Low sample volume part-per billion level ion chromatographic analysis
- Westinghouse Savannah River Company, AIKEN, SC (United States)
ADS has developed an ion chromatographic method which enables low part-per-billion levels of analysis while minimizing liquid waste generation. This method incorporates several recent technical improvements in ion chromatographic instrumentation to achieve a ten- fold increase in sensitivity over existing ion chromatographic methods without additional analysis time or sample pre-concentration. This report outlines the method, establishes the precision and accuracy levels, and discusses the applicability of the method to waste minimization and radiation exposure reduction
- Research Organization:
- Savannah River Site (SRS), Aiken, SC (United States)
- Sponsoring Organization:
- USDOE, Washington, DC (United States)
- DOE Contract Number:
- AC09-89SR18035
- OSTI ID:
- 268552
- Report Number(s):
- WSRC-RP-96-72; ON: DE96060076; TRN: 96:017047
- Resource Relation:
- Other Information: PBD: 27 Feb 1996
- Country of Publication:
- United States
- Language:
- English
Similar Records
Low sample volume part-per-billion level ion chromatographic analysis method
Comparison of sample loops constructed of several different materials for gas chromatographic analysis of parts-per-billion-level organic gas mixtures
Determination of low parts per billion levels of amines in urine by liquid membrane sample cleanup directly coupled to a gas-liquid chromatograph
Conference
·
Tue Dec 31 00:00:00 EST 1996
·
OSTI ID:268552
Comparison of sample loops constructed of several different materials for gas chromatographic analysis of parts-per-billion-level organic gas mixtures
Journal Article
·
Wed Jul 15 00:00:00 EDT 1987
· Anal. Chem.; (United States)
·
OSTI ID:268552
Determination of low parts per billion levels of amines in urine by liquid membrane sample cleanup directly coupled to a gas-liquid chromatograph
Journal Article
·
Fri Jul 01 00:00:00 EDT 1988
· Anal. Chem.; (United States)
·
OSTI ID:268552