Retentive Strength of CAD/CAM-Fabricated All-Ceramic Crowns Luted on Titanium Implant Abutments Using Different Ceramic Materials and Luting Agents: An In Vitro Study

by
Monika Bjelopavlovic
*,
Michael Weyhrauch
,
Herbert Scheller
,
Stefan Wentaschek
and
Karl Martin Lehmann
Department of Prosthetic Dentistry, University Medical Center Mainz, Augustusplatz 2, 55131 Mainz, Germany
*
Author to whom correspondence should be addressed.
Materials 2022, 15(19), 6968; https://doi.org/10.3390/ma15196968
Submission received: 27 July 2022 / Revised: 4 October 2022 / Accepted: 4 October 2022 / Published: 7 October 2022

Abstract

:
Objectives: This study aimed to determine the retentive strength of monolithic hybrid-all-ceramic crowns luted on titanium implant abutments. Material and Methods: In total, 450 crowns (75 each of Mark II, Empress CAD, e.max CAD, Suprinity, Enamic, Celtra Duo) were milled using a CAD/CAM system. The crowns were cemented onto sandblasted titanium implant abutments using five luting agents (Multilink Implant, Variolink II, RelyX Unicem, Fujicem, and Panavia 2.0). After thermocycling was performed (5000 cycles: 5–55 °C, 30-s dwell time), the crowns were removed using a universal testing machine. The location of luting-agent residue on the abutment and inner crown surfaces was evaluated. Analyses of variance (ANOVA) with the Bonferroni correction were performed to evaluate differences of retentive strength depending on the crown material and the kind of the luting agent. Results: The retentive strengths for the different ceramic materials were Vita Mark II: 652N-759N (SD:134N-146N), Empress CAD: 681N-822N (SD: 89N-146N), e.max CAD: 784N-1044N (SD: 109N-176N), Vita Enamic: 716N-1177N (SD: 132N-220N), Vita Suprinity: 867N-1488N (SD: 202N-278N), and Celtra Duo 772N-1335N (SD:151N-229N). After the removal trials, the visual documentation showed different adhesive residue location depending on the ceramic materials. Furthermore, the pull-off force was dependent on the choice of adhesives. No significant differences were found between different luting agents and the ceramic material Vita Mark II and Empress CAD. EmaxCAD showed significant differences with Unicem and FujiCem compared to Panavia, as did VitaSuprinity, VitaEnamic, and Celtra Duo (p < 0.001). Conclusions: The ceramic material used seems to influence the retentive strength and the use of certain luting agents results in a higher retentive strength for some ceramic materials.

1. Introduction

The request for ceramic fixed partial dentures is increasing because all-ceramic materials constitute an appropriate option, optimal esthetics, offering excellent biocompatibility and low thermal conductivity. In reference to this, digital workflows have the potential to reach an outcome with the production of ceramic restorations comparable to conventional procedures from the patient’s point of view [1,2,3]. Over the years, several ceramic systems have been developed and introduced, using various materials and techniques for CAD/CAM fabrication. In this context CAD/CAM systems were optimized regarding an adequate internal and marginal fitting [4,5], esthetics, stability, and functional aspects as relevant success factors. Regarding this a distinction can be made between screw-retained and cemented restorations and also with these systems the production of monolithic restorations are possible, which offer reduced chipping problems compared with restorations based on a core material layered with a veneering ceramic. For this several ceramic classes are used, i.e., feldspathic, glass and oxide ceramic materials and further hybrid-ceramics and different luting systems, namely, adhesive composite materials, non- and self-etching, self and light curing and a glass ionomer cement. These combinations of ceramic restorations and luting agents are not only used on teeth, but also utilized for implant-based restorations. The implant-based restorations could be screw-retained or consist of ceramic suprastructures cemented on abutments [6,7]. Each retention mode offers advantages, so screw retained implant based restorations avoid cement residues and thus biological implant pathologies [8] and guarantees a fixation when the vertical available space for the suprastructure is limited and thus does not offer a sufficiently large cementation surface. However, a fixation concept of a vertical screw retention cannot always be implemented, e.g., with a strong divergence between implant length axis and the axis of the prosthetic restoration regarding esthetic aspects, especially in visually exposed regions as the front teeth. In the case of cemented restorations an important aspect for long term success is a sufficient fixation of the ceramic suprastructures on the abutment platforms. Furthermore, a study showed that the stability of the crown-abutment connection in implants can be dependent on the “ferrule effect”, analogous to natural teeth, and consequently increases in the presence of ferrulization [9]. The fixation data published is as follows: analyzing the effect of contamination of the bond surface [10,11,12], the cement type and abutment height [13], abutment platform configuration [14], airborne-particle abrasion by using alumina [15] or silanisation and laser pretreatment [16] on retention force of the restoration. In addition, it is possibly to be expected that the adhesive interface behavior between the implant surface, the luting agent, and the bond surface of the ceramic restoration could be meaningful for the stabilization of the ceramic restoration. Currently, the adhesion strength between the upper part of the dental implant and the crown is not clear, and this study analyzes and explores previous adhesion.
In the present study the focus was primarily placed on the combination effects using different ceramic materials and luting agents on the retentive strength. With the help of the standardized pull-off test to reach a comparability with previous studies, the quality of the bonding was tested. This target size was accepted as a measure of long-term success using six ceramic materials in combination with five different luting systems regularly used clinically for the fixation of dental restorations.

2. Materials and Methods

For this study, two working hypotheses were formulated:
In terms of retentive strength,
  • all ceramic materials used would behave independently of the type of luting agent applied; and
  • all luting agents would result in equal values independent of the type of ceramic material.
In total, 450 implant abutments (No. 57120, Bego Implant Systems, Bremen, Germany) were screwed onto implant laboratory analogs (No. 56696, Bego Implant Systems, S-RI line, 4.1 mm diameter; 0.5 mm machined shoulder)) using a torque driver (No. 55799, Bego Implant Systems) with 25 Ncm. Seventy-five crowns of a mandibular right first premolar, made of ceramic materials (75 each of Vita Mark II; feldspar ceramic [FSC], Ivoclar Empress CAD; leucite reinforced glass ceramic [LrGC], Ivoclar e.max CAD; lithium disiclicate [LiDS], Vita Enamic; polymer reinforced fine-structure feldspar ceramic [PolyFSP], Vita Suprinity; pre-sintered zirconia reinforced lithium silicate ceramic [PSZirLS], Celtra Duo; fully crystallized zirconia reinforced lithium silicate [FcZirLS]); Table 1), were manufactured using a CAD/CAM system (CEREC inLab version 4.1; Sirona, Bensheim, Germany; CEREC parameters: block size 14 mm, cement space 30 μm, and spacer 0 μm).
The exterior surfaces of the implant abutments were sandblasted (corundum, 50 µm, ~10-mm distance, 1 bar, ~60-s blasting time per abutment), degreased, and bonded with silane (Monobond P; Ivoclar Vivadent Clinical, Schaan, Lichtenstein). The inner surfaces of the crowns were etched with hydrofluoric acid (5%; Ivoclar Vivadent Clinical) and silanized with Monobond P. The abutment and inner crown surfaces were also pretreated with Monobond P. This resulted in 15 different test series. All six ceramic materials and five luting agents used (Table 1) were applied in strict accordance with the manufacturers’ instructions (Table 2). All cement excesses were immediately removed with a sponge. The proprietary glycerine gel of each luting system was subsequently applied at the crown margin. If no such gel was available from the respective manufacturer, an alternative gel (Airblock) was applied. In the case of dual-curing systems crowns were polymerized using a polymerization light (Bluephase; Ivoclar Vivadent Clinical) for 240 s.
To simulate the oral environment, the implant-abutment-crown complexes were placed in a moist container for 30 min at 37 °C in an insulated chamber. After 30 min, specimens were placed in water for 1 week at 37 °C. Thereafter, thermocycling was performed for 5000 cycles (5–55 °C, 30-s dwell time; Willytec; SD Mechatronik GmbH, Feldkirchen-Westerham, Germany). All experiments were performed under a constant room temperature at 21 degrees.
To enable the crowns to be pulled off with the testing machine, it was necessary to create a well-fitting construction that would serve as an adapter between the all-ceramic crowns and the universal testing machine (Model 1425; Zwick Roell GmbH & Co. KG, Ulm, Germany). This resulted in an adapter that had an accurate and homogenous contact with the crowns.
All pull-off trials were conducted with the universal testing machine. The specimens were taken from the water bath prior to the pull-off test and mounted in the machine. The testing machine was adjusted so that the two tensile heads holding the abutment respective to the crown moved strictly vertically at a speed of 1 mm/min. After the crowns had been pulled off, the location of luting-agent residue was evaluated by visual observation.
Retentive strength was recorded, using SPSS 25.0 mean, standard deviations, minimum and maximum were calculated (Table 3). A Bonferroni adjustment was performed, and an analysis of variance was conducted to evaluate statistical differences with alpha adjustment. With level of significance α = 0.00011494.

3. Results

Most of the ceramics used, tended to show higher pull-off forces when using Unicem and FujiCem. No significant differences between the luting agents for the ceramic materials FSC and LrGC were observed (Figure 1). For the material LiDS a significant higher retentive strength for Unicem compared to Panavia was observed (p < 0.0001). For ceramic PolyFSP significant higher retentive strength for the luting agents Multilink implant vs. Panavia 2.0, Fujicem and Unicem vs. Variolink and Unicem and Fujicem vs. Panavia were detected (p < 0.0001). The ceramic PsZirLs showed significantly higher retentive differences between the luting agents Unicem and Fujicem vs. Multilink implant, Fujicem vs. Variolink, and Fujicem vs Panavia (p < 0.0001). For ceramic PcZirLS significant higher retentive strengths were detected between the luting agents Fujicem vs. Multilink and Variolink, Unicem vs. Panavia, and Fujicem vs. Panavia (p < 0.0001).
With the focus on the luting agents the agent Multilink Implant showed a higher retentive strength in the combination with PolyFSP and FcZirLS vs. FSC, PolyFSP and FcZirLS vs. LrGC (p < 0.001); the agent Variolink II showed a higher retentive strength in combination with PolyFSP, PsZirlLs and PcZirLS, compared to FSC, for PsZirLS a higher retentive strength than for LrGC and for LiDS (p < 0.001); when the agent Unicem was used the combination with LiDS, PolyFSP, PsZirLS, and FcZirLS exhibit a higher retentive strength compared to FSC, further the combination of LiDS, PolyFSP, PsZirLS and FcZirLS a higher retentive strength compared to LrGC (p < 0.001), for the combination of PsZirLS a higher retentive strength than for LiDS (p < 0.001); for the luting agent Fujicem the combination with PolyFSP, PsZirLS and FcZirLS a higher retentive strength compared to FSC and LrGC, for PsZirLS and FcZirLS a higher retentive strength compared to LiDS and for PsZirLS compared to PolyFSP; for the luting agent Panavia 2.0 the combination with PSZirLS showed a higher retentive strength compared to FSC, PolyFSP and FcZirLS (p < 0.001) (Appendix A). The location of the luting agent’s residue is documented in Table 4.

4. Discussion

To satisfy patient demand for metal-free fixed partial dentures that provide long-term stability, it is necessary to keep several issues in mind. It is well known that in addition to marginal adaption, high fracture strengths, and sufficient aesthetics, one of the most important aspects for long-term survival of luted fixed partial dentures (i.e., crowns) is the fixation of the restoration achieved as a quality factor for long-term success [17]. Thus, in several studies the bond strength was used as a prognostic factor to evaluate clinical suitability, i.e., for veneer restorations on natural teeth [18]. However, the question arises as to how it behaves with ceramic restorations on titanium abutments. Therefore the experimental designs of those studies, just like in this study, are [19,20] to simulate the clinical situation, i.e., of crowns made of monolithic all-ceramic materials luted on implant abutments, and thus allowing a realistic assessment of the retentive strength as a factor for achievable long-term success when frequently used ceramic materials and luting agents are utilized. With regard to this, the retentive strength depends on several factors: among other things the texture of the retentive surfaces respective to their pretreatment [21,22,23,24], the luting agent or their curing mechanism [25,26], or the geometry of the retentive surfaces [27,28,29,30,31]. So special attention was paid to allow a sufficient dwell time of cemented crowns in 100% humidity before they were placed in water. For some luting agents, placement in water immediately after removing cement excesses can lead to rapid hydrolytic degradation of the bond (for example if the luting agent has not completely cured) and a reduction in the bonding potential of the luting agent [20,32]. This does not absolutely correspond with the clinical situation, in which the cement is in contact with saliva rather than water. However, in this study long-term water storage was not conducted because Ernst et al. [20] showed that it had no further significant effect on retentive strength, whereas thermocycling was carried out to simulate the thermal stress as a relevant factor for reducing bond strength [23,33,34]. In this study it was of value that representative and commonly used ceramic materials and luting agents were used. Different classes of ceramic materials, feldspathic, lithium disilicate, a hybrid ceramic, and ceramic materials with oxide ceramic components were utilized and regarding the luting agents also the different types which are frequently encountered in everyday clinical practice, such as adhesive composite materials, non- and self-etching, self and light curing, and a reinforced glass ionomer cement. Although manufacturers of ceramic materials and professional societies make specific recommendations for the combination of the ceramic materials and luting agents the question arises as to what extent other combinations influence the retentive strength. So, the first working hypothesis, that each ceramic material used would behave independently of the type of luting agent regarding the retentive strength, has only been accepted for the Mark II and Empress CAD ceramics. For the other ceramic materials using this hypothesis had to be rejected. The second hypothesis, that all luting agents would result in equal retentive values independent of the type of ceramic material had to be rejected completely.
To explain the results, it is necessary to analyze the fixation, on the one hand the interface between the ceramic surface and the luting agent and on the other hand between the luting agent and the adhesive surface of the implant abutment. In this regard the location of the residue of the luting agent allows the quality of the bonding to be judged and helps to understand the behavior of releasing the bond during the withdrawal process. Historically dental ceramics were first bonded on natural teeth and based on this there have already been numerous publications and findings on the bonding interface between ceramic material, luting agent, and the surface of natural teeth [34]. Compared to the experimental setup of these studies further studies were conducted where there was a titanium surface instead of human dentin.
The question is, whether there is a context between the location of the adhesive residue and the used luting agents and ceramic materials.
In this study regarding the feldspar ceramic and glass ceramic no significant differences between the luting agents were detected, the residues of the luting agents were primarily in the adhesive surfaces of the ceramic restoration and thus the prevailing adhesive failure was between the titanium surface and the luting agents and that showed a basically good adhesive connection between all luting materials to the feldspathic (Mark II) and glass ceramic (Empress) materials tested. For the lithium disilicate ceramic (LiDS) the luting agent Rely X Unicem showed a significant higher retentive strength compared to Panavia (p < 0.001) whereas the residues were also overwhelmingly in the inner crown surface. For the hybrid dental ceramic PolyFSP more significant differences between the luting agents were detected. So, for the luting agents Multilink implant, Rely X Unicem, and Fujicem a higher retentive strength was detected compared to Panavia 2.0, whereas it is noticeable that in addition to more significant differences between the luting agents for this hybrid ceramic materials more and more residue of the luting agents was located on the titanium abutment surface and less on the inner crown surface. For the pre-sintered zirconia-reinforced lithium silicate ceramic (PsZirLS) and the fully crystalized zirconia-reinforced lithium silicate ceramic (FcZirLS) higher retention values and higher standard deviations for nearly all combinations with the different luting agents were detected and the residue of the luting agents was also increasingly on the abutment surfaces.
With respect to the luting agents the self-polymerizing luting agent Multilink showed significantly higher retentive strength for the ceramic materials Enamic and Celtra Duo compared to the ceramics Mark II and Empress. For the luting agent Variolink as a dual-/light-polymerizing, resin-based dental luting material Enamic, Suprinity, and Celtra offered higher retentive strength compared to Mark II and Suprinity further to Empress. Regarding the dual-polymerizing, self-adhesive resin cement Unicem, the ceramic Empress, Enamic, Suprinity, and Celtra offered a significantly higher retentive strength compared to Mark II, while Enamic, Suprinity, and Celtra are also higher in retentive strength compared to the material Empress. Fujicem as a self-polymerizing, resin-reinforced glass-ionomer luting cement offer sa similar situation with a higher strength for Enamic, Suprinity, and Celtra in comparison to Mark II, and Empress and Suprinity, and Celtra higher than e.max and the material Suprinity higher than Enamic. In contrast the dual-polymerizing, self-adhesive resin cement Panavia showed only for Suprinity a higher retentive strength compared to Mark II.
The interpretation of these results is certainly difficult because despite all standardization of the experimental setup various influencing factors must be considered, essentially the different ceramic materials and luting agents. Because of the standardized setup the reason for this must be due to the factors ceramic material, luting agent, and the combination of them, the interface of the titanium surface, luting agent, and adhesive surface of the ceramic material. It is conspicuous that almost no significant difference between the luting agents in retentive strength for the feldspathic (FSC), the glass (LrGC), and the lithium disilicate (LiDS) material were noted. In summary the ceramic materials Enamic, Suprinity, and Celtra showed, in combination with the different luting agents tested, often relatively higher values for retentive strength compared to the feldspathic Mark II and the glass ceramic Empress. Increasingly more differences between the luting agents were detected for the hybrid (PolyFSP), the pre-sintered (PsZirLS) and fully crystalized (FcZirLS) zirconia-reinforced ceramic. Regarding these ceramic materials tested further studies stated also differences in fracture strength [35] with a higher fracture strength of hybrid (PolyFSP), the pre-sintered (PsZirLS), and fully crystalized (FcZirLS) zirconia-reinforced ceramic compared to the feldspathic (FSC) and the glass (LrGC) materials. It might indicate that the retentive strength could be related to the stability of the ceramic material. A hypothesis is that loosening of the crowns is caused by a failure of the adhesive connection initiated by microcracks that arise first in ceramic materials with lower stability. This could be an explanation for higher retention forces when ceramics with a higher stability are used. However, the question is further to explain the observed significant differences of retentive strength between crowns within each ceramic material made of the high stable materials (PolyFSP), (PsZirLS), and (FcZirLS). Those differences were probably caused by the different luting agents which for the crowns are made of materials mainly responsible for the formation of different adhesive connection patterns between the titanium surface, the layer of the luting agent, and the ceramic surface. However, it must be clarified as to what extent the results depend on individual components of the several luting agents used.
Limitations of this study are as follows:
(1)
The choice of the conical implant with a machined implant shoulder could influence the results and was not compared to another implant system.
(2)
The conditioning (HF) and bonding (Monobond S) are not part of the manufacturer recommended procedure for the cementation. Inclusion of these two steps may affect the comparison of results to other studies. Comparable groups should well be achieved within the study, however with the comparability to other studies being impaired.
In summary, the benefit of this study is that clinically the frequently used numerous ceramic materials and luting agent classes were tested in a standardized laboratory setup with corundum abrasion of the titanium abutments and completely identical testing procedures, with an adjacent consistent pull-off setup without the presence of influencing factors like saliva or blood, that could affect the quality of the fixation of the ceramic crowns. Further it is to be expected that the monolithic ceramic crowns made of the ceramic materials und luting agents used, will exceed physical masticatory forces, however certain combinations of ceramic materials and luting agents offer higher retentive strengths.
Thus, the first hypothesis that all ceramic materials behave independently of the type of luting agent applied was rejected for the materials LiDS, PolyFSP, PsZirLS, and FcZirLS and the second working hypothesis, that the luting agent used would result in equal retentive strength independent of the type of ceramic material was completely rejected.

5. Conclusions

Within the limitations of this study the following conclusions can be drawn:
(1)
The ceramic material used influences the retentive strength of all ceramic crowns luted with various luting agents and
(2)
the combination of certain ceramic materials with certain luting agents results in higher retentive strength.
Further investigation regarding adhesive residues should be the subject of subsequent studies on the effect and the adhesive mechanism of individual combinations of ceramic materials and luting agents.

Author Contributions

Conceptualization, K.M.L., M.W. and H.S.; methodology, K.M.L.; software, S.W., validation, M.B., H.S. and K.M.L.; formal analysis, K.M.L.; investigation, M.W.; resources, H.S.; data curation, M.B., K.M.L. and S.W.; writing—original draft preparation, M.B.; writing—review and editing, M.B. and K.M.L., visualization, M.B.; supervision, K.M.L.; project administration, K.M.L.; funding acquisition, M.W. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Informed consent was obtained from all subjects involved in the study.

Data Availability Statement

Data can be seen in Table 1, Table 2, Table 3 and Table 4.

Conflicts of Interest

The authors declare no conflict of interest.

Appendix A

VITA MARK II [FSC]
Table A1. Comparison of the different adhesives during the pull-off test: Vita Keramik II.
Table A1. Comparison of the different adhesives during the pull-off test: Vita Keramik II.
Multilink Implant Variolink II Rely X Unicem Fuji Cem Panavia 2.0
Multilink Implant 1.0001.0001.0001.000
Variolink II 1.000 1.0000.437 1.000
Rely X Unicem 1.0001.000 0.437 1.000
Fuji Cem 1.0000.437 0.437 1.000
Panavia 2.0 1.0001.0001.0001.000
Ivoclar Empress CAD[LrGC]
Table A2. Comparison of the different adhesives during the pull-off test: Ivoclar Empress CAD.
Table A2. Comparison of the different adhesives during the pull-off test: Ivoclar Empress CAD.
Multilink Implant Variolink II Rely X Unicem Fuji Cem Panavia 2.0
Multilink Implant 0.590 1.0001.0000.047
Variolink II 0.590 1.0001.0001.000
Rely X Unicem 1.0001.000 1.0001.000
Fuji Cem 1.0001.0001.000 1.000
Panavia 2.0 0.047 1.0001.0001.000
Ivoclar e.max CAD [LiDS]
Table A3. Comparison of the different adhesives during the pull-off test: Ivoclar e.max CAD.
Table A3. Comparison of the different adhesives during the pull-off test: Ivoclar e.max CAD.
Multilink Implant Variolink II Rely X Unicem Fuji Cem Panavia 2.0
Multilink Implant 1.0000.003 0.196 0.047
Variolink II 1.000 0.010 0.424 1.000
Rely X Unicem 0.003 0.010 1.0000.000 *
Fuji Cem 0.196 0.424 1.000 0.009
Panavia 2.0 1.0001.0000.000 *0.009
* significant.
Vita Enamic [PolyFSP]
Table A4. Comparison of the different adhesives during the pull-off test: Vita Enamic.
Table A4. Comparison of the different adhesives during the pull-off test: Vita Enamic.
Multilink Implant Variolink II Rely X Unicem Fuji Cem Panavia 2.0
Multilink Implant 1.0000.141 0.098 0.000 *
Variolink II 1.000 0.001 0.001 0.013
Rely X Unicem 0.141 0.001 1.0000.000 *
Fuji Cem 0.098 0.001 1.000 0.000 *
Panavia 2.0 0.000 *0.013 0.000 *0.000 *
* significant.
Vita Suprinity [PsZirLS]
Table A5. Comparison of the different adhesives during the pull-off test: Vita Suprinity.
Table A5. Comparison of the different adhesives during the pull-off test: Vita Suprinity.
Multilink Implant Variolink II Rely X Unicem Fuji Cem Panavia 2.0
Multilink Implant 0.097 0.000 *0.000 *0.956
Variolink II 0.097 0.174 0.001 1.000
Rely X Unicem 0.000 *0.174 0.906 0.011
Fuji Cem 0.000 *0.001 0.906 0.000 *
Panavia 2.0 0.956 1.0000.011 0.000 *
* significant.
Celtra Duo [FcZirLS]
Table A6. Comparison of the different adhesives during the pull-off test: Celtra Duo.
Table A6. Comparison of the different adhesives during the pull-off test: Celtra Duo.
Multilink Implant Variolink II Rely X Unicem Fuji Cem Panavia 2.0
Multilink Implant 1.0000.801 0.000 *0.016
Variolink II 1.000 0.263 0.000 *0.069
Rely X Unicem 0.8010.263 0.053 0.000 *
Fuji Cem 0.000 *0.000 *0.053 0.000 *
Panavia 2.0 0.016 0.069 0.000 *0.000 *
* significant.
Multilink Implant
Table A7. Comparison of the different ceramics during the pull-off test: Multilink Implant.
Table A7. Comparison of the different ceramics during the pull-off test: Multilink Implant.
Vita Mark IIIvoclar Empress CADIvoclar e.max CADVita EnamicVita SuprinityCeltra Duo
Vita Mark II 1.0000.2910.000 *0.0940.000 *
Ivoclar Empress CAD1.000 0.2540.000 *0.0810.000 *
Ivoclar e.max CAD0.2910.254 0.1281.0000.227
Vita Enamic0.000 *0.000 *0.128 0.3841.000
Vita Suprinity0.0940.0811.0000.384 0.642
Celtra Duo0.000 *0.000 *0.2271.0000.642
* significant.
Variolink II
Table A8. Comparison of the different ceramics during the pull-off test: Variolink II.
Table A8. Comparison of the different ceramics during the pull-off test: Variolink II.
Vita Mark IIIvoclar Empress CADIvoclar e.max CADVita EnamicVita SuprinityCeltra Duo
Vita Mark II 0.8320.0130.000 *0.000 *0.000 *
Ivoclar Empress CAD0.832 1.0000.3020.000 *0.024
Ivoclar e.max CAD0.0131.000 1.0000.0011.000
Vita Enamic0.000 *0.3021.000 0.0321.000
Vita Suprinity0.000 *0.000 *0.0010.032 0.382
Celtra Duo0.000 *0.0241.0001.0000.382
* significant.
Rely X Unicem
Table A9. Comparison of the different ceramics during the pull-off test: RelyX Unicem.
Table A9. Comparison of the different ceramics during the pull-off test: RelyX Unicem.
Vita Mark IIIvoclar Empress CADIvoclar e.max CADVita EnamicVita SuprinityCeltra Duo
Vita Mark II 1.0000.000 *0.000 *0.000 *0.000 *
Ivoclar Empress CAD1.000 0.0010.000 *0.000 *0.000 *
Ivoclar e.max CAD0.000 *0.001 1.0000.0011.000
Vita Enamic0.000 *0.000 *1.000 0.3521.000
Vita Suprinity0.000 *0.000 *0.0010.352 0.069
Celtra Duo0.000 *0.000 *1.0001.0000.069
* significant.
Fuji Cem
Table A10. Comparison of the different ceramics during the pull-off test: Fuji Cem.
Table A10. Comparison of the different ceramics during the pull-off test: Fuji Cem.
Vita Mark IIIvoclar Empress CADIvoclar e.max CADVita EnamicVita SuprinityCeltra Duo
Vita Mark II 1.0000.0170.000 *0.000 *0.000 *
Ivoclar Empress CAD1.000 0.0100.000 *0.000 *0.000 *
Ivoclar e.max CAD0.0170.010 0.0220.000 *0.000 *
Vita Enamic0.000 *0.000 *0.022 0.000 *0.247
Vita Suprinity0.000 *0.000 *0.000 *0.000 * 0.291
Celtra Duo0.000 *0.000 *0.000 *0.2470.291
* significant.
Panavia 2.0
Table A11. Comparison of the different ceramics during the pull-off test: Panavia 2.0.
Table A11. Comparison of the different ceramics during the pull-off test: Panavia 2.0.
Vita Mark IIIvoclar Empress CADIvoclar e.max CADVita EnamicVita SuprinityCeltra Duo
Vita Mark II 1.0001.0001.0000.000 *1.000
Ivoclar Empress CAD1.000 1.0001.0000.0191.000
Ivoclar e.max CAD1.0001.000 1.0000.0021.000
Vita Enamic1.0001.0001.000 0.000 *1.000
Vita Suprinity0.000 *0.0190.0020.000 * 0.001
Celtra Duo1.0001.0001.0001.0000.001
* significant.
Figure A1. Schematic representation of the methodology.
Figure A1. Schematic representation of the methodology.
Materials 15 06968 g0a1

References

  1. Kunavisarut, C.; Jarangkul, W.; Pornprasertsuk-Damrongsri, S.; Joda, T. Patient-reported outcome measures (PROMs) comparing digital and conventional workflows for treatment with posterior single-unit implant restorations: A randomized controlled trial. J. Dent. 2022, 117, 103875. [Google Scholar] [CrossRef]
  2. Joda, T.; Ferrari, M.; Bragger, U.; Zitzmann, N.U. Patient Reported Outcome Measures (PROMs) of posterior single-implant crowns using digital workflows: A randomized controlled trial with a three-year follow-up. Clin. Oral Implant. Res. 2018, 29, 954–961. [Google Scholar] [CrossRef] [PubMed]
  3. Michelinakis, G.; Apostolakis, D.; Kamposiora, P.; Papavasiliou, G.; Ozcan, M. The direct digital workflow in fixed implant prosthodontics: A narrative review. BMC Oral Health 2021, 21, 37. [Google Scholar] [CrossRef] [PubMed]
  4. Vichitgomen, J.; Srisawasdi, S. Deep margin elevation with resin composite and resin-modified glass-ionomer on marginal sealing of CAD-CAM ceramic inlays: An in vitro study. Am. J. Dent. 2021, 34, 327–332. [Google Scholar] [PubMed]
  5. Lehmann, K.M.; Weyhrauch, M.; Bjelopavlovic, M.; Scheller, H.; Staedt, H.; Ottl, P.; Kaemmerer, P.W.; Wentaschek, S. Marginal and Internal Precision of Zirconia Four-Unit Fixed Partial Denture Frameworks Produced Using Four Milling Systems. Materials 2021, 14, 2663. [Google Scholar] [CrossRef] [PubMed]
  6. Obermeier, M.; Ristow, O.; Erdelt, K.; Beuer, F. Mechanical performance of cement- and screw-retained all-ceramic single crowns on dental implants. Clin. Oral Investig. 2018, 22, 981–991. [Google Scholar] [CrossRef]
  7. Lamperti, S.T.; Wolleb, K.; Hammerle, C.H.F.; Jung, R.E.; Husler, J.; Thoma, D.S. Cemented versus screw-retained zirconia-based single-implant restorations: 5-year results of a randomized controlled clinical trial. Clin. Oral Implant. Res. 2022, 33, 353–361. [Google Scholar] [CrossRef] [PubMed]
  8. Reda, R.; Zanza, A.; Cicconetti, A.; Bhandi, S.; Guarnieri, R.; Testarelli, L.; Di Nardo, D. A Systematic Review of Cementation Techniques to Minimize Cement Excess in Cement-Retained Implant Restorations. Methods Protoc. 2022, 5, 9. [Google Scholar] [CrossRef] [PubMed]
  9. Cosola, S.; Toti, P.; Babetto, E.; Covani, U.; Penarrocha-Diago, M.; Penarrocha-Oltra, D. In-vitro fatigue and fracture performance of three different ferrulized implant connections used in fixed prosthesis. J. Dent. Sci. 2021, 16, 397–403. [Google Scholar] [CrossRef] [PubMed]
  10. Charasseangpaisarn, T.; Krassanairawiwong, P.; Sangkanchanavanich, C.; Kurjirattikan, A.; Kunyawatyuwapong, K.; Tantivasin, N. The Influence of Different Surface Cleansing Agents on Shear Bond Strength of Contaminated Lithium Disilicate Ceramic: An In Vitro Study. Int. J. Dent. 2021, 2021, 7112400. [Google Scholar] [CrossRef] [PubMed]
  11. Wattanasirmkit, K.; Charasseangpaisarn, T. Effect of different cleansing agents and adhesive resins on bond strength of contaminated zirconia. J. Prosthodont. Res. 2019, 63, 271–276. [Google Scholar] [CrossRef]
  12. Burkhardt, F.; Pitta, J.; Fehmer, V.; Mojon, P.; Sailer, I. Retention Forces of Monolithic CAD/CAM Crowns Adhesively Cemented to Titanium Base Abutments-Effect of Saliva Contamination Followed by Cleaning of the Titanium Bond Surface. Materials 2021, 14, 3375. [Google Scholar] [CrossRef] [PubMed]
  13. Muller, L.; Rauch, A.; Reissmann, D.R.; Schierz, O. Impact of cement type and abutment height on pull-off force of zirconia reinforced lithium silicate crowns on titanium implant stock abutments: An in vitro study. BMC Oral Health 2021, 21, 592. [Google Scholar] [CrossRef] [PubMed]
  14. Kirman, E.A.; Asar, N.V.; Erdem, B.; Turkyilmaz, I. Effects of abutment length and platform size on the retention of implant-supported CAD/CAM crowns using six different cements. J. Dent. Sci. 2022, 17, 601–603. [Google Scholar] [CrossRef]
  15. McLaren, E.A.; Maharishi, A.; White, S.N. Influence of yttria content and surface treatment on the strength of translucent zirconia materials. J. Prosthet. Dent. 2021. [Google Scholar] [CrossRef]
  16. Komar, D.; Bago, I.; Negovetic Vranic, D.; Kranjcic, J.; Brkic, B.; Carek, A. Influence of Different Surface Pretreatments of Zirconium Dioxide Reinforced Lithium Disilicate Ceramics on the Shear Bond Strength of Self-Adhesive Resin Cement. Acta Stomatol. Croat. 2021, 55, 264–279. [Google Scholar] [CrossRef] [PubMed]
  17. Oilo, G. Luting cements: A review and comparison. Int. Dent. J. 1991, 41, 81–88. [Google Scholar]
  18. Gierthmuehlen, P.C.; Jerg, A.; Fischer, J.B.; Bonfante, E.A.; Spitznagel, F.A. Posterior minimally invasive full-veneers: Effect of ceramic thicknesses, bonding substrate, and preparation designs on failure-load and -mode after fatigue. J. Esthet. Restor. Dent. 2022, 34, 145–153. [Google Scholar] [CrossRef] [PubMed]
  19. Ernst, C.P.; Cohnen, U.; Stender, E.; Willershausen, B. In vitro retentive strength of zirconium oxide ceramic crowns using different luting agents. J. Prosthet. Dent. 2005, 93, 551–558. [Google Scholar] [CrossRef]
  20. Ernst, C.P.; Aksoy, E.; Stender, E.; Willershausen, B. Influence of different luting concepts on long term retentive strength of zirconia crowns. Am. J. Dent. 2009, 22, 122–128. [Google Scholar] [PubMed]
  21. Porto, T.S.; Medeiros da Silva, I.G.; de Freitas Vallerini, B.; Fernando de Goes, M. Different surface treatment strategies on etchable CAD-CAM materials: Part II-Effect on the bond strength. J. Prosthet. Dent. 2022. [Google Scholar] [CrossRef]
  22. Dapieve, K.S.; Pereira, G.K.R.; Venturini, A.B.; Daudt, N.; Valcanaia, A.; Bottino, M.C.; Valandro, L.F. Do resin cement viscosity and ceramic surface etching influence the fatigue performance of bonded lithium disilicate glass-ceramic crowns? Dent. Mater. 2022, 38, e59–e67. [Google Scholar] [CrossRef] [PubMed]
  23. Wille, S.; Lehmann, F.; Kern, M. Durability of resin bonding to lithium disilicate using different self-etching and conventional ceramic primers after long-term aging. Dent. Mater. 2022, 38, 444–450. [Google Scholar] [CrossRef] [PubMed]
  24. Robin, C.; Scherrer, S.S.; Wiskott, H.W.; de Rijk, W.G.; Belser, U.C. Weibull parameters of composite resin bond strengths to porcelain and noble alloy using the Rocatec system. Dent. Mater. 2002, 18, 389–395. [Google Scholar] [CrossRef]
  25. Stewart, G.P.; Jain, P.; Hodges, J. Shear bond strength of resin cements to both ceramic and dentin. J. Prosthet. Dent. 2002, 88, 277–284. [Google Scholar] [CrossRef] [PubMed]
  26. Kamada, K.; Yoshida, K.; Atsuta, M. Early bond strength and durability of bond between a ceramic material and chemically-cured or dual-cured resin luting agent. Am. J. Dent. 2001, 14, 85–88. [Google Scholar] [PubMed]
  27. Mou, S.H.; Chai, T.; Wang, J.S.; Shiau, Y.Y. Influence of different convergence angles and tooth preparation heights on the internal adaptation of Cerec crowns. J. Prosthet. Dent. 2002, 87, 248–255. [Google Scholar] [CrossRef] [PubMed]
  28. de Campos, T.N.; Adachi, L.K.; Miashiro, K.; Yoshida, H.; Shinkai, R.S.; Neto, P.T.; Moreira Arantes Frigerio, M.L. Effect of surface topography of implant abutments on retention of cemented single-tooth crowns. Int. J. Periodontics Restor. Dent. 2010, 30, 409–413. [Google Scholar]
  29. Madina, M.M.; Ozcan, M.; Badawi, M.F. Effect of surface conditioning and taper angle on the retention of IPS e.max Press crowns. J. Prosthodont. 2010, 19, 200–204. [Google Scholar] [CrossRef]
  30. Sadig, W.M.; Al Harbi, M.W. Effects of surface conditioning on the retentiveness of titanium crowns over short implant abutments. Implant. Dent. 2007, 16, 387–396. [Google Scholar] [CrossRef]
  31. Bernal, G.; Okamura, M.; Munoz, C.A. The effects of abutment taper, length and cement type on resistance to dislodgement of cement-retained, implant-supported restorations. J. Prosthodont. 2003, 12, 111–115. [Google Scholar] [CrossRef]
  32. Kumari, S.; Jadhav, M.S.; Gupta, V.; Isaac, T.K.; Subramanium, A.; Kumar, V. Microleakage at Implant Abutment and Prosthesis Interface in Cemented Implant-Supported Prosthesis. J. Pharm. Bioallied Sci. 2021, 13 (Suppl. 2), S1679–S1681. [Google Scholar]
  33. Paloco, E.A.; Berger, S.B.; Lopes, M.B.; Favaro, J.C.; Gonini-Junior, A.; Piauilino, A.I.; Borba, A.M.; Guiraldo, R.D. Influence of resin cement and thermocycling on milled lithium disilicate ceramic microshear bond strength. Acta Odontol Lat. 2021, 34, 226–232. [Google Scholar] [CrossRef]
  34. Malysa, A.; Wezgowiec, J.; Grzebieluch, W.; Danel, D.P.; Wieckiewicz, M. Effect of Thermocycling on the Bond Strength of Self-Adhesive Resin Cements Used for Luting CAD/CAM Ceramics to Human Dentin. Int. J. Mol. Sci. 2022, 23, 745. [Google Scholar] [CrossRef]
  35. Weyhrauch, M.; Igiel, C.; Scheller, H.; Weibrich, G.; Lehmann, K.M. Fracture Strength of Monolithic All-Ceramic Crowns on Titanium Implant Abutments. Int. J. Oral Maxillofac. Implant. 2016, 31, 304–309. [Google Scholar] [CrossRef]
Figure 1. Retentive forces of various ceramic materials luted with different luting agents.
Figure 1. Retentive forces of various ceramic materials luted with different luting agents.
Materials 15 06968 g001
Table 1. Ceramic materials and luting agents tested in this study.
Table 1. Ceramic materials and luting agents tested in this study.
Ceramic MaterialsManufacturerManufacturer’s
Site
Specification
Vita Mark II
[FSC]
VITA Zahnfabrik H. Rauter GmbH & Co. KGBad Säckingen,
Germany
Finely structured feldspar ceramic
Empress CAD
[LrGC]
Ivoclar VivadentSchaan,
Liechtenstein
Leucite-reinforced glass ceramic
e.max CAD
[LiDS]
Ivoclar VivadentSchaan,
Liechtenstein
Lithium disilicate glass ceramic
Vita Enamic
[PolyFSP]
VITA Zahnfabrik H. Rauter GmbH & Co. KGBad Säckingen,
Germany
Hybrid dental ceramic with polymer network
86% ceramic network
14% polymer network
Vita Suprinity
[PsZirLS]
VITA Zahnfabrik H. Rauter GmbH & Co. KGBad Säckingen,
Germany
Zirconia-reinforced lithium silicate
pre-sintered
10% zirconia
90% lithium silicate
Celtra Duo
[FcZirLS]
Dentsply Detrey GmbHKonstanz,
Germany
Zirconia-reinforced lithium silicate
fully crystallized
10% zirconia
90% lithium silicate
Multilink
Implant
Ivoclar VivadentSchaan,
Liechtenstein
Self-polymerizing luting composite with light-curing option
Variolink IIIvoclar VivadentSchaan,
Liechtenstein
Dual-/light- polymerizing, resin-based dental luting material
Rely X Unicem3M ESPENeuss,
Germany
Dual-polymerizing, self-adhesive resin cement
GC FujicemGC CorporationTokyo,
Japan
Self-polymerizing, resin-reinforced glass-ionomer luting cement
Panavia 2.0Kuraray Europe GmbHHattersheim am Main,
Germany
Dual-polymerizing, self-adhesive resin cement
Table 2. Conditioning of the ceramic material.
Table 2. Conditioning of the ceramic material.
MaterialConditioningBonding
FSC60 s, hydrofluoric acid 60 s, Monobond S
LrGC60 s, hydrofluoric acid 60 s, Monobond S
LiDS20 s, hydrofluoric acid60 s, Monobond S
PsZirLS60 s, hydrofluoric acid 60 s, Monobond S
PolyFSP60 s, hydrofluoric acid 60 s, Monobond S
FcZirLS30 s, hydrofluoric acid60 s, Monobond S
Table 3. Retentive strength (mean, minimum, maximum, SD in N) divided by ceramic material and luting agent.
Table 3. Retentive strength (mean, minimum, maximum, SD in N) divided by ceramic material and luting agent.
Ceramic MaterialLuting AgentnMeanSdMinMax
FSCMI15685142310885
VII15657145445970
RX15652146430955
GC157591345521010
P15715145485946
LrGCMI15681146290887
VII1577489642925
RX157491755361198
GC1574898579935
P158221316351068
LiDSMI158421764381147
VII158591096911052
RX1510441738251310
GC159701357401137
P15784133458945
PolyFSPMI1510171328421320
VII159191506751227
RX1511691428921398
GC1511772207871518
P157161694581097
PsZirLSMI158672782351411
VII1511092725211569
RX1513322727631935
GC15148821411041737
P1510212026861407
FcZirLSMI1510041705451201
VII159691517661328
RX1511302047901530
GC1513352299741647
P157722095101239
Table 4. Location of luting agents’ residue (MI: Multilink Implant, VII: Variolink II, RX: RelyX Unicem, GC: Fujicem, P: Panavia).
Table 4. Location of luting agents’ residue (MI: Multilink Implant, VII: Variolink II, RX: RelyX Unicem, GC: Fujicem, P: Panavia).
Ceramic MaterialFSCLrGCLiDS
Luting AgentMIVIIRXGCPMIVIIRXGCPMIVIIRXGCP
Location of luting agent residue: implant surface/inner crown surface0%/100%15141501115151521015141489
25%/75%010740108501146
50%/50%000800003000030
75%/25%000000002000000
100%/0%000000000000000
Ceramic MaterialPolyFSPPsZirLSFcZirLS
Luting AgentMIVIIRXGCPMIVIIRXGCPMIVIIRXGCP
Location of luting agent residue: implant surface/inner crown surface0%/100%012010000000000
25%/75%136919314510212852
50%/50%003225021422232
75%/25%020213018210121
100%/0%261102417591014510
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Bjelopavlovic, M.; Weyhrauch, M.; Scheller, H.; Wentaschek, S.; Lehmann, K.M. Retentive Strength of CAD/CAM-Fabricated All-Ceramic Crowns Luted on Titanium Implant Abutments Using Different Ceramic Materials and Luting Agents: An In Vitro Study. Materials 2022, 15, 6968. https://doi.org/10.3390/ma15196968

AMA Style

Bjelopavlovic M, Weyhrauch M, Scheller H, Wentaschek S, Lehmann KM. Retentive Strength of CAD/CAM-Fabricated All-Ceramic Crowns Luted on Titanium Implant Abutments Using Different Ceramic Materials and Luting Agents: An In Vitro Study. Materials. 2022; 15(19):6968. https://doi.org/10.3390/ma15196968

Chicago/Turabian Style

Bjelopavlovic, Monika, Michael Weyhrauch, Herbert Scheller, Stefan Wentaschek, and Karl Martin Lehmann. 2022. "Retentive Strength of CAD/CAM-Fabricated All-Ceramic Crowns Luted on Titanium Implant Abutments Using Different Ceramic Materials and Luting Agents: An In Vitro Study" Materials 15, no. 19: 6968. https://doi.org/10.3390/ma15196968

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop