Abstract
Background: The study aim was to compare the performance of three different methods used for determining urinary glycosaminoglycans (GAG) levels in spot and 24-h urine samples.
Methods: Performance characteristics were studied for cetylpyridinium chloride (CPC), and manual and automated dimethylmethylene blue (DMB) methods.
Results: For automated DMB method, within-run precisions were 9.10% and 1.98%, and between-day precisions were 13.0% and 5.81% in low- and high-urine pools, respectively. The method was linear up to 100 mg/L of GAG concentration. The detection limit of the method was 0.71 mg/L. Mean recovery was 95.7%.
Conclusions: The automated DMB method was found to give better performance characteristics than cetylpyridinium chloride (CPC) and manual DMB methods. It is a fast, cheap, simple and reliable method and can be applied in many diseases in which GAG is used as a screening test.
References
1. Andrade F, Prieto JA, Elorz J, Martín S, Sanjurjo P, Aldámiz-Echevarría L. Stability of urinary glycosaminoglycans in patients with mucopolysaccharidoses. Clin Chim Acta 2008;388:73–7.http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=000252770800012&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=b7bc2757938ac7a7a821505f8243d9f310.1016/j.cca.2007.10.007Search in Google Scholar
2. Mabe P, Valiente A, Soto V, Cornejo V, Raimann E. Evaluation of reliability for urine mucopolysaccharidosis screening by dimethylmethylene blue and Berry spot tests. Clin Chim Acta 2004;345:135–40.10.1016/j.cccn.2004.03.015Search in Google Scholar
3. Whitley CB, Ridnour MD, Draper KA, Dutton CM, Neglia JP. Diagnostic test for mucopolysaccharidosis: I. Direct method for quantifying excessive urinary glycosaminoglycan excretion. Clin Chem 1989;35:374–9.10.1093/clinchem/35.3.374Search in Google Scholar PubMed
4. de Jong JG, Wevers RA, Laarakkers C, Poorthuis BJ. Dimethylmethylene blue-based spectrophotometry of glycosaminoglycans in untreated urine: a rapid screening procedure for mucopolysaccharidosis. Clin Chem 1989;35:1472–7.10.1093/clinchem/35.7.1472Search in Google Scholar
5. Pennock CA. A review and selection of simple laboratory methods used for the study of glycosaminoglycan excretion and the diagnosis of the mucopolysaccharidoses. J Clin Path 1976;29:111–23.10.1136/jcp.29.2.111Search in Google Scholar
6. Kéry V, Orlovska M, Stancikova M, Risko M, Zlnay D. Urinary glycosaminoglycan excretion in rheumatic diseases. Clin Chem 1992;38:841–6.10.1093/clinchem/38.6.841Search in Google Scholar PubMed
7. Clinical and Laboratory Standards Institute. Method comparison and bias estimation using patient samples; approved guideline. CLSI Document EP9-A2, 2nd ed. Wayne (PA): CLSI, 2002.Search in Google Scholar
8. Farndale RW, Buttle DJ, Barrett AJ. Improved quantitation and discrimination of sulphated glycosaminoglycans by use of dimethylmethylene blue. Biochim Biophys Acta 1986;883:173–7.10.1016/0304-4165(86)90306-5Search in Google Scholar
9. Whitley CB, Spielmann RC, Herro G, Teragawa SS. Urinary glycosaminoglycan excretion quantified by an automated semimicro method in specimens conveniently transported from around the globe. Mol Genet Metab 2002;75:56–64.10.1006/mgme.2001.3271Search in Google Scholar PubMed
10. Elmen W, Niebur J, Flanders G, Rutledge J. Measurement of serum hyaluronic acid in patients with rheumatoid arthritis: correlation with disease activity. J Rheumatol 1996;23:6.Search in Google Scholar
11. Goldberg JM, Cotlier E. Specific isolation and analysis of mucopolysaccharides (glycosaminoglycans) from human urine. Clin Chim Acta 1972;41:19–27.10.1016/0009-8981(72)90491-3Search in Google Scholar PubMed
12. Alonso-Fernández JR, Fidalgo J, Colón C. Neonatal screening for mucopolysaccharidoses by determination of glycosaminoglycans in the eluate of urine-impregnated paper: preliminary results of an improved DMB-based procedure. J Clin Lab Anal 2010;24:149–53.http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=000278032500005&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=b7bc2757938ac7a7a821505f8243d9f310.1002/jcla.20375Search in Google Scholar PubMed PubMed Central
©2013 by Walter de Gruyter Berlin Boston