Skip to main content

ORIGINAL RESEARCH article

Front. Neurol., 31 May 2022
Sec. Endovascular and Interventional Neurology

Small and Medium-Sized Aneurysm Outcomes Following Intracranial Aneurysm Treatment Using the Pipeline Embolization Device: A Subgroup Analysis of the PLUS Registry

\nHongyun Zhang,&#x;Hongyun Zhang1,2Li Li,&#x;Li Li1,2Hongqi Zhang,Hongqi Zhang3,4Jianmin LiuJianmin Liu5Donglei SongDonglei Song6Yuanli ZhaoYuanli Zhao7Sheng GuanSheng Guan8Aisha MaimaitiliAisha Maimaitili9Yunyan WangYunyan Wang10Wenfeng FengWenfeng Feng11Yang WangYang Wang12Jieqing WanJieqing Wan13Guohua MaoGuohua Mao14Huaizhang ShiHuaizhang Shi15Bin LuoBin Luo16Qiuji Shao,Qiuji Shao1,2Kaitao Chang,Kaitao Chang1,2Qianqian Zhang,Qianqian Zhang1,2Yingkun He,Yingkun He1,2Peng Zhang,
Peng Zhang3,4*Xinjian Yang
Xinjian Yang16*Tian xiao Li,
Tian xiao Li1,2*
  • 1Cerebrovascular Department of Interventional Center, Henan Provincial People's Hospital, Zhengzhou University People's Hospital, Henan University People's Hospital, Zhengzhou, China
  • 2Neurosurgery Department of Stroke Center, Henan Provincial People's Hospital, Zhengzhou University People's Hospital, Henan University People's Hospital, Zhengzhou, China
  • 3Department of Neurosurgery, International Neuroscience Institute (China-INI), Xuanwu Hospital, Capital Medical University, Beijing, China
  • 4Department of Interventional Neuroradiology, Xuanwu Hospital, Capital Medical University, Beijing, China
  • 5Changhai Hospital Affiliated to Naval Medical University, Shanghai, China
  • 6Shanghai Donglei Brain Hospital, Shanghai, China
  • 7Peking University International Hospital, Beijing, China
  • 8First Affiliated Hospital of Zhengzhou University, Zhengzhou, China
  • 9First Affiliated Hospital of Xinjiang Medical University, Ürümqi, China
  • 10Qilu Hospital of Shandong University, Jinan, China
  • 11Nanfang Hospital, Southern Medical University, Guangzhou, China
  • 12First Affiliated Hospital of Nanchang University, Nanchang, China
  • 13School of Medicine, Renji Hospital, Shanghai Jiao Tong University, Shanghai, China
  • 14Second Affiliated Hospital of Nanchang University, Nanchang, China
  • 15First Affiliated Hospital of Harbin Medical University, Harbin, China
  • 16Beijing Neurosurgical Institute, Beijing Tiantan Hospital, Capital Medical University, Beijing, China

Objective: The purpose of this work was to summarize the real-world safety and efficacy of Pipeline Embolization Device (PED) therapy for small and medium-sized intracranial aneurysms in China.

Methods: Patients from the PED in China post-market multi-center registry study (PLUS) with aneurysms smaller than 12 mm were selected. Radiographic outcomes were assessed using digital subtraction angiography. Clinical outcomes included functional outcomes (modified Rankin Scale, MRS) in the early postoperative period ( ≤ 30 days) and early postoperative complications associated with PED therapy.

Results: A total of 652 patients with a combined 754 aneurysms were included in this study (mean age of 53.9 ± 10.3 years, 68.7% women). Mean aneurysm diameter was 6.78 ± 2.67 mm. Of the 687 stents deployed, 99.7% (685/689) were successfully deployed. In this study, 64.7% (488/754) of aneurysms were treated with only the PED, whereas 35.3% (266/754) were subjected to PED-assisted therapy. Radiographic outcome at the last follow-up (median time: seven months) was available for 64.3% (485/754) of the aneurysms. 82.5% (400/485) of aneurysms demonstrated complete occlusion (Raymond Roy Grade I). 81.4% (395/485) of aneurysms were found to meet the study's primary effectiveness outcome. At the early postoperative period, the mRS score was determined to be 0–2 vs. 3-6 in 98.2% (640/652) vs. 1.8% (12/652) of the cases, respectively. The combined major morbidity and mortality rate was 3.2% (21/652).

Conclusion: In the largest study of PED therapy for small and medium-sized intracranial aneurysms to date, pipeline-assisted coil embolization was chosen more often than multiple stent implantation for aneurysm treatment, demonstrating good results, high surgical success rates, high occlusion rates, and low morbidity and mortality.

Clinical Trial Registration: www.ClinicalTrials.gov, identifier: NCT03831672.

Introduction

Flow-diverting stents provide a new strategy for the treatment of aneurysms. This type of stent is placed in the parent artery at the level of the aneurysm neck to divert blood flow away from aneurysm, thus contributing to the gradual formation of aneurysmal thrombosis. Furthermore, flow-diverting stents provide scaffolding for the growth of endothelial cells in the aneurysm neck (1, 2). In 2011, the US Food and Drug Administration (FDA) approved the use of the flow-diverting Pipeline Embolization Device (PED) for treatment of large or giant wide-neck intracranial aneurysms (IAs) involving the petrous to the superior hypophyseal segments (3). However, large and giant aneurysms account for only a small fraction of all IAs, whereas small and medium-sized unruptured IAs account for ~80% (4). Moreover, the majority of ruptured IAs are smaller than 10 mm, and the annual rupture rate of small aneurysms is 0.5%-2% (4, 5). In the PREMIER study, small and medium-sized ( ≤ 12 mm) wide-neck aneurysms had a high rate of complete occlusion and low rate of associated permanent neurological complications (6). However, the results of PED therapy among large samples of patients with small and medium-sized intracranial aneurysms have rarely been reported. PLUS was a multi-center, panoramic, consecutive, real-world cohort registry study designed to evaluate the safety and effectiveness of PED treatment in the embolization of IAs among the Chinese population (7). Compared with previous PED studies (see Supplementary Table 1), PLUS study provided us with an opportunity to assess the safety and effectiveness of the treatment after flow diversion for small and medium-sized intracranial aneurysms. In this study, we reported the safety and effectiveness of PED treatment after flow diversion for IAs from the PLUS study that were smaller than 12 mm.

Methods

Study Participants

Details of the inclusion and exclusion criteria were described in the original PLUS study (7). In the PLUS study, patients with intracranial aneurysms who were treated with the PED at 14 centers in China between November 2014 and October 2019 were enrolled. A total of 1,171 patients with a combined 1,322 intracranial aneurysms were included in the PLUS study. In this work, we retrospectively evaluated patients with aneurysms smaller than 12 mm in the PLUS study. A total of 652 patients with a combined 754 aneurysms were included in this study. In this study, 652 procedures were performed to treat the 754 aneurysms studied. Both ruptured and unruptured aneurysms are treated.

Baseline Assessments

Baseline neurological assessment was performed prior to surgery using the Modified Rankin Scale (mRS) and the National Institutes of Health Stroke Scale (NIHSS). Preoperative CT angiography, MR angiography, or Digital Substraction Angiography(DSA) was performed to complete the baseline imaging assessment.

Dual Antiplatelet Therapy

Patients received an antiplatelet regimen that included a combination of aspirin (100–300 mg/day) and clopidogrel (75 mg/day) 3–5 days before surgery. Additionally, some patients underwent thromboelastogram (TEG) testing. Patients identified as nonresponsive to clopidogrel were given aspirin (100 mg/day) and ticagrelor (90 mg twice daily). Among each of the centers, the duration of dual-antiplatelet therapy after surgery varied from 3–6 months. After this, Aspirin was used indefinitely.

Procedure Details

The device used in embolization procedures included the PED Classic or PED Flex. The PED was delivered and deployed via a Marksman microcatheter (Medtronic; Irvine, California). The use of coils was determined by the treating physician. Intravenous heparin was administered at 50–100U/kg to achieve an activated clotting time >250s.Aneurysm occlusion was determined using digital subtraction angiography. The degree of aneurysm occlusion was assessed using the Roy Raymond Scale (8).

Study Outcomes

The primary effectiveness outcome was complete occlusion of the target IA (Raymond Scale I) at the last radiographic follow-up without significant (≤ 50%) stenosis of the parent artery. The primary safety outcome included incidence of major stroke (ischemic or hemorrhagic) or neurologic death within 30 days of surgery. The secondary safety results included functional outcomes (modified Rankin Scale, MRS) in the early postoperative period (≤ 30 days).

All early postoperative complications were recorded, including major ischemic stroke (defined as an increase in the National Institutes of Health Stroke Scale [NIHSS]. score >4 lasting more than 7 days), minor neurological stroke (defined as a change in the NIHSS score of ≤ 4 lasting fewer than 7 days with corroborative imaging), transient ischemic attack (TIA, defined as a transient neurological deficit without corroborative imaging), delayed aneurysm rupture (DAR, rupture that occurs after surgery for the treated aneurysm), intracerebral hemorrhage (ICH), subarachnoid hemorrhage (SAH), stent migration, parent artery stenosis, and mortality.

Additional data collected included device deployment success rate and number of PEDs utilized. Dense packing was defined as complete occlusion, and loose packing was defined as incomplete occlusion.

Statistical Analysis

Continuous variables, including demographics, aneurysm characteristics, and procedural characteristics, were represented as means and ranges. Categorical data were presented as numbers and percentages. Statistical analysis was performed using SPSS Version 25 (IBM Corporation; Armonk, NY, USA).

Results

Sample Demographics

Demographic and baseline characteristics are presented in Table 1. A total of 652 patients with a combined 754 aneurysms were included in this study. The mean age of the patients was 53.9 ± 10.3 years, and 68.7% (448/652) were female. The baseline mRS score was 0–2 vs. 3–6 in 99.1% (646/652) vs. 0.9% (6/652) of cases, respectively. Comorbidities included hypertension (42.9%, 280/652), diabetes (6.1%, 40/652), hyperlipidemia (4.4%, 29/652), coronary artery disease (6.0%, 39/652), cerebral atherosclerosis (16.4%, 107/652), alcohol abuse (10.9%, 71/652), and smoking (19%, 124/652).

TABLE 1
www.frontiersin.org

Table 1. Baseline demographic characteristics.

Aneurysm Characteristics

Aneurysm characteristics are presented in Table 2. The median aneurysm size was found to be 6.28 mm (mean 6.78 ± 2.67 mm), and the median neck size was 4 mm (mean 4.41 ± 1.98 mm). There were 83% (626/754) saccular, 3.6% (27/754) fusiform, 9.8% (74/754) dissecting, and 3.6% (27/754) blister aneurysms. Additionally, 3.8% (29/754) were ruptured aneurysms. A total of 629 (83.4%) aneurysms were located along the ICA. Most of the aneurysms were located at the C6 ophthalmic segment (47.7%, 360/629). 4.3% (32/754) of the aneurysms were located in the distal anterior circulation anteries (including the anterior and middle cerebral arteries), and 12.3% (93/754) were in the posterior circulation arteries.

TABLE 2
www.frontiersin.org

Table 2. Aneurysm characteristics.

Procedure

Procedural details are presented in Table 3. Of the 687 stents deployed, 99.7% were successfully deployed, of which 3.8% (26/687) stents were successfully deployed after technical adjustments. Complete wall attachment and complete neck coverage rates were 96.2% (725/754) and 99.7% (752/754), respectively. The mean number of devices used per patient was 1.01 ± 0.12 (range: 1–2). Multiple aneurysms were treated with one PED in 18.8% (142/687) of the patients. The PED Classic was used in 40.5% (278/687) while the PED Flex was used in 59.5% (409/687) of the patients. Of all aneurysms, 64.7% (488/754) were treated with the PED alone, whereas 35.3% (266/754) were treated with PED-assisted coil embolization. Among the aneurysms treated with the PED and coils, dense packing and loose packing accounted for 21.8% (58/266) and 78.2% (208/266), respectively.

TABLE 3
www.frontiersin.org

Table 3. Procedural details.

Primary Effectiveness Outcome

Angiographic outcomes are presented in Table 4. The mean imaging follow-up time was 8.26 ± 5.91 months (median time: 7 months; range: 3–36 months). Radiographic outcome at the last follow-up (median time: 7 months; range: 3–36 months) was available for 64.3% (485/754) of the aneurysms. At the last imaging follow-up, the primary effectiveness outcome was met in 81.4% (395/485) of the cases. Futhermore, 82.5% (400/485) of the aneurysms demonstrated complete occlusion (Raymond–Roy Grade I), and 1.4% (7/485) of parent arteries exhibited stenosis >50%.

TABLE 4
www.frontiersin.org

Table 4. Angiographic outcomes.

Safety Outcomes

The primary safety outcome included incidence of major stroke or neurologic death in the early postoperative period. The combined major morbidity and mortality rate was 3.2% (21/652). Among all patients, early postoperative major ischemic stroke occurred in 1.8% (12/652), major hemorrhagic stroke in 1.2% (8/652), and neurogenic death in 0.8% (5/652). The secondary safety outcomes included functional outcomes (modified Rankin Scale) in the early postoperative period. The mRS score during this period was 0–2 vs. 3–6 in 98.2% (640/652) vs. 1.8% (12/652) of cases, respectively.

Complications

Table 5 presents the early postoperative complications. During the early postoperative period, 1.8% (12/652) of patients experienced major ischemic stroke and 4.0% (26/652) of patients suffered from minor stroke and/or TIA. Early postoperative hemorrhagic complications occurred in 2.0% (13/652) of patients, including SAH in 0.6% (4/652) and ICH in 1.4% (9/652). DAR was observed in 0.6% (4/652) of patients and stent migration occurred in 0.3% (2/652) of cases. The mortality rate was 0.8% (5/652). Overall, poor functional outcomes were observed in 1.8% (12/652) of the patients during the early postoperative period.

TABLE 5
www.frontiersin.org

Table 5. Complications during the early postoperative period.

Discussion

In this study, patients from the PLUS study with aneurysms smaller than 12 mm were selected for analysis. A total of 652 patients harboring a combined 754 aneurysms (≤ 12 mm) were treated with PEDs at 14 centers in China between November 2014 and October 2019. This represents the largest study on PED treatment for small and medium-sized intracranial aneurysms to date.

In the PREMIER study (6), 81.9% of the patients (at 1 year) were Raymond–Roy Grade I, and 76.8% of patients were Raymond–Roy Grade I without major parent vessel stenosis (≤ 50%). In our study, the complete occlusion rate was relatively high compared to that of the PREMIER study, indicating that PEDs were used to treat small and medium-sized intracranial aneurysms with a high complete occlusion rate. At the last imaging follow-up (median time: seven months; range: 3–36 months), the primary effectiveness outcome was met in 81.4% (395/485) of the aneurysms, where 82.5% (400/485) demonstrated complete occlusion (Raymond–Roy grade 1). In a recent multicenter report of PED therapy for small aneurysms (≤ 7 mm), the rate of complete occlusion was 87% at the last imaging follow-up (median time: 6.5 months) (9). The above studies indicate that flow-diverting device have a high occlusion rate in the treatment of small and medium-sized aneurysms.

In line with conventional endovascular techniques, PED usage has similar efficacy and safety in treating small and medium-sized intracranial aneurysms (10). In a recent meta-analysis of stent-assisted coiling of aneurysms (11). the rate of complete occlusion at the final follow-up (9.03 ± 1.03 months) was reported to be between 73 and 88%, which is similar to our results. In addition, low retreatment rate is one of the reasons why PEDs are extensively used for small and medium-sized aneurysms. Malhotra et al. reported that PED treatment may have more health benefits due to its lower rates of retreatment compared to both simple coiling and stent-assisted techniques (12).

In our study, treatment of aneurysms with PEDs resulted in low major morbidity and mortality. Among the patients, early postoperative major ischemic stroke occurred in 1.8% (12/652), major hemorrhagic stroke in 1.2% (8/652), and neurogenic death in 0.8% (5/652). The combined major morbidity and mortality rate was 3.2% (21/652). In the PREMIER study (6), the combined major morbidity and mortality rate was 2.1% (3/140). Therefore, the major morbidity and mortality rates in our study were higher than that of the PREMIER study, which may have been caused by multiple factors. First, the interventional physician experience has an impact on complications. Jabbour et al. reported that the risk of complications decreased significantly as the physician experience increased (13). However, our data are from 2015, and the use of PED therapy for aneurysms is only in preliminary stages in China. Physician experience with PED use for aneurysm treatment is relatively limited, thus increasing the risk for complications. Second, of the aneurysms in our study, 12.3% (93/754) were posterior circulation aneurysms and 4.3% (32/754) were distal anterior circulation IAs. The study from Intre PED reported a higher rate of complications in posterior circulation aneurysms treated with PEDs (14). The complication rate in PED treatment for distal aneurysms is relatively high. A meta-analysis of 572 cases found a procedure-related morbidity of 9% and mortality of 4% (15).

In this study, we observed good device placement success rates. The device was successfully deployed in 99.7% (685/687) of stents, of which 3.8% (26/687) stents were successfully deployed after technical adjustment. In the PREMIER study, successful device deployment was reported in 99.3% (140/141) of cases. Additionally, in our study, 35.3% (266/754) of aneurysms were treated using PED-assisted coil embolization. The mean number of devices used per patient was 1.01 ± 0.12 (range: 1–2). Multiple devices were used in only 1.6% of cases. However, in the PREMIER study, five patients (3.5%) were treated with adjuvant coils, the average number of devices used per patient was 1.1 sie.3 (range: 0–2), and 6.4% of patients required multiple devices. In the Intre PED study, 28.0% of aneurysms ≤ 10 mm were treated using multiple stents (14). In terms of treatment strategies, pipeline-assisted coil embolization was more commonly chosen over multiple stents to treat aneurysms in this study. This choice of surgical strategy is determined by many factors. First, the use of multiple stents increases the financial burden of patients, so more coils are instead selected. Second, the use of multiple stents may affect the patency of branching vessels. Analysis of risk factors for complications suggested that the use of multiple stents increased the risk of complication (16). Third, coils adjunctive with PED are complementary, rather than competing, for cerebral aneurysm treatment (17). In a recent single-center study (18), with a very small subset of the same patients as the present study, it was reported that pipeline-assisted coiling increased the complete occlusion rate without increasing surgical complication rate. In terms of results, the surgical strategy we selected had similar results to the PREMIER study, which was more in line with China's national conditions.

There are some major limitations to this study worth noting. Specifically, this was a multi-center retrospective study, and there are biases in data collection as well as differences in surgical strategy decision among centers.

Conclusion

This was a large sample, multicenter study that evaluated the safety and efficacy of PED use in the treatment of small and medium-sized intracranial aneurysms in China. In the largest study on PED treatment for small and medium-sized intracranial aneurysms to date, pipeline assisted coil embolization was chosen more often than multiple stents, with good results, high surgical success rates, high occlusion rates, and low morbidity and mortality.

Data Availability Statement

The data analyzed in this study is subject to the following licenses/restrictions: To gain access, proposals should be addressed to the corresponding author. Requests to access these datasets should be directed to TL, dr.litianxiao@henu.edu.cn.

Ethics Statement

The studies involving human participants were reviewed and approved by the Ethics Committee of Beijing Tiantan Hospital. Written informed consent to participate in this study was provided by the participants' legal guardian/next of kin.

Author Contributions

PZ, XY, and TL: conception and design. HongqZ, JL, DS, YZ, SG, AM, YuW, WF, YaW, JW, GM, HS, and BL: acquisition of data. QS, KC, QZ, and YH: analysis and interpretation of data. HongyZ: drafting the article. LL: critically revising the article. All authors reviewed submitted version of manuscript.

Funding

This study was sponsored by the Scientific and Technological Project of Henan Province (222102310208) and the Natural Science Foundation of Henan Province (222300420357).

Conflict of Interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher's Note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Supplementary Material

The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fneur.2022.881353/full#supplementary-material

References

1. Ravindran K, Salem MM, Alturki AY, Thomas AJ, Ogilvy CS, Moore JM. Endothelialization following flow diversion for intracranial aneurysms: a systematic review. AJNR Am J Neuroradiol. (2019) 40:295–301. doi: 10.3174/ajnr.A5955

PubMed Abstract | CrossRef Full Text | Google Scholar

2. Kallmes DF, Ding YH Dai D, Kadirvel R, Lewis DA, Cloft HJ. A new endoluminal, flow-disrupting device for treatment of saccular aneurysms. Stroke. (2007) 38:2346–52. doi: 10.1161/STROKEAHA.106.479576

PubMed Abstract | CrossRef Full Text | Google Scholar

3. Becske T, Kallmes DF, Saatci I, McDougall CG, Szikora I., Lanzino G, et al. Pipeline for uncoilable or failed aneurysms: results from a multicenter clinical trial. Radiology. (2013) 267:858–68. doi: 10.1148/radiol.13120099

PubMed Abstract | CrossRef Full Text | Google Scholar

4. Wiebers DO, Whisnant JP, Huston J. 3rd, Meissner I, Brown RD Jr, Piepgras DG, et al. International Study of Unruptured Intracranial Aneurysms Investigators Unruptured intracranial aneurysms: natural history, clinical outcome, and risks of surgical and endovascular treatment. Lancet. (2003) 362:103–10. doi: 10.1016/S0140-6736(03)13860-3

PubMed Abstract | CrossRef Full Text | Google Scholar

5. International Study of Unruptured Intracranial Aneurysms Investigators. Unruptured intracranial aneurysms—risk of rupture and risks of surgical intervention. N Engl J Med. (1998) 339:1725–33. doi: 10.1056/NEJM199812103392401

PubMed Abstract | CrossRef Full Text | Google Scholar

6. Hanel RA, Kallmes DF, Lopes DK, Nelson PK, Siddiqui A, Jabbour P, et al. Prospective study on embolization of intracranial aneurysms with the pipeline device: the PREMIER study 1 year results. J Neurointerv Surg. (2020) 12:62–6. doi: 10.1136/neurintsurg-2019-015091

PubMed Abstract | CrossRef Full Text | Google Scholar

7. Luo B, Kang H, Zhang H, Li T, Liu J., Song D, et al. Pipeline Embolization device for intracranial aneurysms in a large Chinese cohort: factors related to aneurysm occlusion. Ther Adv Neurol Disord. (2020) 13:1756286420967828. doi: 10.1177/1756286420967828

PubMed Abstract | CrossRef Full Text | Google Scholar

8. Roy D, Milot G, Raymond J. Endovascular treatment of unruptured aneurysms. Stroke. (2001) 32:1998–2004. doi: 10.1161/hs0901.095600

PubMed Abstract | CrossRef Full Text | Google Scholar

9. Griessenauer CJ, Ogilvy CS, Foreman PM, Chua MH, Harrigan MR., He L, et al. Pipeline embolization device for small intracranial aneurysms: evaluation of safety and efficacy in a multicenter cohort. Neurosurgery. (2017) 80:579–87. doi: 10.1227/NEU.0000000000001377

PubMed Abstract | CrossRef Full Text | Google Scholar

10. Salem MM, Ravindran K, Enriquez-Marulanda A, Ascanio LC, Jordan N., Gomez-Paz S, et al. Pipeline embolization device versus stent-assisted coiling for intracranial aneurysm treatment: a retrospective propensity score-matched study. Neurosurgery. (2020) 87:516–22. doi: 10.1093/neuros/nyaa041

PubMed Abstract | CrossRef Full Text | Google Scholar

11. Pranata R, Yonas E, Deka H, Vania R, July J. Stent-assisted coiling of intracranial aneurysms using a nitinol-based stent (neuroform atlas): a systematic review and meta-analysis. Cardiovasc Intervent Radiol. (2020) 43:1049–61. doi: 10.1007/s00270-020-02502-9

PubMed Abstract | CrossRef Full Text | Google Scholar

12. Malhotra A, Wu X, Miller T, Matouk CC, Sanelli P, Gandhi D. Comparative effectiveness analysis of Pipeline device versus coiling in unruptured aneurysms smaller than 10 mm. J Neurosurg. (2019) 132:42–50. doi: 10.1093/neuros/nyz130

PubMed Abstract | CrossRef Full Text | Google Scholar

13. Jabbour P, Chalouhi N, Tjoumakaris S, Gonzalez LF, Dumont AS, Randazzo C, et al. The pipeline embolization device: learning curve and predictors of complications and aneurysm obliteration. Neurosurgery. (2013) 73:113–20. doi: 10.1227/01.neu.0000429844.06955.39

PubMed Abstract | CrossRef Full Text | Google Scholar

14. Kallmes DF, Hanel R, Lopes D, Boccardi E, Bonafé A, Cekirge S, et al. International retrospective study of the pipeline embolization device: a multicenter aneurysm treatment study. AJNR Am J Neuroradiol. (2015) 36:108–15. doi: 10.3174/ajnr.A4111

PubMed Abstract | CrossRef Full Text | Google Scholar

15. Yan Y, Zhu D, Tang H, Huang Q. Safety and efficacy of flow diverter treatment for aneurysm in small cerebral vessels: a systematic review and meta-analysis. World Neurosurg. (2018) 115:54–64. doi: 10.1016/j.wneu.2018.04.009

PubMed Abstract | CrossRef Full Text | Google Scholar

16. Brinjikji W, Lanzino G, Cloft HJ, Siddiqui AH, Boccardi E, Cekirge S, et al. Risk factors for ischemic complications following pipeline embolization device treatment of intracranial aneurysms: results from the IntrePED Study. AJNR Am J Neuroradiol. (2016) 37:1673–8. doi: 10.3174/ajnr.A4807

PubMed Abstract | CrossRef Full Text | Google Scholar

17. Lin N, Brouillard AM, Krishna C, Mokin M, Natarajan SK., Sonig A, et al. Use of coils in conjunction with the pipeline embolization device for treatment of intracranial aneurysms. Neurosurgery. (2015) 76:142–9. doi: 10.1227/NEU.0000000000000579

PubMed Abstract | CrossRef Full Text | Google Scholar

18. Zhang Q, Shao Q, Chang K, Zhang H, He Y, Andrade-Barazarte H, et al. Safety and efficacy of coils in conjunction with the pipeline flex embolization device for the treatment of cerebral aneurysms. Front Neurol. (2021) 12:651465. doi: 10.3389/fneur.2021.651465

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: endovascular, pipeline embolization device, small and medium-sized, intracranial aneurysm, multi-center study, complications, outcome

Citation: Zhang H, Li L, Zhang H, Liu J, Song D, Zhao Y, Guan S, Maimaitili A, Wang Y, Feng W, Wang Y, Wan J, Mao G, Shi H, Luo B, Shao Q, Chang K, Zhang Q, He Y, Zhang P, Yang X and Li Tx (2022) Small and Medium-Sized Aneurysm Outcomes Following Intracranial Aneurysm Treatment Using the Pipeline Embolization Device: A Subgroup Analysis of the PLUS Registry. Front. Neurol. 13:881353. doi: 10.3389/fneur.2022.881353

Received: 10 March 2022; Accepted: 25 April 2022;
Published: 31 May 2022.

Edited by:

Bharathi Dasan Jagadeesan, University of Minnesota Twin Cities, United States

Reviewed by:

Karan Topiwala, University of Minnesota Twin Cities, United States
Hannes Nordmeyer, Independent Researcher, Essen, Germany
Bu-Lang Gao, Second Hospital of Hebei Medical University, China
Guglielmo Pero, Niguarda Ca' Granda Hospital, Italy

Copyright © 2022 Zhang, Li, Zhang, Liu, Song, Zhao, Guan, Maimaitili, Wang, Feng, Wang, Wan, Mao, Shi, Luo, Shao, Chang, Zhang, He, Zhang, Yang and Li. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Peng Zhang, xwzhangpeng@126.com; Xinjian Yang, yangxinjian@voiceoftiantan.org; Tian xiao Li, dr.litianxiao@henu.edu.cn

These authors have contributed equally to this work

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.