Abstract
An inductively coupled plasma−optical emission spectrometry method was optimized and validated for the determination of major elements (Ca, K, Mg, Na and P) in cultivated freshwater fish (rainbow trout Oncorhynchus mykiss). The method was validated by analysis of a Certified Reference Material, consisting in a frozen tissue homogenate from lake trout (Salvelinus namaycush namaycush). The linearity of this method was very good, as evidenced by the coefficients of correlation (r) for calibration graphs that were higher than 0.9999 in all cases and by linearity test (response factor <5% and relative calibration graph slope <2%). Accuracy, expressed as relative recovery (%) in comparison with certified concentration ranged from 100 to 109%, and precision, expressed as residual standard deviation (%) ranged from 1.2 to 6.5% (repeatability) and from 1.0 to 9.6% (reproducibility). The limit of quantification ranged from 4 ng/mL (Ca and Mg) to 203 ng/mL (P). The optimized method was applied to major element determination in skin and muscle samples from rainbow trout fillets.
Similar content being viewed by others
References
MAGRAMA, Ministerio de Agricultura, Alimentación y Medio Ambiente. Producción de Acuicultura. Madrid, Spain, 2012.
Hunter, B.J. and Roberts, D.C.K., Nutr. Res., 2000, vol. 20, p. 1047.
Burger, J. and Gochfeld, M., Environ. Res., 2009, vol. 109, p. 343.
Rebolé, A., Velasco, S., Rodríguez, M.L., Treviño, J., Alzueta, C., Tejedor, J.L., and Ortiz, L.T., Food Chem., 2015, vol. 174, p. 614.
Mayneris-Perxachs, J., Bondia-Pons, I., Serra-Majem, L., and Castellote, A.I., Food Chem., 2010, vol. 119, p. 54.
Lal, S.P., in Fish Nutrition, Halver, J.E., Ed., San Diego, CA: Academic, 1989, p. 220.
Lal, S.P., in Fish and Fishery Products: Composition, Nutritive Properties and Stability, Ruiter, A., Ed., Wallingford, CN: CAB Int., 1995, p. 187.
Liang, Y. and Wong, M.H., Arch. Environ. Contam. Toxicol., 2000, vol. 39, p. 506.
Mermet, J.M., J. Anal. Atom. Spectrom., 2005, vol. 20, p. 11.
Moreda-Piñeiro, J., Alonso-Rodríguez, E., López-Maia, P., Muñategui-Lorenzo, S., Prada-Rodríguez, D., Moreda-Piñeiro, A., and Bermejo-Barrera, P., Anal. Chim. Acta, 2007, vol. 598, p. 95.
Larrea-Marin, M.T., Pomares-Afonso, M.S., Gómez-Juaristi, M., Sánchez-Muniz, F.J., and Ródenas de la Rocha, S., J. Food Compos. Anal., 2010, vol. 23, p. 814.
International Conference on Harmonization of Technical Requirements for Registration of Pharmaceuticals for Human Use, 1996.
A Laboratory Guide to Method Validation and Related Topics, Eurachem Working Group, 1998. https://www. eurachem.org/index.php/publications/guides/mv.
International Union of Pure and Applied Chemistry: Compendium of Chemical Terminology, Cambridge, UK, 1997, 2nd ed.
Thomsen, P., Schatzlein, D., and Mercuro, D., Spectroscopy, 2003, vol. 18, p. 112.
Castro, M., Gascón, S., Pujol, M., Sans, J.M., and Vicente, L., Validación de métodos analíticos. Asociación Española de Farmacéuticos de la Industria, Monografia AEFI, Sección Catalana, Comisión de Normas de Buena Fabricación y Control de la Calidad, Hewlett Packard, 1989.
Veeramachaneni, M. and Jayavarapu, K.R., J. Adv. Pharm. Tech. Res., 2013, vol. 3, p. 516.
Shearer, K.D., Can. J. Fish. Aquat. Sci., 1984, vol. 41, p. 1592.
Brucka-Jastrzębska, E., Kawczuga, D., Rajkowska, M., and Protasowicki, M., J. Elem., 2009, vol. 14, no. 3, p. 437.
Aburto, N.J., Ziolkovska, A., Hooper, L., Elliott, P., Cappuccio, F.P., and Meerpohl, J.J., Br. Med. J., 2013, vol. 346.
He, F.J., Li, J., and Macgregor, G.A., Br. Med. J., 2013, vol. 346.
Intersalt Cooperative Research Group, Br. Med. J., 1988, vol. 297, p. 319.
Sacks, F.M., Svetkey, L.P., Vollmer, W.M., Appel, L.J., Bray, G.A., Harsha, D., Obarzanek, E., Conlin, P.R., Miller, E.R., Simons-Morton, D.G., Karanja, N., and Lin, P.H., New Engl. J. Med., 2001, vol. 344, p. 3.
Du, S., Batis, C., Wang, H., Zhang, B., Zhang, J., and Popkin, B.M., Am. J. Clin. Nutr., 2014, vol. 99, p. 334.
Author information
Authors and Affiliations
Corresponding author
Additional information
The article is published in the original.
Rights and permissions
About this article
Cite this article
Velasco, S., Ortiz, L.T., Rodríguez, M.L. et al. Validation of an inductively coupled plasma-optical emission spectrometry method for the determination of major elements in farmed rainbow trout (Oncorhynchus mykiss). J Anal Chem 72, 966–971 (2017). https://doi.org/10.1134/S106193481709012X
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1134/S106193481709012X