Impact of collimator leaf width and treatment technique on stereotactic radiosurgery and radiotherapy plans for intra- and extracranial lesions

Abstract

Background

This study evaluated the dosimetric impact of various treatment techniques as well as collimator leaf width (2.5 vs 5 mm) for three groups of tumors – spine tumors, brain tumors abutting the brainstem, and liver tumors. These lesions often present challenges in maximizing dose to target volumes without exceeding critical organ tolerance. Specifically, this study evaluated the dosimetric benefits of various techniques and collimator leaf sizes as a function of lesion size and shape.

Methods

Fifteen cases (5 for each site) were studied retrospectively. All lesions either abutted or were an integral part of critical structures (brainstem, liver or spinal cord). For brain and liver lesions, treatment plans using a 3D-conformal static technique (3D), dynamic conformal arcs (DARC) or intensity modulation (IMRT) were designed with a conventional linear accelerator with standard 5 mm leaf width multi-leaf collimator, and a linear accelerator dedicated for radiosurgery and hypofractionated therapy with a 2.5 mm leaf width collimator. For the concave spine lesions, intensity modulation was required to provide adequate conformality; hence, only IMRT plans were evaluated using either the standard or small leaf-width collimators.

A total of 70 treatment plans were generated and each plan was individually optimized according to the technique employed. The Generalized Estimating Equation (GEE) was used to separate the impact of treatment technique from the MLC system on plan outcome, and t-tests were performed to evaluate statistical differences in target coverage and organ sparing between plans.

Results

The lesions ranged in size from 2.6 to 12.5 cc, 17.5 to 153 cc, and 20.9 to 87.7 cc for the brain, liver, and spine groups, respectively. As a group, brain lesions were smaller than spine and liver lesions. While brain and liver lesions were primarily ellipsoidal, spine lesions were more complex in shape, as they were all concave. Therefore, the brain and the liver groups were compared for volume effect, and the liver and spine groups were compared for shape. For the brain and liver groups, both the radiosurgery MLC and the IMRT technique contributed to the dose sparing of organs-at-risk(OARs), as dose in the high-dose regions of these OARs was reduced up to 15%, compared to the non-IMRT techniques employing a 5 mm leaf-width collimator. Also, the dose reduction contributed by the fine leaf-width MLC decreased, as dose savings at all levels diminished from 4 – 11% for the brain group to 1 – 5% for the liver group, as the target structures decreased in volume. The fine leaf-width collimator significantly improved spinal cord sparing, with dose reductions of 14 – 19% in high to middle dose regions, compared to the 5 mm leaf width collimator.

Conclusion

The fine leaf-width MLC in combination with the IMRT technique can yield dosimetric benefits in radiosurgery and hypofractionated radiotherapy. Treatment of small lesions in cases involving complex target/OAR geometry will especially benefit from use of a fine leaf-width MLC and the use of IMRT.

Background

Stereotactic intracranial radiosurgery (SRS) and extracranial body radiosurgery and radiotherapy (SBRT) are characterized by ablative, high dose irradiation of target structures. Complex targets, such as spine metastases, brain lesions abutting the brain stem, and liver lesions present challenges in maximizing the dose to the target volume while not exceeding the critical organ tolerances [1–20]. This study evaluates the benefits of a dedicated radiosurgery system with a fine leaf-width collimator for these different groups of patients. More specifically, this study evaluates the dosimetric benefits of this system as a function of target size and shape complexity.

Materials and methods

Patient Data

This retrospective study included 15 cases (5 each for brain, liver and spine sites) treated with SRS/SBRT at our institution. As shown in Additional file 1, these lesions ranged in size from 2.6 – 12.5 cc, 17.5 – 153.1 cc, and 20.9 – 87.7 cc for the brain, liver, and spine groups, respectively. The volumes selected were intended to represent the ranges of the target volumes typically encountered for these sites. The selected cases all involved lesions next to or within critical structures, i.e., lesions next to the brainstem for the brain group, lesions within the liver for the liver group, and tumors near/abutting the spinal cord for the spine group. The brain and liver lesions were mostly ellipsoidal, with brain lesions having smaller volumes. The spine lesions were more complex in shape, as they were all concave; volumes were similar to the liver lesions. Therefore, the brain and the liver groups were compared for volume effect and the liver and the spine groups were compared for shape effect. Figure 1 displays the rendering of the target volumes and critical structures in 3D space for these three sites.

Figure 1
figure 1

Examples of lesions and adjacent critical structures/organs for the brain (a), the liver(b), and the spine (c) groups. The 3D rendering of the geometrical relationships is shown in (d) for the brain case (left), liver case (middle) and spine case (right).

Treatment planning

Treatment plans were designed with the Novalis Tx radiosurgery system (BrainLab AG, Munich, Germany and Varian Medical Systems, Palo Alto, CA) versus the standard Varian 2100Clinac system (Varian Medical Systems, Palo Alto, CA), which served as the baseline for comparison. The Novalis Tx radiosurgery system became commercially available in early 2008, and was equipped with a newly designed micro-mulitleaf collimator (HD120 MLC) system, replacing the standard Millennium MLC system. In contrast to previous micro-MLC systems [2, 21–24], the Novalis Tx radiosurgery MLC is within the gantry housing, instead of being an add-on tertiary system [2, 21–24]. The mounting of an add-on tertiary MLC not only prolongs the treatment procedure, but also reduces the clearance between the gantry and couch, therefore limiting the freedom to select certain beam angles. The Novalis Tx HD120 MLC has 32 leaf pairs in the center, each a with leaf width of 2.5 mm (projected at the isocenter) and 14 leaf pairs on each side (a total of 28) with a leaf width of 5 mm; thus, the total number of leaves is 120. Our initial measurements also showed a sharper penumbra. The HD120 MLC leaf side penumbra was 2.5 mm vs. 2.8 mm with the standard MLC, and the HD120 MLC leaf end penumbra was 2.8 mm vs. 3.6 mm with the standard MLC[25]. For conformal static treatment (3D) and dynamic conformal arc treatment (DARC) treatments, the maximum dose rate with the Novalis Tx system is 1000 MU/min versus 600 MU/min with the Clinac system, allowing faster radiation delivery.

For each case, the normal structures and OARs were contoured by a physician with expertise in SRS/SBRT. The target volumes were contoured by the same physician. For the brain lesions, the planning target volume (PTV) was generated by expanding the contrast-enhancing T1-weighted MRI volume by 1-mm, except at the junctions of the tumor and brainstem where no expansion was added. The liver lesions were treated with deep inhale breath-hold technique and cone-beam CT (CBCT) guidance, and the PTV was obtained by expanding the lesion volume by 5 mm right-left and anterior-posterior and 7 mm superior-inferior. The PTV expansion for spine lesions was also non-uniform, usually 3 mm except at the lesion and cord interface, where 0 – 1 mm expansion was used.

Treatment plans were generated using three techniques: 3D, DARC, and IMRT. For the brain and liver lesions, plans using all three techniques were designed. For each lesion and planning technique, plans were generated using both the standard MLC and the HD120 MLC. All the spine lesions exhibited concave shapes, and the 3D and the DARC techniques could not provide clinically adequate conformality. Hence, only IMRT plans were generated for the spine lesions, again using both the standard MLC and HD120 MLC. Typically, the 3D plans used 6–12 beams, DARC plans used 4 – 7 arcs and IMRT plans used 4–12 static beams. The multiple arcs for the brain lesions were mainly designed to take advantage of different couch angles and those for the liver lesions utilized different beam weightings through different sections of anatomy. Each plan was individually optimized according to the treatment techniques selected. Beam angles (non-coplanar or coplanar) were chosen to minimize doses to the critical structures and to achieve high dose fall-off around the target at the same time. For IMRT planning, planning objectives included dose uniformity to the PTV and dose constraints for the OARs as well as dose falloff at the target boundary. For each plan, 90% of the prescription dose covered at least 97% of the PTV.

In summary, a total of 70 treatment plans were designed and each plan was individually optimized according to the technique employed. For the purpose of the dosimetric analysis in this study, the prescription dose was set to a nominal 12.5 Gy in a single fraction for the brain lesions, three 12-Gy fractions totaling 36 Gy for the liver lesions, and 18 Gy in a single fraction for the spine lesions, respectively.

Dosimetric evaluation parameters and statistical analysis

Each treatment plan was evaluated with respect to target coverage criteria and OAR sparing criteria. For targets, the mean PTV doses, as well as the minimum and maximum doses to the PTVs, were compared. The maximum dose (Dmax) was defined as the maximum dose value that covers 1% of the target volume (i.e. D1) and the minimum dose (Dmin) was defined as the minimum dose value that covers 99% of the target volume (i.e. D99). The dosimetric metrics for the OARs were Dmax, Dmean, D5 and D1 for the brainstem; Dmean, D10 and D30 for the liver; and Dmax, Dmean and D10 for the spine cases, respectively.

We generalized the shape into two basic categories: the round/ellipsoidal shapes which are representative of all the targets in the brain and the liver group, and the concave shapes which are representative of all the spine cases, as shown in figure 1. We also generalized the volumes into two major categories, the small volume sizes which are generally seen in the intracranial group and the large volume sizes which are generally seen in the extracranial group, as shown in Additional file 1.

Both the treatment technique and the MLC system can impact the dosimetric outcome for the brain and the liver groups. Therefore, the Generalized Estimating Equation technique (GEE) [26, 27] was used to separate their individual influences and to analyze their interactions for dosimetry outcomes. The analysis started by testing for the two-factor interaction, followed either by testing for the individual factor effect if no interaction was detected at critical value of 0.05, or by conducting the pair-wise group comparison if the interaction was significant at the 0.05 level. For the spine lesions, a paired-t test was performed to evaluate the difference between the HD120 MLC and the standard MLC on dosimetric parameters for IMRT.

Results

PTV coverage

For each case, the PTV coverage was essentially equivalent among different plans. Additional file 2 lists the dosimetric indices for all cases. The small standard deviations for all plans indicate consistency in applying the treatment techniques to achieve the optimal dose coverage; i.e., the prescription isodose covered at least 97% of the target volume and the dose heterogeneity inside the target was kept at about 10%. For each individual case, the Dmin was within 2% and Dmax was within 2.5% among different plans. Using the 3D treatment technique with standard MLC as baseline, Figure 2 displays the percentage difference between plans using other treatment techniques and the HD120 MLC system and the baseline plans.

Figure 2
figure 2

The percentage difference of PTV coverage between plans using other treatment techniques and MLC systems and the baseline plans, the 3D treatment technique with standard MLC.

Compared to the standard MLC system, the HD120 MLC improved the dose homogeneity for the brain group, with larger Dmin (p < 0.01) and smaller Dmax (p < 0.01). In contrast, the HD120 MLC had no significant impact on Dmin, Dmax, or Dmean values for either the liver group or the spine group. The use of intensity modulation also significantly reduced the Dmax values for the liver group (p <0.01). Although statistically significant, all the above differences were quite small (< 2%) and may likely have little clinical significance.

OAR dose sparing

Again, using the 3D treatment technique with the standard MLC as the baseline, Figure 3 displays the percentage difference between the baseline and plans using other treatment techniques and/or the HD120 MLC system

Figure 3
figure 3

The percentage difference of OAR sparing between plans using other treatment techniques and MLC systems and the baseline plans, the 3D treatment technique with standard MLC.

For the brain group, the HD120 MLC and the use of intensity modulation contributed jointly to dose sparing of the brainstem, as a positive interaction of these two factors was detected with GEE (p = 0.003). First, the intensity modulation technique showed significant sparing for the brainstem, compared to other treatment techniques (Additional file 3). Specifically, IMRT produced the greatest dose reduction in the high-dose region, as reflected by the D1 and D5 doses. The reduction in D1 was about 10% and 9%, and in D5 was about 18% and 14%, when 3D and DARC techniques, respectively, were replaced by the IMRT technique. Secondly, for the IMRT technique the HD120 MLC also reduced all brainstem doses (p = 0.04) relative to the standard MLC. The improved field shaping with the HD120 MLC also helped to reduce the D5 dose for all techniques by 3 – 9% (p = 0.003). Third, the HD120 MLC combined with IMRT jointly benefited dose reduction in the middle dose range, reflected by the D10 dose, (p = 0.001). As a result, the mean dose to the brainstem was reduced by 35%, between the best (IMRT with the HD120 MLC) and the worst (3D with standard MLC) plans.

Similarly, the HD120 MLC and the IMRT techniques contributed jointly to the dose sparing for the liver group (Additional file 4). The HD120 MLC reduced doses at both D10 and D30, as well as the mean dose (p < 0.01). The dose reduction (at all levels) attributable to the HD120 MLC was between 3 – 5% with the 3D and DARC techniques and between 1 – 2% with the IMRT technique. When comparing treatment techniques, IMRT plans were significantly better than either 3D or DARC plans. The IMRT technique improved the dose sparing at all levels, as the D10, D30 and Dmean indices were reduced by over 12% over the 3D/DARC techniques.

For the spine group, using the HD120 MLC substantially improved cord sparing (p < 0.01) as the ability to map the dose to the concave-shaped target improved (Additional file 5). Figure 4 displays dose distribution for a spine case. As shown, the dose fall off is much steeper at the target-cord junction for the smaller leaves. The overall dose reduction was 14%, 19% and 29% for the D1, D10 and Dmean dose indices, respectively.

Figure 4
figure 4

Example dose distribution of a spine case, IMRT plans with (a) standard MLC system and (b) with HD120 MLC system. The black circles indicate the regions where dose fall off being significantly different between the two plans.

Effect of target shape and volume size

The effect of target volume was compared with the brain and the liver group. As stated earlier, the targets were in general round or ellipsoidal for both the brain and liver lesions, and the most significant differences between these two groups were the target and OAR volumes. The lesions in the brain were several times smaller than those in the liver, and the size of the brainstem was also several times smaller than the liver. The dose reduction (combining all levels) contributed by the HD120 MLC decreased, from 4 – 11% to 1 – 5%, as target sizes increased from the brain group to the liver group.

The effect of target shape was compared with the spine and the liver group. While the target volumes were similar for the liver and the spine groups, the target shapes were much more complex in the spine group (concave vs. round or ellipsoidal for the liver). In contrast to the moderate contribution for liver lesions, the HD120 MLC significantly improved cord sparing for the spine group. By using the 120HD MLC, the dose to the cord was reduced on average by 19% to 14% at high to middle dose levels (D1 to D10, respectively) for the spine plans.

Discussion

This study was designed to provide similar target coverage for all plans. The prescription dose was set to the isodose line at the periphery of the PTV that covered at least 97% of the target for all plans. Using this strategy, the DVH to the target volumes was very similar using different treatment planning techniques, yielding dose indices that varied less than 2.5% for the same lesion.

The results from this study suggested that the degree of improved organ sparing varied with target size and shape. The targets in the brain and liver groups are similar in shape (round or ellipsoidal) differed substantially in volume (small vs. large). The lesions in the brain were several times smaller than those in the liver, and the volume of the brainstem was also several times smaller than that of the liver itself. The HD120 MLC yields a better match of the beam aperture to the target projection; however, its benefits become less noticeable as the target/OAR becomes larger. Therefore, the dose reduction contributed by the HD120 MLC decreased, from 4 – 11% to 1 – 5%, as target sizes increased between the brain and liver groups. On the other hand, the targets in the liver and spine groups were of roughly the same size but fell into different shape groups (round vs. concave, respectively). The concave lesion shape presented a challenge for conventional techniques to provide adequate target coverage and optimal organ sparing. Clinically, all spine SBRT lesions in our institution are planned with IMRT, owing to the ability to manipulate the intensity at virtually the voxel level using this technique. Since spinal cords were always adjacent to the target volume, the ability to manipulate the radiation beam with greater precision via the high-definition MLC leaves helped reduce the doses the cord received. Therefore, in contrast to the moderate benefit for liver lesions, the HD120 MLC significantly improved cord sparing for the spine group, realizing a 14%–19% dose reduction in D1 and D10, respectively.

Conclusion

The finer HD120 MLC in combination with IMRS provides significant dosimetric benefits for SRS/SBRT. Sparing of the OARs is dependent on the lesion and critical organ size and shape complexity. Small lesions (such as brain lesions treated with SRS) and complex target/OAR geometry (such as the spine lesions encountered in SBRT) will benefit most from the finer-leaf collimator and treatment planning capabilities provided by a dedicated radiosurgery system, compared to larger and more rounded or regularly shaped target volumes. Prospective clinical trials with comprehensive data collection should be conducted to determine whether these dosimetric advantages translate into clinically significant benefits.

References

  1. Monk JE, Perks JR, Doughty D, Plowman PN: Comparison of a micro-multileaf collimator with a 5-mm-leaf-width collimator for intracranial stereotactic radiotherapy. Int J Radiat Oncol Biol Phys 2003, 57: 1443-1449.

    Article  PubMed  Google Scholar 

  2. Jin JY, Yin FF, Ryu S, Ajlouni M, Kim JH: Dosimetric study using different leaf-width MLCs for treatment planning of dynamic conformal arcs and intensity-modulated radiosurgery. Med Phys 2005, 32: 405-411. 10.1118/1.1842911

    Article  PubMed  Google Scholar 

  3. Perks JR, El-Hamri K, Blackburn TP, Plowman PN: Comparison of radiosurgery planning modalities for acoustic neuroma with regard to conformity and mean target dose. Stereotact Funct Neurosurg 2005, 83: 165-171. 10.1159/000089987

    Article  CAS  PubMed  Google Scholar 

  4. Perks JR, St George EJ, El Hamri K, Blackburn P, Plowman PN: Stereotactic radiosurgery XVI: Isodosimetric comparison of photon stereotactic radiosurgery techniques (gamma knife vs. micromultileaf collimator linear accelerator) for acoustic neuroma – and potential clinical importance. Int J Radiat Oncol Biol Phys 2003, 57: 1450-1459.

    Article  PubMed  Google Scholar 

  5. Dvorak P, Georg D, Bogner J, Kroupa B, Dieckmann K, Potter R: Impact of IMRT and leaf width on stereotactic body radiotherapy of liver and lung lesions. Int J Radiat Oncol Biol Phys 2005, 61: 1572-1581.

    Article  PubMed  Google Scholar 

  6. Flickinger JC, Loeffler JS, Larson DA: Stereotactic radiosurgery for intracranial malignancies. Oncology (Williston Park) 1994, 8: 81-86.

    CAS  Google Scholar 

  7. Larson DA, Flickinger JC, Loeffler JS: The radiobiology of radiosurgery. Int J Radiat Oncol Biol Phys 1993, 25: 557-561.

    Article  CAS  PubMed  Google Scholar 

  8. Solberg TD, Boedeker KL, Fogg R, Selch MT, DeSalles AA: Dynamic arc radiosurgery field shaping: a comparison with static field conformal and noncoplanar circular arcs. Int J Radiat Oncol Biol Phys 2001, 49: 1481-1491.

    Article  CAS  PubMed  Google Scholar 

  9. Benedict SH, Cardinale RM, Wu Q, Zwicker RD, Broaddus WC, Mohan R: Intensity-modulated stereotactic radiosurgery using dynamic micro-multileaf collimation. Int J Radiat Oncol Biol Phys 2001, 50: 751-758.

    Article  CAS  PubMed  Google Scholar 

  10. Cardinale RM, Benedict SH, Wu Q, Zwicker RD, Gaballa HE, Mohan R: A comparison of three stereotactic radiotherapy techniques; ARCS vs. noncoplanar fixed fields vs. intensity modulation. Int J Radiat Oncol Biol Phys 1998, 42: 431-436.

    Article  CAS  PubMed  Google Scholar 

  11. Yin FF, Ryu S, Ajlouni M, Zhu J, Yan H, Guan H, Faber K, Rock J, Abdalhak M, Rogers L, et al.: A technique of intensity-modulated radiosurgery (IMRS) for spinal tumors. Med Phys 2002, 29: 2815-2822. 10.1118/1.1521722

    Article  PubMed  Google Scholar 

  12. Nelson JW, Yoo DS, Sampson JH, Isaacs RE, Larrier NA, Marks LB, Yin FF, Wu QJ, Wang Z, Kirkpatrick JP: Stereotactic Body Radiotherapy for Lesions of The Spine and Paraspinal Regions. Int J Radiat Oncol Biol Phys 2008.

    Google Scholar 

  13. Chang BK, Timmerman RD: Stereotactic body radiation therapy: a comprehensive review. Am J Clin Oncol 2007, 30: 637-644. 10.1097/COC.0b013e3180ca7cb1

    Article  PubMed  Google Scholar 

  14. Chang EL, Shiu AS, Mendel E, Mathews LA, Mahajan A, Allen PK, Weinberg JS, Brown BW, Wang XS, Woo SY, et al.: Phase I/II study of stereotactic body radiotherapy for spinal metastasis and its pattern of failure. J Neurosurg Spine 2007, 7: 151-160. 10.3171/SPI-07/08/151

    Article  PubMed  Google Scholar 

  15. Timmerman RD, Kavanagh BD, Cho LC, Papiez L, Xing L: Stereotactic body radiation therapy in multiple organ sites. J Clin Oncol 2007, 25: 947-952. 10.1200/JCO.2006.09.7469

    Article  PubMed  Google Scholar 

  16. Kavanagh BD, McGarry RC, Timmerman RD: Extracranial radiosurgery (stereotactic body radiation therapy) for oligometastases. Semin Radiat Oncol 2006, 16: 77-84. 10.1016/j.semradonc.2005.12.003

    Article  PubMed  Google Scholar 

  17. Timmerman RD, Forster KM, Chinsoo Cho L: Extracranial stereotactic radiation delivery. Semin Radiat Oncol 2005, 15: 202-207. 10.1016/j.semradonc.2005.01.006

    Article  PubMed  Google Scholar 

  18. Ryu S, Jin R, Jin JY, Chen Q, Rock J, Anderson J, Movsas B: Pain control by image-guided radiosurgery for solitary spinal metastasis. J Pain Symptom Manage 2008, 35: 292-298. 10.1016/j.jpainsymman.2007.04.020

    Article  PubMed  Google Scholar 

  19. Jin JY, Chen Q, Jin R, Rock J, Anderson J, Li S, Movsas B, Ryu S: Technical and clinical experience with spine radiosurgery: a new technology for management of localized spine metastases. Technol Cancer Res Treat 2007, 6: 127-133.

    Article  PubMed  Google Scholar 

  20. Ryu S, Jin JY, Jin R, Rock J, Ajlouni M, Movsas B, Rosenblum M, Kim JH: Partial volume tolerance of the spinal cord and complications of single-dose radiosurgery. Cancer 2007, 109: 628-636. 10.1002/cncr.22442

    Article  PubMed  Google Scholar 

  21. Huq MS, Das IJ, Steinberg T, Galvin JM: A dosimetric comparison of various multileaf collimators. Phys Med Biol 2002, 47: N159-170. 10.1088/0031-9155/47/12/401

    Article  PubMed  Google Scholar 

  22. Fiveash JB, Murshed H, Duan J, Hyatt M, Caranto J, Bonner JA, Popple RA: Effect of multileaf collimator leaf width on physical dose distributions in the treatment of CNS and head and neck neoplasms with intensity modulated radiation therapy. Med Phys 2002, 29: 1116-1119. 10.1118/1.1481515

    Article  CAS  PubMed  Google Scholar 

  23. Georg D, Dieckmann K, Bogner J, Zehetmayer M, Potter R: Impact of a micromultileaf collimator on stereotactic radiotherapy of uveal melanoma. Int J Radiat Oncol Biol Phys 2003, 55: 881-891.

    Article  PubMed  Google Scholar 

  24. Kubo HD, Wilder RB, Pappas CT: Impact of collimator leaf width on stereotactic radiosurgery and 3D conformal radiotherapy treatment plans. Int J Radiat Oncol Biol Phys 1999, 44: 937-945.

    Article  CAS  PubMed  Google Scholar 

  25. Chang Z, Wang Z, Wu QJ, Yan H, Bowsher J, Zhang J, Yin FF: Dosimetric Characteristics of Novalis Tx System with High Definition Multi-leaf Collimator. Med Phys 2008, in press.

    Google Scholar 

  26. Zeger SL, Liang KY: Longitudinal data analysis for discrete and continuous outcomes. Biometrics 1986, 42: 121-130. 10.2307/2531248

    Article  CAS  PubMed  Google Scholar 

  27. Zeger SL, Liang KY, Albert PS: Models for longitudinal data: a generalized estimating equation approach. Biometrics 1988, 44: 1049-1060. 10.2307/2531734

    Article  CAS  PubMed  Google Scholar 

Download references

Acknowledgements

The authors wish to thank the Engineers and Physicists of Varian Medical Systems and BrainLAB for their technical assistance.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Q Jackie Wu.

Additional information

Competing interests

The authors declare that they have no competing interests.

Authors' contributions

All authors read and approved the final manuscript.

QJW, ZW, FFY and JPK designed the study and the analysis, generated the treatment plans, performed the analysis drafted and revised the manuscript.

ML participated in the study design, statistically analysis and revised the manuscript.

JM and ZC participated in the study design and revised the manuscript.

CH participated in the study design, provided technique assistance and revised the manuscript.

Electronic supplementary material

Authors’ original submitted files for images

Rights and permissions

This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Reprints and permissions

About this article

Cite this article

Wu, Q.J., Wang, Z., Kirkpatrick, J.P. et al. Impact of collimator leaf width and treatment technique on stereotactic radiosurgery and radiotherapy plans for intra- and extracranial lesions. Radiat Oncol 4, 3 (2009). https://doi.org/10.1186/1748-717X-4-3

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/1748-717X-4-3

Keywords