Defective Endothelialization of Pipeline Embolization Device after Flow Diverter Therapy: An Autopsy Case Report
Hidenori OISHIYumiko MISHIMAKenji YATOMIKosuke TERANISHIKazumoto SUZUKITakashi FUJII
Author information
JOURNAL OPEN ACCESS FULL-TEXT HTML

2021 Volume 8 Issue 1 Pages 33-37

Details
Abstract

The authors report a rare autopsy case. A 59-year-old woman underwent flow diverter (FD) therapy using a pipeline embolization device (PED) for a large paraclinoid internal carotid artery aneurysm. Follow-up magnetic resonance (MR) examinations were performed 6 months after the treatment. Although the T2-weighted images showed progressive thrombosis of the aneurysm, the silent MR angiography (MRA) clearly showed the residual blood flow within the aneurysm. The patient committed suicide 2 months after the follow-up MR examinations. An autopsy specimen showed a small section of the defective membranes with the PED that matched the entry point of residual blood flow seen clearly in the silent MRA. Macroscopic photograph and hematoxylin and eosin stained sections showed defective endothelialization. In contrast, complete endothelialization was observed in membranes covering the PED. The autopsy findings after FD therapy showed defective endothelialization that perfectly matched and corroborated the silent MRA findings.