Advanced photon source proposal for upgrading the radiation safety of x-ray labs
There are two adjacent x-ray labs in building 360, each having two entrance doors. Lab A240 has two x-ray machines and lab A248 has one. All machines are equipped with sliding safety windows and microswitches to monitor the state of the windows - open or closed. Two modes of operation are possible. (1) Secure Mode in which all safety windows are closed as indicated by the microswitches. This satisfies the interlock system, allowing the high voltage power supply to be turned on. (2) Bypass Mode in which the interlock system is overridden by a key-controlled selector switch and high voltage can be turned on with machine hutch window(s) open. The bypass mode is potentially unsafe because it is possible for an operator to leave a running instrument unattended while the windows are open. Thus, it is possible for someone entering the lab to expose themselves to x-rays.