1995 Volume 18 Issue 10 Pages 1311-1314
Barnidipine is a 1, 4-dihydropyridine calcium antagonist. HPLC was conducted on a polybutadiene coated alumina column using an alkaline mobile phase and an electrochemical detector to determine the content of this drug in serum and plasma. A good linear relationship between barnidipine concentration and peak height was found in 5-500ng/ml with a correlation coefficient of 0.998. The detection limit was 1 ng/ml. The within-day and day-to-day variations were examined for control human serum. Relative standard deviation of within-day assay for serum spiked with 10 ng/ml barnidipine·HCl was 6.9% and the recovery was 104%. A pharmacokinetic study was made in which the time course of barnidipine in dog plasma was followed.