Article

How South Africa Used National Cycle Threshold (Ct) Values to Continuously Monitor SARS-CoV-2 Laboratory Test Quality

1
Wits Diagnostic Innovation Hub, Faculty of Health Science, University of the Witwatersrand, Johannesburg 2093, South Africa
2
Division of Medical Virology, Faculty of Heath Sciences, University of Cape Town, Cape Town 7700, South Africa
3
The National Health Laboratory Service, Johannesburg, Private Bag X8, Sandringham 2131, South Africa
4
The National Priority Program of the National Health Laboratory Service, Johannesburg, Private Bag X8, Sandringham 2131, South Africa
5
Department of Pathology, Faculty of Medicine and Health Sciences, Stellenbosch University, Stellenbosch 7600, South Africa
6
Division of Chemical Pathology, Faculty of Heath Sciences, University of Cape Town, Cape Town 7700, South Africa
*
Author to whom correspondence should be addressed.
Diagnostics 2023, 13(15), 2554; https://doi.org/10.3390/diagnostics13152554
Submission received: 30 May 2023 / Revised: 24 July 2023 / Accepted: 29 July 2023 / Published: 1 August 2023
(This article belongs to the Special Issue Highlights of Molecular Laboratory Diagnostics in South Africa)

Abstract

:
The high demand for SARS-CoV-2 tests but limited supply to South African laboratories early in the COVID-19 pandemic resulted in a heterogenous diagnostic footprint of open and closed molecular testing platforms being implemented. Ongoing monitoring of the performance of these multiple and varied systems required novel approaches, especially during the circulation of variants. The National Health Laboratory Service centrally collected cycle threshold (Ct) values from 1,497,669 test results reported from 6 commonly used PCR assays in 36 months, and visually monitored changes in their median Ct within a 28-day centered moving average for each assays’ gene targets. This continuous quality monitoring rapidly identified delayed hybridization of RdRp in the Allplex™ SARS-CoV-2 assay due to the Delta (B.1.617.2) variant; S-gene target failure in the TaqPath™ COVID-19 assay due to B.1.1.7 (Alpha) and the B.1.1.529 (Omicron); and recently E-gene delayed hybridization in the Xpert® Xpress SARS-CoV-2 due to XBB.1.5. This near “real-time” monitoring helped inform the need for sequencing and the importance of multiplex molecular nucleic acid amplification technology designs used in diagnostics for patient care. This continuous quality monitoring approach at the granularity of Ct values should be included in ongoing surveillance and with application to other disease use cases that rely on molecular diagnostics.

1. Introduction

South Africa (population of ~60.6 million people) [1] has one of the highest HIV and TB prevalence rates in the world: 8.45 million people live with HIV (13.9% of the population) [1] and a TB incident rate of 513/100,000 is reported [2]. Healthcare is funded through the government for 84% of the population (public sector) and the remainder is funded privately through individuals, medical schemes, and insurance companies [3]. Laboratory services are provided through networks of private and public sector laboratories, with the latter comprising 256 National Health Laboratory Service (NHLS) facilities positioned centrally and at district level across the nine provinces.
On 5 March 2020, South Africa reported their first case of COVID-19 [4] and both public and private laboratories’ [5] virology services had to rapidly scale their SARS-CoV-2 nucleic acid amplification technology (NAAT); the primary method of diagnosing infection with SARS-CoV-2 [6,7]. Within one month, however, the country’s SARS-CoV-2 testing demands required expanding molecular testing to the national priority program (NPP) of the NHLS. The NPP has capacity for 10 million molecular tests/annum, which predominantly support HIV viral load (VL) monitoring [8] and TB molecular diagnostics [9]. Testing is performed in 17 VL laboratories (11 also provide HIV early infant diagnostics) equipped with cobas® [Roche Molecular, Pleasanton, CA, USA] and Alinity m [Abbott Molecular, Des Plains, IL, USA] platforms) and 175 GeneXpert (Cepheid, Sunnyvale, CA, USA) testing laboratories, which service ~3800 primary health care facilities. Test kit demands for the NPP’s closed platforms (until 2021), however, could not be met due to the inability of suppliers to ship to South Africa as demands escalated in their countries of manufacture. SARS-CoV-2 testing therefore had to expand to a third network of laboratories. This involved a SARS-CoV-2 surge program initiated by the South African Medical Research Council (SAMRC) [10] with solidarity funding to support selected academic laboratories, councils, and research institutes to conduct SARS-CoV-2 testing. A rapid laboratory electronic assessment tool was developed to collect information on site location, ISO 15189 status, Health Professions Council of South Africa status of key staff, SARS-CoV-2 readiness (available testing platforms) and implemented quality management systems (trained staff, use of electronic requisition forms, use of barcodes, PPE stock, kit and supply storage and procurement processes, laboratory information (LIS) and IT support on-site). Thirty-two sites were assessed by the NHLS’ Quality Assurance Division and eleven were selected for SARS-CoV-2 surge testing. Testing in these laboratories commenced in June 2020 and continued until January 2021.
Test results (SARS-CoV-2 detected, SARS-CoV-2 not detected and test unsuccessful) from the private laboratories, the surge testing sites and the NHLS (including NPP) were reported as cases by the National Institute of Communicable Diseases (NICD), who provide central COVID-19 epidemiology and surveillance reporting nationally [11] and regionally [12]. Overall, the private laboratories performed ~55% of South Africa’s SARS-CoV-2 diagnostics (as reported in week 34 of 2021 [13]). A link was also established to the genomics network with five laboratories identified across South Africa capable of performing SARS-CoV-2 sequencing [14]. Throughout the five COVID-19 infectious waves in South Africa [15], specimens were shared between the diagnostic laboratories and genomics group to detect SARS-CoV-2 variants, including variants of concern (VOC).
Laboratories are required to participate in external proficiency testing to monitor pre-analytical, analytical and post-analytical performance. However, after decades of quality indicators, a paradigm shift should be towards indicators of total quality [16]. Figure 1 briefly outlines some of the processes required for quality management prior to selecting suitable laboratory tests, implementing tests in the field and then data environments for result reporting. During large scale pandemic testing, such as during COVID-19, external quality assessment programs were implemented [17,18,19]. However, these would not easily identify quality issues timeously. The South African NHLS approached this challenge by including the cycle threshold (Ct) values for each gene target of the SARS-CoV-2 tests performed in their laboratories that identified the presence of SARS-CoV-2 (and therefore diagnostic for COVID-19) in their LIS interface. The NHLS has a single national LIS, TrakCare (InterSystems, Cambridge, MA, USA), where all SARS-CoV-2 diagnostic systems are interfaced, and Ct values accessed for analysis from a central data warehouse (Netezza, IBM, USA-based server in Johannesburg).
A Ct value is intrinsic to PCR assays and is a measure of the amount of target nucleic acid in the specimen [20]. SARS-CoV-2 test manufacturers provide information on which SARS-CoV-2 gene regions their assays target and provide the Ct values in the test output comma separator value (.csv) files. Therefore, in addition to central monitoring of daily test volumes, qualitative test results and indicators to track changes in South Africa’s COVID-19 epidemic, the NHLS’ centrally collected SARS-CoV-2 Ct values were analyzed.
The aim of this study was therefore to highlight how the Ct values from the ensemble of closed and open testing platforms and assays in the NHLS laboratories were used for continuous quality monitoring (CQM) of diagnostic assay performance.

2. Materials and Methods

Specimens received in the NHLS laboratories for SARS-CoV-2 testing were collected using nasopharyngeal and oropharyngeal swabs. These were transported dry during shortages of universal transport medium (UTM) and were cut and placed into phosphate buffered saline upon arrival and processed according to standard operating procedures. Specimens were registered in the LIS and tested across 205 NHLS laboratories using the locally available platforms and testing protocols. Test results, including Ct values (when SARS-CoV-2 was detected) were accessed from the CDW through an extract, transform and load process to generate a .csv file for analysis. Data did not include patient unique identifiers, and hence analyses performed included longitudinal follow-up testing. Ethics approval was obtained from the University of the Witwatersrand Human Research Ethics Committee (number M160978), Johannesburg, South Africa. Data files were analyzed and data visualized using STATA 14 (StataCorp. 2015, Stata Statistical Software: Release 14. College Station, TX, USA: StataCorp LP) and Tableau 2020.3 (Tableau Software. 2020. Seattle, WA, USA: Tableau). Bar charts depicted daily positive test numbers and line graphs represented changes in median Ct values within a 28-day centered moving average. Changes in patterns of the moving average of the median Ct values of one or more gene targets could reflect an assay performance issue (quality or even lot variability) or due to the introduction of a SARS-CoV-2 mutation that impacts PCR efficiency. The latter could lead to primer/probe reduced hybridization (where the median Ct value would increase above other unaffected targets) or complete drop-out (where the median Ct value would decrease substantially due to the recording of zero Ct values in contrast to unaffected target Ct values). These changes were monitored across all assays and their gene targets, and no Ct values were modified or removed to ensure consistent daily monitoring of raw data outputs that could inform critical alerts in near-real time. Ten SARS-CoV-2 molecular assays were recorded in the CDW database. However, 94.2% of tests were performed on six assays, which form the focus of this analysis. These assays are classified into open laboratory testing platforms: Allplex™ SARS-CoV-2 (SeeGene Inc., Seoul, Korea); TaqPath™ COVID-19 (Thermo Fisher Scientific, Waltham, MA, USA) and closed platforms: cobas® SARS-CoV-2 (Roche Molecular, Pleasanton, CA, USA); Xpert® Xpress SARS-CoV-2 (Cepheid, Sunnyvale, CA, USA), RealTime SARS-CoV-2 (Abbott Molecular, Des Plains, IL, USA) and Alinity m (Abbott Molecular, Des Plains, IL, USA). Table 1 details the assay gene targets and range in Ct values when a specimen is reported positive for the detection of SARS-CoV-2. All assays report their gene targets in different fluorescent channels and can individually be identified, except the RealTime SARS-CoV-2 and the Alinity m, where both gene targets are reported in a single fluorescent channel and neither target can be individually identified.

3. Results

The six SARS-CoV-2 assays utilized by the NHLS were implemented at different times during the pandemic and their uptake at a province level varied as outlined in Figure 2. Across 36 months of testing (March 2020–March 2023), the Allplex™ SARS-CoV-2 and the Xpert® Xpress SARS-CoV-2 contributed 48% of the test results. The Xpert® Xpress SARS-CoV-2 assay is the only testing platform used in all provinces, which is a consequence of the placement of this platform by the NPP for use in the national TB diagnostic program. In contrast, the TaqPath™ COVID-19 assay was predominantly only used in the Gauteng Province. The RealTime SARS-CoV-2 and Alinity m assays only contributed ~10% towards testing, despite these platforms used for the HIV diagnostics program of NPP. At least 66% of all testing was reported from the three most densely populated provinces (Gauteng, KwaZulu-Natal, and the Western Cape).
A total of 8,573,872 tests were performed by NHLS laboratories during the study time period and 1,572,098 (18.33%) of them reported the presence of SARS-CoV-2, and 1,497,669 (95.2%) of these included Ct values for their gene targets. Ct values would not have been captured where pathology results were entered manually by some academic partners whose systems were not interfaced with the NHLS LIS. The daily median Ct for each SARS-CoV-2 gene target for the open and closed platforms is described in the following sections. Common to all assays (and visualized across all plots), is the decrease in median Ct (increase in SARS-CoV-2 viral concentration) during an infection wave and vice versa after an infection wave.

3.1. The NHLS SARS-CoV-2 Open Testing Platforms

3.1.1. Allplex™ SARS-CoV-2 (SeeGene)

The Allplex™ SARS-CoV-2 assay was the first test to be implemented by NHLS in March 2020 and contributed a national test count of 2,095,588 million tests (26%) after 36 months. Although this assay was used by all provinces, 91% of the results were generated from the Gauteng, Western Cape, KwaZulu-Natal and Eastern Cape provinces. Figure 3 highlights that the E-gene generated the lowest Ct values followed by the RdRp and N-gene Ct values. The changes in 28-day centered moving average of the median Ct for all three genes mirror (duplicate) each other’s visual pattern except during South Africa’s third COVID-19 wave, which peaked in July 2021. The RdRp median Ct increased above the N-gene Ct. This increase in Ct (reduced PCR performance) was reported to be a result of the B.1.617.2 (Delta) variant [21] which had a highly conserved nonsynonymous mutation (G15451A) exclusively within the RdRp gene, and thereby negatively affecting the RdRp PCR efficiency of the Allplex™ SARS-CoV-2 assay.
This was continuously monitored by calculating the relative change in Ct between RdRp and E and any test result where the Ct of RdRp-E > 3.5 indicated the presence of the Delta variant. During the Delta wave, up to 62% of all positive specimens tested using the Allplex™ SARS-CoV-2 assay reported this phenomenon. No changes in this assays’ PCR efficiency were noted beyond the impact from the Delta variant. However, SeeGene introduced a new Allplex™ SARS-CoV-2 assay version that included the S-gene target, but this target is reported in the same PCR fluorescent channel as RdRp and therefore neither target’s PCR efficiency is discernable. Towards the end of the study period (March 2023), the Allplex™ SARS-CoV-2 testing volumes were reduced to ~42/day, making daily monitoring less reliable.

3.1.2. TaqPath™ COVID-19 (Thermo Fisher Scientific)

The TaqPath™ COVID-19 assay was implemented in May 2020 and contributed to 1,660,419 million tests (21%) performed by the NHLS. Implementation, however, was not national, and 50% of testing was performed in the Gauteng Province. Figure 4 shows the Ct values of ORF1ab, N and S-genes generally mirrored each other for the first two waves. However, the S-gene pattern changed during the 3rd, 4th and 5th waves, with the median Ct of the S-gene target much lower than the ORF1ab and N-genes. This phenomenon was due to S-gene target failure (SGTF) [22], brought about by deletion of amino acids 69 and 70 in B.1.1.7 (Alpha) and B.1.1.529 (Omicron) spike genes. This yielded a distinct absent S-gene and hence no amplification during PCR and Ct values reported as zero in the TaqPath™ COVID-19 .csv file, which contributed to an overall low median S-gene Ct value. The TaqPath™ COVID-19 testing rates decreased to <4 tests/day at the end of March 2023, making continuous quality monitoring less reliable.

3.2. The NHLS SARS-CoV-2 Closed Testing Platforms

3.2.1. Xpert® Xpress SARS-CoV-2 (Cepheid)

The Xpert® Xpress SARS-CoV-2 assay was implemented by the NHLS in March 2020 and contributed to 2,073,844 million tests (26%) after 36 months. This assay was used in all provinces across South Africa, however, Gauteng (21%), KwaZulu-Natal (17%), and the Western Cape (16%) accounted for more than half (54%) of the total Xpert® Xpress SARS-CoV-2 test results. Figure 5 highlights a change in the NHLS’ testing algorithm implemented by their Xpert® Xpress SARS-CoV-2 testing sites on 15 September 2021. Concerns were raised by the NHLS virology expert committee on reporting a positive SARS-CoV-2 result when Ct values of either or both gene targets approached the threshold of 45. The following algorithm was therefore implemented within the NHLS LIS: where a single E or N2 has a Ct > 38 or where both E and N2 Ct > 40, these specimens are reported as “inconclusive” (internal memo Dr M.P. da Silva). This change is evident in Figure 5 with greater mirroring (less variability in the Ct trends) between both gene targets beyond this date. Figure 5 further shows that the E-gene generated lower Ct values than the N2-gene for 33 of the 36 months. Changes in the 28-day centered moving average of the median Ct for both the E and N2-genes mirror each other with the exception of the period after January 2023, where the median E-gene Ct values increase. This phenomenon is due to XBB.1.5. VOC, which affects the E-gene coverage dropping by 1% due to two mismatches (personal communication from Cepheid medical affairs) that delay PCR hybridization. As the Xpert® Xpress SARS-CoV-2 continues to be used in NHLS testing sites (400 tests/day at the end of March 2023), the proportion of specimens with E-gene Ct > N2-gene Ct can therefore be used to monitor the circulation of XBB.1.5. This was evident among 92% (28-day moving average) of Xpert® Xpress SARS-CoV-2 test results at the end of March 2023.

3.2.2. cobas® SARS-CoV-2 (Roche Molecular)

The cobas® SARS-CoV-2 assay was implemented in April 2020 and contributed 1,357,991 million tests (17%) after 36 months to the NHLS’ test volumes. Although the assay was implemented across all provinces in South Africa; the Eastern Cape (22%), KwaZulu-Natal (21%), and the Western Cape (15%) contributed 60% of the total testing volumes. Figure 6 clearly shows that changes in the 28-day centered moving average of the median Ct is perfectly aligned for both gene targets (E and ORF1ab). Their mirrored pattern remains throughout the 36 months of testing and this assay appears unaffected by any VOCs. SARS-CoV-2 testing on this platform however ceased in the NHLS in July 2022.

3.2.3. RealTime SARS-CoV-2 and ALINITY m SARS-CoV-2 (Abbott Molecular)

The RealTime SARS-CoV-2 and ALINITY m SARS-CoV-2 assays, from a single supplier, were implemented by the NHLS in May 2020 and November 2020 respectively. The RealTime SARS-CoV-2 assay contributed 409,654 tests (5%) while the ALINITY m SARS-CoV-2 assay contributed 479,536 tests (6%) to the NHLS’ national testing volumes during the 36 months. A significant portion (62% of RealTime SARS-CoV-2 tests and 81% of ALINITY m SARS-CoV-2) of these tests were conducted by two provinces (KwaZulu-Natal and Gauteng). Both these assays target the RdRP and N gene regions, but neither region can be distinguished by a Ct value as both targets fluoresce in a single channel as outlined in Figure 7. SARS-CoV-2 testing continues to be reported on the ALINITY m SARS-CoV-2 platform (65 tests/day at the end of March 2023), however, testing was discontinued on the RealTime SARS-CoV-2 in July 2022.

4. Discussion

The Ct is a variable that correlates with the amount of target RNA in a specimen [23], and during COVID-19 the SARS-CoV-2 viral load in the respiratory tract was reported to align with an individual’s disease progression [6,24,25,26]. In this analysis, we showed that continuous quality monitoring of median Ct’s of SARS-CoV-2 gene targets at a national (population) level could rapidly identify changes in assay PCR gene target performance, due to genetic mutations of SARS-CoV-2 and assist in changing laboratory reporting algorithms. Our Ct analyses were therefore provided in weekly updates (in the form of written brief and weekly zoom meetings) to the South African scientific community involved in the COVID-19 responses. Changes in Ct values over time could therefore be an additional parameter included in good laboratory practices’ quality management systems, and especially within the context of gene mutations that could impact molecular assay performance.
Although the CQM of Ct values across the gene targets and across assays was limited to only the NHLS testing laboratories (<50% of South Africa’s SARS-CoV-2 testing), 1,497,669 Ct values were analyzed over 36 months. This “big” data also reduced the impact of variables (central limit theorem) such as specimen quality, PCR inhibitory substances, variable transport media, variable front-end extraction or extraction-free technologies and NAAT reaction volume, to name a few, that is reported as complexities in SARS-CoV-2 laboratory testing [27].Quality control practices at a laboratory level (EQA or inter-laboratory exercises) to assess the intrinsic laboratory quality and capacity are still necessary to be implemented.
Our findings show that of the six SARS-CoV-2 molecular assays implemented at scale across the NHLS, the Ct value analysis by gene target rapidly identified three assays’ targets affected by VOCs: (i) the B.1.617.2 (Delta) variant affected the Allplex™ SARS-CoV-2 assay in the RdRp target region [21]; (ii) the B.1.1.7 (Alpha) and the B.1.1.529 (Omicron) affected the TaqPath™ COVID-19 assay in the S-gene target region [15]; and (iii) the XBB.1.5 affected the Xpert® Xpress SARS-CoV-2 in the E-gene target region. The cobas® SARS-CoV-2 assay appeared not to be affected by circulating VOCs based on no changes identified in their gene target Ct value trends. The inability to distinguish the Ct values for the multiple gene targets in the RealTime SARS-CoV-2 and ALINITY m SARS-CoV-2 made the CQM less helpful in identifying changes in the assay quality potentially due to VOC but was a stable marker outlining the expected patterns in Ct change over the course of COVID-19 that other assays could be compared. Despite several assays’ gene target performance being affected by VOCs, their multiplex nature [28] (at least two viral genes targeted to increase the probability to identify the virus at low viral load and in the presence of viral mutations), enabled the assays to continue being used as the primary diagnostic for patient care [29,30].
Through the Ct CQM, the loss or change in target performance was, however, also quickly identified as an advantage in monitoring the spread of VOC [15,22] in “near” real-time with only a two-day lag period between specimen receipt and authorization. This also included monitoring the recovery of Ct values of affected gene targets, with the best example shown by the TaqPath™ COVID-19 where the B.1.1.7 (Alpha) caused the SGTF during wave 3 (which was, however, rapidly replaced by the B.1.617.2 (Delta) variant). Hence, the TaqPath™ COVID-19 S-gene target performance could inform changes in circulating variants.
Although the NHLS’ multi-assay implementation approach was governed by test demands and availability of platforms and reagents, it did prove possible to monitor such a multi-assay program through the unique centralized LIS. This in turn also highlighted some limitations, such as the TaqPath™ COVID-19 assay not being implemented nationally, and >50% of test results reported from only the Gauteng Province. Findings such as the SGTF therefore could not be extrapolated to regions where this assay was not being used.
Overall, this study is the first, to our knowledge, that highlights CQM using national program laboratory Ct values from multiple SARS-CoV-2 assays. The data strongly show that variables of molecular test results can be a key part of laboratory quality management. It also highlights the multidisciplinary approach to CQM with the need to understand molecular technology, the need to understand the role of diagnostics in clinical and laboratory practices and the need to understand big data analytics and visualization. This study also highlights the value molecular diagnostics “near-real-time” analysis has in informing the need for sequencing. The introduction of rapid SARS-CoV-2 antigen tests, and self-tests however severely limits this value and ongoing molecular surveillance should be maintained. This system’s approach to quality management and program performance monitoring therefore should also be investigated for other disease use cases, such as TB and HIV, where molecular technology is the primary diagnostic test.

5. Conclusions

In conclusion, this study demonstrates the potential of Continuous Quality Monitoring (CQM) using Ct values of SARS-CoV-2 gene targets at a national level to rapidly identify changes in SARS-CoV-2 laboratory assay performance, particularly in response to genetic mutations of the virus. Despite some limitations, such as the regional distribution of certain assays, the approach using centralized laboratory information emphasizes the significance of molecular diagnostics in informing clinical decisions and surveillance efforts but does require multidisciplinary collaboration between clinical, molecular and data scientists. Furthermore, the study suggests the potential application of this system’s approach to quality management in other disease contexts where molecular technology serves as a primary diagnostic tool. Overall, this research underscores the value of continuous monitoring in pandemic responses and the need to maintain molecular surveillance.

Author Contributions

Authors contributed in the following ways: L.E.S., N.-y.H., and G.D. contributed to conceptualization. L.E.S., G.D., and J.T. performed the formal analyses, methodology and data visualization. H.V. and K.M. were responsible for data curation. L.H., P.M., M.P.d.S., W.P., and K.M. validated the data findings. L.E.S. and W.S.S. oversaw funding acquisition and project administration. L.E.S. prepared the original manuscript draft. N.-y.H., G.D., W.P., K.M., and W.S.S. reviewed the manuscript and contributed to the writing. All authors have read and agreed to the published version of the manuscript.

Funding

Research reported in this publication was supported by the South African Medical Research Council with funds received from the Department of Science and Innovation. The content and findings reported are the sole deduction, view and responsibility of the researcher and do not reflect the official position and sentiments of the SAMRC or the Department of Science and Innovation. Research reported in this publication was also supported by the National Institute of Allergy and Infectious Diseases of the National Institutes of Health under Award Number R01AI152126. The content is solely the responsibility of the authors and does not necessarily represent the official views of the National Institutes of Health.

Institutional Review Board Statement

The study was conducted in accordance with the Declaration of Helsinki, and those accessing the data were approved by the Institutional Review Board (or Ethics Committee) of The Human Research Ethics Committee of the University of the Witwatersrand, Johannesburg, South Africa (protocol code M160978 9 October 2020) under the program ILDAC (Integrated Laboratory Data Analysis for Care) coordinated by M.P.d.S. and W.S.

Informed Consent Statement

The molecular analyses of the study did not contain unique patient identifiers and the Ct laboratory variable was considered no greater than minimal risk by the IRB. Given the practicalities of retrospectively contacting everyone in this large laboratory database, waiver of consent was considered appropriate by the IRB. All methods were carried out in accordance with relevant guidelines and regulations.

Data Availability Statement

The National Health Laboratory Service data analyzed in this study is unavailable due to patient privacy and data restrictions. Data resides on the CDW for pathology services. Aggregated data may be accessible through the AARQA (Academic Affairs Research Office and Quality Assurance) division of the NHLS under leadership of KM.

Acknowledgments

The authors wish to thank the many NHLS laboratories who performed the SARS-CoV-2 tests analyzed in this study, the Virology Expert Committee of the NHLS and the SAMRC supported South African SARS-CoV-2 Variant Group whose technical, scientific and clinical insights in molecular diagnostics enabled this study to continue and add value to patient care in South Africa during the pandemic.

Conflicts of Interest

The authors declare no conflict of interest. The funders had no role in the design of the study; in the collection, analyses, or interpretation of data; in the writing of the manuscript; or in the decision to publish the results.

References

  1. Mid-Year Population Estimates. 2022. Available online: https://www.statssa.gov.za/publications/P0302/P03022022.pdf (accessed on 29 May 2023).
  2. Global Tuberculosis Programme. 2022. Available online: https://www.who.int/teams/global-tuberculosis-programme/data#profiles (accessed on 29 May 2023).
  3. M’Bouaffou, F.; Buch, E.; Olorunju, S.; Thsehla, E. Perceived knowledge of scheme members and their satisfaction with their medical schemes: A cross-sectional study in South Africa. BMC Public Health 2022, 22, 1700. [Google Scholar] [CrossRef]
  4. Zhu, N.; Zhang, D.; Wang, W.; Li, X.; Yang, B.; Song, J.; Zhao, X.; Huang, B.; Shi, W.; Lu, R.; et al. A Novel Coronavirus from Patients with Pneumonia in China, 2019. N. Engl. J. Med. 2020, 382, 727–733. [Google Scholar] [CrossRef]
  5. Baxter, C.; Abdool Karim, Q.; Abdool Karim, S.S. Identifying SARS-CoV-2 infections in South Africa: Balancing public health imperatives with saving lives. Biochem. Biophys. Res. Commun. 2021, 538, 221–225. [Google Scholar] [CrossRef]
  6. Sethuraman, N.; Jeremiah, S.S.; Ryo, A. Interpreting Diagnostic Tests for SARS-CoV-2. JAMA 2020, 323, 2249–2251. [Google Scholar] [CrossRef]
  7. Tang, Y.W.; Schmitz, J.E.; Persing, D.H.; Stratton, C.W. Laboratory Diagnosis of COVID-19: Current Issues and Challenges. J. Clin. Microbiol. 2020, 58, 10–128. [Google Scholar] [CrossRef]
  8. Bor, J.; Gage, A.; Onoya, D.; Maskew, M.; Tripodis, Y.; Fox, M.P.; Puren, A.; Carmona, S.; Mlisana, K.; MacLeod, W. Variation in HIV care and treatment outcomes by facility in South Africa, 2011–2015: A cohort study. PLoS Med. 2021, 18, e1003479. [Google Scholar] [CrossRef]
  9. Stevens, W.S.; Scott, L.; Noble, L.; Gous, N.; Dheda, K. Impact of the GeneXpert MTB/RIF Technology on Tuberculosis Control. Microbiol. Spectr. 2017, 5, 389–410. [Google Scholar] [CrossRef]
  10. South Africa Strengthens Its Fight against COVID-19 through International Solidarity and the BRICS Cooperation. 2020. Available online: https://www.samrc.ac.za/press-releases/south-africa-strengthens-its-fight-against-covid-19-through-international-solidarity (accessed on 29 May 2023).
  11. COVID-19 Special Reports. 2020. Available online: https://www.nicd.ac.za/diseases-a-z-index/disease-index-covid-19/surveillance-reports/covid-19-special-reports/ (accessed on 29 May 2023).
  12. Salyer, S.J.; Maeda, J.; Sembuche, S.; Kebede, Y.; Tshangela, A.; Moussif, M.; Ihekweazu, C.; Mayet, N.; Abate, E.; Ouma, A.O.; et al. The first and second waves of the COVID-19 pandemic in Africa: A cross-sectional study. Lancet 2021, 397, 1265–1275. [Google Scholar] [CrossRef]
  13. COVID19 Testing Summary South Africa: Week 34 2021. 2021. Available online: https://www.nicd.ac.za/wp-content/uploads/2021/09/COVID-19-Testing-Summary-Week-34-2021.pdf (accessed on 29 May 2023).
  14. Msomi, N.; Mlisana, K.; de Oliveira, T.; Network for Genomic Surveillance in South Africa Writing Group. A genomics network established to respond rapidly to public health threats in South Africa. Lancet Microbe 2020, 1, e229–e230. [Google Scholar] [CrossRef] [PubMed]
  15. Tegally, H.; Moir, M.; Everatt, J.; Giovanetti, M.; Scheepers, C.; Wilkinson, E.; Subramoney, K.; Makatini, Z.; Moyo, S.; Amoako, D.G.; et al. Emergence of SARS-CoV-2 Omicron lineages BA.4 and BA.5 in South Africa. Nat. Med. 2022, 28, 1785–1790. [Google Scholar] [CrossRef]
  16. Plebani, M. Quality in laboratory medicine: 50years on. Clin. Biochem. 2017, 50, 101–104. [Google Scholar] [CrossRef]
  17. Sung, H.; Han, M.G.; Yoo, C.K.; Lee, S.W.; Chung, Y.S.; Park, J.S.; Kim, M.N.; Lee, H.; Hong, K.H.; Seong, M.W.; et al. Nationwide External Quality Assessment of SARS-CoV-2 Molecular Testing, South Korea. Emerg. Infect. Dis. 2020, 26, 2353–2360. [Google Scholar] [CrossRef] [PubMed]
  18. Matheeussen, V.; Corman, V.M.; Donoso Mantke, O.; McCulloch, E.; Lammens, C.; Goossens, H.; Niemeyer, D.; Wallace, P.S.; Klapper, P.; Niesters, H.G.; et al. International external quality assessment for SARS-CoV-2 molecular detection and survey on clinical laboratory preparedness during the COVID-19 pandemic, April/May 2020. Eurosurveillance 2020, 25, 2001223. [Google Scholar] [CrossRef]
  19. Gorzer, I.; Buchta, C.; Chiba, P.; Benka, B.; Camp, J.V.; Holzmann, H.; Puchhammer-Stockl, E.; Aberle, S.W. First results of a national external quality assessment scheme for the detection of SARS-CoV-2 genome sequences. J. Clin. Virol. 2020, 129, 104537. [Google Scholar] [CrossRef]
  20. Bustin, S.A.; Mueller, R. Real-time reverse transcription PCR (qRT-PCR) and its potential use in clinical diagnosis. Clin. Sci. 2005, 109, 365–379. [Google Scholar] [CrossRef] [Green Version]
  21. Valley-Omar, Z.; Marais, G.; Iranzadeh, A.; Naidoo, M.; Korsman, S.; Maponga, T.; Hussey, H.; Davies, M.A.; Boulle, A.; Doolabh, D.; et al. Reduced amplification efficiency of the RNA-dependent-RNA-polymerase (RdRp) target enables tracking of the Delta SARS-CoV-2 variant using routine diagnostic tests. medRxiv 2021. [Google Scholar] [CrossRef]
  22. Scott, L.; Hsiao, N.Y.; Moyo, S.; Singh, L.; Tegally, H.; Dor, G.; Maes, P.; Pybus, O.G.; Kraemer, M.U.G.; Semenova, E.; et al. Track Omicron’s spread with molecular data. Science 2021, 374, 1454–1455. [Google Scholar] [CrossRef] [PubMed]
  23. Hughes, K.; Rao, S.N.; Manissero, D. The Role of Cycle Threshold Values in Infectious Disease Diagnostics. 2021. Available online: https://www.mlo-online.com/continuing-education/article/21206319/the-role-of-cycle-threshold-values-in-infectious-disease-diagnostics (accessed on 29 May 2023).
  24. Salvatore, P.P.; Dawson, P.; Wadhwa, A.; Rabold, E.M.; Buono, S.; Dietrich, E.A.; Reses, H.E.; Vuong, J.; Pawloski, L.; Dasu, T.; et al. Epidemiological Correlates of PCR Cycle Threshold Values in the Detection of SARS-CoV-2. Clin. Infect. Dis. 2020, 72, e761–e767. [Google Scholar] [CrossRef]
  25. Rao, S.N.; Manissero, D.; Steele, V.R.; Pareja, J. A Systematic Review of the Clinical Utility of Cycle Threshold Values in the Context of COVID-19. Infect. Dis. Ther. 2020, 9, 573–586. [Google Scholar] [CrossRef]
  26. Tom, M.R.; Mina, M.J. To Interpret the SARS-CoV-2 Test, Consider the Cycle Threshold Value. Clin. Infect. Dis. 2020, 71, 2252–2254. [Google Scholar] [CrossRef]
  27. Scott, L.E.; Noble, L.D.; Singh-Moodley, A.; Kahamba, T.; Hardie, D.R.; Preiser, W.; Stevens, W.S. Challenges and complexities in evaluating severe acute respiratory syndrome coronavirus 2 molecular diagnostics during the COVID-19 pandemic. Afr. J. Lab. Med. 2022, 11, 1429. [Google Scholar] [CrossRef] [PubMed]
  28. Elnifro, E.M.; Ashshi, A.M.; Cooper, R.J.; Klapper, P.E. Multiplex PCR: Optimization and application in diagnostic virology. Clin. Microbiol. Rev. 2000, 13, 559–570. [Google Scholar] [CrossRef] [PubMed]
  29. Lai, C.K.C.; Lam, W. Laboratory testing for the diagnosis of COVID-19. Biochem. Biophys. Res. Commun. 2021, 538, 226–230. [Google Scholar] [CrossRef] [PubMed]
  30. Laboratory Testing for Coronavirus Disease 2019 (COVID-19) in Suspected Human Cases: Interim Guidance. 2 March 2020. Available online: https://apps.who.int/iris/handle/10665/331329 (accessed on 29 May 2023).
Figure 1. An overview of quality management systems required for laboratory diagnostics applied in patient care and the key role of laboratory information systems. This diagram highlights South Africa’s spread of care facilities (orange) across the nine provinces (listed on the South African map). The purple shading reflects population density at a district (municipality) level. The names of the nine provinces are listed on the map. Each instrument placed in the field within the NHLS is connected to a LIS (TrakCare), with data extracted, transformed and loaded and stored within a central data warehouse (CDW). This is then accessible for post-market surveillance and in the case of this study for continuous quality monitoring of the Ct values obtained from each test performed.
Figure 1. An overview of quality management systems required for laboratory diagnostics applied in patient care and the key role of laboratory information systems. This diagram highlights South Africa’s spread of care facilities (orange) across the nine provinces (listed on the South African map). The purple shading reflects population density at a district (municipality) level. The names of the nine provinces are listed on the map. Each instrument placed in the field within the NHLS is connected to a LIS (TrakCare), with data extracted, transformed and loaded and stored within a central data warehouse (CDW). This is then accessible for post-market surveillance and in the case of this study for continuous quality monitoring of the Ct values obtained from each test performed.
Diagnostics 13 02554 g001
Figure 2. Proportion of molecular SARS-CoV-2 testing performed by NHLS laboratories between March 2020 and March 2023 across the six commonly used laboratory testing systems (accounting for 94.2% NHLS test results). The testing proportions are stratified by province and conditional formatting applied as colored bars.
Figure 2. Proportion of molecular SARS-CoV-2 testing performed by NHLS laboratories between March 2020 and March 2023 across the six commonly used laboratory testing systems (accounting for 94.2% NHLS test results). The testing proportions are stratified by province and conditional formatting applied as colored bars.
Diagnostics 13 02554 g002
Figure 3. The number of Allplex™ SARS-CoV-2 tests (daily) reporting the presence of SARS-CoV-2 in red bars between March 2020 and March 2023 (primary vertical axis) overlaid with the median Ct values from each gene target in line plots. The median Ct value of the 28-day centered moving average for each gene target is represented on the secondary vertical axis. The key highlights the assay specific gene targets. A change in performance of the RdRp-gene target median Ct is visible over July 2021 (circled).
Figure 3. The number of Allplex™ SARS-CoV-2 tests (daily) reporting the presence of SARS-CoV-2 in red bars between March 2020 and March 2023 (primary vertical axis) overlaid with the median Ct values from each gene target in line plots. The median Ct value of the 28-day centered moving average for each gene target is represented on the secondary vertical axis. The key highlights the assay specific gene targets. A change in performance of the RdRp-gene target median Ct is visible over July 2021 (circled).
Diagnostics 13 02554 g003
Figure 4. The number of TaqPath™ COVID-19 tests (daily) reporting the presence of SARS-CoV-2 in red bars between March 2020 and March 2023 (primary vertical axis) overlaid with the median Ct values from each gene target in line plots. The median Ct value of the 28-day centered moving average for each gene target is represented on the secondary vertical axis. The key highlights the assay specific gene targets. The impact of the B.1.1.7 (wave 3) and B.1.1.529 (waves 4 and 5) in causing the SFTG is evident by the S-gene median Ct diverging from the ORF1ab and N-gene’s Ct (circles).
Figure 4. The number of TaqPath™ COVID-19 tests (daily) reporting the presence of SARS-CoV-2 in red bars between March 2020 and March 2023 (primary vertical axis) overlaid with the median Ct values from each gene target in line plots. The median Ct value of the 28-day centered moving average for each gene target is represented on the secondary vertical axis. The key highlights the assay specific gene targets. The impact of the B.1.1.7 (wave 3) and B.1.1.529 (waves 4 and 5) in causing the SFTG is evident by the S-gene median Ct diverging from the ORF1ab and N-gene’s Ct (circles).
Diagnostics 13 02554 g004
Figure 5. The number of Xpert® Xpress SARS-CoV-2 tests (daily) reporting the presence of SARS-CoV-2 in red bars between March 2020 and March 2023 (primary vertical axis) overlaid with the median Ct values from each gene target in line plots. The median Ct value of the 28-day centered moving average for each gene target is represented on the secondary vertical axis. The key highlights the assay specific gene targets. A new specimen result reporting algorithm was implemented in September 2021 (dotted vertical line). The effect of the XBB.1.5. is evident in the last few months of 2023 (circle), which causes a delayed hybridization in the E-gene PCR.
Figure 5. The number of Xpert® Xpress SARS-CoV-2 tests (daily) reporting the presence of SARS-CoV-2 in red bars between March 2020 and March 2023 (primary vertical axis) overlaid with the median Ct values from each gene target in line plots. The median Ct value of the 28-day centered moving average for each gene target is represented on the secondary vertical axis. The key highlights the assay specific gene targets. A new specimen result reporting algorithm was implemented in September 2021 (dotted vertical line). The effect of the XBB.1.5. is evident in the last few months of 2023 (circle), which causes a delayed hybridization in the E-gene PCR.
Diagnostics 13 02554 g005
Figure 6. The number of cobas® SARS-CoV-2 tests (daily) reporting the presence of SARS-CoV-2 in red bars between March 2020 and March 2023 (primary vertical axis) overlaid with the median Ct values from each gene target in line plots. The median Ct value of the 28-day centered moving average for each gene target is represented on the secondary vertical axis. The key highlights the assay specific gene targets. Both gene target Ct curves perfectly mirror each other throughout the testing period.
Figure 6. The number of cobas® SARS-CoV-2 tests (daily) reporting the presence of SARS-CoV-2 in red bars between March 2020 and March 2023 (primary vertical axis) overlaid with the median Ct values from each gene target in line plots. The median Ct value of the 28-day centered moving average for each gene target is represented on the secondary vertical axis. The key highlights the assay specific gene targets. Both gene target Ct curves perfectly mirror each other throughout the testing period.
Diagnostics 13 02554 g006
Figure 7. The number of RealTime SARS-CoV-2 (a) and ALINITY m SARS-CoV-2 (b) tests (daily) reporting the presence of SARS-CoV-2 in red bars between March 2020 and March 2023 (primary vertical axis) overlaid with the median Ct values from each gene target in line plots. The median Ct value of the 28-day centered moving average for each gene target is represented on the secondary vertical axis. The key highlights the assay specific gene targets, which for these assays, both targets are reflected in a single fluorescent channel (visualized as blue) and cannot be differentiated. Testing on the RealTime SARS-CoV-2 discontinued in July 2022.
Figure 7. The number of RealTime SARS-CoV-2 (a) and ALINITY m SARS-CoV-2 (b) tests (daily) reporting the presence of SARS-CoV-2 in red bars between March 2020 and March 2023 (primary vertical axis) overlaid with the median Ct values from each gene target in line plots. The median Ct value of the 28-day centered moving average for each gene target is represented on the secondary vertical axis. The key highlights the assay specific gene targets, which for these assays, both targets are reflected in a single fluorescent channel (visualized as blue) and cannot be differentiated. Testing on the RealTime SARS-CoV-2 discontinued in July 2022.
Diagnostics 13 02554 g007
Table 1. Description of the six commonly used SARS-CoV-2 open and closed testing platforms within NHLS laboratories.
Table 1. Description of the six commonly used SARS-CoV-2 open and closed testing platforms within NHLS laboratories.
Assay Name
(Manufacturer)
Platform (Type)Gene
Target(s)
Ct Range 2
Allplex SARS-CoV-2
(SeeGene Inc, Seoul, Republic of Korea)
BioRad CFX96Touch, Applied Biosystems (open)RdRp, N, E, S 16–40
TaqPath COVID-19
(Thermo Fisher Scientific, Waltham, MA, USA)
Applied Biosystems (open)N, Orf1ab, S5–43
Cobas SARS-CoV-2
(Roche Molecular, Pleasanton, CA, USA)
Cobas 6800/8800 (closed)E, Orf1ab12–42
Xpert Xpress SARS-CoV-2
(Cepheid, Sunnyvale, CA, USA)
GeneXpert (GX) 1, 4, 16 or Infinity-48/80 (closed)N2, E11–45
RealTime SARS-CoV-2 3
(Abbott Molecular, Abbott Park, IL, USA)
m2000sp and m2000rt (closed)RdRp, N2–31
Alinity m SARS-CoV-2 AMP Kit (Abbott Molecular, Abbott Park, IL, USA) 3Alinity m system (closed)RdRp, N5–42
1 version 2 included S-gene targets but in the same fluorescent channel as RdRp. 2 Ranges as captured through the LIS and aggregated within the CDW environment. These Ct ranges are established by the assay manufacturers, and specific to each assay, and several available in the assay package insert. The LIS reports these ranges as they are drawn from each assays’ LIS interfaced file. 3 The two Abbott assays report their gene targets in a single fluorescence channel.
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Scott, L.E.; Hsiao, N.-y.; Dor, G.; Hans, L.; Marokane, P.; da Silva, M.P.; Preiser, W.; Vreede, H.; Tsoka, J.; Mlisana, K.; et al. How South Africa Used National Cycle Threshold (Ct) Values to Continuously Monitor SARS-CoV-2 Laboratory Test Quality. Diagnostics 2023, 13, 2554. https://doi.org/10.3390/diagnostics13152554

AMA Style

Scott LE, Hsiao N-y, Dor G, Hans L, Marokane P, da Silva MP, Preiser W, Vreede H, Tsoka J, Mlisana K, et al. How South Africa Used National Cycle Threshold (Ct) Values to Continuously Monitor SARS-CoV-2 Laboratory Test Quality. Diagnostics. 2023; 13(15):2554. https://doi.org/10.3390/diagnostics13152554

Chicago/Turabian Style

Scott, Lesley Erica, Nei-yuan Hsiao, Graeme Dor, Lucia Hans, Puleng Marokane, Manuel Pedro da Silva, Wolfgang Preiser, Helena Vreede, Jonathan Tsoka, Koleka Mlisana, and et al. 2023. "How South Africa Used National Cycle Threshold (Ct) Values to Continuously Monitor SARS-CoV-2 Laboratory Test Quality" Diagnostics 13, no. 15: 2554. https://doi.org/10.3390/diagnostics13152554