger than the background rate. The incident source activity at 64.5 c m and 128.5 c m distance is 7 ± 4 Bq and 2 ± 1 Bq , respectively.
Performing the same analysis for the two most intense 241 Am γ -lines allows to establish whether this limit can be improved taking calorimetric information into account. For the 26.3 k e V line a falling trend for the detected rate is observed over the full distance range (Figure 11b). Comparing to the rate measured when no source is present the point at 64.5 c m is already compatible within five standard deviations. For 32.5 c m distance the detected activity is 4 ± 1 m Hz , corresponding to an incident activity of 1.0 ± 0.6 Bq . In the case of the 59.5 k e V γ line, the 32.5 c m point is already compatible with the background rate for this energy window within five standard deviations. The detected rate at 16.5 c m of 1.5 ± 0.1 m Hz , corresponding to an incident activity of 57 ± 33 Bq , is thus the minimal detected rate, different from the background rate in this energy window.

5.2.3. Detection Efficiency of the Neo Scmos

A comparison between Figure 11a,b allows to calculate both the intrinsic and absolute efficiency of the detector. These are defined as
ϵ intrinsic = number of particles recorded number of particles incident on the detector
ϵ absolute = number of particles recorded number of particles emitted by source = ϵ intrinsic · ϵ G ,
where ϵ G is the geometric acceptance, which is given by Equation (10). The ratio of the recorded rate (Figure 11b) and the incident activity (Figure 11a) yield the intrinsic efficiency of the Neo sCMOS camera for γ -rays of an 241 Am source. This is shown in Figure 12a for the 26.3 k e V and 59.5 k e V peaks as a function of the camera-to-source distance. For completeness, the absolute efficiencies for these two γ -rays are shown in Figure 12b.
By definition ϵ intrinsic has to be independent of the distance. The results of a constant fit to the data are shown in Figure 12a. However, a weak distance dependence of ϵ intrinsic is observed; this is likely caused by a slight discrepancy between the calculated geometric acceptance and the actual source geometry. However, all points are within error-bars compatible with a constant as expected from Equation (12). The absolute efficiency’s dependence on the distance in Figure 12b is expected, due its dependence on the geometric efficiency (Equation (13)). The corresponding intrinsic efficiencies for the 26.3 k e V and the 59.5 k e V peaks are 0.08(2)% and 0.0011(2)%, respectively. Thus, the efficiency to detect a 26.3 k e V γ -ray is about a factor of 80 higher than the efficiency to detect a γ -ray with an energy of 59.5 k e V . To check this ratio for consistency the material composition and thickness of the sensor would need to be known, which is however not the case. Figure 13b in the next section shows photon absorption efficiencies calculated from the attenuation coefficients in Reference [30]. Depending on the CMOS thickness, the difference in photo-absorption for 59.5 k e V and 26.3 k e V can easily reach a factor 80. However, the efficiency at the 59.5 k e V γ -line appears lower than expected from photo-absorption cross sections in Si.

5.2.4. Cmos Sensor Thickness

Based on the efficiencies determined in the previous section one can estimate the thickness of the CMOS chip used in the Neo sCMOS camera. A toy Monte Carlo (MC) simulation of the 241 Am spectrum is used together with the photon attenuation coefficient from Reference [30] to calculate, for each photon, the energy-dependent photon absorption probability in silicon. This step is repeated for different silicon thicknesses. For each thickness we create a spectrum of the photons absorbed in Si—these photons are the ones which would be measured by a chip as in the Neo sCMOS. The ratio of the unattenuated 241 Am toy MC spectrum divided by the spectra of absorbed photons is constructed. Finally, these ratios are compared to the ratio of the measured count rate at the γ peak energies divided by the incident activity for the respective γ line.
For the toy MC, all possible γ -rays and X-rays of 241 Am decays as listed in Reference [31] are taken into account. These are used to create a probability density function for photons emitted during 241 Am decays. Each γ - and X-ray line is represented by a Gaußian peak where the amplitude is proportional to the yield per decay. The ε peak is set to the respective X-ray or γ -ray energy and the σ peak is set to 2% of the peak energy to approximate the measurement for the Neo sCMOS in Section 5.1.2. Creating 1 M decays from this probability density function results in the No att. cluster charge spectrum in Figure 13a. This unattenuated spectrum does not take material effects into account.
Figure 13a shows also spectra of photons absorbed in silicon layers with a thickness from 1 μ m to 5 μ m . Toy MC spectra for absorbed photons and the measured spectrum (e.g., Figure 7b) are similar. The ordering of the different peaks’ height is the same except for the fact that the largest measured peak is at 18 k e V , while the peak with the most counts in the simulation is the peak at 14 k e V . In Section 5.1.1 the detection efficiency is observed to drop at lower energies—most notably making it impossible to detect a peak below 10   k e V , cf. Figure 7c and Figure 8. The efficiency turn-on responsible for this behaviour may also affect the peak height of the 14 k e V peak and lead to the non-observation of the small peak visible at 11   k e V in Figure 13a.
Figure 13b shows photon absorption efficiency curves calculated from the coefficients in Reference [30] for 1 μ m , 2 μ m and 4 μ m silicon layer thickness. Furthermore, the intrinsic efficiency values calculated in Section 5.2.3 are shown (Figure 12a). The two points follow roughly the trend expected by photon absorption in silicon, although the efficiency measured for the 59.5 k e V peak is lower than expected from any absorption efficiency curve. The actual peak height of any γ -line is given by the absorption probability as well as by the capability of the active material to contain the full energy deposit. The latter is not contained in the toy MC simulation and this may explain that the 59.5 k e V does not fit with the displayed curves. The 26.3 k e V favours a silicon layer thickness of 2 μ m . This scale of a few μ m is on the same order of magnitude as the pixel dimension and seems reasonable for commercial CMOS chips, with a typical silicon-dioxide layer thickness of less than 10 μ m [6,7]. Since the silicon thickness is not the only contribution to the Neo sCMOS intrinsic efficiency, the estimate in this section can only be seen as a lower limit to the actual sensor thickness.

5.3. Geant4 Simulations of the Neo Scmos Detector

To extend the limited knowledge of the sensor geometry beyond the estimations based on the toy MC simulations, a Geant4 MC simulation study is carried out to estimate the impact of the window, the micro lens array and sensor thickness on the γ - and X-ray absorption.
Geant4 [16,32,33] version 10.5 patch 1 is used to simulate the primary particles and the production of the resulting secondary particles, for the tracking of all particles through the detector geometry, and to assess the energy deposition in the sensitive detector parts. This work employs physics lists which follow mainly the Shielding physics list of the above mentioned Geant4 version with one physics modification; namely G 4 EmStandardPhysics _ option 4 . G 4 EmStandardPhysics _ option 4 is used instead of G 4 EmStandardPhysics , because the former is more accurate for low-energy electromagnetic interactions.

5.3.1. Detector Geometry and Simulated Particles

The sensor is modelled as a silicon CMOS layer, behind a glass layer to represent the entrance window, behind an acrylic layer to represent the micro lens array. For this simulation, the material of the window is chosen to be SiO 2 . The camera specifications indicate an organic material for the micro lens. Thus, for the material of this volume element, acrylic C 5 H 8 O 2 is chosen. For the thicknesses of each volume element, values in the range of O ( 1 - 1000   μ m ) are used. This study varies the thicknesses of the silicon and glass window. As discussed in Section 5.1, 210 Pb decays either by a combined γ - and β -decay (84(3)%), where the γ -ray has an energy of 46.5 k e V and an average β energy of 4.2 k e V , or by a pure β -decay (16(3)%) with an average β energy of 16.2 k e V . When simulating electrons of this energy impinging on the detector with a 200 μ m glass window and 5 μ m of Si thickness, there are zero hits on the active region out of 10 6 simulated events. Since the energy of the other β is even lower, the signature 210 Pb γ -ray of 46.5 k e V is the main focus of this simulation work whereas β -particles are not included.
The simulated particle source is located 14.06 m m from the outermost layer of the surface, and fires mono-energetic γ - and X-rays directly at the detector.
To construct the energy variable used to compare with data, the simulated ionisation energy deposition in the Si layer is recorded for each incident particle, and then smeared according to the energy resolution function
σ peak ε peak = p 0 + p 1 ε fit
fitted to the data in Figure 9b.

5.3.2. Analysis of the Simulated Spectra

A set of different thicknesses for the silicon layer and the glass window are simulated for comparison with the data. The list of thicknesses simulated ranges from 2–5 μ m for the Si, and 200–2000 μ m for the window, listed in Table 6. Figure 14 shows example spectra of the energy deposition in the Si layer for 200 μ m (Figure 14a), 1000 μ m (Figure 14b) and 2000 μ m (Figure 14c) glass thickness and a constant 4 μ m micro lens thickness. Colours represent different thicknesses of the silicon layer varying in between 2 μ m and 5 μ m while the thickness of other volume elements is kept constant. There are two visible differences between each simulated template:
  • The total number of events registered in the silicon layer increases with Si thickness which is caused by more particles being absorbed by a thicker Si layer.
  • The number of events in the photo-peak and Compton continuum increases and decreases, respectively, as the silicon thickness increases. This is because the fraction of events for which the incident photons’ energy is fully contained in the Si rises as the thickness increases.
The information regarding the number of events in the photo-peak can be used to characterise the thickness of the silicon layer in conjunction with the amount of events in the Compton continuum. The ratio η is used as a variable to compare the simulated templates with data:
η = ε peak σ peak ε peak + σ peak N events d N 0 ε peak σ peak N events d N ,
where ε peak stands again for the energy of the photo-peak, σ peak for the peak’s standard deviation and N events for the number of events. The exact value of the photo-peak and the energy resolution of the detector need to be known, to precisely define the integration limits in (15). We take ε peak to be 46.539 ± 0.001 k e V [23] and calculate σ peak from the energy resolution for this energy using Equation (14) to be 0.595 k e V . η -values for different silicon layer and window thickness calculated from the simulation results can be found in Table 6.

5.3.3. Comparison of Simulation with the Experimental Data

The thickness of the Neo sCMOS active layer, assuming Si as material, is estimated by comparing the spectra in Figure 14 and the photopeak ratio η with the data.
When this variable is calculated for the data (black line in Figure 14), it is found to be η data = ( 2.7 ± 0.2 ) × 10 2 . Comparing η data with the values calculated from simulated templates (cf. Table 6), it can be seen clearly that a combination of 4 μ m silicon layer and 1000 μ m glass window comes closest to this value with η = ( 2.8 ± 0.2 ) × 10 2 .
As another check, one can investigate the spectral shape of simulation with respect to experimental data. Above 30 k e V , the spectral shape of the data are closest to the spectrum of the 2 μ m Si thickness simulation. Below that, the shape matches of the measured spectrum lies between the spectrum simulated for a 3 μ m and 4 μ m thick Si layer. All templates except for the 4 μ m and 5 μ m ones show a discontinuity before the photo-peak as the energies of the particles increase, which is an expected result due to thinner silicon layers being less efficient in stopping γ -rays and containing the particles energy.

5.3.4. Closing Remarks

The close match of η -values and the spectral shape between the data and the Geant4 simulation suggest a silicon thickness between 2 μ m and 4 μ m for a window of 1000 μ m , where the comparison to η data places the thickness at the high end of this range. Table 6 shows that for this geometry only ( 1.4 ± 0.1 ) × 10 3 % of the γ -rays of 210 Pb are fully absorbed in the silicon layer, that is, contribute to the photo-peak. In Section 5.2.3 the intrinsic efficiencies for the 26.3 k e V and 59.5 k e V 241 Am -peak are determined to be ( 80 ± 20 ) × 10 3 % and ( 1.1 ± 0.2 ) × 10 3 % , respectively. The efficiency for the 59.5 k e V γ -line is compatible with the value simulated here for 46.5 k e V γ -rays. Given the fact that the measured value is at a higher energy than the 46.54 k e V of the simulated γ -rays, the Geant4 simulation most likely underestimates the Neo sCMOS detection efficiency at 46.5 k e V slightly or the actual Si thickness is larger than 4 μ m . This is, because the efficiency to detect 46.54 k e V photons is expected to be larger than the one for higher energy photons.
Both the Geant4 simulation of 210 Pb γ -rays and the toy MC simulation of 241 Am photons place the silicon layer thickness in the range between 2 μ m and 4 μ m , cf. Section 5.2.4, Figure 12b. While the full Geant4 simulation gives more parameters to compare between the data and the simulation, the toy MC simulation is significantly faster as it runs in an instant. The agreement between the two gives confidence to use the less sophisticated toy MC simulation in instances where a full fledged Geant4 simulation is not easily accessible.

6. Summary and Discussion

An Oxford instruments Neo 5.5 scientific CMOS [15] camera is examined as a detector for photons in the X-ray and low γ -ray energy regime. This camera is designed to image photons of optical wavelengths. The analysis (Section 4) of camera images identifies clusters (Section 4.2) – contiguous pixels with high charge values (unit: ADU ) – which are due to energy deposits of radiation impinging on the camera chip.
Requiring the cluster size to be larger than 2 pixel allows to sufficiently reject most of the clusters due to background radiation. We note a trend towards larger cluster sizes for increasing photon energy (Section 4.2.4).
The relation of the cluster charge in ADU to e V is measured to be 2.467 ± 0.007 e V /ADU or 0.405 ± 0.001 ADU/ e V (Figure 9a, Section 5.1.1). This relation is linear without an offset in the energy range from 13.8 k e V to 59.5 k e V . It is in reasonable agreement with 2.446 e V /ADU, which is the expected value based on the supplier specified gain value of 0.67 electron/ADU and the energy of 3.65 e V [28] needed to create an electron-hole-pair in Si. The energy resolution is for the most part slightly better than 2% (Section 5.1.2). We have not been able to measure any peaks below 10   k e V , which is most likely due to the photon absorption in glass for energies 10   k e V (cf. the comments on 55 Fe in Section 5.1).
The rate of background events detected without the presence of any source is measured to be 20.4 ± 0.8 m Hz (Section 5.2). By increasing the distance between the camera and an 241 Am source we reduce the activity incident on the camera chip. The lowest detectable rate measured 5 σ above background is 40 ± 3 m Hz , which corresponds to an incident activity of 7 ± 4 Bq . Taking calorimetric information into account and integrating only around the 26.3 k e V and 59.5 k e V γ -peaks, the minimal detectable rate is 4 ± 1 m Hz and 1.5 ± 0.1 m Hz , respectively, which corresponds to incident activities of 1.0 ± 0.6 Bq and 57 ± 33 Bq (Section 5.2.2), respectively.
Comparing the measured rates and incident source activity allows to determine the intrinsic efficiency of the Neo sCMOS camera at the two 241 Am γ -lines. They are found to be 0.08(2)% and 0.0011(2)% for 26.3   k e V and the 59.5   k e V peaks, respectively (Section 5.2.3). The efficiency drop from the lower to the higher energy follows roughly the drop of the photon-absorption efficiency in silicon of a few μ m thickness as function of energy. The absolute value of the efficiency for 59.5 k e V is lower than expected for a silicon sensor with a few μ m thickness, when only photo-absorption coefficients are considered (Section 5.2.4). Geant4 simulations (Section 5.3) show that for thin silicon layers the fraction of an absorbed photon’s energy contained in the silicon decreases with increasing energy of the photon. Thus, lowering the fraction of counts in the photopeak and the detection efficiency measured at the energy of the photopeak. The increasing volume, in which a photon deposes its energy, for increasing photon energy fits the measurement of larger cluster sizes for high energy γ -lines (Section 4.2.4). The Geant4 simulations as well as the toy MC simulations indicate a thickness in the order of 2 μ m to 4 μ m for Neo sCMOS silicon layer.

Lead Detection Capabilities Assuming Lead-210 as a Trace Isotope

The ratio of the radioactive isotopes in P b to stable P b isotopes is required to estimate the capability of the Neo sCMOS to detect lead in drinking water. In Reference [9] the 210 Pb to stable lead ratio is used to monitor the lead intake of plants. They measure 96 ± 9 Bq / m g for 210 Pb / Pb in rainwater in London, corresponding to a ratio of 34 ppb of 210 Pb / Pb (molar mass of lead: 207.2   g /mol, half-life of lead: 22.3 yr). However, the origin of the stable lead and 210 Pb are not necessarily the same. In Reference [5] several Pb samples of different age – from ancient lead to recently produced lead – are analysed for their 210 Pb , 232 Th and 238 U contents. They find 210 Pb / Pb ratios from 0.09 Bq / k g to 68.7 Bq / k g ( 3.9 × 10 8 ppb to 2.4 × 10 5 ppb) in their lead samples.
In the following we will estimate the incident activity on the Neo sCMOS, produced by 10 ppb lead in water. Doing so the 210 Pb / Pb ratio of Reference [9] will be used, since it results from a measurement of Pb in water. However, the results in Reference [5] have to be kept in mind as caveat.
With the 210 Pb / Pb ratio of 34 ppb, the WHO limit of 10 ppb [1] of lead in drinking water translates to a fraction of 3.4 × 10 16 parts 210 Pb to one part of water. One gram of water contains at this ratio 9.9 × 10 5 210 Pb atoms, which initially decay at a rate of 0.96 m Bq ( R 210 Pb decay ). A sample of water containing lead could be placed on the camera’s window – in 1.75 c m distance from the silicon chip. Considering 1 g of water as point source, the incident rate ( R 210 Pb incident ) is only 0.001 m Hz after taking the geometric acceptance in Equation (10) into account as well as the fact that the γ -yield for the 45 k e V γ is only 4%:
R 210 Pb incident = R 210 Pb decay · ϵ G 1.8   c m · 210 Pb γ yield = 0.96 × 10 3   Bq · 3 ( 1 ) % · 4 % = ( 1.1 ± 0.4 ) × 10 6 Bq .
R 210 Pb incident has to be compared to the measured value of the incident 241 Am source activity of 1.0 ± 0.6 Bq and 57 ± 33 Bq at the 26.3 k e V and 59.5 k e V lines, respectively (Section 5.2.3). With 46   k e V the 210 Pb γ -line is located between these two energies. Thus, the sensitivity of the Neo sCMOS is a factor of 10 6 to 10 7 too low to detect the decay radiation of trace amounts of 210 Pb occurring with 10 ppb lead in 1 g of water. The fraction is even lower, given that the above calculations assume a point source and 1 g of water measures a 1 c m 3 .
Independent from the actual 210 Pb / Pb ratio for commercial lead or lead in the water one can estimate the fraction of 210 Pb per mass (e.g., 210 Pb / Pb or 210 Pb per gram of water) which corresponds to a R 210 Pb incident as the measured, minimal incident rate. Considering the same parameters as for Equation (16) and using the two 241 Am γ -line energies we estimate a sensitivity between 0.3 ± 0.2 ppb and 166 ± 11 ppb 210 Pb per mass. This sensitivity is by itself not enough to reach the WHO limit, but makes the Neo sCMOS a competitive radiation detector.
Two more points need to be mentioned: the first estimate depends highly on the 210 Pb / Pb ratio; and, water can be concentrated by boiling it, potentially leaving heavy metals behind. In case of ashing of plants and acrylic [9,13], it has been found that heavy metals stay behind after the process. Furthermore, 210 Pb is not the only trace isotope potentially occurring together with stable lead. In case the radio-isotope to stable lead ratio is larger than the 210 Pb / Pb in Reference [9], the γ yield of these radio-isotope decays is larger than the one of 210 Pb – the same is true when the water’s volume has been reduced.

7. Outlook and Conclusions

Based on the results above, we are currently examining how well 210 Pb is retained when water is concentrated by volume reducing it through boiling off. Furthermore, while our results do not allow to conclude that we can measure 210 Pb in low enough concentrations as needed to detect 10 ppb of lead, our study has shown that a CMOS sensor optimised for optical wavelengths is well suited as γ - and X-ray detector for low energies in the range from 10   k e V to 60   k e V . Follow up studies will establish the performance of commercial CMOS sensors and whether they are suitable for radio assay of materials.

Author Contributions

Conceptualization, A.D. (Alexander Deisting) and J.M.; methodology, A.D. (Alexander Deisting) and A.D. (Adriana Dias) and J.M. and J.W.; software, A.D. (Alexander Deisting) and A.D. (Adriana Dias) and C.T.; validation, A.D. (Alexander Deisting) and A.D. (Adriana Dias) and C.T.; formal analysis, A.D. (Alexander Deisting) and A.D. (Adriana Dias) and C.T.; investigation, A.D. (Alexander Deisting); resources, J.M.; data curation, A.D. (Alexander Deisting) and A.D. (Adriana Dias); Writing—Original draft preparation, A.D. (Alexander Deisting) and A.D. (Adriana Dias) and C.T.; Writing—Review and editing, A.A.-A. and X.B. and C.C. and M.A.C.-P. and A.D. (Alexander Deisting) and A.D. (Adriana Dias) and J.C.D. and F.F.-P. and E.A.G. and A.G.M. and J.O.G.-P. and J.M.-A. and D.J.M.-L. and M.M.M. and J.M. and S.P. and S.J.M.P. and P.S. and C.T. and E.V.-J. and J.W.; visualization, A.D. (Alexander Deisting) and A.D. (Adriana Dias) and C.T.; supervision, J.M. and S.J.M.P. and J.W.; project administration, J.M. and E.V.-J.; funding acquisition, X.B. and J.M. and E.V.-J. All authors have read and agreed to the published version of the manuscript.

Funding

This research was funded by STFC Global Challenges Research Fund (Foundation Awards, Grant ST/R002908/1), by STFC Grant ST/T506382/1, by DGAPA UNAM grants PAPIIT-IT100420 and PAPIIT-IN108020, and by CONACyT grants CB-240666 and A1-S-8960.

Acknowledgments

The Authors wish to thank Ian Murray, Royal Holloway, University of London for his technical support.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Guidelines for Drinking-Water Quality, 4th Edition, Incorporating the 1st Addendum. Available online: https://www.who.int/water_sanitation_health/publications/drinking-water-quality-guidelines-4-including-1st-addendum/en/ (accessed on 23 September 2020).
  2. Pure Earth. Fact Sheet: Lead. Available online: https://www.pureearth.org/lead-fact-sheet/ (accessed on 23 September 2020).
  3. Caravanos, J.; Dowling, R.; Téllez-Rojo, M.M.; Cantoral, A.; Kobrosly, R.; Estrada, D.; Orjuela, M.; Gualtero, S.; Ericson, B.; Rivera, A.; et al. Blood Lead Levels in Mexico and Pediatric Burden of Disease Implications. Ann. Glob. Health 2014, 80, 269–277. [Google Scholar] [CrossRef] [PubMed]
  4. Drinking Water and Health: A Guide for Public and Environmental Health Professionals and for Those in the Water Industry in Northern Ireland. Available online: https://www.niwater.com/sitefiles/resources/pdf/guidancedocumentondrinkingwaterandhealth13.06.04.pdf (accessed on 23 September 2020).
  5. Orrell, J.L. Assay methods for238U,232Th, and210Pb in lead and calibration of210Bi bremsstrahlung emission from lead. J. Radioanal. Nucl. Chem. 2016, 309, 1271–1281. [Google Scholar] [CrossRef] [Green Version]
  6. Pérez, M.; Lipovetzky, J.; Haro, M.S.; Sidelnik, I.; Blostein, J.J.; Bessia, F.A.; Berisso, M.G. Particle detection and classification using commercial off the shelf CMOS image sensors. Nucl. Instrum. Methods Phys. Res. Sect. Accel. Spectrom. Detect. Assoc. Equip. 2016, 827, 171–180. [Google Scholar] [CrossRef]
  7. Servoli, L.; Biagetti, D.; Passeri, D.; Gattuso, E.S. Characterization of standard CMOS pixel imagers as ionizing radiation detectors. J. Instrum. 2010, 5, P07003. [Google Scholar] [CrossRef]
  8. Appleby, P. Three decades of dating recent sediments by fallout radionuclides: A review. Holocene 2008, 18, 83–93. [Google Scholar] [CrossRef]
  9. Yanga, H.; Appleby, P.G. Use of lead-210 as a novel tracer for lead (Pb) sources in plants. Sci. Rep. 2016, 6, 21707. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  10. Amaudruz, P.A.; Baldwin, M.; Batygov, M.; Beltran, B.; Bina, C.; Bishop, D.; Bonatt, J.; Boorman, G.; Boulay, M.; Broerman, B.; et al. Design and construction of the DEAP-3600 dark matter detector. Astropart. Phys. 2019, 108, 1–23. [Google Scholar] [CrossRef] [Green Version]
  11. Nantais, C.M. Radiopurity Measurement of Acrylic for the DEAP-3600 Dark Matter Experiment. Master’s Thesis, Queen’s University, Kingston, ON, Canada, 2014. [Google Scholar]
  12. Widorski, M. Development of a Detection System to Measure Low Levels of 210Pb using β/γ Coincidence. Master’s Thesis, Royal Holloway University of London, London, UK, 2013. [Google Scholar]
  13. Milton, G.; Kramer, S.; Deal, R.; Earle, E. Ultra trace analysis of acrylic for 232Th and 238U daughters. Appl. Radiat. Isot. 1994, 45, 539–547. [Google Scholar] [CrossRef]
  14. Dr. Love Disputes EPA’s Position on FlintWater Safety: New FACHEP Research Reports a 100 Rate and 50 % Mortality Rate from Flint POU Filters. Available online: http://flintwaterstudy.org (accessed on 23 September 2020).
  15. Neo 5.5 sCMOS Specifications. Available online: https://pdf.medicalexpo.com/pdf/andor-technology-plc/neo-55-scmos/90363-177117.html (accessed on 23 September 2020).
  16. Agostinelli, S.; Allison, J.; Amako, K. Geant4—A simulation toolkit. Nucl. Instrum. Methods Phys. Res. Sect. Accel. Spectrometers Detect. Assoc. Equip. 2003, 506, 250–303. [Google Scholar] [CrossRef] [Green Version]
  17. Aguilar-Arevalo, A. Search for low-mass WIMPs in a 0.6 kg day exposure of the DAMIC experiment at SNOLAB. Phys. Rev. D 2016, 94, 082006. [Google Scholar] [CrossRef] [Green Version]
  18. Instruction sheet 554 800. Available online: https://www.ld-didactic.de/documents/en-US/GA/GA/5/554/554800e.pdf (accessed on 23 September 2020).
  19. Flexible Image Transport System (FITS), Version 3.0. Available online: https://www.loc.gov/preservation/digital/formats/fdd/fdd000317.shtml?__cf_chl_captcha_tk__=bc6cd77850ccaf05824c7ec7f6fb556b523d4a70-1601193275-0-AUn8UK_eKqsAwaEdoZTPoduYbWoEpd4HhBD4ePnjUsi7tup_9Eey5KHrD3Fhokbb2fJrrYn14MP2vXVwo9du6MDrZavmkyLnPxok5aYW7lawzJGj20B-XDYmANP5EVgDJS5XiIT6pYr8MOk_hAkM4RVYk9aSflmo5zYC9IIL8NJtJqq_-wkVt4rHcj6r1ssc_Kv-NCnkDZ3ybxrCcRSZavOAuJV7Qw-XGSQ08EaOKUsZ4BEXLFGRt1fZ9Ve-h_nXhKmEsw9OWpdN-pdCM4BTjoCwUfkedeDVvlpDy1-RveDJ0YS7hxMh1cuG_Ib-ZAXeabWCAYp7Fg9puNbQI4JI6XoFbKCvPZT8Mp-0yZUogZkkNtWqTWlmRV_DeupB-nUYZFAtHj56lsjvkMvFzV4zPynWfRdXFaAFp0xpGy64qGL46zk6zZBkWWo2TYw-OKLPwO_P0Fo4vSE8qRc6nKTy1mAErrR1pSjVlNx5xV8zQYbB8XNxEAQp-h1zd9R_-Qh2UfmuTVfwraiodKDzc5vjiGaHIUI2xgG-P9RfqlLkWIG6 (accessed on 23 September 2020).
  20. Brun, R.; Rademakers, F. ROOT: An object oriented data analysis framework. Nucl. Instrum. Meth. A 1997, 389, 81–86. [Google Scholar] [CrossRef]
  21. Chong, C.S.; Ibrahim, S.E.; Ahmad, S.M.K.; Abdul, A.T. Gamma Ray Spectrum of Am 241 in a Backscattering Geometry Using a High Purity Germanium Detector. In Proceedings of the International Nuclear Conference: A New era In Nuclear Science and Technology—The Challenge of the 21st Century, Kuala Lumpur, Malaysia, 27–28 October 1997. [Google Scholar]
  22. Demir, D.; Eroğlu, M.; Turşucu, A. Studying of characteristics of the HPGe detector for radioactivity measurements. J. Instrum. 2013, 8, P10027. [Google Scholar] [CrossRef]
  23. LiveChart of Nuclides—Advanced version. Available online: https://www.iaea.org/resources/databases/livechart-of-nuclides-advanced-version (accessed on 23 September 2020).
  24. Stopping-Power & Range Tables for Electrons, Protons, and Helium Ions. Available online: https://www.nist.gov/pml/stopping-power-and-range-tables-electrons-protons-and-helium-ions-version-history (accessed on 23 September 2020).
  25. Tables of X-ray Mass Attenuation Coefficients and Mass Energy-Absorption Coefficients 1 keV to 20 MeV for Elements Z = 1 to 92 and 48 Additional Substances of Dosimetric Interest. Available online: https://www.nist.gov/pml/X-ray-mass-attenuation-coefficients (accessed on 23 September 2020).
  26. XCOM: Photon Cross Sections Database. Available online: https://www.nist.gov/pml/xcom-photon-cross-sections-database (accessed on 23 September 2020).
  27. Dark Matter Time Projection Chamber (DMTPC) Optical TPC for Low Energy Particle Tracking. Available online: http://dmtpc.mit.edu/twiki/pub/Projects/DmtpcPublications/GDruittMasterThesis.pdf (accessed on 23 September 2020).
  28. Kolanoski, H.; Wermes, N. Teilchendetektoren; Springer: Berlin/Heidelberg, Germany, 2016. [Google Scholar] [CrossRef]
  29. Gow, R.D.; Renshaw, D.; Findlater, K.; Grant, L.; McLeod, S.J.; Hart, J.; Nicol, R.L. A Comprehensive Tool for Modeling CMOS Image-Sensor-Noise Performance. IEEE Trans. Electron Devices 2007, 54, 1321–1329. [Google Scholar] [CrossRef]
  30. X-ray Mass Attenuation Coefficients. Available online: https://www.nist.gov/pml/X-ray-mass-attenuation-coefficients (accessed on 23 September 2020).
  31. WWW Table of Radioactive Isotopes. Available online: http://nucleardata.nuclear.lu.se/toi/ (accessed on 23 September 2020).
  32. Allison, J.; Amako, K.; Apostolakis, J.; Arce, P. Recent developments in Geant4. Nucl. Instrum. Methods Phys. Res. Sect. Accel. Spectrom. Detect. Assoc. Equip. 2016, 835, 186–225. [Google Scholar] [CrossRef]
  33. Allison, J.; Amako, K.; Apostolakis, J.; Araujo, H.; Arce Dubois, P.; Asai, M.; Barrand, G.; Capra, R.; Chauvie, S.; Chytracek, R.; et al. Geant4 developments and applications. IEEE Trans. Nucl. Sci. 2006, 53, 270–278. [Google Scholar] [CrossRef] [Green Version]
Figure 1. (a) Photograph of the source holder and the Neo sCMOS camera in the dark-box. (b) The Neo sCMOS in the LD Didactic X-ray apparatus (554 800)—The blue lines are for the water cooling circuit, which is not necessary for operation in the dark box.
Figure 1. (a) Photograph of the source holder and the Neo sCMOS camera in the dark-box. (b) The Neo sCMOS in the LD Didactic X-ray apparatus (554 800)—The blue lines are for the water cooling circuit, which is not necessary for operation in the dark box.
Sensors 20 05746 g001
Figure 2. Raw frames (without any correction) recorded when (a) no source of radiation is present, that is, background (BKG), and (b) when the camera is irradiated with an 241 Am source. A zoom in x and y of b is shown latter in this section. All images are zoomed on the intensity scale, visible in the colour-bar to the right of each image (Which means values larger than 200 are displayed as 200.).
Figure 2. Raw frames (without any correction) recorded when (a) no source of radiation is present, that is, background (BKG), and (b) when the camera is irradiated with an 241 Am source. A zoom in x and y of b is shown latter in this section. All images are zoomed on the intensity scale, visible in the colour-bar to the right of each image (Which means values larger than 200 are displayed as 200.).
Sensors 20 05746 g002
Figure 3. Histograms of the (a) raw pixel values C of all pixels in the frame in Figure 2a (BKG C), Figure 2b ( 241 Am C), and of the C col sub values calculated for the data in these frames using Equation (3). (b) Histogram of the C col sub values after time-series subtraction (T, form Equation (4)).
Figure 3. Histograms of the (a) raw pixel values C of all pixels in the frame in Figure 2a (BKG C), Figure 2b ( 241 Am C), and of the C col sub values calculated for the data in these frames using Equation (3). (b) Histogram of the C col sub values after time-series subtraction (T, form Equation (4)).
Sensors 20 05746 g003
Figure 4. Pair wise subtracted frames (cf. Section 4.1.2). The same frames as in Figure 2 are shown to illustrate this next step in the background rejection procedure. The left frame (a) shows background data obtained with no source present and the right frame (b) a frame obtained when the camera is irradiated with an 241 Am source. More high intensity points in the 241 Am frame than in the BKG frame are visible, when comparing the two plots.
Figure 4. Pair wise subtracted frames (cf. Section 4.1.2). The same frames as in Figure 2 are shown to illustrate this next step in the background rejection procedure. The left frame (a) shows background data obtained with no source present and the right frame (b) a frame obtained when the camera is irradiated with an 241 Am source. More high intensity points in the 241 Am frame than in the BKG frame are visible, when comparing the two plots.
Sensors 20 05746 g004
Figure 5. Detail view (a) of the frame in Figure 2b—raw frame without corrections applied—and (b) of the frame in Figure 4b, pair wise subtracted frame. (c) Identified clusters in the zoomed image shown in (b) using k seed = 10 and k skirt = 3 as parameters for the cluster finding thresholds, cf. Section 4.2, Equation (7). The scale to the right of the image shows the cluster number.
Figure 5. Detail view (a) of the frame in Figure 2b—raw frame without corrections applied—and (b) of the frame in Figure 4b, pair wise subtracted frame. (c) Identified clusters in the zoomed image shown in (b) using k seed = 10 and k skirt = 3 as parameters for the cluster finding thresholds, cf. Section 4.2, Equation (7). The scale to the right of the image shows the cluster number.
Sensors 20 05746 g005
Figure 6. The plots show cluster charge and maximal charge spectra for 241 Am data in the left column (a,c,e), and for background data in the right column (b,d,f). As energy unit kilo Analogue to DigitalUnit ( ADU ) , that is, k ADU , is used. The live-time of the camera during the 241 Am data taking is 2470 s where the camera is radiated with the corresponding source, while the live-time for the background data taking as 2945 s . The first row shows spectra containing all clusters found in the frames of the respective data taking runs, the second row shows cluster charge spectra grouped by cluster size in logarithmic scale while third row shows all data displayed in the first column, which passed a size > 2 cut. Note the larger binning in the second column.
Figure 6. The plots show cluster charge and maximal charge spectra for 241 Am data in the left column (a,c,e), and for background data in the right column (b,d,f). As energy unit kilo Analogue to DigitalUnit ( ADU ) , that is, k ADU , is used. The live-time of the camera during the 241 Am data taking is 2470 s where the camera is radiated with the corresponding source, while the live-time for the background data taking as 2945 s . The first row shows spectra containing all clusters found in the frames of the respective data taking runs, the second row shows cluster charge spectra grouped by cluster size in logarithmic scale while third row shows all data displayed in the first column, which passed a size > 2 cut. Note the larger binning in the second column.
Sensors 20 05746 g006
Figure 7. (a) Combined plot showing an overview of the different data sets acquired with the Neo sCMOS. The live-time of the background ( 241 Am , 55 Fe and 210 Pb , respectively) measurement is 29450 s (24700 s , 41800 s and 30400 s , respectively). The spectra are shown on a log scale for better visibility since the rates of the sources vary as does the observed event rate. (b,c,d) shows spectra for the respective sources. These spectra have been scaled to a live-time of 47500 s and are subtracted with the background spectrum scaled to the same live-time. Shaded regions represent the statistical error. For all plots a cluster size > 2 pixels is required. The spectra in (b,d) illustrate furthermore the peaks fitted with Gaußian functions, cf. Section 5.1.1. (Note that only the Gaußians are plotted, and not the additionally fitted backgrounds).
Figure 7. (a) Combined plot showing an overview of the different data sets acquired with the Neo sCMOS. The live-time of the background ( 241 Am , 55 Fe and 210 Pb , respectively) measurement is 29450 s (24700 s , 41800 s and 30400 s , respectively). The spectra are shown on a log scale for better visibility since the rates of the sources vary as does the observed event rate. (b,c,d) shows spectra for the respective sources. These spectra have been scaled to a live-time of 47500 s and are subtracted with the background spectrum scaled to the same live-time. Shaded regions represent the statistical error. For all plots a cluster size > 2 pixels is required. The spectra in (b,d) illustrate furthermore the peaks fitted with Gaußian functions, cf. Section 5.1.1. (Note that only the Gaußians are plotted, and not the additionally fitted backgrounds).
Sensors 20 05746 g007
Figure 8. (a) Spectrum recorded with the molybdenum X-ray tube (cf. Section X-ray Data Taking) as well as fitted curves to establish the position of the K α and K β peak. (b) The same data as in (a) is shown together with data recorded when a Zr or a Cu filter is placed between the X-ray tube with Mo target and the Neo sCMOS. For this plot all data has been normalised to a live-time of 50 m s and a cluster size > 2 pixels is required. Shaded regions indicate the statistical error.
Figure 8. (a) Spectrum recorded with the molybdenum X-ray tube (cf. Section X-ray Data Taking) as well as fitted curves to establish the position of the K α and K β peak. (b) The same data as in (a) is shown together with data recorded when a Zr or a Cu filter is placed between the X-ray tube with Mo target and the Neo sCMOS. For this plot all data has been normalised to a live-time of 50 m s and a cluster size > 2 pixels is required. Shaded regions indicate the statistical error.
Sensors 20 05746 g008
Figure 9. (a) Comparison between the expected and the measured peak and edge energies ( ε peak ) in Table 4. One σ peak of the peak is used as uncertainty for ε peak and the red line through the points is a fit without an additional axis intercept. (b) Measured energy resolution ( σ peak / ε peak ) as function of the measured peak position. The boxes indicate to what spectrum a given peak belongs to.
Figure 9. (a) Comparison between the expected and the measured peak and edge energies ( ε peak ) in Table 4. One σ peak of the peak is used as uncertainty for ε peak and the red line through the points is a fit without an additional axis intercept. (b) Measured energy resolution ( σ peak / ε peak ) as function of the measured peak position. The boxes indicate to what spectrum a given peak belongs to.
Sensors 20 05746 g009
Figure 10. Geometric acceptance of the experimental set-up with the Neo sCMOS: Shown is the analytical function for a point source in Equation (10) and values from a toy Monte Carlo resembling the actual source geometry.
Figure 10. Geometric acceptance of the experimental set-up with the Neo sCMOS: Shown is the analytical function for a point source in Equation (10) and values from a toy Monte Carlo resembling the actual source geometry.
Sensors 20 05746 g010
Figure 11. (a) Simulated incident activity on the camera chip for different camera-to-source distances for the full source activity, and for the fractional activities corresponding to the 26.3 k e V and 59.5 k e V   γ -lines. (b) The measured rates are calculated using Equation (11) and integrating over the full 241 Am spectra in Figures (c,d), or integrating only over the energy region of the 26.3 k e V and 59.5 k e V peaks in the same spectra. The measured background rate is established by integrating over the full spectrum recorded in absence of any source and in ± 5 σ peak windows around the mentioned peak energies. The shaded areas in the left half of the plot below the lines of the respective background rates show the five and 1.28 standard deviation ( σ ) regions around the background rate in yellow and green, respectively. (c,d) Measured 241 Am spectra at different distances with a live time of 3800 s . The data has to pass the cluster size > 2 pixels cut and is not background subtracted. The background data sample in both plots is the same as it has been shown previously (e.g., Figure 6f) and is normalised to the same live-time.
Figure 11. (a) Simulated incident activity on the camera chip for different camera-to-source distances for the full source activity, and for the fractional activities corresponding to the 26.3 k e V and 59.5 k e V   γ -lines. (b) The measured rates are calculated using Equation (11) and integrating over the full 241 Am spectra in Figures (c,d), or integrating only over the energy region of the 26.3 k e V and 59.5 k e V peaks in the same spectra. The measured background rate is established by integrating over the full spectrum recorded in absence of any source and in ± 5 σ peak windows around the mentioned peak energies. The shaded areas in the left half of the plot below the lines of the respective background rates show the five and 1.28 standard deviation ( σ ) regions around the background rate in yellow and green, respectively. (c,d) Measured 241 Am spectra at different distances with a live time of 3800 s . The data has to pass the cluster size > 2 pixels cut and is not background subtracted. The background data sample in both plots is the same as it has been shown previously (e.g., Figure 6f) and is normalised to the same live-time.
Sensors 20 05746 g011
Figure 12. (a) Intrinsic and (b) absolute efficiency for the 26.3 k e V and the 59.5 k e V peaks. The bands in (a) correspond to the result of fitting a constant to the data as discussed in Section 5.2.3 with its error bars. The two full sample points are extracted from data in Figure 7b in the same manner as all the other points are extracted from the data in Figure 11.
Figure 12. (a) Intrinsic and (b) absolute efficiency for the 26.3 k e V and the 59.5 k e V peaks. The bands in (a) correspond to the result of fitting a constant to the data as discussed in Section 5.2.3 with its error bars. The two full sample points are extracted from data in Figure 7b in the same manner as all the other points are extracted from the data in Figure 11.
Sensors 20 05746 g012
Figure 13. (a) 241 Am γ -ray and X-ray simulation for 1 M decays. The spectrum with no material effects taken into account (No att.) is scaled by a factor of 1 / 1000 to improve the readability of the plot. The other spectra show the photons which are absorbed by a silicon layer of given thickness. (b) Photon absorption efficiencies in silicon for different silicon layer thickness as well as the measured intrinsic efficiency of the Neo sCMOS at the two 241 Am γ -ray energies.
Figure 13. (a) 241 Am γ -ray and X-ray simulation for 1 M decays. The spectrum with no material effects taken into account (No att.) is scaled by a factor of 1 / 1000 to improve the readability of the plot. The other spectra show the photons which are absorbed by a silicon layer of given thickness. (b) Photon absorption efficiencies in silicon for different silicon layer thickness as well as the measured intrinsic efficiency of the Neo sCMOS at the two 241 Am γ -ray energies.
Sensors 20 05746 g013
Figure 14. Simulated spectra of 46.5 k e V γ -rays shot towards the detector with varying active silicon layer thicknesses along with constant micro lens thickness of 4 μ m with following glass window thicknesses: (a) 200 μ m , (b) 1000 μ m and (c) 2000 μ m . Photopeak of each individual simulated spectrum is normalised to the photopeak of experimental data ( 210 Pb ). The γ -rays’ origin from a point source centred on the detector over the window, cf. Section 5.3.1.
Figure 14. Simulated spectra of 46.5 k e V γ -rays shot towards the detector with varying active silicon layer thicknesses along with constant micro lens thickness of 4 μ m with following glass window thicknesses: (a) 200 μ m , (b) 1000 μ m and (c) 2000 μ m . Photopeak of each individual simulated spectrum is normalised to the photopeak of experimental data ( 210 Pb ). The γ -rays’ origin from a point source centred on the detector over the window, cf. Section 5.3.1.
Sensors 20 05746 g014
Table 1. Default settings of Neo sCMOS camera during all measurement runs. Runs with different settings are explicitly noted.
Table 1. Default settings of Neo sCMOS camera during all measurement runs. Runs with different settings are explicitly noted.
SettingValue
Readout binning 4 × 4
Exposure time10 s
Camera to source distance 1.75   c m
Number of exposures/data taking run100
Pixel Readout Rate (inverse row time)200 M Hz
Dynamic Range16-bit
ModeLow noise/high well capacity
Electronic Shuttering ModeRolling
Table 2. Expected lines in the decay spectrum of 241 Am based on data from References [21,23]. Np X-rays in the energy region from 11.87 k e V to 22.4 k e V are expected to make up for another 37% of intensity [23]. The X-ray energies are approximate and are composed of several overlapping lines. Therefore no intensities are given, since these require assumptions on the detectors energy resolution.
Table 2. Expected lines in the decay spectrum of 241 Am based on data from References [21,23]. Np X-rays in the energy region from 11.87 k e V to 22.4 k e V are expected to make up for another 37% of intensity [23]. The X-ray energies are approximate and are composed of several overlapping lines. Therefore no intensities are given, since these require assumptions on the detectors energy resolution.
EnergySourceIntensity
13.8   k e V Np: L α 1 , X-ray
17.8   k e V Np: L β 1 , X-ray
20.8   k e V Np: L γ 1 , X-ray
26.3   k e V 241 Am : γ 2.3%
59.5   k e V 241 Am : γ 35.9%
Table 3. Approximate ranges of γ -rays and β s (electrons) for typical decay energies in Silicon. For the γ -rays the attenuation length is calculated from the attenuation cross section given in Reference [26] using the density of silicon-dioxide. The same density is used to calculate the electron range from the CSDA (“continuous stopping down approximation”) range for electrons given in Reference [24].
Table 3. Approximate ranges of γ -rays and β s (electrons) for typical decay energies in Silicon. For the γ -rays the attenuation length is calculated from the attenuation cross section given in Reference [26] using the density of silicon-dioxide. The same density is used to calculate the electron range from the CSDA (“continuous stopping down approximation”) range for electrons given in Reference [24].
Energy [keV]Attenuation Length/
Range in Silicon
γ -/X-ray β
10111 μ m 1.2   μ m
15365 μ m 2.5   μ m
457 m m 16 μ m
6012 m m 28 μ m
10020 m m 66 μ m
100059 m m 2 m m
Table 4. The table lists the fit results (peak position ε peak and standard deviation σ ) of Gaußian peak fits in Figure 7b,d and Figure 8a as well as the absorption edge read off from Figure 8b. (Since the properties of the absorption edge are not determined by a fit, the corresponding values in the “Zr edge” column are labelled n.a. for non applicable.). The expected energies have been extracted from Reference [23] and the Zr absorption edge from Reference [26].
Table 4. The table lists the fit results (peak position ε peak and standard deviation σ ) of Gaußian peak fits in Figure 7b,d and Figure 8a as well as the absorption edge read off from Figure 8b. (Since the properties of the absorption edge are not determined by a fit, the corresponding values in the “Zr edge” column are labelled n.a. for non applicable.). The expected energies have been extracted from Reference [23] and the Zr absorption edge from Reference [26].
Source Radiation 241 Am 210 Pb Mo X-ray Tube
γ X-ray (Np) γ X-rayZr Edge
Expected energy [ k e V ] 26.359.513.817.820.846.517.419.618.0
Fit range ε min [ k ADU ] 9.8723.494.996.017.7818.371n.a.
ε max [ k ADU ] 11.4524.56.127.99.9919.6415n.a.
Fit result ε peak [ k ADU ] 10.6524.135.547.18.3718.947.087.857.3
σ peak [ ADU ] 13327695124199239163353n.a.
χ 2 / N dof 1.221.051.612.21.320.774.62n.a.
Table 5. The lowest incident activities and the corresponding measured rates obtained with the Neo sCMOS. See Section 5.2.2 and Figure 11b. Values for two cases—either a 5 σ or a 90% Confidence Level (CL) condition—are quoted.
Table 5. The lowest incident activities and the corresponding measured rates obtained with the Neo sCMOS. See Section 5.2.2 and Figure 11b. Values for two cases—either a 5 σ or a 90% Confidence Level (CL) condition—are quoted.
Incident ActivityMeasured Rate
5 · σ 90% CL 5 · σ 90% CL
full spectrum 7 ± 4   Bq 2 ± 1   Bq 40 ± 3   m Hz 27 ± 3   m Hz
26.3 k e V -peak 1.0 ± 6   Bq 1.0 ± 6   Bq 4 ± 1   m Hz 4 ± 1   m Hz
59.5 k e V -peak 57 ± 33   Bq 57 ± 33   Bq 1.5 ± 1   m Hz 1.5 ± 1   m Hz
Table 6. Corresponding full absorption efficiencies and η -values of 46.5 keV γ -rays for silicon layers of different thicknesses.
Table 6. Corresponding full absorption efficiencies and η -values of 46.5 keV γ -rays for silicon layers of different thicknesses.
Silicon Thickness Glass Window Full Absorption η 10 2
μ m Thickness μ m Efficiency 10 3 %
22000.13 ± 0.040.41 ± 0.11
10000.09 ± 0.030.29 ± 0.10
20000.10 ± 0.030.35 ± 0.01
32000.49 ± 0.071.1 ± 0.2
10000.40 ± 0.061.0 ± 0.1
20000.41 ± 0.061.1 ± 0.2
42001.7 ± 0.13.3 ± 0.3
10001.4 ± 0.12.8 ± 0.2
20001.4 ± 0.13.1 ± 0.3
52003.4 ± 0.25.7 ± 0.3
10003.3 ± 0.25.8 ± 0.3
20002.6 ± 0.25.0 ± 0.3

Share and Cite

MDPI and ACS Style

Aguilar-Arevalo, A.; Bertou, X.; Canet, C.; Cruz-Pérez, M.A.; Deisting, A.; Dias, A.; D’Olivo, J.C.; Favela-Pérez, F.; Garcés, E.A.; González Muñoz, A.; et al. Dosimetry and Calorimetry Performance of a Scientific CMOS Camera for Environmental Monitoring. Sensors 2020, 20, 5746. https://doi.org/10.3390/s20205746

AMA Style

Aguilar-Arevalo A, Bertou X, Canet C, Cruz-Pérez MA, Deisting A, Dias A, D’Olivo JC, Favela-Pérez F, Garcés EA, González Muñoz A, et al. Dosimetry and Calorimetry Performance of a Scientific CMOS Camera for Environmental Monitoring. Sensors. 2020; 20(20):5746. https://doi.org/10.3390/s20205746

Chicago/Turabian Style

Aguilar-Arevalo, Alexis, Xavier Bertou, Carles Canet, Miguel Angel Cruz-Pérez, Alexander Deisting, Adriana Dias, Juan Carlos D’Olivo, Francisco Favela-Pérez, Estela A. Garcés, Adiv González Muñoz, and et al. 2020. "Dosimetry and Calorimetry Performance of a Scientific CMOS Camera for Environmental Monitoring" Sensors 20, no. 20: 5746. https://doi.org/10.3390/s20205746

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop