Evaluation of eight commercial Zika virus IgM and IgG serology assays for diagnostics and research

  • Swee Ling Low,

    Roles Data curation, Formal analysis, Methodology, Visualization, Writing – original draft, Writing – review & editing

    Affiliation Environmental Health Institute, National Environment Agency, Singapore, Singapore

  • Yee Sin Leo,

    Roles Resources

    Affiliation National Centre for Infectious Disease, Singapore, Singapore

  • Yee Ling Lai,

    Roles Resources

    Affiliation Environmental Health Institute, National Environment Agency, Singapore, Singapore

  • Sally Lam,

    Roles Resources

    Affiliation Blood Services Group, Health Sciences Authority, Singapore, Singapore

  • Hwee Huang Tan,

    Roles Resources

    Affiliation Blood Services Group, Health Sciences Authority, Singapore, Singapore

  • Judith Chui Ching Wong,

    Roles Writing – review & editing

    Affiliation Environmental Health Institute, National Environment Agency, Singapore, Singapore

  • Li Kiang Tan,

    Roles Conceptualization, Funding acquisition, Supervision

    Affiliation Environmental Health Institute, National Environment Agency, Singapore, Singapore

  • Lee Ching Ng

    Roles Conceptualization, Funding acquisition, Supervision, Writing – review & editing

    Ng_Lee_Ching@nea.gov.sg

    Affiliations Environmental Health Institute, National Environment Agency, Singapore, Singapore, School of Biological Sciences, Nanyang Technological University, Singapore, Singapore

Abstract

Several commercial Zika virus (ZIKV) serology assays have been developed since the recognition of ZIKV outbreaks as a Public Health Emergency of International Concern in 2016. However, test interpretation for ZIKV serology can be challenging due to antibody cross-reactivity with other flaviviruses like dengue virus (DENV). Therefore, we sought to evaluate the performance of eight commercially available ZIKV IgM and IgG assays across three testing platforms, namely, immunochromatographic tests (ICT), ELISAs and immunofluorescence tests (IIFT). The test panel comprised of 278 samples, including acute and convalescent sera or plasma from ZIKV-confirmed, DENV-confirmed, non-ZIKV and non-DENV patients, and residual sera from healthy blood donors. The ZIKV IgM and IgG serology assays yielded higher test sensitivities of 23.5% - 97.1% among ZIKV convalescent samples as compared to 5.6% - 27.8% among ZIKV acute samples; the test specificities were 63.3% - 100% among acute and convalescent DENV, non-DENV samples. Among the ELISAs and IIFTs, the Diapro ZIKV IgM ELISA demonstrated high test sensitivity (96%) and specificity (80%) when tested on early convalescent samples, while the Euroimmun ZIKV IgG ELISA yielded the highest test specificity of 97% - 100% on samples from non-ZIKV patients and healthy blood donors. For rapid ICTs, the LumiQuick IgM rapid ICT yielded low test sensitivity, suggesting its limited utility. We showed that commercial ZIKV IgM and IgG serology assays have differing test performances, with some having moderate to high test sensitivities and specificities when used in a dengue endemic setting, although there were limitations in IgG serology.

Introduction

Zika virus (ZIKV) is a mosquito-borne virus that belongs to the family Flaviviridae, genus Flavivirus. The virus was first isolated from a rhesus monkey in the Zika forest of Uganda [1], with only few and sporadic cases previously reported in Africa and Asia [2]. In 2007, a large ZIKV outbreak occurred on Yap Island, Federated States of Micronesia, where the disease was reported to be mild [3]. Subsequently, ZIKV infections received significant global public health attention when they were associated with severe neurological complications such as Guillain-Barré syndrome (GBS) in adults and congenital Zika syndrome (includes microcephaly and other severe birth defects) in neonates, in epidemics in French Polynesia and Brazil, respectively [48]. This prompted the World Health Organization (WHO) to declare the ZIKV epidemic as a Public Health Emergency of International Concern (PHIC) in February 2016 [9]. Since then, several countries reported autochthonous transmission of ZIKV [10] including Singapore in August 2016 [11, 12]. As of March 2017, 84 countries reported new introductions or re-introductions of ZIKV following the WHO classification [13].

ZIKV is primarily transmitted by infected Aedes species mosquitoes, mainly Aedes aegypti, which also transmits dengue virus (DENV). Both viruses share similar geographical distribution, transmission cycles and symptoms of infection [14]. The presence of Aedes aegypti and Aedes albopictus mosquitoes places Singapore at risk of DENV and ZIKV epidemics, with local transmission of both viruses being reported [11, 1517]. While DENV is endemic in Singapore and cases are reported throughout the year [15], there was only one recorded ZIKV outbreak in 2016, with 458 reported cases [11, 12]. Following which, sporadic ZIKV cases were reported in 2018 and 2019 [18], suggesting that there could be low levels of transmission in Singapore. The threat of future Zika outbreaks in Singapore remains due to the presence of competent vectors [16, 17] and the low levels of immunity against ZIKV in the population (unpublished data, EHI).

Recommended test guidelines for ZIKV include the use of both molecular and serology methods [1921]. A definitive ZIKV infection in a symptomatic patient is based on ZIKV RNA detection in whole blood, serum or urine [2224], while the detection of ZIKV-specific antibodies can support ZIKV diagnosis when tests are carried out after the viremia or viruria stage of illness. ZIKV IgM is usually detectable by Day 4 post infection and persists in the first 12 weeks of infection [22, 25], although persistence beyond 12 weeks has also been observed [21]. ZIKV IgG, which is detected after IgM but persists after infection, can be used as a marker for seroconversion and seroprevalence. The interpretation of ZIKV serology results (both IgM and IgG) may be challenging due to antibody cross-reactivity and non-specific reactivity between ZIKV and other flaviviruses such as DENV [2529]. In scenarios where cross-reactive antibodies are present, the plaque reduction neutralization test (PRNT) is recommended as a confirmatory test to differentiate antibodies to related viruses [21]. However, the PRNT is labor-intensive and has a low sample throughput, which renders it impractical for use in routine diagnostics.

Several commercial serological assays utilizing different methods have been developed since the recognition of ZIKV as a public health threat. There is a need to assess their performance to identify suitable tests for serological diagnosis or serosurveys of ZIKV infections. In a dengue endemic country like Singapore, where dengue seroprevalence is approximately 50% in the resident population [30, 31], it is also important to evaluate the impact of cross-reactivity of DENV antibodies when using ZIKV serological assays in an endemic area. Therefore, in this study, we evaluated the performance of eight ZIKV serology assays [one ZIKV IgM/IgG rapid immunochromatograhic tests (ICT), four ZIKV IgM ELISAs, one ZIKV IgM immunofluorescence Test (IIFT) and two ZIKV IgG ELISA assays] for their potential use.

Materials and methods

Sample panel

The test panel comprised of 278 sera or plasma samples from three categories (Set A, Set B and Set C), as summarized in Table 1. Briefly, Set A consisted of 78 acute sera (≤5 days’ post fever onset) collected from patients suspected of having ZIKV or DENV infections in 2016. The samples were collected at primary healthcare clinics in Singapore and sent to the Environmental Health Institute (EHI) of the National Environment Agency (NEA) for diagnostic testing. The ZIKV and DENV infections were confirmed by real-time ZIKV RT-PCR analysis of urine [25] and the rapid Standard Diagnostics (SD) BIOLINE Dengue Duo Non-structural (NS) 1 Ag/IgM/IgG combo test (Standard Diagnostics, Gyeonggi-do, Republic of Korea), respectively.

Set B consisted of 121 convalescent plasma samples from patients suspected of having DENV or ZIKV infection in two separate studies conducted at Tan Tock Seng Hospital (TTSH), Singapore [32, 33]. Plasma samples were collected from 2010 to 2012, and from August 2016 to September 2016, for the DENV and ZIKV studies, respectively. ZIKV infections were confirmed by real-time ZIKV RT-PCR analysis of urine [25] while DENV infections were confirmed by real-time DENV RT-PCR analysis or DENV NS-1 Ag ELISA (Bio-Rad Platelia NS1 Ag ELISA, Bio-Rad Laboratories, Marnes-la-Coquette, France). The convalescent plasma samples were further grouped into two categories, samples that were collected between 7 to 14 days’ (Set B1, n = 57) and those that were collected between 23 to 34 days’ (Set B2, n = 64) post onset of fever. Among the dengue confirmed patients (n = 15), six and nine patients were identified as primary and secondary DENV infections respectively, using the previously described laboratory criteria [34]. In addition to the dengue patients, samples from an equal number of non-dengue patients (n = 15) were included for specificity testing.

Lastly, to evaluate the performance of the ZIKV IgG ELISAs for use in serosurveys, 79 residual serum samples obtained from healthy blood donors (Set C) were included. Among these, there were 19 samples with both ZIKV and DENV neutralizing antibodies (NAb), 30 samples with DENV NAb only, and 30 non-ZIKV, non-DENV samples as determined by PRNT (described below). Written informed consent was obtained from the blood donors for use of the residual serum samples for research (DSRB Ref: 2013/00791).

This study was reviewed and approved by the NEA-EHI Institutional Review Board (IRB 019/2016). All samples were anonymized before used.

Serologic analyses

Acute and convalescent dengue patient samples, convalescent non-dengue patient samples, and dengue negative healthy blood donor samples were tested for ZIKV NAb using the previously described PRNT method [30], all other samples were tested for both DENV and ZIKV NAb. Briefly, for the virus-antibody binding reaction, equal volumes (100 μL) of diluted sera (heat-inactivated and diluted 10-fold from 1:10 to 1:1000) and ZIKV or DENV (800 plaque forming units/mL) were added into 96-well plates and incubated at 37°C for an hour. Fifty microliters of the virus–antibody mixture was added into 24-well plates containing Baby Hamster Kidney (BHK) cells (seeded at 1 × 105 cells/ mL) in triplicates. The plates were incubated overnight at 37°C to allow virus adsorption onto BHK cells and were overlaid with 1% carboxymethyl cellulose (CMC) medium the following day. After 4 days of incubation at 37°C, viral plaques were fixed with 20% formalin and stained using 0.1% naphthol blue solution. The plaques were counted and the 50% end-point plaque reduction neutralization titers (PRNT50) were computed using the log probit method as described by Russell et al. [35]. Antibody titers were expressed as the reciprocal of the end point dilution. The DENV and ZIKV strains used for the PRNT are listed in S1 Table. Samples with ZIKV PRNT50 ≥1:30 to both ZIKV strains and DENV PRNT50 ≥1:30 to any of the four serotypes indicated the presence of neutralizing activity against the two viruses respectively. The DENV and ZIKV PRNT results of the sample panels are described in S1 Text.

The commercial kits evaluated in this study are listed in Table 2. The tests were performed and results were interpreted according to the manufacturers’ recommendations. The result outcomes for all ELISAs were positive, equivocal/borderline and negative, except for the InBios ZIKV IgM ELISA which had four result outcome categories: presumptive Zika positive, possible Zika positive, presumptive other flavivirus positive, and negative.

thumbnail
Table 2. List of ZIKV IgM and IgG serologic assays evaluated.

https://doi.org/10.1371/journal.pone.0244601.t002

Data analysis

Diagnostic accuracy indices were calculated for the ZIKV serology assays [36]. The sensitivity, specificity with 95% Wald confidence intervals (CIs) and the Area under Receiver Operating Characteristics Curve (AUROCC) values with 95% CI were determined. The AUROCC is a numerical value outcome (between 0 and 1) that assesses both the test sensitivity and specificity and could be used to determine and compare the diagnostic accuracy of tests. A perfect diagnostic test has an AUROCC value of 1.0 whereas a non-discriminating test has an area of ≤0.5 [37, 38]. The Chi-square test was used for testing significant differences in ZIKV seropositivity obtained by the ZIKV serology assays in primary and secondary DENV samples. Statistical significance was set at p < 0.05. All analyses were performed using SPSS (Version 23, IBM) and R software [39].

Results

Performance of ZIKV IgM and IgG tests on acute and convalescent samples from ZIKV-confirmed, DENV-confirmed, and non-ZIKV and non-DENV patients

When acute samples were tested (Set A, n = 78), ZIKV IgM test sensitivities for rapid ICT and ELISA/IIFT tests ranged from 5.6% to 27.8% (Fig 1A), with the InBios ZIKV IgM ELISA yielding the highest ZIKV IgM test sensitivity of 27.8% (95% CI: 9.69%-53.48%) and highest AUROCC value of 0.631 (Table 3). Test specificities for the ZIKV IgM assays ranged from 86.7% to 100% (Fig 1A), with NovaLisa® ZIKV IgM ELISA yielding the highest specificity. The low test sensitivities observed using ZIKV acute samples could be due to the time of collection, whereby antibodies have not been produced <5 days after fever onset. This was further explained by the absence of ZIKV NAb in almost 30% of ZIKV acute samples (S1 Text).

thumbnail
Fig 1.

(A) Sensitivity and specificity of ZIKV IgM rapid ICT, ELISAs and IIFT using acute samples from ZIKV-confirmed, DENV-confirmed, and non-ZIKV and non-DENV patients (Set A; n = 78). Sensitivity and specificity of ZIKV IgM/IgG rapid ICT, IgM ELISAs/IIFT and IgG ELISAs using (B) 7–14 days’ (Set B1; n = 57) and (C) 23–34 days’ (Set B2; n = 64) convalescent samples from ZIKV-confirmed, DENV-confirmed, and non-DENV patients. Confidence intervals (black bars) were constructed using Wald’s method.

https://doi.org/10.1371/journal.pone.0244601.g001

thumbnail
Table 3. Test performance characteristics (AUROCC) in acute and convalescent patient samples.

https://doi.org/10.1371/journal.pone.0244601.t003

For acute DENV and non-DENV, non-ZIKV samples, the test specificities of the ZIKV IgM assays were 73.3% - 100% and 96.7% - 100%, respectively (Fig 2A). The presence of cross-reactive DENV antibodies in the DENV-confirmed samples could have resulted in positive ZIKV results, leading to the lower test specificities. The AUROCC analysis revealed overall low values (AUROCC values of 0.461 to 0.631) for ZIKV IgM tests in acute samples (Table 3).

thumbnail
Fig 2.

(A) Specificity of ZIKV IgM rapid ICT, ELISAs and IIFT using acute samples from DENV-confirmed (n = 30), and non-ZIKV and non-DENV (n = 30) patients. (B) Specificity of ZIKV IgM/IgG rapid ICT, IgM ELISAs/IIFT and IgG ELISAs using 7–34 days’ convalescent samples from DENV-confirmed (n = 30), and non-DENV patients (n = 30) (Set B). Confidence intervals (black bars) were constructed using Wald’s method. LumiQuick QuickProfile™ ZIKV IgM & IgG rapid ICTs refer to a combo test kit.

https://doi.org/10.1371/journal.pone.0244601.g002

As expected, test sensitivities improved markedly for ZIKV convalescent samples. Using these samples, the test sensitivities of ZIKV IgM component of rapid ICT and IgM ELISA/IIFT tests ranged from 51.9% to 96.3% for Set B1 (7–14 days’; n = 27) and 23.5% to 94.1% for Set B2 (23–34 days’; n = 34) (Fig 1B and 1C, respectively). Among the tests, the Diapro ZIKV IgM ELISA yielded the highest ZIKV IgM test sensitivity of 94.1% - 96.3%, with a high specificity of 80% - 86.7% and high AUROCC values of 0.881 and 0.904 (Fig 1B and 1C, Table 3).

With regard to ZIKV IgG assays, the test sensitivities for rapid ICT and ELISA ranged from 70.4% to 81.5% in Set B1 and were at 97.1% in Set B2 (Fig 1B and 1C, respectively). For samples collected 7–14 days’ post-fever onset (Set B1), LumiQuick’s ICT yielded the highest ZIKV IgG test sensitivity of 81.5% (95%CI: 61.92%-93.70%) (Fig 1B) and had a high AUROCC value of 0.891 (Table 3), while LumiQuick’s ICT, Diapro’s and Euroimmun’s ELISAs yielded the highest ZIKV IgG test sensitivities of 97.1% (95% CI: 84.67–99.93%) for samples collected 23–34 days’ post-fever onset (Set B2) (Fig 1C).

The test specificities of the ZIKV IgM and IgG tests ranged from 63.3% to 100% in Set B1 (Fig 1B) and 76.7% to 100% in Set B2 (Fig 1C) with NovaLisa® ZIKV IgM ELISA, achieving the highest specificity of 100%. Using the DENV convalescent samples in Set B, ZIKV IgM tests generally achieved higher test specificities as compared to ZIKV IgG tests (IgM: 40% - 100%; IgG: 63.3% - 90%) (Fig 2B). For non-DENV convalescent samples, the test specificities of the ZIKV IgM and IgG tests were 90% - 100% and 100%, respectively (Fig 2B). The AUROCC values of the ZIKV IgM and IgG tests ranged from 0.582 to 0.952 (Table 3). The 2 X 2 contingency table of the ZIKV IgM and IgG tests’ results are available as S2 Table.

Performance of ZIKV IgG ELISAs on ZIKV and non-ZIKV sera samples from healthy blood donors

The utility of ZIKV IgG ELISAs in ZIKV seroprevalence studies (serosurveys) was further evaluated using residual sera from healthy blood donors that have been tested by PRNT (Set C, n = 79). The test sensitivities of the ZIKV IgG ELISAs ranged from 52.6% to 84.2%, with Diapro ZIKV IgG ELISA yielding the highest sensitivity (Table 4). The test specificities ranged from 90.0% to 100%, with Euroimmun ZIKV IgG ELISA yielding the highest specificity (100%). Among 60 samples that comprised of an equal number of DENV NAb-only samples and DENV/ZIKV NAb-absent samples, Diapro ZIKV IgG ELISA incorrectly identified six of these samples as ZIKV IgG positive, resulting in reduced test specificity (90%). The Euroimmun ZIKV IgG ELISA did not detect any of these samples as ZIKV IgG positive and had 100% specificity.

thumbnail
Table 4. Results of ZIKV IgG ELISA using healthy blood donor samples.

https://doi.org/10.1371/journal.pone.0244601.t004

Comparison of ZIKV seropositivity in primary and secondary DENV convalescent samples

Studies have shown that DENV-positive sera, especially those from secondary DENV infections, may cross-react with ZIKV [27, 28, 40, 41]. In this study, the ZIKV IgM and IgG assays tested 0% - 41.7% of primary DENV samples (n = 12) and 0% - 72.2% of secondary DENV samples (n = 18) to be positive for ZIKV antibodies. Similar to previous reports, a higher proportion of samples were observed to be positive for ZIKV antibodies in secondary DENV convalescent samples (n = 12; χ2 = 5.03–7.74, P<0.05) as compared to primary DENV samples (n = 18) (Table 5), using the Diapro ZIKV IgM and IgG ELISA, and Euroimmun ZIKV IgM IIFT assays. Conversely, NovaLisa® ZIKV IgM ELISA did not detect any of the primary and secondary DENV samples as positives. The serological differentiation between ZIKV and DENV in regions with a high proportion of secondary DENV infections remains challenging and a larger sample size may be required to validate the above observations.

thumbnail
Table 5. ZIKV seropositivity in convalescent primary and secondary DENV patient samples.

https://doi.org/10.1371/journal.pone.0244601.t005

Discussion

We compared the test performance of one ZIKV IgM/IgG rapid ICT, four ZIKV IgM ELISAs, one ZIKV IgM IIFT and two ZIKV IgG ELISAs using characterized sample panels from a dengue endemic setting. Moreover, the gold standard PRNT was used as a reference in this study to determine the seroconversion in ZIKV convalescent samples collected from ZIKV RT-PCR-confirmed patients, the presence of ZIKV NAb in healthy blood donor samples, and the presence of probable cross-reactive DENV NAb. It is advantageous that PRNT is able to establish the presence of ZIKV antibodies.

The performance of the assays for use in routine testing, high throughput screening, and serosurveys was assessed. For rapid ICTs, the LumiQuick ZIKV IgM rapid ICT obtained low test sensitivities (5.6% - 55.6%), suggesting its limited utility; there are also limited commercial ZIKV ICTs available. This contrasts with the availability of many robust DENV rapid ICTs that have high sensitivities [42]. Regarding test reliability, the LumiQuick ZIKV rapid ICT initially gave nine invalid test results but was resolved upon retesting.

For high throughput screening, the Diapro ZIKV IgM ELISA yielded test sensitivities of 94.1%– 96.3% and specificities of 80% - 86.7% among ZIKV, DENV and non-DENV convalescent samples. The sensitivity values met the proposed WHO Zika target product profile’s acceptable sensitivity of >95% but the specificity values fell short of the acceptable specificity of >98% for the diagnosis of patients with active infection [43]. The test sensitivity results were consistent with other studies that reported sensitivities of 69% - 100% [44, 45] but test specificities were lower than the reported specificity of 96% [45]. The evaluation data of the ZIKV IgM ELISA/IIFT tests obtained from this study supplemented the limited reports on test specificities in dengue endemic settings and supported the test performance results in prior studies. For instance, the Euroimmun ZIKV IgM ELISA yielded high test specificities of 93.3% - 96.7% but lower test sensitivities of 23.5% - 51.9%, this corroborating with the results from other studies (test specificity >90%; test sensitivity 19.7% - 58.8%) [4648]. For the NovaLisa® ZIKV IgM ELISA, higher test specificities were reported by Basile et al. (98%, n = 155) [49] and for this study (100%, n = 120) as compared to results reported by Safronetz et al. [50]. These authors reported reduced test specificities of 66% in negative samples (n = 25) and 70% in DENV-positive samples (n = 10); the difference in performance could be due to the fewer number of samples used.

The use of a specific ZIKV IgG assay for serosurveys can greatly reduce the false positivity rate in a dengue endemic setting. Although this study demonstrated that the ZIKV IgG tests generally achieved lower test specificities compared to ZIKV IgM tests (IgG: 76.7% - 96.7%; IgM: 63.3% - 100% when DENV and non-DENV convalescent samples were used), the Euroimmun ZIKV IgG ELISA, however, exhibited high test specificities in this and other studies (90.0% - 96.7%, and 100%, respectively) [46, 51]. Furthermore, our results showed that its test specificity was 87% when DENV convalescent patient samples were used, and 100% when both the primary DENV patient samples and DENV NAb-positive healthy blood donor samples were used. Notably, the Euroimmun ZIKV IgG ELISA has been used in ZIKV serosurveys conducted in Laos, Martinique, Bolivia and Taiwan [5255].

A key focus of the study was the assessment of test specificities for the ZIKV IgM and IgG assays using DENV-confirmed patient samples. Reduced specificities of the ZIKV IgM and IgG tests were observed among DENV convalescent samples, likely due to the increase in cross-reactive DENV antibodies in the DENV convalescent samples. Notably, the basal force of infection in Singapore is approximately 1% and dengue infections (and in turn, our samples) could be primary infections [31]. Although other studies and our results suggest higher cross-reactivity among samples from secondary infections [27, 28, 40, 41], the observation of cross-reactivity among DENV samples from primary infections as well emphasizes the need for specific, yet sensitive assays. In this regard, our study distinguished different test specificities among the commercial ZIKV IgM and IgG assays and provided further information on the assays’ performance.

This study highlighted two key challenges of ZIKV serology testing. Firstly, seropositivity of the ZIKV serologic assays was observed in some DENV convalescent samples, due to the presence of cross-reactive DENV antibodies. This finding corroborated with other studies [41, 56] where an in-house ZIKV NS1 Blockade of Binding assay found 17 of 86 secondary DENV samples obtained from Nicaraguan positive for ZIKV [41]. In another study, the analyses of DENV-immune samples collected less than 12 weeks post infection revealed IgG cross-reactivity with the ZIKV-Envelope Domain I and III antigens, although the level of cross-reactivity decreased in samples collected after more than 12 weeks post infection [56]. The use of PRNT could offer a partial solution to the serological differentiation of DENV and ZIKV infections. Similar to other studies [3, 25, 57, 58], specific ZIKV NAb responses were detected in primary ZIKV infections, but extensive cross-reactive ZIKV and DENV NAb responses were detected in secondary ZIKV infections.

Secondly, low ZIKV antibody levels may also complicate the interpretation of serological test results [59]. Interestingly, negative results were consistently obtained using the ZIKV IgM and IgG tests for convalescent samples of one ZIKV-confirmed patient (RT-PCR positive). The ZIKV NAb titres (PRNT50 titers: 31–43) were lower than those for DENV NAb [PRNT50 titers: DENV-1 (140–180); DENV-2 (340–450); DENV-3 (<100) and DENV-4 (<100)] for convalescent samples from the patient, although the acute sample tested negative for DENV NS1, IgM and IgG, therefore ruling out a recent DENV infection. Therefore, laboratory diagnoses should always be interpreted with all available test results, and correlated with clinical findings.

In this study, test specificities of the recombinant ZIKV NS1-based Euroimmun ZIKV ELISAs, Diapro ZIKV ELISAs, and LumiQuick ZIKV IgM/IgG rapid ICT were higher than that of the InBios ZIKV IgM ELISA, which utilizes a recombinant ZIKV envelope glycoprotein as the antigen. The difference in performance of the ZIKV serology tests could be attributed to the ZIKV antigen used in detection. Antibodies to NS1 antigens appear to be more virus-specific when compared to envelope glycoprotein antigens, likely due to the electrostatic differences in the NS1 surface loop among flaviviruses such as DENV, ZIKV and West Nile virus [29, 60, 61]. Tests that utilizes ZIKV-infected cells such as the Euroimmun ZIKV IgM IIFT also had lower test specificities. This observation corresponds with previous reports that antigen preparations from whole viruses yielded high level of antibody cross-reactivity between ZIKV and DENV [3, 25, 62]. Of the remaining tests, the NovaLisa® ELISA did not specify the ZIKV antigen which is used. In general, if ZIKV serology tests are used in an area where there is no known DENV transmission, tests with high sensitivities would be preferred, and follow-up tests using a more specific method could be done to confirm the results. On the contrary, ZIKV serology tests with high specificities would be preferred if they are performed in areas with known DENV transmission.

This study evaluated a considerable number of ZIKV serologic assays and provided invaluable information of their test performance in a dengue endemic setting.

Supporting information

S1 Table. DENV and ZIKV strains used in PRNT.

https://doi.org/10.1371/journal.pone.0244601.s001

(DOCX)

S2 Table. 2 x 2 contingency table for ZIKV IgM and IgG tests on acute and convalescent ZIKV and non-ZIKV samples.

https://doi.org/10.1371/journal.pone.0244601.s002

(DOCX)

Acknowledgments

We thank the staff members of EHI disease surveillance team, and the TTSH Dengue and Zika cohort study for their effort in obtaining and processing the clinical samples; and Sze Sze Chua and Kooi Sim Ng at the Blood Services Group, for their assistance in providing anonymized residual blood donor samples.

References

  1. 1. Dick GW, Kitchen SF, Haddow AJ. Zika virus. I. Isolations and serological specificity. Trans R Soc Trop Med Hyg. 1952;46(5):509–20. Epub 1952/09/01. pmid:12995440.
  2. 2. Hayes EB. Zika virus outside Africa. Emerg Infect Dis. 2009;15(9):1347–50. Epub 2009/10/01. pmid:19788800; PubMed Central PMCID: PMC2819875.
  3. 3. Duffy MR, Chen TH, Hancock WT, Powers AM, Kool JL, Lanciotti RS, et al. Zika virus outbreak on Yap Island, Federated States of Micronesia. N Engl J Med. 2009;360(24):2536–43. Epub 2009/06/12. pmid:19516034.
  4. 4. de Araujo TVB, Ximenes RAA, Miranda-Filho DB, Souza WV, Montarroyos UR, de Melo APL, et al. Association between microcephaly, Zika virus infection, and other risk factors in Brazil: final report of a case-control study. Lancet Infect Dis. 2017. Epub 2017/12/16. pmid:29242091.
  5. 5. Schuler-Faccini L, Ribeiro EM, Feitosa IM, Horovitz DD, Cavalcanti DP, Pessoa A, et al. Possible Association Between Zika Virus Infection and Microcephaly—Brazil, 2015. MMWR Morb Mortal Wkly Rep. 2016;65(3):59–62. Epub 2016/01/29. pmid:26820244.
  6. 6. Oehler E, Watrin L, Larre P, Leparc-Goffart I, Lastere S, Valour F, et al. Zika virus infection complicated by Guillain-Barre syndrome—case report, French Polynesia, December 2013. Euro Surveill. 2014;19(9). Epub 2014/03/15. pmid:24626205.
  7. 7. Cao-Lormeau VM, Blake A, Mons S, Lastere S, Roche C, Vanhomwegen J, et al. Guillain-Barre Syndrome outbreak associated with Zika virus infection in French Polynesia: a case-control study. Lancet. 2016;387(10027):1531–9. Epub 2016/03/08. pmid:26948433; PubMed Central PMCID: PMC5444521.
  8. 8. Dos Santos T, Rodriguez A, Almiron M, Sanhueza A, Ramon P, de Oliveira WK, et al. Zika Virus and the Guillain-Barre Syndrome—Case Series from Seven Countries. N Engl J Med. 2016;375(16):1598–601. Epub 2016/11/01. pmid:27579558.
  9. 9. Heymann DL, Hodgson A, Sall AA, Freedman DO, Staples JE, Althabe F, et al. Zika virus and microcephaly: why is this situation a PHEIC? Lancet. 2016;387(10020):719–21. Epub 2016/02/16. pmid:26876373.
  10. 10. Hills SL, Fischer M, Petersen LR. Epidemiology of Zika Virus Infection. J Infect Dis. 2017;216(suppl_10):S868–S74. Epub 2017/12/22. pmid:29267914.
  11. 11. Singapore Zika Study G. Outbreak of Zika virus infection in Singapore: an epidemiological, entomological, virological, and clinical analysis. Lancet Infect Dis. 2017;17(8):813–21. Epub 2017/05/22. pmid:28527892.
  12. 12. Ministry of Health S. Vector-borne diseases, Communicable Diseases Surveillance in Singapore 2016. 2016.
  13. 13. World Health Organization. Situation Report: Zika Virus, Microcephaly, Guillain-Barré Syndrome, 10 March 2017. 2017.
  14. 14. Musso D, Gubler DJ. Zika Virus. Clin Microbiol Rev. 2016;29(3):487–524. Epub 2016/04/01. pmid:27029595; PubMed Central PMCID: PMC4861986.
  15. 15. Rajarethinam J, Ang L-W, Ong J, Ycasas J, Hapuarachchi HC, Yap G, et al. Dengue in Singapore from 2004 to 2016: Cyclical Epidemic Patterns Dominated by Serotypes 1 and 2. The American journal of tropical medicine and hygiene. 2018;99(1):204–10. Epub 2018/05/29. pmid:29848407.
  16. 16. Lee C, Vythilingam I, Chong CS, Abdul Razak MA, Tan CH, Liew C, et al. Gravitraps for management of dengue clusters in Singapore. Am J Trop Med Hyg. 2013;88(5):888–92. Epub 2013/03/13. pmid:23478581; PubMed Central PMCID: PMC3752753.
  17. 17. Ong J, Liu X, Rajarethinam J, Yap G, Ho D, Ng LC. A novel entomological index, Aedes aegypti Breeding Percentage, reveals the geographical spread of the dengue vector in Singapore and serves as a spatial risk indicator for dengue. Parasit Vectors. 2019;12(1):17. Epub 2019/01/10. pmid:30621762; PubMed Central PMCID: PMC6325748.
  18. 18. Ministry of Health S. Weekly Infectious Diseases Bulletin 2019 [26 June 2019].
  19. 19. Oduyebo T, Polen KD, Walke HT, Reagan-Steiner S, Lathrop E, Rabe IB, et al. Update: Interim Guidance for Health Care Providers Caring for Pregnant Women with Possible Zika Virus Exposure—United States (Including U.S. Territories), July 2017. MMWR Morb Mortal Wkly Rep. 2017;66(29):781–93. Epub 2017/07/28. pmid:28749921.
  20. 20. World Health Organization. Laboratory testing for Zika virus infection—Interim guidance. 2016. Available from: http://apps.who.int/iris/bitstream/10665/204671/1/WHO_ZIKV_LAB_16.1_eng.pdf?ua=1.
  21. 21. CDC U. Updated Guidance for US Laboratories Testing for Zika Virus Infection 2017. Available from: https://www.cdc.gov/zika/laboratories/lab-guidance.html.
  22. 22. Landry ML, St George K. Laboratory Diagnosis of Zika Virus Infection. Arch Pathol Lab Med. 2017;141(1):60–7. Epub 2016/10/21. pmid:27763787.
  23. 23. Gourinat AC, O'Connor O, Calvez E, Goarant C, Dupont-Rouzeyrol M. Detection of Zika virus in urine. Emerg Infect Dis. 2015;21(1):84–6. Epub 2014/12/23. pmid:25530324; PubMed Central PMCID: PMC4285245.
  24. 24. Terzian ACB, Estofolete CF, Alves da Silva R, Vaz-Oliani DCM, Oliani AH, Brandao de Mattos CC, et al. Long-Term Viruria in Zika Virus-Infected Pregnant Women, Brazil, 2016. Emerg Infect Dis. 2017;23(11):1891–3. Epub 2017/10/20. pmid:29048293; PubMed Central PMCID: PMC5652423.
  25. 25. Lanciotti RS, Kosoy OL, Laven JJ, Velez JO, Lambert AJ, Johnson AJ, et al. Genetic and serologic properties of Zika virus associated with an epidemic, Yap State, Micronesia, 2007. Emerg Infect Dis. 2008;14(8):1232–9. Epub 2008/08/06. pmid:18680646; PubMed Central PMCID: PMC2600394.
  26. 26. Rabe IB, Staples JE, Villanueva J, Hummel KB, Johnson JA, Rose L, et al. Interim Guidance for Interpretation of Zika Virus Antibody Test Results. MMWR Morb Mortal Wkly Rep. 2016;65(21):543–6. Epub 2016/06/03. pmid:27254248.
  27. 27. Priyamvada L, Quicke KM, Hudson WH, Onlamoon N, Sewatanon J, Edupuganti S, et al. Human antibody responses after dengue virus infection are highly cross-reactive to Zika virus. Proc Natl Acad Sci U S A. 2016;113(28):7852–7. Epub 2016/06/30. pmid:27354515; PubMed Central PMCID: PMC4948328.
  28. 28. Dejnirattisai W, Supasa P, Wongwiwat W, Rouvinski A, Barba-Spaeth G, Duangchinda T, et al. Dengue virus sero-cross-reactivity drives antibody-dependent enhancement of infection with zika virus. Nat Immunol. 2016;17(9):1102–8. Epub 2016/06/25. pmid:27339099; PubMed Central PMCID: PMC4994874.
  29. 29. Stettler K, Beltramello M, Espinosa DA, Graham V, Cassotta A, Bianchi S, et al. Specificity, cross-reactivity, and function of antibodies elicited by Zika virus infection. Science. 2016;353(6301):823–6. Epub 2016/07/16. pmid:27417494.
  30. 30. Low SL, Lam S, Wong WY, Teo D, Ng LC, Tan LK. Dengue seroprevalence of healthy adults in Singapore: serosurvey among blood donors, 2009. Am J Trop Med Hyg. 2015;93(1):40–5. Epub 2015/05/28. pmid:26013376; PubMed Central PMCID: PMC4497902.
  31. 31. Tan LK, Low SL, Sun H, Shi Y, Liu L, Lam S, et al. Force-of-infection and True Infection Rate of Dengue in Singapore–its implication on dengue control and management. American Journal of Epidemiology. 2019. pmid:31062837
  32. 32. Ho HJ, Wong JGX, Mar Kyaw W, Lye DC, Leo YS, Chow A. Diagnostic Accuracy of Parameters for Zika and Dengue Virus Infections, Singapore. Emerg Infect Dis. 2017;23(12):2085–8. Epub 2017/11/18. pmid:29148371; PubMed Central PMCID: PMC5708236.
  33. 33. Ng DHL, Ho HJ, Chow A, Wong J, Kyaw WM, Tan A, et al. Correlation of clinical illness with viremia in Zika virus disease during an outbreak in Singapore. BMC Infectious Diseases. 2018;18(1):301. pmid:29973158
  34. 34. Pok KY, Lai YL, Sng J, Ng LC. Evaluation of nonstructural 1 antigen assays for the diagnosis and surveillance of dengue in Singapore. Vector Borne Zoonotic Dis. 2010;10(10):1009–16. Epub 2010/04/30. pmid:20426686; PubMed Central PMCID: PMC2992696.
  35. 35. Russell PK, Nisalak A, Sukhavachana P, Vivona S. A Plaque Reduction Test for Dengue Virus Neutralizing Antibodies. The Journal of Immunology. 1967;99(2):285. pmid:6031202
  36. 36. Bossuyt PM, Reitsma JB, Bruns DE, Gatsonis CA, Glasziou PP, Irwig L, et al. STARD 2015: an updated list of essential items for reporting diagnostic accuracy studies. BMJ: British Medical Journal. 2015;351.
  37. 37. Okeh UM OC. Evaluating Measures of Indicators of Diagnostic Test Performance: Fundamental Meanings and Formulars. J Biomet Biostat 2012;3(1).
  38. 38. Zweig MH, Campbell G. Receiver-operating characteristic (ROC) plots: a fundamental evaluation tool in clinical medicine. Clinical Chemistry. 1993;39(4):561–77. pmid:8472349
  39. 39. RCoreTeam. R: A language and environment for statistical computing. Vienna, Austria: R Foundation for Statistical Computing; 2014.
  40. 40. Swanstrom JA, Plante JA, Plante KS, Young EF, McGowan E, Gallichotte EN, et al. Dengue Virus Envelope Dimer Epitope Monoclonal Antibodies Isolated from Dengue Patients Are Protective against Zika Virus. MBio. 2016;7(4). Epub 2016/07/21. pmid:27435464; PubMed Central PMCID: PMC4958264.
  41. 41. Balmaseda A, Stettler K, Medialdea-Carrera R, Collado D, Jin X, Zambrana JV, et al. Antibody-based assay discriminates Zika virus infection from other flaviviruses. Proc Natl Acad Sci U S A. 2017;114(31):8384–9. Epub 2017/07/19. pmid:28716913; PubMed Central PMCID: PMC5547631.
  42. 42. Jang WS, Kwak SY, May WL, Yang DJ, Nam J, Lim CS. Comparative evaluation of three dengue duo rapid test kits to detect NS1, IgM, and IgG associated with acute dengue in children in Myanmar. PLoS One. 2019;14(3):e0213451. Epub 2019/03/14. pmid:30865680; PubMed Central PMCID: PMC6415848.
  43. 43. World Health Organization. Target Product Profiles for better diagnostic tests for Zika Virus Infection 2016 [cited 2020 22 Oct]. Available from: https://www.who.int/blueprint/what/research-development/zika-tpp.pdf?ua=1.
  44. 44. Pasquier C, Joguet G, Mengelle C, Chapuy-Regaud S, Pavili L, Prisant N, et al. Kinetics of anti-ZIKV antibodies after Zika infection using two commercial enzyme-linked immunoassays. Diagn Microbiol Infect Dis. 2018;90(1):26–30. Epub 2017/11/07. pmid:29107414.
  45. 45. Matheus S, Talla C, Labeau B, de Laval F, Briolant S, Berthelot L, et al. Performance of 2 Commercial Serologic Tests for Diagnosing Zika Virus Infection. Emerg Infect Dis. 2019;25(6):1153–60. Epub 2019/05/21. pmid:31107211; PubMed Central PMCID: PMC6537740.
  46. 46. Steinhagen K, Probst C, Radzimski C, Schmidt-Chanasit J, Emmerich P, van Esbroeck M, et al. Serodiagnosis of Zika virus (ZIKV) infections by a novel NS1-based ELISA devoid of cross-reactivity with dengue virus antibodies: a multicohort study of assay performance, 2015 to 2016. Euro Surveill. 2016;21(50). Epub 2016/12/23. pmid:28006649; PubMed Central PMCID: PMC5291135.
  47. 47. L'Huillier AG, Hamid-Allie A, Kristjanson E, Papageorgiou L, Hung S, Wong CF, et al. Evaluation of Euroimmun Anti-Zika Virus IgM and IgG Enzyme-Linked Immunosorbent Assays for Zika Virus Serologic Testing. J Clin Microbiol. 2017;55(8):2462–71. Epub 2017/06/02. pmid:28566316; PubMed Central PMCID: PMC5527425.
  48. 48. Granger D, Hilgart H, Misner L, Christensen J, Bistodeau S, Palm J, et al. Serologic Testing for Zika Virus: Comparison of Three Zika Virus IgM-Screening Enzyme-Linked Immunosorbent Assays and Initial Laboratory Experiences. J Clin Microbiol. 2017;55(7):2127–36. Epub 2017/04/28. pmid:28446573; PubMed Central PMCID: PMC5483914.
  49. 49. Basile AJ, Goodman C, Horiuchi K, Sloan A, Johnson BW, Kosoy O, et al. Multi-laboratory comparison of three commercially available Zika IgM enzyme-linked immunosorbent assays. J Virol Methods. 2018;260:26–33. Epub 2018/07/03. pmid:29964076.
  50. 50. Safronetz D, Sloan A, Stein DR, Mendoza E, Barairo N, Ranadheera C, et al. Evaluation of 5 Commercially Available Zika Virus Immunoassays. Emerg Infect Dis. 2017;23(9):1577–80. Epub 2017/07/01. pmid:28665268; PubMed Central PMCID: PMC5572859.
  51. 51. Huzly D, Hanselmann I, Schmidt-Chanasit J, Panning M. High specificity of a novel Zika virus ELISA in European patients after exposure to different flaviviruses. Euro Surveill. 2016;21(16). Epub 2016/04/30. pmid:27126052.
  52. 52. Guey Chuen P, Tzu-Chuan H, Hsin IS, Chia-Hua L, Pei-Wen H, Chih-Huan C, et al. Seroprevalence of Zika and Dengue Virus Antibodies among Migrant Workers, Taiwan, 2017. Emerging Infectious Disease journal. 2019;25(4). pmid:30882327
  53. 53. Gallian P, Cabie A, Richard P, Paturel L, Charrel RN, Pastorino B, et al. Zika virus in asymptomatic blood donors in Martinique. Blood. 2017;129(2):263–6. Epub 2016/11/10. pmid:27827826.
  54. 54. Saba Villarroel PM, Nurtop E, Pastorino B, Roca Y, Drexler JF, Gallian P, et al. Zika virus epidemiology in Bolivia: A seroprevalence study in volunteer blood donors. PLOS Neglected Tropical Diseases. 2018;12(3):e0006239. pmid:29513667
  55. 55. Pastorino B, Sengvilaipaseuth O, Chanthongthip A, Vongsouvath M, Souksakhone C, Mayxay M, et al. Low Zika Virus Seroprevalence in Vientiane, Laos, 2003–2015. Am J Trop Med Hyg. 2019;100(3):639–42. Epub 2019/01/30. pmid:30693859.
  56. 56. Premkumar L, Collins M, Graham S, Liou GA, Lopez CA, Jadi R, et al. Development of envelope protein antigens to serologically differentiate Zika from dengue virus infection. J Clin Microbiol. 2017. Epub 2017/12/22. pmid:29263206.
  57. 57. Montoya M, Collins M, Dejnirattisai W, Katzelnick LC, Puerta-Guardo H, Jadi R, et al. Longitudinal Analysis of Antibody Cross-neutralization Following Zika Virus and Dengue Virus Infection in Asia and the Americas. The Journal of Infectious Diseases. 2018;218(4):536–45. pmid:29618091
  58. 58. Tan CH, Tan LK, Hapuarachchi HC, Lai YL, Wong PSJ, Yap G, et al. Viral and Antibody Kinetics, and Mosquito Infectivity of an Imported Case of Zika Fever Due to Asian Genotype (American Strain) in Singapore. Viruses. 2018;10(1). Epub 2018/01/19. pmid:29346272.
  59. 59. St George K, Sohi IS, Dufort EM, Dean AB, White JL, Limberger R, et al. Zika Virus Testing Considerations: Lessons Learned from the First 80 Real-Time Reverse Transcription-PCR-Positive Cases Diagnosed in New York State. J Clin Microbiol. 2017;55(2):535–44. Epub 2016/12/09. pmid:27927917; PubMed Central PMCID: PMC5277524.
  60. 60. Song H, Qi J, Haywood J, Shi Y, Gao GF. Zika virus NS1 structure reveals diversity of electrostatic surfaces among flaviviruses. Nat Struct Mol Biol. 2016;23(5):456–8. Epub 2016/04/19. pmid:27088990.
  61. 61. Speer SD, Pierson TC. VIROLOGY. Diagnostics for Zika virus on the horizon. Science. 2016;353(6301):750–1. Epub 2016/08/20. pmid:27540152.
  62. 62. Keasey SL, Pugh CL, Jensen SM, Smith JL, Hontz RD, Durbin AP, et al. Antibody Responses to Zika Virus Infections in Environments of Flavivirus Endemicity. Clin Vaccine Immunol. 2017;24(4). Epub 2017/02/24. pmid:28228395; PubMed Central PMCID: PMC5382833.