Dose rate and dose robustness for proton transmission FLASH-RT treatment in lung cancer

Shouyi WeiShouyi Wei1Haibo LinHaibo Lin1Sheng HuangSheng Huang1Chengyu ShiChengyu Shi2Weijun XiongWeijun Xiong1Huifang ZhaiHuifang Zhai1Lei HuLei Hu1Gang YuGang Yu1Robert H. PressRobert H. Press1Shaakir HasanShaakir Hasan1Arpit M. ChhabraArpit M. Chhabra1J. Isabelle ChoiJ. Isabelle Choi1Charles B. Simone IICharles B. Simone II1Minglei Kang*Minglei Kang1*
  • 1New York Proton Center, New York, NY, United States
  • 2City of Hope, Orange County, Irvine, CA, United States

Purposes: To evaluate the plan quality and robustness of both dose and dose rate of proton pencil beam scanning (PBS) transmission FLASH delivery in lung cancer treatment.

Methods and materials: An in-house FLASH planning platform was used to optimize 10 lung cancer patients previously consecutively treated with proton stereotactic body radiation therapy (SBRT) to receive 3 and 5 transmission beams (Trx-3fds and Trx-5fds, respectively) to 34 Gy in a single fraction. Perturbation scenarios (n=12) for setup and range uncertainties (5 mm and 3.5%) were introduced, and dose-volume histogram and dose-rate-volume histogram bands were generated. Conventional proton SBRT clinical plans were used as a reference. RTOG 0915 dose metrics and 40 Gy/s dose rate coverage (V40Gy/s) were used to assess the dose and dose rate robustness.

Results: Trx-5fds yields a comparable iCTV D2% of 105.3%, whereas Trx-3fds resulted in inferior D2% of 111.9% to the clinical SBRT plans with D2% of 105.6% (p<0.05). Both Trx-5fds and Trx-3fds plans had slightly worse dose metrics to organs at risk than SBRT plans. Trx-5fds achieved superior dosimetry robustness for iCTV, esophagus, and spinal cord doses than both Trx-3fds and conventional SBRT plans. There was no significant difference in dose rate robustness for V40Gy/s coverage between Trx-3fds and Trx-5fds. Dose rate distribution has similar distributions to the dose when perturbation exists.

Conclusion: Transmission plans yield overall modestly inferior plan quality compared to the conventional proton SBRT plans but provide improved robustness and the potential for a toxicity-sparing FLASH effect. By using more beams (5- versus 3-field), both dose and dose rate robustness for transmission plans can be achieved.

Introduction

Currently, in proton FLASH radiotherapy, single-energy transmission beams are the most feasible solution without any significant beamline modifications. In a cyclotron system, a very high beam current can be achieved by bypassing the energy degraders in the beamline, allowing for ultra-high dose rates and the potential for a FLASH effect (13). As proton beams shoot through the body instead of being stopped as in the conventional proton treatment plans, the dosimetric performances are also expected to be very different. Several studies have investigated proton pencil beam scanning (PBS) transmission planning in the lung (1, 2), head and neck (4, 5), brain (6), and liver patients (7, 8). All of those studies have shown promising dosimetric outcomes using transmission beams, which helps to inform the characteristics of proton transmission beams for treatment planning. The first human clinical trials, FAST-01 for bone metastases, used PBS single-beam to deliver 8 Gy/fraction for a group of 10 enrolled patients (9). The planned FAST-02 human trial for thoracic bone metastases will similarly use transmission beams, which are being considered in several future clinical trials (10). Thus, the robustness of dose and dose rate of transmission delivery are essential characteristics for making informative decisions regarding the FLASH effect.

Another critical aspect of FLASH transmission planning is evaluating FLASH dose rate coverage, especially in the organs at risk (OARs). Currently, there is no universal way of quantifying dose rate in proton PBS because it is unclear whether average or instantaneous dose rate is more relevant to the FLASH effect, and since the spot scanning procedure further confounds the dose rate definition. Previously, several groups have formed unique frameworks for studying the proton PBS dose rate in their transmission plans (2, 5, 11). Among the different methods, the average dose rate defined by the Folkerts et al. (11) is associated with time and is a relatively more conservative dose rate metric than others (2).

Lung cancer treatments have perhaps the highest toxicity of all sites treated and thus the most significant potential benefits from FLASH-RT (1215). And the ultrafast dose delivery (beam time <1 s) may make the motion-related challenges less critical compared to conventional dose rate PBS (16). Previous in vivo FLASH experimental results showed evidence that FLASH-RT reduced the occurrence of lung fibrosis (17), spared lung progenitor cells, and limited the incidence of radio-induced senescence (18). Thus far, two studies have investigated the FLASH lung transmission plans (1, 2), both under stereotactic body radiation therapy (SBRT) dose fractionations. While these studies indicated current machine settings are capable of delivering transmission plans that achieve better quality than photon VMAT plan quality and high OAR FLASH dose rate coverage, they did not elucidate important planning considerations, especially in terms of the number of beams and beam arrangement, which may further affect the plan quality and FLASH dose rate coverage.

Moreover, plan robustness is critical in conventional proton RT plans using the Bragg peaks; however, as transmission plans use the plateau region that is least affected by range uncertainties, the plan robustness could become superior (16). Nevertheless, robustness in proton transmission FLASH plans has not been reported to date to our knowledge. This study aims to provide the first investigation to evaluate the sensitivity to the plan parameters, such as the number of beams and beam arrangement, as well as setup and range uncertainties, in terms of both dose and dose rate outcomes. Lastly, as dosimetric differences between single-energy FLASH transmission plans and conventional multi-energy intensity-modulated proton therapy (IMPT) plans have not been well characterized, we expect a comparison study might also provide a necessary reference to the future clinical practice for transmission FLASH planning. To achieve this goal, we optimized lung FLASH transmission plans with a varying number of beams and beam arrangements to assess the plan robustness for both dose and dose rate in this study. The conventional IMPT plans were used as references to better understand the plan quality of transmission planning.

Methods and materials

Treatment planning and dose rate quantification

The dose rate quantification was implemented into an in-house tool using the average dose rate (ADR) (2, 11), with a minimum spot time of 2 ms and a scanning speed of 10 mm/ms between adjacent spots. The beam delivery is presumed under a spot scanning mode, and the spot dwelling time and the scanning time between spots are both considered to calculate the averaged field dose rate. The details of the dose rate implementation and relations between beam currents and dose rates have previously been described (2).

Ten consecutive lung patients previously treated with proton SBRT at our institute were selected for this study which was approved by an internal research board (IRB). The tumor size ranged from 24.4-194.4 cm3, and the median size was 86.65 cm3. An iCTV was used to account for the respiratory motion based on the 4D CT images (2). A margin of 8 mm was isotropically extended from iCTV for PTV as a field target to place the spots for plan optimization. In transmission planning, only the flat region of a Bragg peak curve will be used to deposit doses to targets, and multiple-field optimization (MFO) with beams shooting through different angles is essential to achieve dosimetric conformity. The inverse planning system will provide the planning freedoms such as beam angles, spot weightings, spot placement, etc., to reach the planning goals and dose constraints. Thus, beam arrangement is an important factor in determining the plan quality and robustness. In theory, more fields with equal separations will help to improve the dosimetric distribution. In reality, the delivery efficiency will not allow an unlimited number of fields; according to the previous studies (2, 5, 6, 16, 19), 3-7 fields are sufficient to achieve optimal dose and dose rate metrics for target and OARs. Here, clinically possible beam arrangements with 5-field (Trx-5fds) and 3-field (Trx-3fds) beam arrangements were studied.

The transmission beams deliver dose via the plateau region rather than the Bragg peaks, the plan quality is less sensitive to range uncertainties and setup errors compared to the Bragg peak plans. Therefore, our current in-house transmission FLASH planning system includes MFO but not robustness optimization. The in-house planning platform based on the matRad framework uses a pencil beam convolution superposition (PCS) for dose calculation (20), and the dose grid was 1 mm. Eclipse treatment planning system used PCS version 16.1 as the dose calculation engine, and a 2.5 mm dose grid was used. The conventional multiple-energy MFO plans were generated using robust optimization in Eclipse, which uses worst-case scenario robust optimization (21). Twelve perturbation scenarios with a combination of 6 setup shifts of 5 mm in cardinal axes (± 5 mm) and ±3.5% CT HU to RSP (relative stopping power) conversion uncertainties were assessed (22). Twelve different dose distributions were calculated based on twelve perturbations. The objective function value for each iteration was calculated based on the worst-case dose result. Conventional multiple-energy proton IMPT plans planned by professional dosimetrists and reviewed by radiation oncologists served as references to the transmission FLASH plans. Following our current clinical practice, the global maximum dose is less than 115% of the prescribed doses for both FLASH and conventional plans.

Plan quality, dose rate, and robustness analysis

Dosimetric performances were evaluated following the RTOG 0915 guidelines for lung treatment planning. The dosimetric metrics of CTV D2%, spinal cord D1.2cc, D0.35 cc, Dmax, lung-GTV V7Gy, V7.4Gy, and heart D15cc, Dmax were evaluated. The V40Gy/s was used to quantify the dose rate performance for FLASH plans, which indicates the voxels ratio that received doses with a dose rate > 40 Gy/s in OARs and target structures. A dose rate volume histogram (DRVH) was also used to evaluate the dose rate distribution vs. volume for a region of interest (2).

Similarly, a clinically applied robustness evaluation method was implemented for dose and dose rate robustness analysis based on the twelve perturbation scenarios. The nominal plans were normalized to 100% prescribed dose covering 95% target volume. The conventional second worst-case and median scenarios for the D95% were employed to evaluate plan robustness. The max-min dose distribution is the maximum dose differences, i.e., maximum minus minimum dose values, among the 12 perturbations from the nominal plans. To visualize the dose rate volume histograms with uncertainty bands, the bandwidth (BW) of a dose rate metric was used to evaluate the robustness, with an example shown below:

BWV40Gy/s =V40Gy/s, upper boundV40Gy/s, lower bound(1)

Here the bandwidth of V40Gy/s is given by the difference between the upper and lower bounds of V40Gy/s from the 12 perturbation scenarios of DRVHs. Similar to the max-min dose distribution, the max-min dose rate distribution was also calculated to identify the locations where the dose rate has the most uncertainties with perturbation existing.

Results

Plan quality assessment - dosimetric performances

Two patients were selected to demonstrate the differences in the treatment plans of the transmission and multiple-energy PBS plans in terms of dose distribution and DVHs (Figure 1). Patient #1 has a tumor located on the right mid lung, and the results are shown in Figures 1A–D, and the 2D dose distributions for different types of plans are visually very different. As displayed in Figures 1A, B a transmission beam can completely shoot through the body, the Bragg peak is outside of the patient, and all of the normal tissue and OARs in the beam path receive similar doses to the target for a given beam. Multiple-field shooting from different angles can overlap at the target to give the target the full prescribed doses and to increase conformality. In contrast, the OARs beyond the target only receive a relatively small fraction of the prescribed dose. Compared to conventional multiple-energy proton PBS plans, transmission plans generate dose spillage to the heart and healthy lung tissues beyond the target in the beam paths. As the Trx-5fds plans used equal beam separation of 72 degrees, the dose spillage is spread equally to the body. While in the Trx-3fds, less low dose spillage to the right lung is achieved compared to the 5-field arrangement as displayed in the DVHs. However, as the prescription doses are the same, the average dose per beam in the Trx-3fds plans tends to be higher than that of the Trx-5fds plans, which increases the doses in healthy tissues irradiated by the Trx-3fds beams. In contrast, the conventional IMPT plans use multiple-energy Bragg peaks to deliver conformable doses to the target, and the dose spillage to normal tissues is much less than transmission plans. Similar planning outcomes can be observed for patient #2 (Figures 1E–H), whose tumor is located in the left anterior lung close to the chest wall. Given the tumor’s location, 3 beams deliver less dose to normal tissues and can completely spare the contralateral lung relatively to 5-field plans. The 5-beam plan, however, splits the doses to a large area of tissue and achieves better target conformity.

FIGURE 1
www.frontiersin.org

Figure 1 The 2D dose distributions and DVH comparisons for 2 selected patients. (A, E) Trx-5fds plan 2D dose distribution, (B, F) Trx-3fds plan 2D dose distribution, (C, G) conventional PBS plan dose distribution, (D, H) DVHs of the Trx-5fds, Trx-3fds, and conventional IMPT-SBRT plans.

The statistics of the dose performances of the 10 lung patients are summarized in Table 1. All dose metrics were averaged for treatment plans of Trx-3fds, Trx-5fds, and conventional PBS plans, respectively. Trx-5fds yields comparable target uniformity, while Trx-3fds results in inferior (p<0.05) target uniformity compared to the conventional IMPT plans. Trx-3fds achieve improved results in most OARs dose metrics, including the spinal cord, lung-GTV, and esophagus, than Trx-5fds. The dose metrics of the heart are comparable using 3- and 5-field arrangements in FLASH planning. Trx-5fds results in worse dose performances in all OARs except for the heart than the IMPT plans. While still modestly inferior, the Trx-3fds dose metrics are improved to be closer to the IMPT plans relative to the Trx-5fds dose metrics.

TABLE 1
www.frontiersin.org

Table 1 Summary of dose metrics of RT structures in Trx-5fds, Trx-3fds and IMPT plans.

Plan quality assessment - dose rate performances

The 2D averaged dose rate distribution, and DRVHs of Trx-3fds plans for the two lung patients’ plans are shown in Figures 2A–H. The field edges give higher dose rate distributions consistent with the findings in (2, 11) for both patient cases. The streak-like patterns in the dose rate distributions also reflect the PBS spot scanning patterns for the ADR calculation methods, consistent with (2, 11). For the DRVHs, as observed for patient #1, the FLASH coverage can be as high as 80% for most ROIs, as the tumor location is close to the spinal cord and esophagus, which receive very high exit dose with high FLASH dose rate coverage. For patient #2, the FLASH coverages in the heart and spinal cord drop, reflecting that only the field penumbrae cover such RT structures. Although they correspond to a large dose rate at the edges, most volume beyond the hot edges into the penumbrae falls to lower dose rate values, causing the overall drop in the FLASH coverages. The statistics of the V40Gy/s of the 10 lung patients are summarized in Figure 3. Except for the spinal cord, the differences in V40Gy/s in all the other RT structures are comparable between the Trx-5fds and Trx-3fds cases.

FIGURE 2
www.frontiersin.org

Figure 2 (A–C) Dose rate distributions for beam 1, 2 and 3 for patient #1, (D) DRVHs of the Trx-3fds plan, (E–G) dose rate distributions for beam 1, 2 and 3 for patient #2, (H) DRVHs for the Trx-3fds plan.

FIGURE 3
www.frontiersin.org

Figure 3 V40Gy/s statistics of 10 patients for target and OARs. The dots represent each patient, the line inside the box represents the mean value, and the error bars outside the box represent the 25-75th percentile of the scattered values.

Plan does robustness assessment

The max-min dose distributions between the perturbation plans and the initial plan were calculated for Trx-5fds, Trx-3fds, and conventional IMPT plans, as well as the DVH bands for each plan, are shown in Figure 4 for the 2 selected patients. As the transmission plans place Bragg peaks outside of the patient’s body, most of the uncertainty in the dose distribution arises from the edge of the fields, especially at the rim of the target where the fields overlap. The max-min uncertainties distribute more evenly around targets in 5-field plans than 3-field plans. While in conventional IMPT plans, as Bragg peaks are used to deliver doses to the targets, the uncertainty occurs at the lateral field edge and the distal edges. We can also observe from these two patient cases that the robustness achieved with Trx-5fds is better than Trx-3fds and conventional PBS plans in the target DVH bands (Figures 4B, D, F, H, J, L), with narrower CTV bands indicated.

FIGURE 4
www.frontiersin.org

Figure 4 Max-min dose distributions from 12 perturbations for Trx-5fds (A, G), Trx-3fds (C, I) and conventional PBS plans (E) and (K) for patient 1 and 2. DVH bands from 12 perturbations for Trx-5fds (B, H), Trx-3fds (D, J) and conventional IMPT plans (F, L).

The second worst-case scenario, media, and max of D95% were investigated as indicators for the robustness assessment. The statistics of the CTV D95% of the different plans for the 10 patients are presented in Figure 5. Trx-5fds result in the narrowest distribution with the least CTV D95% variation compared to the other two types of plans. Trx-3fds result in a worse distribution in D95% for the perturbations than conventional IMPT plans in general but are presented with fewer outliners in this case. The better robustness performance of Trx-5fds can benefit from the beam arrangement and number of beams, i.e., 5 fields are evenly distributed 72 degrees apart, thereby minimizing the setup errors effect in different directions. As Trx-3fds have fewer fields and the angles are optimized to avoid lung, heart, and other critical organs, the uncertainties from setup errors become more significant than those of Trx-5fds. For the median and max of D95%, all 3 types of plans achieve relatively comparable results.

FIGURE 5
www.frontiersin.org

Figure 5 The D95% of iCTV statistics distributions for the 10 patients for the second-worst cases, median and max values. The dots represent each patient, the line inside the box represents the mean value, and the error bars outside the box represent the 25-75th percentile of the scattered values.

Plan dose rate robustness evaluation

Similar to the max dose uncertainties distribution, the max dose rate distributions between the perturbation plans and the initial plan were calculated for Trx-5fds and Trx-3fds. As shown in Figure 6, the largest uncertainty in dose rate from the 12 perturbation scenarios arising from the field edges, which has a similar pattern to the dose case as shown in Figure 4.

FIGURE 6
www.frontiersin.org

Figure 6 Max-min dose rate distributions in 2D from 12 different perturbation scenarios for lung patient #1 (A) and patient #2 (B) were demonstrated by the Trx-3fds plans.

The DRVH bands for each OARs are shown in Figures 7, 8 for 2 selected patients. The iCTV and lung-GTV bands are very narrow for both patients, with bandwidths of V40Gy/s less than 2%. Large variations from perturbations can be consistently observed in the spinal cord and heart for both 3-field and 5-field transmission plans, partially covered by the transmission beams in the penumbra regions. Using 3-field can completely avoid the esophagus, thus, there are no DRVHs for 3-field plans shown in Figures 7E, 8E. Another significant difference between the 2 patients can be observed from the spinal cord DRVH for 3-field transmission plans. Patient #1 has a much narrower dose rate uncertainty in the spinal cord than that patient #2. This can be explained as the main portion of beams in patient #1 shooting through the spinal cord while the penumbra regions of the beams shoot through the spinal cord for patient #2.

FIGURE 7
www.frontiersin.org

Figure 7 DRVH bands for different RT structures for Trx-3fds and Trx 5fds for patient #1, target iCTV (A) and OARs spinal cord, lung-GTV, heart, and esophagus (B–E).

FIGURE 8
www.frontiersin.org

Figure 8 DRVH bands for different RT structures for Trx-3fds and Trx 5fds for patient #2, target iCTV (A) and OARs spinal cord, lung-GTV, heart, and esophagus (B–E).

Figure 9 is the V40Gy/s robustness of 10 patients of different RT structures. The V40Gy/s BWs for iCTV and lung-GTV for both Trx-3fds and Trx-5fds plans are < 2% on average, although those of Trx-5fds plans appear to be slightly narrower. V40Gy/s BWs vary differently for the spinal cord, esophagus, and heart, which tend to have more considerable dose rate variations when perturbations exist. This phenomenon is mainly due to the relative location of these OARs to the target, causing it to be in the middle of beam paths or penumbrae. In general, Trx-5fds appears to be slightly more dose rate robust than Trx-3fds, with ~1%, 2%, and 5% on average in the heart, spinal cord, and esophagus, respectively.

FIGURE 9
www.frontiersin.org

Figure 9 V40Gy/s robustness of 10 patients of different RT structures. The dots represent each of the 10 patients, the line inside the box represents the mean value, and the error bars outside and the box represent the 25-75th percentile of the scattered values.

Discussion

This study simulates realistic planning settings in FLASH treatment planning characterized with feasibility in machine delivery based on beam current, energy, and minimum MU/spot, and practical planning considerations characterized by practical dose prescription and beam arrangement, including number and angles. We further assessed different plan quality regarding dose and dose rate performances as well as robustness in comparison with conventional IMPT plans using Bragg peaks with multiple energy layers.

The in-house planning platform allows inverse planning and, therefore, improves the transmission plans’ quality given the OAR doses and target uniformity. Despite the FLASH effect, which is currently the most standard approach, this study still considers the dose metrics imposed by conventional dose rate plans for the transmission plans. While this could be subject to changes in the future as more biological or clinical trial evidence of FLASH is to be uncovered, right now, assuming the same dose constraints may enable us to evaluate the “lower bound” of the FLASH plans’ quality, providing benchmarks for future clinical planning practices.

Another significant and novel contribution of this work is that we evaluated the robustness of the transmission plans in terms of both dose and dose rate. Previously, van Marlen et al.’s study (1) indicated that transmission plans using plateau regions do not involve the range uncertainties as in conventional Bragg peak-based plans and, therefore, can achieve better robustness. Our results show that with 5 transmission beams, better plan robustness can be achieved in target coverage. Still, with 3 beams, such robustness can be degraded to a similar or slightly worse level than the conventional IMPT. One thing to note is that our current in-house FLASH planning system does not include robustness optimization, once implemented, which may further improve the robustness of transmission plans. The value of robustness optimization in FLASH planning requires further investigation. For dose rate robustness, the uncertainty of setup and range perturbations in V40Gy/s is especially small (< 3% for lung-GTV). The results vary for other OARs such as the esophagus, heart, and spinal cord, given their relative anatomical location to the target. For example, we expect the larger differences in the heart, spinal cord, and esophagus since we tend to spare these OARs with the beam penumbras that may have larger dose gradients. Such penumbra regions also correspond to lower doses that may not be able to trigger the FLASH effect, given the recent biological observations that a dose threshold > 5 Gy may need to be met (2327). The lack of dose rate robustness and low dose deposition will require careful thought in future treatment planning practice for FLASH, especially when OARs are present in such regions. Beam numbers and angles are critical parameters for treatment planning optimization, which exert a crucial impact on the planning outcomes. This study indicates that more transmission beams might reduce the sensitivity to setup errors to achieve a better dose rate robustness. Due to the nature of the transmission beams, when more beams are used to deliver target doses, the low dose spillage may be increased to OARs, and the tradeoff between dose rate robustness and low dose spillage should be balanced in designing transmission FLASH plans.

Similar to the photon beams, the transmission FLASH always has the exit dose. PBS field employs scanning spots to deliver the prescribed dose. The depth dose of a broad PBS field is very different from the photon beam that follows an exponential attenuation beyond the buildup region. The depth dose of the plateau region slightly increases with depth at the first 30 cm in water for a 250 MeV proton beam. Considering the beam divergence and scattering effect, the entrance dose of the proton beam is very close to that of the exit dose. In contrast, the entrance dose is always much higher than the exit dose beyond the maximum dose depth for a photon beam.

We chose the same minimum MU/spot of 400 for both Trx-5fds and Trx-3fds plans, which resulted in around 80% OARs FLASH coverage in most cases. The ADR was used for quantifying the FLASH dose rate coverage in our treatment plans, which is a conservative definition in proton PBS planning (2). Previous studies also demonstrated that when using ADR, the FLASH sparing effect can be observed in the skin of mice at a dose rate of ~ 65 Gy/s (28). The minimum MU/spot choice could be flexible given the machine’s delivery limit, especially the beam current and allowed minimum spot time. Here we chose a feasible cyclotron current for the Varian ProBeam system, and under this current, the FLASH coverage for both plans is close. We demonstrated earlier in (7, 8) that a less minimum spot time in the FLASH spot map delivery will allow smaller MUs in the treatment plans at the fixed beam current, likely resulting in superior plan quality to ones with greater minimum spot times.

Motion is an important factor for lung treatment plans, especially given the large fractionation doses in SBRT (2931). For FLASH plans, as the delivery of the spot maps is reduced to be less than one second, the motion within a fraction is minimal, but an accurate coverage of the dose in the target requires timely monitoring of the phase of the motion, thereby posing a particular challenge in FLASH delivery for moving targets. Nevertheless, the time intervals between beams are not trivial, giving rise to motion-related uncertainties in the treatment. Our study used iCTV methods to generate a more accurate target volume for treatment planning. However, the interplay effect was not included in the treatment planning study. Multiple methods have been mentioned to reduce the motion-caused interplay effect, such as using an abdominal compression belt (32) or deep inspiration breath-hold (DIBH) (33, 34). To what extent the interplay will affect the dose and dose rate accuracy warrant more detailed investigation.

Our work has demonstrated the dosimetric advantages of clinical IMPT plans over transmission plans, with superior OAR doses, comparable target uniformity, and slightly inferior plan robustness in target coverage. Significant efforts have been made to bring the Bragg-peak-based plans for FLASH applications (16, 35), including the Spread-out of Bragg peak (SOBP) and single-energy Bragg-peak plans. While this may present new technical challenges in FLASH, we expect such approaches will fill the gaps between the transmission plans and clinical IMPT plans for better FLASH treatment outcomes.

Conclusion

Using the highest energy from the cyclotron system to deliver FLASH-RT is practical for clinical application. The MFO inversed optimization can achieve quality treatment plans for transmission beams. Transmission plans yield overall modestly inferior plan quality compared to the conventional proton SBRT plans but provide improved robustness with 5 beams while allowing for the potential for achieving a FLASH effect in this lung cancer study. By introducing more beams (5- versus 3-field), both dose and dose rate robustness for the transmission plans can be achieved.

Data availability statement

The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.

Ethics statement

The studies involving human participants were reviewed and approved by Western IRB. Written informed consent for participation was not required for this study in accordance with the national legislation and the institutional requirements.

Author contributions

Conceptualization, MK, SW, CBSII, and HL; methodology, MK and SW, software, SW; validation, MK and SW; formal analysis, SW; investigation, MK and SW; resources, JIC, CBSII, and HL; data curation, MK and HL; writing—original draft preparation, SW and MK; writing—review and editing, all authors.; visualization, MK and SW; supervision, CBSII and HL; project administration, HL; funding acquisition, HL. All authors have read and agreed to the published version of the manuscript.

Funding

The authors declare that this study received funding from the Varian Research Grant. The funder was not involved in the study design, collection, analysis, interpretation of data, the writing of this article or the decision to submit it for publication.

Acknowledgments

The authors would like to thank Dr. Michael Folkerts and Krystian Wrobel for valuable discussions and suggestions. This study was partly supported by Varian, a Siemens Healthineers Company.

Conflict of interest

HL, JIC, and CBSII report an honorarium from Varian Medical Systems. This study was partly funded by Varian, a Siemens Healthineers Company.

The remaining authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

References

1. van Marlen P, Dahele M, Folkerts M, Abel E, Slotman BJ, Verbakel WFAR. Bringing FLASH to the clinic: treatment planning considerations for ultrahigh dose-rate proton beams. Int J Radiat Oncol Biol Phys (2020) 106:621–9. doi: 10.1016/j.ijrobp.2019.11.011

PubMed Abstract | CrossRef Full Text | Google Scholar

2. Kang M, Wei S, Choi JI, Simone CB 2nd, Lin H. Quantitative assessment of 3D dose rate for proton pencil beam scanning FLASH radiotherapy and its application for lung hypofractionation treatment planning. Cancers (Basel) (2021) 13(14):3549. doi: 10.3390/cancers13143549

PubMed Abstract | CrossRef Full Text | Google Scholar

3. Wei S, Shi C, Chen C, Huang S, Press RH, Choi JI, et al. Recent progress in pencil beam scanning FLASH proton therapy: a narrative review. Ther Radiol Oncol (2022). doi: 10.21037/tro-22-1

CrossRef Full Text | Google Scholar

4. van Marlen P, Dahele M, Folkerts M, Abel E, Slotman BJ, Verbakel W. Ultra-high dose rate transmission beam proton therapy for conventionally fractionated head and neck cancer: Treatment planning and dose rate distributions. Cancers (2021) 13(8):1859. doi: 10.3390/cancers13081859

PubMed Abstract | CrossRef Full Text | Google Scholar

5. van de Water S, Safai S, Schippers JM, Weber DC, Lomax AJ. Towards FLASH proton therapy: The impact of treatment planning and machine characteristics on achievable dose rates. Acta Oncol (2019) 58:10:1463–9. doi: 10.1080/0284186X.2019.1627416

CrossRef Full Text | Google Scholar

6. Verhaegen F, Wanders R-G, Wolfs C, Eekers D. Considerations for shoot-through FLASH proton therapy. Phys Med Biol (2021) 66(6):06NT01. doi: 10.1088/1361-6560/abe55a

PubMed Abstract | CrossRef Full Text | Google Scholar

7. Wei S, Lin H, Choi JI, Press RH, Lazarev S, Kabarriti R, et al. FLASH radiotherapy using single-energy proton PBS transmission beams for hypofractionation liver cancer: Dose and dose rate quantification. Front Oncol (2022) 11:813063 doi: 10.3389/fonc.2021.813063

PubMed Abstract | CrossRef Full Text | Google Scholar

8. Wei S, Lin H, Shi C, Xiong W, Chen C, Huang S, et al. Use of single-energy proton pencil beam scanning Bragg peak for intensity-modulated proton therapy FLASH treatment planning in liver hypofractionated radiation therapy. Med Phys (2022). doi: 10.1002/mp.15894

PubMed Abstract | CrossRef Full Text | Google Scholar

9. ClinicalTrials.gov. Identifier NCT04592887, feasibility study of FLASH radiotherapy for the treatment of symptomatic bone metastases (FAST-01). Bethesda, MD, USA: National Library of Medicine (US (2020).

Google Scholar

10. FlashForward consortium (Accessed November 5, 2021).

Google Scholar

11. Folkerts M, Abel E, Busold S, Perez J, Krishnamurthi V, Ling CC. A framework for defining FLASH dose rate for pencil beam scanning. Med Phys (2020) 47:6396–404. doi: 10.1002/mp.14456

PubMed Abstract | CrossRef Full Text | Google Scholar

12. Simone CB 2nd, Rengan R. The use of proton therapy in the treatment of lung cancers. Cancer J (2014) 20(6):427–32. doi: 10.1097/PPO.0000000000000080

PubMed Abstract | CrossRef Full Text | Google Scholar

13. Simone CB 2nd. Thoracic radiation normal tissue injury. Semin Radiat Oncol (2017) 27(4):370–7. doi: 10.1016/j.semradonc.2017.04.009

PubMed Abstract | CrossRef Full Text | Google Scholar

14. Verma V, Simone CB 2nd, Werner-Wasik M. Acute and late toxicities of concurrent chemoradiotherapy for locally-advanced non-small cell lung cancer. Cancers (Basel) (2017) 9(9):120. doi: 10.3390/cancers9090120

CrossRef Full Text | Google Scholar

15. Chow R, Kang M, Wei S, Choi JI, Press RH, Hasan S, et al. FLASH radiation therapy: Review of the literature and considerations for future research and proton therapy FLASH trials. Appl Radiat Oncol (2021) 10:15–21.

Google Scholar

16. Kang M, Wei S, Choi JI, Lin H, Simone CB 2nd. A universal range shifter and range compensator can enable proton pencil beam scanning single-energy Bragg peak FLASH-RT treatment using current commercially available proton systems. Int J Radiat Oncol Biol Phys (2022) 113(1):203–13. doi: 10.1016/j.ijrobp.2022.01.009

PubMed Abstract | CrossRef Full Text | Google Scholar

17. Favaudon V, Caplier L, Monceau V, Pouzoulet F, Sayarath M, Fouillade C, et al. Ultrahigh dose-rate FLASH irradiation increases the differential response between normal and tumor tissue in mice. Sci Transl Med (2014) 6(245):245ra93. doi: 10.1126/scitranslmed.3008973

PubMed Abstract | CrossRef Full Text | Google Scholar

18. Fouillade C, Curras-Alonso S, Giuranno L, Quelennec E, Heinric S, Bonnet-Boissinot S, et al. FLASH irradiation spares lung progenitor cells and limits the incidence of radio-induced senescence. Clin Cancer Res (2020) 26:1497–1506. doi: 10.1158/1078-0432.CCR-19-1440

PubMed Abstract | CrossRef Full Text | Google Scholar

19. Gao H, Lin B, Lin Y, Fu S, Langen K, Liu T, et al. Simultaneous dose and dose rate optimization (SDDRO) for FLASH proton therapy. Med Phys (2020) 47:6388–95. doi: 10.1002/mp.14531

PubMed Abstract | CrossRef Full Text | Google Scholar

20. Wieser HP, Cisternas E, Wahl N, Ulrich S, Stadler A, Mescher H, et al. Development of the opensource dose calculation and optimization toolkit matrad. Med Phys (2017) 44:2556–68. doi: 10.1002/mp.12251

PubMed Abstract | CrossRef Full Text | Google Scholar

21. Liu W, Zhang X, Li Y, Mohan R. Robust optimization of intensity modulated proton therapy. Med Phys (2012) 39(2):1079–91. doi: 10.1118/1.3679340

PubMed Abstract | CrossRef Full Text | Google Scholar

22. Kang M, Hasan S, Press RH, Yu F, Abdo M, Xiong W, et al. Using patient-specific bolus for pencil beam scanning proton treatment of periorbital disease. J Appl Clin Med Phys (2021) 22(1):203–9. doi: 10.1002/acm2.13134

PubMed Abstract | CrossRef Full Text | Google Scholar

23. MacKay R, Burnet N, Lowe M, Rothwell B, Kirkby N, Kirkby K, et al. FLASH radiotherapy: Considerations for multibeam and hypofractionation dose delivery. Radiother Oncol (2021) 164:122–7. doi: 10.1016/j.radonc.2021.09.011

PubMed Abstract | CrossRef Full Text | Google Scholar

24. Adrian G, Konradsson E, Lempart M, Bäck S, Ceberg C, Petersson K. The FLASH effect depends on oxygen concentration. Br J Radiol (2020) 93(1106):20190702. doi: 10.1259/bjr.20190702

PubMed Abstract | CrossRef Full Text | Google Scholar

25. Wilson P, Jones B, Yokoi T, Hill H, Vojnovic B. Revisiting the ultrahigh dose rate effect: implications for charged particle radiotherapy using protons and light ions. Br J Radiol (2012) 85:e933–9. doi: 10.1259/bjr/17827549

PubMed Abstract | CrossRef Full Text | Google Scholar

26. Bourhis J, Montay-Gruel P, Jorge PG, Bailat C, Petit B, Ollivier J, et al. Clinical translation of FLASH radiotherapy: Why and how? radiother. Oncol (2019) 139:11–7. doi: 10.1016/j.radonc.2019.04.008

CrossRef Full Text | Google Scholar

27. Krieger M, van de Water S, Folkerts MM, Mazal A, Fabiano S, Bizzocchi N, et al. A quantitative FLASH effectiveness model to reveal potentials and pitfalls of high dose rate proton therapy. Med Phys (2022) 49(3):2026–38. doi: 10.1002/mp.15459

PubMed Abstract | CrossRef Full Text | Google Scholar

28. Cunningham S, McCauley S, Vairamani K, Speth J, Girdhani S, Abel E, et al. FLASH proton pencil beam scanning irradiation minimizes radiation-induced leg contracture and skin toxicity in mice. Cancers (Basel) (2021) 13(5):1012. doi: 10.3390/cancers13051012

PubMed Abstract | CrossRef Full Text | Google Scholar

29. Kang M, Huang S, Solberg TD, Mayer R, Thomas A, Teo BK, et al. A study of the beam-specific interplay effect in proton pencil beam scanning delivery in lung cancer. Acta Oncol (2017) 56(4):531–40. doi: 10.1080/0284186X.2017.1293287

PubMed Abstract | CrossRef Full Text | Google Scholar

30. Lin L, Kang M, Huang S, Mayer R, Thomas A, Solberg TD, et al. BSPTV with 4D CT for PBS proton therapy of thoracic tumors. J Appl Clin Med Phys (2015) 16(6):281–92. doi: 10.1120/jacmp.v16i6.5678

CrossRef Full Text | Google Scholar

31. Molitoris JK, Diwanji T, Snider JW 3rd, Mossahebi S, Samanta S, Onyeuku N, et al. Optimizing immobilization, margins, and imaging for lung stereotactic body radiation therapy. Transl Lung Cancer Res (2019) 8(1):24–31. doi: 10.21037/tlcr.2018.09.25

PubMed Abstract | CrossRef Full Text | Google Scholar

32. Lin L, Souris K, Kang M, Glick A, Lin H, Huang S, et al. Evaluation of motion mitigation using abdominal compression in the clinical implementation of pencil beam scanning proton therapy of liver tumors. Med Phys (2017) 44(2):703–12. doi: 10.1002/mp.12040

PubMed Abstract | CrossRef Full Text | Google Scholar

33. Keall PJ, Mageras G, Balter JM, Emery RS, Forster KM, Jiang SB, et al. The management of respiratory motion in radiation oncology report of AAPM task group 76a. Med Phys (2006) 33:3874–900. doi: 10.1118/1.2349696

PubMed Abstract | CrossRef Full Text | Google Scholar

34. Chang JY, Zhang X, Knopf A, Li H, Mori S, Dong L, et al. Consensus guidelines for implementing pencil-beam scanning proton therapy for thoracic malignancies on behalf of the PTCOG thoracic and lymphoma subcommittee. Int J Radiat Oncol (2017) 99:41–50. doi: 10.1016/j.ijrobp.2017.05.014

CrossRef Full Text | Google Scholar

35. Wei S, Lin H, Choi JI, Simone CB 2nd, Kang M. A novel proton pencil beam scanning FLASH RT delivery method enables optimal OAR sparing and ultra-high dose rate delivery: A comprehensive dosimetry study for lung tumors. Cancers (Basel) (2021) 13(22):5790. doi: 10.3390/cancers13225790

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: proton FLASH-RT, ultra-high dose rate, pencil beam scanning, hypofractionation, lung stereotactic body radiation therapy, robustness, transmission proton beam

Citation: Wei S, Lin H, Huang S, Shi C, Xiong W, Zhai H, Hu L, Yu G, Press RH, Hasan S, Chhabra AM, Choi JI, Simone CB 2nd and Kang M (2022) Dose rate and dose robustness for proton transmission FLASH-RT treatment in lung cancer. Front. Oncol. 12:970602. doi: 10.3389/fonc.2022.970602

Received: 16 June 2022; Accepted: 27 July 2022;
Published: 15 August 2022.

Edited by:

Xuanfeng Ding, William Beaumont Hospital, United States

Reviewed by:

Haijiao Shang, RaySearch Medical Device Co., LTD., China
Shikui Tang, Texas Center for Proton Therapy, United States

Copyright © 2022 Wei, Lin, Huang, Shi, Xiong, Zhai, Hu, Yu, Press, Hasan, Chhabra, Choi, Simone and Kang. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Minglei Kang, mleikang@gmail.com

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.