A comparison of hydrophobic polyurethane and polyurethane peripherally inserted central catheter: results from a feasibility randomized controlled trial

Abstract

Background

To evaluate the feasibility of an efficacy trial comparing a hydrophobic polyurethane peripherally inserted central catheter (PICC) with a standard polyurethane PICC.

Methods

This pilot randomised controlled trial (RCT) was conducted between May 2017 and February 2018. Adult participants (n = 111) were assigned to hydrophobic polyurethane PICC with proximal valve (intervention) or a polyurethane PICC with external clamp (standard care). Primary outcome was trial feasibility including PICC failure. Secondary outcomes were central line-associated bloodstream infection, local infection, occlusion, thrombosis, fracture and dislodgement, phlebitis, local or systemic allergic reaction, and PICC dwell time.

Results

All feasibility outcomes were achieved, apart from eligibility criteria. In total, 338 patients were screened, 138 were eligible (41%), and of these 111 were randomised (80%). Patients received the allocated PICC in 106 (95%) insertions. No patients withdrew from the study and there was no missing data. PICC failure was 24% (13/55) in the intervention group and 22% (12/55) in the standard care group (p = 0.820). PICC failure per 1000 PICC days was 16.3 in the intervention group and 18.4 in the control group (p = 0.755). The average dwell time was 12 days in the intervention and 8 days in the control group.

Conclusions

This study demonstrates the feasibility of an efficacy trial of PICC materials in an adult population, once adjustments were made to include not only in-patients, but also patients being discharged to the Hospital in the Home service.

Trial registration

Australia and New Zealand Clinical Trials Registry ACTRN12616001578493. Prospectively registered on 16 November 2016. The trial protocol was published a priori (Kleidon et al., Vasc Access 3:15–21, 2017).

Peer Review reports

Background

Peripherally inserted central catheters (PICCs) are used for intravenous (IV) fluids, medications, and blood products and for blood sampling to prevent frequent phlebotomy [1, 2]. PICCs are the most frequently inserted central venous access device outside of the intensive care unit [3], and their appropriate use [4, 5] enables treatment in diverse settings, including inpatient, outpatient, and community-based [2]. Unfortunately, about one third of PICCs fail prior to completion of treatment [6], often necessitating removal and replacement [7], due to mechanical (blockage, dislodgement, vein thrombosis, rupture) or infective (local or bloodstream infections) complications [6]. PICC complications reduce patient satisfaction, prolong hospitalisation, increase healthcare costs, and risk mortality [8,9,10,11,12].

PICC material and designs have evolved from the first silicone PICCs with external clamps in the 1970s [13] to modern materials and characteristics which include the following: trimmable or un-trimmable catheters, silicone or polyurethane (power injectable and non-power injectable), and anti-microbial or heparin bonding [14]. PICCs are available in various sizes (1 to 6 Fr), configured with single or multiple lumens, open- or closed-ended, with or without external clamps [14]. Despite this abundance of choice, there is a paucity of evidence comparing PICC features. In a recent scoping review [15] of 178 randomised controlled trials (RCTs) in central venous access devices in the past decade, only five studies compared PICC materials [16,17,18,19,20] and two PICC types [21, 22].

A recent addition to the market, BioFlo® is a hydrophobic polyurethane PICC with a surface-modifying macromolecule (Endexo®) that enhances the biocompatibility of medical devices [23]. This durable surface modification occurs during the extrusion moulding manufacturing process. A small amount of polymer/macromolecule is added to the polyurethane/carbothane® to provide hydrophobic properties to the PICC [23, 24]. Earlier unpublished pre-clinical data suggested the Endexo® technology suppressed procoagulant conformation, reduced platelet adhesion, inhibited platelet activation in the presence of blood, and reduced bacteria adhesion and encrustation in the presence of bacteria [23]; however, no anti-infective claim is made by the BioFlo® manufacturers [25]. In addition to surface modification, BioFlo® has a direction-specific pressure-activated safety valve situated at the proximal hub of the PICC, designed to reduce retrograde blood flow into the PICC during normal venous pressures [25]. This technology demonstrates some effectiveness at reducing thrombus accumulation in laboratory settings, but its clinical effectiveness is less established [14].

A recent pilot RCT in paediatric patients [26] compared a polyurethane, power-injected PICC with an external clamp to a hydrophobic polyurethane PICC (BioFlo®). Failure was halved in the hydrophobic polyurethane group (8/72; 11%) compared to standard care (16/74; 22%) (p = 0.087), which when expressed as an incident rate was 12.6 and 7.3 per 1000 PICC days (incident rate ratio 0.58; 95% confidence interval (CI) 0.21–1.43; p = 0.172). However, this has not been tested in the adult population. Therefore, the primary aim of our feasibility RCT was to compare the effectiveness of a hydrophobic polyurethane PICC with a standard polyurethane PICC to prevent PICC failure. We hypothesised that conducting an RCT would be feasible.

Methods

Design

This pilot parallel RCT with 1:1 allocation compared a hydrophobic polyurethane PICC with pressure-activated proximal valve to a polyurethane PICC with external clamp.

Study setting

This trial recruited between May 2017 and February 2018 at the Royal Brisbane and Women’s Hospital (RBWH) in Brisbane, Australia, a large quaternary hospital. During the study, PICCs at this hospital were inserted in the Department of Medical Imaging, which provided a specialist diagnostic imaging and radiology service, supporting the care and treatment of patients. Patients with a PICC were either be treated in a traditional hospital ward or in the Hospital in the Home (HITH) service, which allowed patients to receive their hospital intravenous treatment in their own homes.

Sample

The target sample size was 110 participants, 50 participants per group plus 10% for potential attrition, as determined by pilot trial sample size recommendations [27]. Patients were consecutively recruited and randomised if they met the inclusion criteria: PICC to be inserted in the Department of Medical Imaging for fluid or medication administration, predicted hospital inpatient admission > 24 h, 18 years or older, and provided informed consent. Due to slow recruitment, the inclusion criteria were expanded to include patients who were transferred to the HITH service within 24 h of PICC insertion. Patients were excluded if they had a current bloodstream infection (classified as a positive blood culture within 48 h prior to PICC insertion); had an allergy to the study product; the PICC was to be inserted through diseased, burned, or scarred skin; the PICC was inserted in other departments or units in the hospital; they could not provide consent without an interpreter; had a previous enrolment in the study; or had an existing central venous access device including pulmonary artery catheters.

Participants were randomised to receive either (1) standard care—polyurethane, power-injectable PICC with external clamp (Arrow International, Reading, Pennsylvania)—or (2) the intervention—hydrophobic polyurethane PICC with a proximal valve (BioFlo® with Endexo® and PASV®, AngioDynamics Inc., Queensbury).

Outcomes

Primary outcomes

Feasibility

The primary outcome was feasibility of conducting a large, definitive RCT comparing hydrophobic polyurethane PICC with simple polyurethane to prevent PICC failure and complications in adults. Feasibility was assessed as a composite analysis of elements [27,28,29] of the following: (i) eligibility (> 70% of screened patients were eligible), (ii) recruitment (> 70% of eligible patients agreed to enrol), (iii) retention and attrition (< 15% of participants are lost to follow-up or withdrew from the study), (iv) protocol adherence (> 80% of participants received their randomly assigned study product), (v) missing data (< 10% of endpoint data were missed during data collection), and (vi) patient and healthcare professional satisfaction and acceptability. Patient satisfaction with the PICC was collected as pain at removal on an 11-point Likert scale (0 being no pain and 10 being worst imaginable pain). PICC inserters were asked to rate their levels of satisfaction with insertion equipment (0 being very dissatisfied and 10 being very satisfied) and ease of PICC insertion (0 being very difficult and 10 being very easy) on an 11-point Likert scale. For staff satisfaction and acceptability, nurses were asked to rate their difficulty during PICC removal on an 11-point Likert scale (0 being no difficulty and 10 being significant difficulty); and (vii) sample size estimates for a definitive trial.

PICC failure

PICC failure was defined as the following complications associated with PICC removal: (i) central line-associated bloodstream infection (CLABSI) [30], (ii) local infection [30], (iii) device occlusion [31, 32], (iv) venous thrombosis [33], and (v) PICC fracture or dislodgement [31, 34]. The outcome of device failure was an objective measure, assigned by clinical staff (not research staff or investigators). The published protocol by Kleidon and colleagues [35] defines PICC failure in further detail.

Secondary outcomes

In addition to individual complications (CLABSI, local infection, occlusion, thrombosis, fracture, and dislodgement), secondary outcomes included the following: (i) phlebitis, defined as any sign of chemical, mechanical, or infective phlebitis, determined by patient complaint of pain and nurse examining PICC site; (ii) safety endpoints (local or systemic allergic reactions); and (iii) PICC dwell time in days.

Study procedures

A research nurse (ReN) screened for eligible patients. Patients who agreed to participate were randomised via computer-generated randomisation immediately before PICC insertion via a web-based service (https://randomisation.griffith.edu.au/) to ensure allocation concealment until study entry. Patients were randomly assigned in a 1:1 ratio with randomly varied block sizes of two and four. The ReN provided the randomised study product to the inserting clinician. Data were collected using Research Electronic Data Capture [36]. Demographic and clinical data were collected at recruitment. The ReN visited inpatients twice weekly and the HITH nurse visited daily until the PICC was removed due to treatment completion, PICC failure, or until 4 weeks (maximum follow-up). Primary and secondary outcomes were collected by the ReN from the medical records, clinical staff, and patient assessment. Outcome data were collected at 48 h post PICC removal. The infection control physician (adjudicating infection outcomes), radiologist (adjudicating thrombosis outcomes), and data analyst were blinded to the study allocation. Infection and mortality data were collected at 48 h post PICC removal. Participants were followed for 4 weeks, unless the PICC was removed earlier. It was not possible to blind the PICC inserters, patients, or healthcare professionals caring for the patients enrolled in the study.

PICC procedures

Prior to the study, the intervention PICC was not used in the hospital. Pre-study training for PICC inserters, inpatient, and HITH staff was provided by the manufacturer in the month preceding study commencement and for the first month after recruitment of the first patient. The ReN provided ongoing training and support for PICC inserters and clinical staff caring for the PICC.

PICCs were inserted by a registered nurse, radiographer, or medical officer; experience ranged from novice to expert. PICC insertion occurred in either digital fluoroscopy or the angiography suite. Asepsis was maintained using maximal barrier precautions including use of 2% chlorhexidine gluconate in 70% isopropyl alcohol skin preparation (Soluprep; 3M) [37]. Ultrasound was used for all initial venous assessment and initial vein puncture. A modified Seldinger technique was used for the remainder of the PICC insertion. Fluoroscopy was used to confirm optimal catheter tip placement in the lower third of the superior vena cava or cavo-atrial junction. When necessary, 10 mL of iodinated contrast and digital subtraction angiography was used to assist accurate tip placement during complicated PICC insertion. Dressing and securement were achieved with a chlorhexidine impregnated foam disc (Biopatch®; Ethicon, New Jersey), a sterile semi-permeable, adhesive transparent dressing (IV3000 10 × 12cm; Smith and Nephew; Hull), and a stabilisation device (StatLock® PICC Plus Stabilization Device; Bard, Georgia) [37].

Post insertion care and maintenance was standardised with the PICC dressing and needleless connectors (SmartSite™; BD, Utah) replaced weekly or as required if soiled or lifting [38]. Inpatients who had continuous infusions had their IV administration sets replaced every 72 h, excluding blood products or chemotherapy, which were replaced on completion of the infusion. Intermittent infusions were flushed and locked with 0.9% sodium chloride. Complications were identified and monitored by the treating team. Occlusion was managed with a thrombolytic agent (e.g. urokinase). The decision to take blood cultures and remove PICCs was made by the medical officer treating the patient. HITH patients were visited daily to have their elastomeric device (Infusor™; Baxter, Illinois) and antibiotic replaced. If HITH patients required treatment to manage a PICC complication, such as administration of a thrombolytic agent, they were re-admitted to the ward.

Statistical analyses

Data were exported to Stata 15 for cleaning and analysis. Data cleaning of outlying figures and missing and implausible data were undertaken prior to analysis. Missing data were not imputed. The patient was the unit of measurement, and all randomly assigned patients were analysed on an intention-to-treat basis [39]. Descriptive statistics (frequencies and percentages) were used to ascertain the primary outcome of feasibility for the larger trial. Incidence rates (per 1000 PICC days) and rate ratios, including 95% confidence intervals, were calculated. The comparability of groups at baseline was described across demographic, clinical, and device characteristics. Kaplan-Meier survival curves (with log-rank tests) were used to compare PICC failure between study groups over time. Associations between failure and baseline characteristics were described by calculating hazard ratios. Multivariable Cox regression was not performed due to the relatively low number of outcomes. p values of < 0.05 were considered statistically significant.

Results

Participant and PICC characteristics

The participant and PICC characteristics are described in Table 1. Over 90% of patients recruited to this study were medical or surgical patients. Demographic and PICC insertion characteristics were balanced between the groups. The PICC inserters were more experienced at inserting the standard care PICC: in only 1/52 (2%) PICC insertions had the PICC inserter never inserted a standard care PICC compared to 16/56 (29%) in the intervention group, and three quarters of the PICC inserters (39/52; 75%) had inserted more than 11 standard care PICCs compared to 3/56 (5%) in the intervention group.

Table 1 Participant and insertion characteristics at insertion

Feasibility outcomes

Apart from eligibility criteria, all feasibility outcomes were achieved. In total, 338 patients screened, 138 were eligible (41%), and of these 111 were randomised (80%). Figure 1 outlines the reasons for unsuccessful achievement of the eligibility criteria. Additionally, Fig. 1 displays the flow of patients through the study confirming > 80% participants received their randomly assigned study product. Of the five who did not receive the intervention (106/111 received the allocated treatment, 95%), four received the standard care PICC. No patients withdrew from the study; therefore, the retention and attrition criteria of < 15% were met. Thirteen (12%) patients were censored on transfer to another hospital (seven in the standard care and six in the intervention group). There was no missing data for the primary outcome. The research team did not have ethical approval to request clinical information for these patients transferred to another hospital.

Fig. 1
figure 1

CONSORT 2010 flow diagram

PICC inserter satisfaction with ease of insertion with the insertion equipment was rated higher in the standard care group (10.0 in the standard care group and 7.0 in the intervention) (Table 1). Although the number of insertion attempts was balanced between the groups, there were more difficult insertions in the intervention group (20; 36% in the intervention group compared to 13; 24% in the standard care group). Patient satisfaction and acceptability was collected as pain on PICC removal. Of the twenty-six patients asked, no one reported pain at removal (Table 2).

Table 2 End points

PICC failure

In total, 25 out of 110 patients (23%) experienced PICC failure: of these, two patients (one in each group) had completed their treatment and 23 had not. Presented as absolute numbers, 13/55 (24%) PICCs failed in the intervention group and 12/55 (22%) in the standard care group (p = 0.820). The risk ratio was 1.08 (0.54–2.16), and the risk difference was 1.8% (− 13.8–17.5). PICC failure per 1000 PICC days was 16.3 (95% CI 9.5–28.1) in the intervention group and 18.4 (95% CI 10.5–32.5) in the control group (p = 0.755). Failure was most commonly due to dislodgement, followed by occlusion. See Tables 2 and 3, also Fig. 2. The average dwell time was 12 days in the intervention and 8 days in the control group (Table 3). During the study, six adverse events were recorded: one skin reaction (in the intervention group), four patients were transferred to the intensive care unit (two in each group), and three patients died (one in the intervention and two in the standard care group). In all cases, the patients’ deterioration was not related to the PICC.

Table 3 Failure rates and survival analysis
Fig. 2
figure 2

Kaplan-Meier survival curve

Discussion

This study demonstrates the feasibility of an efficacy trial of PICC materials in an adult population, once adjustments were made to include patients to be discharged to the HITH service. No lead-in time with the intervention PICC was possible outside of the trial; therefore, PICC inserters were inexperienced with the new device. This study included a broad selection of medical/surgical patients who required a PICC for treatment in the hospital and/or as a HITH patient in their own homes. Patients ranged from being ambulant requiring long-term antibiotics to patients with major surgery, such as pelvic exenterations. To further improve eligibility, future trials might consider surrogate decision maker and telephone consent as 21 out of the 338 (6%) patients approached were too confused to consent. Despite these challenges, it is important that an adequately powered trial comparing the effectiveness of hydrophobic PICCs to other PICC technologies is conducted, as PICC failure remains unacceptably high. Previous systematic reviews have highlighted that one-third of PICCs fail [6] with comparable incidence seen in our study. Clinicians and policy makers need to take urgent steps to investigate potential improvements in PICC outcomes since failure disrupts patient treatment due to delay in insertion of a new PICC and multiple PICCs can increase patient complications and decrease the quality of a patient’s experience [8,9,10,11,12]. PICCs are associated with morbid complications, such as CLABSI and deep vein thrombosis; thus, it is essential researchers generate further evidence to guide clinicians to select the most appropriate PICC materials to reduce these potentially fatal adverse events and improve the overall quality of care provided to patients [14, 40].

This pilot RCT compared a hydrophobic polyurethane PICC with proximal valve with a power-injectable polyurethane PICC with external clamp to reduce PICC failure and complications in an adult population. This study followed the same methodology as the paediatric protocol [35] and published study [26] and was reported in line with the CONSORT guidelines. A quarter of PICCs failed (25/110; 23%) before treatment completion. The primary objective of this study was feasibility (rather than reducing PICC failure); and although the incident rate ratio of 0.89 favoured the intervention, PICC failure was not significantly different between the groups (12/55; 22% in the standard care and 13/55; 24% in the intervention group).

To date, six studies, published in peer reviewed journals and as conference abstracts, have compared the BioFlo® PICC with non-hydrophobic PICCs: one paediatric pilot RCT [26], one adult RCT [41], one quasi-experimental clinical evaluation [42], and three retrospective cohort studies [43,44,45]. With the exception of the paediatric pilot RCT [26], it is difficult to assess the risk of bias in the other studies published to date as they are conference abstracts. All but one [41] study demonstrated an improvement in PICC-associated thrombosis and occlusion with BioFlo®. Musial and Hamad [43] reported in their economic evaluation that despite the reduction in occlusion with BioFlo® PICC, when the increased cost of BioFlo® was compared to cost savings from reduced use of the thrombolytic treatment alteplase (Cathflo Activase, Genentec Inc., San Francisco, CA), there was no economic benefit, but they did not consider costs of lost treatment time and extended inpatient stay. Currently, there is insufficient clinical data to definitively demonstrate the potentially beneficial hydrophobic properties of BioFlo® PICC in clinical practice, thus further large trials have commenced.

Limitations of this study include the length of follow-up of patients. We followed our patients for up to of 4 weeks rather than until PICC removal. This might explain why only four incidences of occlusion resulting in PICC failure occurred (see Table 2). Another limitation was that only patients showing clinical symptoms of thrombus were referred for an ultrasound for confirmation. This may explain the low rate of PICC-associated thrombus in this cohort. Future studies should consider routine ultrasounds as approximately two-thirds of PICC-associated thrombi are asymptomatic [46]. Additionally, it was not possible to blind the PICC inserters, patient or other healthcare professionals. Despite these limitations, intervention fidelity was strong and the potential for reproducibility in future trials through the publication of the trial protocol [35] and reporting of the study and PICC procedures, which allows for generalisability of the study results.

The results of this study should be interpreted with caution as it reports the results of a pilot RCT that recruited 110 patients. A full-scale adequately powered efficacy trial is required to test the statistical hypotheses of the efficacy of a hydrophobic polyurethane PICC with proximal valve compared to a plain polyurethane PICC with external clamp to reduce PICC failure and complications. A multi-centre trial recruiting adults and children would ensure these results are generalizable beyond this single centre and demonstrate external validity and safeguarding internal validity with comparable groups at baseline.

Conclusions

In conclusion, this study demonstrated that it is feasible to conduct an RCT comparing the efficacy of hydrophobic polyurethane PICCs (with pressure-activated proximal valves) with polyurethane PICCs (with external clamps) in an adult population receiving treatment at a large quaternary referral hospital and/or HITH service. It is important that future studies evaluate cost-effectiveness, including inserter and patient acceptability, to robustly evaluate the impact of different PICC materials on failure. Additionally, this study demonstrated the importance of providing a sufficient lead-in time to train PICC inserters before the commencement of data collection as lack of familiarity with an interventional product could impact study results.

Availability of data and materials

The datasets generated and/or analysed during the current study are not publicly available due to our Human Research and Ethics Committee approval but are available from the corresponding author on reasonable request.

Abbreviations

CLABSI:

Central line-associated bloodstream infection

HITH:

Hospital in the Home

PASV®:

Pressure-activated safety valve

PICC:

Peripherally inserted central catheter

RBWH:

Royal Brisbane and Women’s Hospital

RCT:

Randomised controlled trial

ReN:

Research nurse

References

  1. Cancer Nurses Society of Australia. Central venous access devices: principles for nursing practice and education. Sydney: Cancer Nurses Society of Australia; 2007.

  2. Loveday HP, Wilson JA, Pratt RJ, Golsorkhi M, Tingle A, Bak A, et al. epic 3: National evidence-based guidelines for preventing healthcare-associated infections in NHS hospital in England. J Hosp Infect. 2014;8651:S1–S70.

    Article  Google Scholar 

  3. Chopra V, Flanders SA, Saint S. The problem with peripherally inserted central catheters. J Am Med Assoc. 2012;308(15):1527–8.

    Article  CAS  Google Scholar 

  4. Moureau N, Trick N, Nifong T, Perry C, Kelley C, Carrico R, et al. Vessal health and preservation (part 1): a new evidence-based approach to vascular access selection and management. J Vasc Access. 2012;13(3):351–6.

    Article  PubMed  PubMed Central  Google Scholar 

  5. Chopra V, Flanders SA, Saint S, Woller SC, O'Grady NP, Sadfar N, et al. The Michigan Appropriateness Guide for Intravenous Catheters (MAGIC): results from a multispecialty panel using the RAND/UCLA appropriateness method. Ann Intern Med. 2015;163(6):S1–S39.

    Article  PubMed  Google Scholar 

  6. Rickard CM, Marsh NM, Webster J, Gavin NC, Chan RJ, McCarthy AL, et al. Peripherally InSerted CEntral catheter dressing and securement in patients with cancer: the PISCES trial. Protocol for a 2x2 factorial, superiority randomised controlled trial. BMJ Open. 2017;7:e015291.

    Article  PubMed  PubMed Central  Google Scholar 

  7. Jennings K, Cann T, Smyth W. Peripherally inserted central catheter complications highlight the need for ongoing support: results of a chart audit. Healthc Infect. 2011;16(3):95–9.

    Article  Google Scholar 

  8. Ng F, Mastoroudes H, Paul E, Davies N, Tibballs J, Hochhauser D, et al. A comparison of Hickman line- and Port-a-Cath-associated complications in patients with solid tumours undergoing chemotherapy. Clin Oncol (R Coll Radiol). 2007;19(7):551–6.

    Article  CAS  Google Scholar 

  9. Lim MY, Al-Kali A, Ashrani AA, Begna KH, Elliott MA, Hogan WJ, et al. Comparison of complication rates of Hickman catheters versus peripherally inserted central catheters in patients with acute myeloid leukemia undergoing induction chemotherapy. Leuk Lymphoma. 2013;54(6):1263–7.

    Article  PubMed  Google Scholar 

  10. Turcotte S, Dube S, Beauchamp G. Peripherally inserted central venous catheters are not superior to central venous catheters in the acute care of surgical patients on the ward. World J Surg. 2006;30(8):1605–19.

    Article  PubMed  Google Scholar 

  11. Cowl CT, Weinstock JV, Al-Jurf A, Ephgrave K, Murray JA, Dillon K. Complications and cost associated with parenteral nutrition delivered to hospitalized patients through either subclavian or peripherally-inserted central catheters. Clin Nutr. 2000;19(4):237–43.

    Article  CAS  PubMed  Google Scholar 

  12. Napalkov P, Felici DM, Chu LK, Jacobs JR, Begelman SM. Incidence of catheter-related complications in patients with central venous or hemodialysis catheters: a health care claims database analysis. BMC Cardiovasc Disord. 2013;13:86.

    Article  PubMed  PubMed Central  Google Scholar 

  13. Gow KW, Tapper D, Hickman RO. Between the lines: the 50th anniversary of long-term central venous catheters. Am J Surg. 2017;213(5):837–48.

    Article  PubMed  Google Scholar 

  14. Ullman AJ, Bulmer AC, Dargaville TR, Rickard CM, Chopra V. Antithrombogenic peripherally inserted central catheter: overview of efficacy and safety. Expert Re Med Devices. 2019;16(1):25–33.

  15. Takashima M, Ray-Barruel G, Ullman A, Keogh S, Rickard CM. Randomized controlled trials in central vascular access devices: a scoping review. PLoS One. 2017;12(3):e0174164.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  16. Itkin M, Mondshein JI, Stavropoulos SW, Shlansky-Goldberg RD, Soulen MC, Trerotola SO. Peripherally inserted central catheter thrombosis - reverse tapered versus nontapered catheters: a randomized controlled study. J Vasc Interv Radiol. 2014;25(1):85–91.

    Article  PubMed  Google Scholar 

  17. Miyagaki H, Nakajima K, Hara J, Yamasaki M, Kurokawa Y, Miyata H, et al. Performance comparison of peripherally inserted central venous catheters in gastrointestinal surgery: a randomized controlled trial. Clin Nutr. 2012;31(1):48–52.

    Article  PubMed  Google Scholar 

  18. Pittiruti M, Emoli A, Porta P, Marche B, DeAngelis R, Scoppettuolo G. A prospective, randomized comparison of three different types of valved and nonvalved peripherally inserted central catheters. J Vasc Access. 2014;15(6):519–23.

    Article  PubMed  Google Scholar 

  19. Alport B, Burbridge B, Lim H. Bard PowerPICC Solo2 vs Cook Turbo-Ject: a tale of two PICCs. Can Assoc Radiol J. 2012;63(4):323–8.

    Article  PubMed  Google Scholar 

  20. Ong CK, Venkatesh SK, Lau GB, Wang SC. Prospective randomized comparative evaluation of proximal valve polyurethane and distal valve silicone peripherally inserted central catheters. J Vasc Interv Radiol. 2010;21(8):1191–6.

    Article  PubMed  Google Scholar 

  21. Identifying complications of central venous catheters: infection, thrombosis, occlusion. Prescrire Int. 2009;18(102):173–4. https://pubmed.ncbi.nlm.nih.gov/19746562/.

  22. Hosseini MB, Jodeiri B, Mahallei M, Abdoli-Oskooi S, Safari A, Salimi Z. Early outcome of peripherally inserted central catheter versus peripheral IV line in very low birth weight neonates. Feyz J Kashan Univ Med Sci. 2014;17(6):561–7.

    Google Scholar 

  23. Interface Biologics. Surface modification technology platform. Available from: http://www.interfacebiologics.com/endexo.htm. Accessed 28 Aug 2019.

  24. Lareau R, Bell B, Santerre JP, Ho J, inventors; AngioDynamics Inc., assignee. Catheteres with high-purity fluopolymer additives patent 8876797. 2014.

    Google Scholar 

  25. AngioDynamics. Vascular Access BioFlo PICC. Available from: https://www.angiodynamics.com/products/23/BioFlo-PICC/. Accessed 28 Aug 2019.

  26. Kleidon T, Ullman AJ, Zhang L, Mihala G, Chaseling B, Schoutrop J, et al. How does your PIC COMPARE? A pilot randomized controlled trial comparing various PICC materials in pediatrics. J Hosp Med. 2018;13(8):517–25.

    PubMed  Google Scholar 

  27. Hertzog MA. Considerations in determining sample size for pilot studies. Res Nurs Health. 2008;31(2):180–91.

    Article  PubMed  Google Scholar 

  28. Thabane L, Ma J, Chu R, Cheng J, Ismaila A, Rios LP, et al. A tutorial on pilot studies: the what, why and how. BMC Med Res Methodol. 2010;10(1):1.

    Article  PubMed  PubMed Central  Google Scholar 

  29. Lancaster GA, Dodd S, Williamson PR. Design and analysis of pilot studies: recommendations for good practice. J Eval Clin Pract. 2004;10(2):307–12.

    Article  PubMed  Google Scholar 

  30. Centers for Disease Control and Prevention. National Healthcare Safety Network Patient Safety Component Manual. 2018. Available from: https://www.cdc.gov/nhsn/pdfs/pscmanual/pcsmanual_current.pdf.

    Google Scholar 

  31. Chow LML, Friedman JN, MacArthur C, Restrepo R, Temple M, Chait PG, et al. Peripherally inserted central catheter (PICC) fracture and embolozation in the pediatric population. Pediatrics. 2003;142(2):141–4.

    Article  Google Scholar 

  32. Smith SN, Moureau N, Vaughn VM, Boldenow T, Kaatz S, Grant PJ, et al. Patterns and predictors of peripherally inserted central catheter occlusion: the 3P-O study. J Vasc Interv Radiol. 2017;28(5):749–56.e2.

    Article  PubMed  Google Scholar 

  33. Lobo BL, Vaidean G, Broyles J, Reaves AB, Shorr RI. Risk of venous thromboembolism in hospitalized patients with peripherally inserted central catheters. J Hosp Med. 2009;4(7):417–22.

    Article  PubMed  Google Scholar 

  34. Chopra V, Kuhn L, Ratz D, Flanders SA, Krein SL. Vascular nursing experience, practice knowledge, and beliefs: results from the Michigan PICC1 survey. J Hosp Med. 2016;11(4):269–75.

    Article  PubMed  Google Scholar 

  35. Kleidon P, Ullman A, Northfield S, Zhang L, Mihala G, Gavin N, et al. Peripherally Inserted Central Catheter OutcoMes PolyurethAne veRsus Endexo: the PICCOMPARE trial. Protocol for a randomised controlled trial. Vasc Access. 2017;3(1):15–21.

    Google Scholar 

  36. Harris PA, Taylor R, Thielke R, Payne J, Gonzalez N, Conde JG. Research electronic data capture (REDCap) - a metadata-driven methodology and workflow process for providing translational research informatics support. J Biomed Inform. 2009;42(2):377–81.

    Article  PubMed  Google Scholar 

  37. Metro North Hospital and Health Service. Peripherally inserted central catheter (PICC) insertion resource package. Brisbane: Metro North Hospital and Health Service; 2019.

    Google Scholar 

  38. Metro North Hospital and Health Service. Central venous access devices (CVAD) management - paediatric and adult resource package. Brisbane: Metro North Hospital and Health Service; 2019.

    Google Scholar 

  39. Rothwell PM. Factors that can affect the external validity of randomised controlled trials. PLos Clin Trials. 2006;1(1):e9.

    Article  PubMed  PubMed Central  Google Scholar 

  40. Kramer RD, Rogers MAM, Conte M, Mann J, Saint S, Chopra V. Are antimicrobial peripherally inserted central catheters associated with reduction in central line–associated bloodstream infection? A systematic review and meta-analysis. Am J Infect Control. 2017;45(2):108–14.

    Article  PubMed  Google Scholar 

  41. Yoon H, Drabkin M, Loya M, Patel C, Saif A, Shah S. Prospective randomized evaluation of complications with Endexo PICC Technology (PRECEPT). J Vasc Interv Radiol. 2016;27(3):S283.

    Article  Google Scholar 

  42. Hill J. Clinical evaluation of the Navilyst PICC; 2017.

    Google Scholar 

  43. Musial ER, Hamad L, Wang C, Hare R. Alteplase use in surface-modified peripherally inserted central catheters in a National Cancer Institute-designated comprehensive cancer center: a pharmacoeconomic analysis. J Vasc Access. 2016;21(1):39–43.

    Article  Google Scholar 

  44. Simcock L, Salvador J, Patel B, McInerney K. Reduced incidence of clinically evident PICC-related DVT in sarcoma patients. Copenhagen: World Congress of Vascular Access; 2018.

    Book  Google Scholar 

  45. Pain J. Reducing peripherally inserted central catheter related upper extremity deep vein thrombosis (UEDVT) and occlusions - a retrospective case controlled study. Copenhagen: World Congress of Vascular Access; 2018. p. NP1–NP85.

    Google Scholar 

  46. Wall C, Moore J, Thachill J. Catheter-related thrombosis: a practical approach. J Intensive Care Soc. 2016;17(2):160–7.

    Article  PubMed  Google Scholar 

Download references

Acknowledgments

The authors thank the patients at the RBWH for participating in this research study. We appreciate the support from the nurses, radiographers, and medical officers in the Department of Medical Imaging and the nurses from the Vascular Access Surveillance and Education and the participating wards for their support.

Funding

AngioDynamics (the BioFlo® PICC manufacturer) provided partial funds to undertake this research with an unrestricted donation to Griffith University (but not to the study authors). Queensland Health provided in kind support to fund the remainder of the trial. The funders had no role in the study design, collection, analysis, or interpretation of the data, writing of the report, or decision to submit the article for publication.

Author information

Authors and Affiliations

Authors

Contributions

NG contributed to the acquisition, analysis, and interpretation of data and drafted the first version; TK contributed to the design of the study and the analysis and interpretation of data; EL contributed to the analysis and interpretation of data; COB contributed to the acquisition, analysis, and interpretation of data; AU contributed to the design of the study and the analysis and interpretation of data; SN contributed to the design of the study; GM contributed to the analysis and interpretation of data; NR contributed to the analysis and interpretation of data; NM contributed to the analysis and interpretation of data; CR contributed to the design of the study and the analysis and interpretation of data. All authors substantially revised the publication and have read and approved the submitted version. All authors agree to be personally accountable for the author’s own contributions and to ensure that questions related to the accuracy or integrity of any part of the work, even ones in which the author was not personally involved, are appropriately investigated, resolved, and the resolution documented in the literature.

Corresponding author

Correspondence to Nicole C. Gavin.

Ethics declarations

Ethics approval and consent to participate

The trial was approved by the Children’s Health Queensland Hospital and Health Service Human Research Ethics Committee (HREC/15/QRCH/164) and Griffith University Human Research Ethics Committee (Ref No. 2016/077). Written informed consent was obtained from all participants prior to entry into the study.

Consent for publication

Not applicable

Competing interests

NG: none; TK: related to the submitted project, TK reports an investigator-initiated research grant from Angiodynamics. TK reports investigator-initiated research grants and speaker fees provided to Griffith University from 3M Medical, Angiodynamics, Baxter, Becton Dickinson, Centurion Medical, Cook Medical, Medical Specialties Australia, Smiths Medical, and Vygon (unrelated to the current project); EL: Griffith University has received consultancy payment for an educational lecture from 3M on behalf of EL; an investigator-initiated research grant from Cardinal Health (formerly Medtronic); and a conference scholarship attendance supported by Angiodynamics (unrelated to the submitted work) to support EL’s research. EL is a recipient of a Higher Degree Research Scholarship from the Australian Government Research Training Program; COB: none; AU: related to the submitted project, AU reports an investigator-initiated research grant from Angiodynamics. AU reports investigator-initiated research grants and speaker fees provided to Griffith University from 3M Medical, Becton Dickinson and Cardinal Health (unrelated to the current project). AU is currently supported by an Australian Government, National Health and Medical Research Council Fellowship, via Griffith University; SN: related to the submitted project, SN reports an investigator-initiated research grant from Angiodynamics; GM: related to the submitted project, GM reports an investigator-initiated research grant from Angiodynamics. GM is a recipient of a Higher Degree Research Scholarship from the Australian Government Research Training Program; NR: none; NM: NM’s previous employer (Griffith University) has received on her behalf: investigator-initiated research grants and unrestricted educational grants from Becton Dickinson and Cardinal Health and a consultancy payment provided to Griffith University from Becton Dickinson for clinical feedback related to catheter placement and maintenance (unrelated to the current project); CM: Griffith University has received unrestricted investigator-initiated research or educational grants on my behalf from product manufacturers (BD-Bard; Cardinal Health,). Griffith University has received consultancy payments on my behalf from manufacturers (3M, BBraun, BD-Bard).

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Gavin, N.C., Kleidon, T.M., Larsen, E. et al. A comparison of hydrophobic polyurethane and polyurethane peripherally inserted central catheter: results from a feasibility randomized controlled trial. Trials 21, 787 (2020). https://doi.org/10.1186/s13063-020-04699-z

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13063-020-04699-z

Keywords