18 May 2022 Investigation of functional near-infrared spectroscopy signal quality and development of the hemodynamic phase correlation signal
Uzair Hakim, Paola Pinti, J. Adam Noah, Xian Z. Zhang, Paul Burgess, Antonia Hamilton, Joy Hirsch, Ilias Tachtsidis
Author Affiliations +
Abstract

Significance: There is a longstanding recommendation within the field of fNIRS to use oxygenated (HbO2) and deoxygenated (HHb) hemoglobin when analyzing and interpreting results. Despite this, many fNIRS studies do focus on HbO2 only. Previous work has shown that HbO2 on its own is susceptible to systemic interference and results may mostly reflect that rather than functional activation. Studies using both HbO2 and HHb to draw their conclusions do so with varying methods and can lead to discrepancies between studies. The combination of HbO2 and HHb has been recommended as a method to utilize both signals in analysis.

Aim: We present the development of the hemodynamic phase correlation (HPC) signal to combine HbO2 and HHb as recommended to utilize both signals in the analysis. We use synthetic and experimental data to evaluate how the HPC and current signals used for fNIRS analysis compare.

Approach: About 18 synthetic datasets were formed using resting-state fNIRS data acquired from 16 channels over the frontal lobe. To simulate fNIRS data for a block-design task, we superimposed a synthetic task-related hemodynamic response to the resting state data. This data was used to develop an HPC-general linear model (GLM) framework. Experiments were conducted to investigate the performance of each signal at different SNR and to investigate the effect of false positives on the data. Performance was based on each signal’s mean T-value across channels. Experimental data recorded from 128 participants across 134 channels during a finger-tapping task were used to investigate the performance of multiple signals [HbO2, HHb, HbT, HbD, correlation-based signal improvement (CBSI), and HPC] on real data. Signal performance was evaluated on its ability to localize activation to a specific region of interest.

Results: Results from varying the SNR show that the HPC signal has the highest performance for high SNRs. The CBSI performed the best for medium-low SNR. The next analysis evaluated how false positives affect the signals. The analyses evaluating the effect of false positives showed that the HPC and CBSI signals reflect the effect of false positives on HbO2 and HHb. The analysis of real experimental data revealed that the HPC and HHb signals provide localization to the primary motor cortex with the highest accuracy.

Conclusions: We developed a new hemodynamic signal (HPC) with the potential to overcome the current limitations of using HbO2 and HHb separately. Our results suggest that the HPC signal provides comparable accuracy to HHb to localize functional activation while at the same time being more robust against false positives.

1.

Introduction

Functional near-infrared spectroscopy (fNIRS) is a noninvasive, optical neuroimaging modality that measures and maps the tissue concentrations changes of oxy- and deoxyhemoglobin (ΔHbO2 and HHb), providing an estimate of the brain hemodynamic and oxygenation changes secondary to neuronal activation. Theoretically, brain activation in fNIRS is considered as a statistically significant increase in HbO2 and a decrease in HHb,1 based on the principle of neurovascular coupling, where brain activation is marked by an initial dip in HHb followed by a large increase of HbO2 in the blood to support the increased neurons’ metabolic demand. This increase in blood volume, which is composed of an increase in HbO2 and a smaller decrease in HHb, is called the hemodynamic response.

fNIRS offers some advantages over functional magnetic resonance imaging (fMRI), which by many is considered one of the gold standard neuroimaging methods. It has a higher temporal resolution, greater portability, increased tolerance to movements, and relatively lower cost. In particular, fNIRS allows more ecologically valid experimental tasks and wider application in populations such as infants and toddlers.

However, fNIRS presents some limitations. The emitted light must travel through multiple extra-cerebral layers (i.e., scalp, skull, and CSF) before reaching the brain and reflecting back onto the detector on the surface of the head. As a result, the light is attenuated through the various layers by the changes in the oxygen levels in the blood. This can be a significant limitation when fNIRS is used to investigate brain functions, which also leads to substantial changes in systemic physiology, such as partial pressure of CO2 in the blood, heart rate, blood pressure, or autonomic nervous system activity.26 These physiological responses can lead to changes in oxygenation independently of the neuronal activation, in both the cerebral and extracerebral layers. The end effect of this issue is the possibility of false positives and negatives arising when inferring functional brain activity from fNIRS signals.

This issue has been a longstanding problem in fNIRS analysis, brought to light by Obrig and Villringer.7 They specified that while an increase in HbO2 is part of the definition of neuronal activation, it may also reflect a change in blood pressure or an increase in skin blood volume. This was reinforced by Kirlilna et al.8 with quantitative results showing that the HbO2 signal is much more strongly affected by the systemic physiology. In review paper, Scholkmann et al.9 discussed this in more detail, and separated the fNIRS signal into six components: cerebral and extracerebral each with evoked and nonevoked neuronal and systemic components. Based on the analysis from Tachtsidis and Scholkmann,2 HHb seems to be more robust to systemic interferences than HbO2, and hence arguably more robust against false positives and negatives; however, HHb is not completely robust against false positives and negatives, therefore using it does not solve the issue of false positives and negatives. The reduced sensitivity to systemic interference is likely due to its smaller amplitude changes compared to HbO2, which in turn also means less statistical power. On the contrary, HbO2 is a high contrast signal, with larger task-related changes, but it seems to be more sensitive to physiological changes and thus more prone to false positives/negatives. Obrig and Villringer7 provided one possible way to account for this problem, which is to use the HbO2 signal in conjunction with HHb when inferring functional brain activity with fNIRS. This was reiterated again by Tachtsidis and Scholkmann again, 13 years later,2 where the authors state that it is of immediate importance for studies to report results from both HbO2 and HHb to allow better physiological interpretation of results. Most recently, the fNIRS best practices paper10 again recommended that both HbO2 and HHb need to be included when reporting and visualizing statistical results.

To investigate which fNIRS signals researchers used most often for inferring brain activation in the field of cognitive neuroscience, we looked at peer-reviewed studies published in the field from 2017 to 2020. A search was carried out on the ScienceDirect database using the keywords “fNIRS” and “cognitive neuroscience” (repeated with the abbreviation fNIRS) between 2017 and 2020, although the search parameters capture papers, they are not exhaustive of the field, as such this was not a systematic review. About 132 papers were reviewed once papers examining resting-state connectivity were excluded. After inspecting the methods and results section of each paper for which fNIRS-derived signal was used to assess the presence of significant task-related brain activity, we found that 55% of papers only used HbO2, and 3% of papers used HHb, and 26% of papers used both HbO2 and HHb. Figure 1 displays the percentages of each signal used and a full table can be found in Table S4 in the Supplemental Material.

Fig. 1

Percentage of studies using each signal. Peer-reviewed studies from 2017 to 2020 in the field of cognitive neuroscience using fNIRS were reviewed. Resting-state papers were excluded.

NPH_9_2_025001_f001.png

The majority of the studies draw their conclusions based on an analysis of HbO2 only. This is usually because neuronal activation leads to a larger increase in HbO2, since neurons require more oxygen to cope with the energetic requirements, and a smaller decrease in HHb, giving HbO2 changes higher contrast and hence superior sensitivity in the evaluation of activity from baseline to task periods.1113 In turn, this makes it easier to get statistically significant results using HbO2 and so studies are more likely to base their analysis on this signal only.

In contrast with this, only four studies used the HHb signal alone. Three of these studies provided justifications for their choice, such as the HHb ability to maximize spatial specificity,14 and its robustness to artifacts caused by extra-cerebral perfusion.15,16 These benefits of HHb have been quantified by studies investigating the physiological components of HHb and HbO2,3,8,17 while the basis of BOLD fMRI has been well established18 and its similarity to HHb has been shown Refs. 1920.21. However, HHb typically has a lower magnitude and so there is less statistical power and larger sample sizes might be needed to observe a significant effect.

Promisingly, 34 studies2255 did use both signals in some way; however, a lack of a common statistical framework to analyze both signals simultaneously has led to variation in how these signals were used. For example,29 considered activation as an “increase in HbO2 and/or a decrease in HHb.” In this case, both signals were analyzed statistically but interpretation based on only one signal. The authors found no significant activation in the HHb data and neuroscientific interpretations were based only on HbO2. In comparison (Wriessnegger, et al, 2017), conducted statistical analysis on the maxima of the concentration changes of HbO2 and the minima of HHb, after averaging over each trial per task.

Similar to de Klerk et al. (2018),29 Jackson et al. (2019)34 conducted four separate ANOVAs for two experimental conditions and HbO2 and HHb and presented results such that an increase in HbO2 or a decrease in HHb is used to determine activation. In this case, the authors analyzed both signals, but the interpretation was based on the signals separately. The lack of a framework to analyze and interpret both HbO2 and HHb has caused variability in methods and interpretations between studies.

One possible method of overcoming the problem of no framework to analyze both signals is to use one signal that has been derived using HbO2 and HHb. There are currently three such signals in the fNIRS field. Total hemoglobin (HbTot=HbO2+HHb), which can be used as an indication of blood volume changes. HbTotal is expected to increase during neuronal activity with a smaller amplitude than HbO2. From the reviewed studies only two studies used HbTotal.56 justified their use of the signal as it incorporates the characteristics of neurovascular coupling from HbO2 and HHb, while57 did not provide any reasoning for its use. Hemoglobin difference (HbDiff=HbO2HHb) allows a measurement of oxygenation and identification of the mismatch between HbO2 and HHb. During neuronal activation, the signal should increase with an amplitude higher than HbO2, since HHb should be negative. Only one study used this signal.58 The authors justified its use using the same reasoning as Ref. 56 but did not provide specific reasoning for using HbDiff rather than HbTotal. Although these signals do indeed combine information from HbO2 and HHb, systemic interference occurring in either HbO2 or HHb will still affect them. A typical marker of systemic interference is HHb increasing at the same time as HbO2. In this case, the HbTotal will increase with an amplitude larger than HbO2. In turn, this will lead to false-positive activations being interpreted. The HbDiff signal will show an amplitude lower than HbO2 in this case, causing false negatives since the amplitudes will be lower than expected.

Finally, the correlation-based signal improvement (CBSI) signal59 is a good example of a possible way to combine the HbO2 and HHb signals for the identification of brain activity, while trying to account for the brain physiological response to activation. The signal scales the correlation of HbO2 and HHb such that it enforces the anticorrelation and reduces any positive correlations. To do this it incorporates a scaling factor, which is the ratio of the standard deviations of HbO2 and HHb. In this paper, the authors show using experimental data that removes motion spikes, improves the CNR, and provides better spatial specificity, enforcing the idea that using the relationship of HbO2 and HHb to form a single signal can provide better localization of activity. Thirteen studies utilized the signal.6072 The main justification was its ability to reduce motion and systemic artifacts; however Ref. 61 also specified its assumption of cortical activation being reflected by simultaneous increases and decreases in HbO2 and HHb. The key difference between the CBSI signal and HbDiff and HbTotal is the CBSI is formed based on an underlying characteristic of the hemodynamic response function (HRF) that is the anticorrelation between HbO2 and HHb.

As highlighted by the literature review, the majority of papers utilize HbO2 when analyzing fNIRS data, despite calls for the analysis of both HbO2 and HHb dating back to 2003. Nonetheless, there is heterogeneity in which fNIRS-derived signal is used to infer functional activation. In the studies using both signals, there is no standard framework for analyzing both signals together. In this paper, we aim to explore these issues. We provide initial investigations on which fNIRS-derived signal can predict brain activity most accurately. Additionally, we introduce and develop a novel signal that encodes phase information from HbO2 and HHb, named the hemodynamic phase correlation (HPC) signal. To achieve these goals, we compare the performance of the most used fNIRS-derived signals, namely the HbO2, HHb, CBSI, HbTotal, and HbDiff, alongside the newly proposed HPC signal.

Based on the recommendation of Refs. 2, 7, 9, and 10, the HPC signal utilizes information coming from HbO2 and HHb simultaneously. The HPC signal quantifies the phase relationship between HbO2 and HHb at each timepoint of the fNIRS recording. Specifically, the HPC uses the fundamental antiphase relationship between HbO2 and HHb to identify functional brain activity.

In this paper, we first present the development of the HPC signal. We then test the accuracy of HbO2, HHb, HbTotal, HbDiff, CBSI, and the HPC in identifying functional brain activity using a synthetic dataset of 18 subjects with 16 channels and experimental fNIRS data acquired from 128 adult participants during a finger-thumb tapping task using 134 channels.

2.

Method

2.1.

fNIRS Synthetic Data Generation

The generation of the synthetic data is shown in Fig. 2. The function “spm_hrf” part of the spm12 toolbox was used to generate a synthetic HRF with the default parameters. The function uses two gamma functions to form the HRF; a positive one that models the response and has a peak at 6 s after onset, and a negative one to model the undershoot, which has a minimum at 16 s. In particular, this is the canonical waveform reflecting the HbO2 response to a single functional event. The HHb signal is typically negative and with a smaller amplitude than the HbO2.73 The HHb signal, therefore, was formed by multiplying the canonical HRF amplitude (i.e., the HbO2 response) by 13 assuming HHb to have an amplitude 3 times smaller than HbO2. This procedure was repeated with three different HbO2 and HHb amplitudes [i.e., strong (RMS=4.8×107), mild (RMS=2.9×107), small (RMS=1.7×107) responses) to vary the SNR. To create the task-related component, a boxcar with a 40 s active condition and a 40 s rest condition was created to model a block-design task, and then convolved with the HRF. The task-related component was then added to resting state fNIRS data that approximates systemic noise. Actual systemic noise in an experiment may be highly correlated with task-block onsets and the method used here may not represent the task-induced noise but serves as an adequate approximation to it.

Fig. 2

Synthetic data formation. Resting-state fNIRS data were obtained from the Hitachi WOT system using 16 channels at 5-Hz sampling frequency.

NPH_9_2_025001_f002.png

Resting-state fNIRS data were collected using the Hitachi wearable optical tomography (WOT, Hitachi High-Technologies Corporation, Japan) system at a sampling frequency of 5 Hz. Light is emitted via six light-sources at 705 and 830 nm. Light sources and detectors were spaced 3-cm apart, with one pairing constituting a single channel. Data were collected from 18 subjects with a 16-channels configuration across the frontal lobe. To ensure uniformity in channel positioning across participants, the 10/20 system was used, with channel 8 placed at the FpZ point and channels 8 and 9 aligned to the Nasion–Inion line. Data were collected for 10 min, while participants were sitting in a chair with their eyes closed. The channel configuration is shown in Fig. S11 in the Supplemental Material.

The raw fNIRS data were first converted to optical densities and then to concentration changes using the modified Beer–Lambert law.74 This data was then added to the task-related HRF and preprocessing was carried out on the output. Motion correction was applied using the PCA method from Ref. 75 with the default parameters, and a fourth-order Butterworth band-pass filter with frequency cut-offs at 0.01 and 0.2 Hz, these parameters were determined to be adequate for attenuating systemic interference by the filter analysis in Ref. 76. The filtered HbO2 and HHb signals were used to form the HbDiff, HbTot, CBSI,59 and HPC signals. The peak amplitudes of each of the resulting signals for each amplitude variation are shown in Table 1. Following the process outlined in Fig. 2, the synthetic data consisted of 18 datasets with 16 channels each.

Table 1

Peak amplitudes for synthetic data conventional signals.

Amplitude 1 (μ mol)Amplitude 2 (μ mol)Amplitude 3 (μ mol)
HbO20.80.50.3
HHb−0.27−0.17−0.1
CBSI0.80.50.3
HbDiff1.070.670.4
HbTot0.530.330.2

2.2.

fNIRS Experimental Data Acquisition and Preprocessing

Experimental data were provided by the Brain Function Laboratory, Yale University.77 Hemodynamic signals were acquired using the LABNIRS System from Shimadzu Corporation (LABNIRS, Shimadzu Corporation, Japan). Data from 128 participants were used with mean age 25±7.6 years of age, 50% were female and 93% were right-handed. Signals were acquired from 134 channels covering the entire head (Fig. S12 in the Supplemental Material). Channels were localized using a 3D digitizing system and projected into the MNI space. Anatomical locations are shown in Table S5 in the Supplemental Material. The system acquired data at a sampling frequency of 8.1 Hz. The sources emit light at three wavelengths of 780, 805, and 830 nm. Caps were selected based on participant head size, with larger heads having a source–detector separation of 3 cm and smaller heads at 2.75 cm, to have comparable channel locations across subjects. Both caps had identical optode configurations. Assessment was made visually on the day, if the cap was to lose or tight a different sized cap was used. The experimental protocol was a right-handed finger thumb tapping task, chosen as it is a well-established protocol and serves as a reliable fiducial marker for signal comparisons. Participants were positioned 70 cm from a screen displaying instructions. A block design was used with alternating finger-thumb tapping and rest periods of 15 s each, repeated 6 times, for a total duration of 3 min.

Task periods included visual presentation of the numbers 1, 2, 3, or 4 in random order shown in black font in the center of a white screen. Numbers indicated which right-hand finger to firmly press against the right thumb and were presented every 0.75 s for a total of 20 numbers per active epoch. Participants were asked to respond as quickly and accurately as possible. In the case of mistakes, participants were told to continue without correcting. Rest epochs consisted of viewing a black crosshair presented in the center of a white screen. During rest epochs, participants were instructed to remain stationary without moving their fingers, clear their mind, and focus their eyes on the center of the screen. Participants were also told, in general, to refrain from taking deep breaths, keep their body still, and refrain from fidgeting and touching their face. The data were exported from the system as raw voltages and processed using functions from the Homer2 toolbox.78 The raw voltages were converted to optical density using the hmrIntensity2OD function and then motion-corrected using hmrMotionCorrectWavelet.79 The output was bandpass filtered using a Butterworth filter with order 5 and cut-offs at 0.01 and 0.2 Hz.76 Finally, hmrOD2Conc was used with differential pathlength factors of 6 for each wavelength. The final preprocessed HbO2 and HHb signals were used to compute the remaining signals, shown in Table 2.

Table 2

Equations used to form CBSI, HbDiff, and HbTotal.

SignalFormula
CBSI12(HbO2αHHb)
HbDiffHbO2HHb
HbTotHbO2+HHb

2.3.

Hemodynamic Phase Correlation (HPC) Computation

The method outlined by Ref. 80 was used to form the HPC signal. The method transforms the time series data of HbO2 and HHb to the time-frequency domain and the phase correlation is then derived, the formation is shown in Fig. 3.

Fig. 3

Generation of HPC signal from its component signals, (a) HbO2 and (b) HHb. A task frequency of 0.025 Hz was used. Denoted by the shaded areas in the time series and highlighted by a white box in the spectrograms. Continuous wavelet transform applied to both signals (c) and (d). Cross wavelet transform is computed (e) and the phase extracted is shown in (f), which is converted to a normalized value, the HPC, in (g).

NPH_9_2_025001_f003.png

The first equation in Fig. 3 refers to the continuous wavelet transform where s is the scale, u is the displacement, Ψ is the mother wavelet used, and * refers to the complex conjugate. The mother wavelet used was the Morlet wavelet, in agreement with other fNIRS studies using wavelets.3,4,81,82 Once both CWTs are computed, they are used to compute the cross wavelet transform83 by multiplying the CWT of HbO2 with the complex conjugate of HHb. The output is a complex value with an amplitude A, and local phase θ, which is calculated using the third equation in Fig. 3. θ describes the phase correlation between the two signals at each frequency, for all time points in radians and ranges between π and +π. One disadvantage of using solely the θ values is that the values of π and +π symbolize the same thing, 180 deg out of phase: an anticorrelation between the two datasets. This can lead to a misevaluation of the data, especially if inspecting visually. For example, in Fig. 3(F), the row of data occurring at 0.015 Hz shows distinctive areas of +π and π, in actual terms, these pertain to the same thing, an anticorrelated phase. By calculating the cosine of these values, we can use this information in a more useful way. The values now range from 1 to +1 (anticorrelated, to correlated) representing the information in a more interpretable way. The parameter n in Eq. (4) of Fig. 3 is an odd integer >0 and can be altered to sharpen the response of the HPC by reducing the smearing of an area where there is a strong positive or negative correlation. In this case, the value of n was left at 1 as at this value there was no smearing and the output provided responses that were expected. The output of this is shown in Fig. 3(g).

The y-axis of the spectrograms shows the scale parameter alongside the frequency. The scale refers to how the mother wavelet is scaled when fitting the signal. The scale can be converted to a “pseudofrequency” using the following equation:84

Eq. (1)

Fa=Fca·Δ,
where “a” is the scale, Fa is the pseudofrequency of a corresponding scale “a,” Fc is the center frequency of the wavelet in Hz and Δ is the sampling period. As we are determining the frequency based on the scale, in the text only frequencies will be used to avoid over-complication; however, in the figures both scale and frequency are shown.

Since the method utilizes wavelet transforms, the issue of edge-effect artifacts is present. These occur when the scaled wavelet extends beyond the edges of the observation interval. These regions are delineated in all spectrograms by the dashed-white line, known as the cone-of-influence (COI). Regions within the COI are potentially affected by edge effects. However, as shown in Figs. 5 and 6, the task frequencies used for the simulated data minimally overlap with the COI. Additionally, the experimental data is a higher frequency than the simulated and since the COI is larger for lower frequencies (low frequency = high scale = extends past signal), the higher the frequency the less the signal exists within the COI. However, in the case where the task frequency has a large overlap within the COI, the user should state how much of the signal is overlapping within the COI.

2.3.1.

HbO2 decreasing and HHb increasing (reversal pattern)

A possible source of error using the HPC signal may arise when the HbO2 signal is decreasing and the HHb is increasing during functional activation. Since this is still antiphase, the HPC will detect it as activation, although it is not. This reversal hemodynamic pattern may be of physiological interest, but it is a confounding factor in the HPC-based detection of functional activation. To account for this, we developed an add-on function able to detect functional blocks where this reversed response pattern occurs. This method is based on the Hbdiff signal (HbDiff=HbO2HHb). During functional activation, where HbO2 is increasing (positive) and HHb is decreasing (negative), the HbDiff signal should be positive; instead, in the case of HbO2 being negative and HHb positive (i.e., the reversed response pattern), HbDiff is negative. Therefore, detecting the negative trend of HbDiff can be used to determine where and when the reversal pattern occurs.

The add-on function proposed here first computes the HbDiff signal within each task block and then computes the area of the regions bounded by the HbDiff signal and the x axis. In this way, we quantify the ratio between the areas of the positive and negative regions of the HbDiff signal during the functional stimulation period. If the HbDiff signal during the stimuli block is dominated by the reversed response pattern, there should be much more negative area than positive, resulting in a ratio <0.5; once this is detected, the affected block can be automatically removed from further analyses or manually inspected and removed by the user. The 0.5 ratio threshold was empirically chosen by using our synthetic data while on purpose implementing hemodynamic reversal response patterns. This ratio was a good compromise in accurately distinguishing a true hemodynamic reversal response pattern versus post-stimulus hemodynamic undershoots, where HbO2 and HHb are reversed as part of the cerebral hemodynamic response to stimuli.

2.4.

Data Analysis

Synthetic and experimental data were analyzed using the general linear model (GLM). This method was developed for the analysis of fMRI data by Ref. 85 and extended to fNIRS. It expresses the fNIRS signal as a linear combination of several variables, known as regressors, plus an error term. The GLM is the preferable method of analysis as it considers the entire time series of the data and allows the inclusion of multiple variables to explain the measured signal (see Ref. 1 for a full review and explanation of the principles of the GLM). For HbO2, HHb, CBSI, HbTot, HbDiff, the design matrix included the task regressor (either synthetic 40-s task blocks or 15-s finger tapping blocks) created through the convolution of the HRF with the boxcar representing the task onsets. For the experimental data task onset times were acquired using triggers implemented with the fNIRS acquisition software. The HbO2 version of the HRF was used for all signals apart from HHb, where a negative HRF was used, and the HPC where a specific HPC-RF was used (see Sec. 2.4.1).

2.4.1.

HPC specific data analysis

Since the HPC exists in the time-frequency domain and does not resemble a typical fNIRS signal, a specific framework for its application to the GLM was developed using the synthetic data. The data was formed using synthetic data with a peak HbO2 amplitude of 0.8×106  Mol and peak HHb amplitude of 0.27×106  Mol, since it has the highest SNR and the framework derived from it will have the strictest parameters for use with the experimental data.

In the design matrix of standard fNIRS analysis (i.e., HbO2/HHb), the HRF convolved with a task-related boxcar function is used, called here the standard HRF (sHRF). In the case of HPC analysis, this method would not be suitable since the signal is not directly comparable to the HRF, rather it is a representation of the relationship between HbO2 and HHb. Therefore, to use the GLM with the HPC signal, a representation of this relationship should be used in the design matrix (comparable to standard analysis), we denote this as the HPC-RF. The HPC-RF is equal to the HPC of the sHRF of HbO2 and HHb. Since the sHRF of HHb is the negative of HbO2 the HPC-RF is infact 1 at all points. This is in line with the hypothesis that brain activation can be expressed as an antiphase relationship between HbO2 and HHb.

As the HPC exists in the time-domain frequency, a frequency must be chosen to be used in the GLM as the measured signal. For this paper, since both synthetic and experimental data was determined using a constant-frequency block-design task, we hypothesized that the HPC frequency that should be used would correspond to the task frequency used, i.e., for the synthetic data, it would be at 0.025 Hz. This was confirmed using the process shown in Fig. 4.

Fig. 4

Pipeline used to determine which frequency should be used as the measured signal in the GLM. HPC was computed for all 16 channels of all 18 datasets. The GLM was used to output β values for all frequencies of all HPC transforms. The index of the maximum β value was determined. The index corresponds to the frequency at which that β value was found. The most commonly occurring index was found for each dataset, across channels. The task periods are highlighted in gray in the time data. The outlined area in the HPC spectrograms correspond to frequencies of 0.0234, 0.0250, and 0.0268 Hz. All three are outlined to account for the indirect scale to frequency conversion. The corresponding task-frequency is highlighted in yellow for the HPC spectrograms.

NPH_9_2_025001_f004.png

Due to the nature of the dynamic brain system, we would expect there to be an underlying, base-level of antiphase correlations in the experimental data arising from noise and confounding factors. The effect of this noise on the results would manifest as widespread activations. To account for this base-level of antiphase correlation an activation threshold was determined and used for the experimental data analysis. The threshold would say that a certain level of anticorrelation is required to be reached before considering it to be activation, which should be above 0. It was determined in the β space and represents the lowest value which could be activation. This is possible since the HPC has defined bounds, +1 to 1, and so a specific level of activation can be defined. To decide this threshold, β values were computed from the synthetic data with added activation. Group statistics were determined and signals showing significant activation at p0.01 were considered. Of these considered signals, the smallest value was chosen as the activation threshold. The determined threshold value was used as the hypothesized mean for computing t-values using the experimental data.

The full workflow using the HPC signal is shown in Fig. 5. The development of the HPC framework was verified using the synthetic data and the results are shown in Sec. 3.1.

Fig. 5

Full workflow of analysis using HPC signal. (a) The HPC is computed using the preprocessed measured HbO2 and HHb signals for a given channel. The relevant task frequency is identified, and the HPC signal according to that frequency (or scale) is extracted from the full spectrogram. (b) To compute the HPC-RF the standard HRF of HbO2 and HHb are first computed using the task timings, the HPC transform is then applied using the standard HRFs as inputs. (B.i) The reversal artifact detection function can be used at this stage to evaluate blocks that may contain the reversed artifact. Users have the option to automatically remove these blocks from the HPC transform, or to visually inspect blocks using the output as a guide and then remove from analysis. (c) The standard GLM is used with the Measured HPC at a task frequency as the measured parameter (Y in GLM) and the HPC-RF as the input to the design matrix (X in GLM). Once the β values are computed standard statistical analysis can be carried out, and a threshold β value can be applied, the authors suggest a value of βThresh0.3.

NPH_9_2_025001_f005.png

2.5.

Performance Evaluation

The beta values from the synthetic and experimental data were normalized to t-values calculated using Eq. (2). Where x^ refers to the sample mean, μ is a hypothesized mean, σ is the standard deviation and n is the sample size

Eq. (2)

t=x^μσn.

For the conventional signals in both synthetic and experimental data, a one-sample, two-tailed t-test was carried out using a hypothesized mean of 0.

The synthetic data performance was evaluated on how large the calculated t-value was for each signal. For the synthetic HPC data, the hypothesized mean was set to 0.

For the experimental data, performance was based on how well a signal could localize activation to the motor cortex region of the brain. To do this, the active conditions of the task were modeled, and β values were obtained for all channels of each participant. For the conventional signals, the task-timings were used to form a task-boxcar function and convolved with the relevant HRF to form the task-related regressor modeling the hemodynamics response. For the HPC signal, the HPC-RF discussed in Sec. 2.4.1 was used, which does not require task timings since the task frequency is known. T-values were then computed based on β values from the GLM using Eq. (2). For the computation of the t-values for the experimental data HPC signal, a one-sample one-tailed t-test was used, and the hypothesized mean was set to the determined activation threshold, which was informed by the synthetic data analysis. Multiple comparisons correction was carried out using the false discovery rate (FDR), using the method outlined in Ref. 86.

2.6.

Synthetic Data Experiments

Three experiments were conducted with the synthetic dataset to evaluate the performance of each signal. The first experiment aimed at evaluating the performance of each signal in localizing functional brain activity at different SNRs. To this goal, synthetic data were generated with three different amplitudes to provide datasets with varying SNR.

The additional two experiments involved evaluating signals’ performance when activation was removed from signals by reversing the sign of the HbO2 and then the HHb task-HRF. This was done to remove the anticorrelated characteristic typical of brain activation and to understand how each signal reflects nonactivation depending on which chromophore (HbO2 or HHb) is affected. The synthetic data was adapted for this purpose by reversing the polarity of the task-related HRF, in the first case the HbO2 was made negative and in the second case, the HHb was made positive. In both cases, this was done incrementally to accentuate the trending behavior of all signals when the polarity is reversed.

3.

Results

3.1.

HPC Specific Data Analysis

An example of the spectrogram corresponding to the HPC of HbO2 and HHb is shown in Fig. 6, using the synthetic data from dataset 3, channel 11. A clear anticorrelation can be seen at the 0.025 Hz task frequency when the activation is encoded into the time series, highlighted by the yellow box on each spectrogram.

Fig. 6

Example HPC output. The plots on the left show the resting state fNIRS data (see Fig. 1) and the corresponding HPC signal. The right shows the same data after the resting state fNIRS data has been added to the synthetic task HRFs with a task frequency of 0.025 Hz. In both HPC, the task frequency is highlighted by a yellow box and zoomed in below. The time data also show the task blocks shaded in gray.

NPH_9_2_025001_f006.png

The first set of results is from the HPC-specific analysis described in Sec. 2.4.1. The frequency component that should be used in the analysis of the HPC signal is shown in Fig. 7(a) and the evaluation of the HPC-based activation threshold is included in Fig. 7(b).

Fig. 7

(a) Frequency check results. (b) Activation threshold results; the points correspond to the means as explained in Fig. 4, and the error-bars represent standard error of the mean. The red line in (a) displays the expected task related frequency (0.025 Hz), while (b) displays the chosen threshold based on the results.

NPH_9_2_025001_f007.png

To determine which frequency from the HPC signal to use the synthetic data with task-component occurring at 0.025 Hz was used. This frequency is shown in Fig. 7(a) by the red line. The frequencies between 0.022 and 0.031 Hz displayed the best fit for 61% of signals. This provided a quantitative basis to determine which frequencies to use in the GLM for the HPC analysis in all following experiments. To account for the discrepancies present in scale-frequency conversion, the frequency present in the HPC closest to the task frequency (3 d.p.) was chosen, the frequency immediately preceding and following was also chosen and the three were averaged together.

To determine the threshold β, the minimum β value from the GLM analysis on the synthetic data was chosen. As activation was encoded in all channels of all datasets the minimum β value found would correspond to the minimum value that activation can occur at, while taking noise into account. To ensure that the activation was present and determinable while including the noise, the beta values from the HbO2 and the HHb signals were tested for significance at p<0.01 and the minimum β value of the channels showing significance was chosen as the threshold. The minimum value found was 0.32, to account for variations the threshold was chosen to be 0.3, shown by the red line in Fig. 7(b). This value would be used as the hypothesized mean of the experimental data.

3.2.

Synthetic Data

Figures 8 and 9 display the results of the experiments conducted using the synthetic dataset. For both Figs. 8 and 9, the raw beta values can be seen in the Supplemental Material (Fig. S13 in the Supplemental Material). In all cases, a higher t-value indicates a signal which has a closer fit with the task-related HRF used in the GLM. Figure 8 shows the results for varying SNR. Amplitude 3 is the signal with the lowest SNR. For all signals, the t-values decrease as SNR reduces, as expected. Table 3 shows the numerical values for Fig. 8. The legend entries in Figs. 9(a) and 9(b) refer to how much of the time series is encoded with activation. At 10% only 10% of the time series is encoded with activation by a positive HbO2 and a negative HHb. The SNR used for these tests was amplitude 1, as such the results for 90% relate closely to the results for amplitude 1 in Fig. 8. The results when HHb HRF is made positive are shown in Fig. 9(a), and when HbO2 is negative in Fig. 9(b).

Fig. 8

Percentage deviation of all signals when varying the amplitude of the task-related component of the synthetic data. Amplitude 1 has the highest signal-noise ratio. T-values were computed using a one-sample two-tailed t-test, with a hypothesized mean of 0.

NPH_9_2_025001_f008.png

Fig. 9

(a) Results from reversing the sign of the HHb HRF and (b) results from reversing the sign of the HbO2 signal. The percentages listed in the legend refer to what percentage of the time series was convolved with a task-related HRF, 90% corresponds to near full activation. T-values were computed using a one-sample two-tailed t-test with hypothesized mean equal to 0.

NPH_9_2_025001_f009.png

Table 3

Numeric t-values for Fig. 8.

HbO2HHbCBSIHbDiffHbTotHPC
Amp 132.436.435.636.624.436.9
Amp 223.627.129.727.916.421.9
Amp 314.916.921.918.110.112.2

The HbO2 result in Fig. 9(a) is constant at 32.4, expected as the HbO2 data are not altered. Comparatively, the HHb signal ranges from 32.1 at 90% to 43 at 10%, representing the intended effect of reversing the sign of HHb. CBSI displays similar behavior, t-values ranging from 35.5 at 10% to 1.8 at 90%. HbDiff ranges from 35.9 to 24.2. HbTotal displayed the opposite effect, at 10% the t-value was 38.3, while at 90% it was 25.5. Finally, the HPC ranges from 31.3 to 30.7.

As expected for Fig. 9(b), the HHb response is constant at 36.4, as the data for HHb were not altered. The HbO2 ranges from 29.6 at 90% activation to 39.5. CBSI displays the same pattern as shown in Fig. 9(a), ranging from 34.0 at 90% activation to 5.1 at 10% activation. HbDiff ranges from 34.98 to 24.39. HbTotal ranges from 19.8 to 46.9. Finally, the HPC signal ranges from 30.6 to 26.9, showing the same pattern as Fig. 9(b).

3.3.

Experimental Data

The group results for the finger-thumb tapping task are shown in Fig. 10. The views shown are axial view, where the top of each image would be the front of the body. The p-value threshold was set at p0.01 after applying FDR correction.

Fig. 10

All signals, FDR corrected p-value0.01. (a) HbO2, (b) HHb, (c) HbDiff, (d) HbTotal, (e) CBSI, and (f) HPC. The experimental task was a right-handed finger-thumb tapping task. Expected channel activations were in the left primary motor cortex.

NPH_9_2_025001_f010.png

The HbO2 signal found 96 channels with activation. The mean t-score across these channels was 5.13 (σ=±0.176) and the mean p-value 9.28×1004 (σ=±2.11×1004). In contrast, HHb showed 1 channel with activation with a t-score of 4.18 and p-value of 0.007, localized to the left primary motor cortex. CBSI and HbDiff found 48 channels being active, in slightly different locations. CBSI had a mean t-score of 4.56 (σ=±0.200) compared with HbDiff with 4.60 (σ=±0.207). Similarly, CBSI had mean p-value of 0.002 (σ=±4.03×1004) compared with HbDiff with p-value 0.0015 (σ=±2.541×1004). HbTotal showed the most activated channels, with 131 below the p-value threshold, with mean p-value of 1.80×1004 (σ=±9.89×1005) and mean t-score at 6.55 (σ=±0.144). Finally, the HPC displayed similar performance to HHb showing two activated channels with a mean p-value 0.0035 and mean t-score 4.10 (σ=±0.092).

4.

Discussion

The HPC signal is formed based on the fundamental characteristic of antiphase correlation between HbO2 and HHb present in the hemodynamic response. Thus, the signal is encoded with information from both HbO2 and HHb. This key design aspect has two benefits: first, by incorporating information from both chromophores we improve the statistical robustness of the results; second, we reduce confusion in the field in how to analyze and interpret both signals by combining them into one signal.

We have first assessed the performance of HbO2, HHb, CBSI, HbDiff, HbTotal, and HPC using synthetic data. The results are shown in Figs. 8 and 9. The first test (Fig. 8) showed that the CBSI signal performed the most consistently, with the highest t-values for amplitudes 2 and 3, but slightly worse than HHb, HbDiff, and HPC for amplitude 1. This result suggests that the CBSI is the most robust against a reducing SNR since it is the least affected by the decreasing SNR from amplitudes 2 and 3.

The HPC signal provided the highest t-value when the SNR was high, suggesting that the HPC is able to indicate brain activation. However, the HPC seems to be the most susceptible to the decreasing SNR, indicated by a large fall in t-value from amplitude 1 to 2. By decreasing the amplitudes of HbO2 and HHb the noise in the signal will dominate and dilute the antiphase relationship between HbO2 and HHb. This result was the driving force behind the requirement for an activation threshold.

We have also carried out an analysis with the synthetic data by forcibly introducing false positives into the time series data in order to further demonstrate that the use of a single chromophore is not robust enough against false positives which may arise due to systemic physiology, artifacts, or otherwise. The results are shown in Figs. 9(a) and 9(b). The ideal signal, in this case, would reflect the degree of activation and perform uniformly, independent of whether it was HbO2 or HHb that was altered. In this test, the signals which were formed using the anticorrelation present in brain activation (CBSI, HPC) achieved both of these requirements. Since they are encoded with the specific characteristic of the relationship between HbO2 and HHb, it seems they are better suited to determine activation based on a GLM fit. An interesting point however is that although the CBSI does determine low activation, the HPC is able to determine an “inverse” activation, where it can determine that either HbO2 or HHb are in the opposite polarity. In comparison, the two single chromophores were unable to distinguish false positives. Similarly, the effect on a single chromophore had repercussions on the t-value of HbTotal and HbDiff depending on which chromophore was altered. HbTotal performed the worst of all the signals, in the case where HHb was altered HbTotal displayed the opposite of the expected trend; at low activation, it showed a high t-value. This is due to the HHb signal becoming positive and, therefore, the formation leads to a larger signal. Similarly, HbDiff was affected differently depending on the chromophore being altered however it displayed the expected trend of high t-value at high activation and decreasing t-value with activation.

The difference between the CBSI, HPC and HbDiff, HbTotal stands in their formation. CBSI and HPC are derived on the basis of a physiological characteristic of brain activation (i.e., anticorrelated patterns of HbO2 and HHb). Thus, as long as that specific characteristic is present in the fNIRS data, the CBSI and HPC are better suited to detect brain activation.

To our knowledge, this is the first study that has evaluated and compared the performance of the most commonly used signals in the field using the GLM approach. The results from the synthetic analysis suggest that the use of a signal formed from the combination of HbO2 and HHb is more robust against a decreasing SNR and the possibility of false positives in the time series. As shown by either the HPC or CBSI performing the best when the SNR was varied and also being able to uniformly and adequately reflect the degree of activation in the following two tests. This method of deriving functional activation from the combination of HbO2 and HHb was mentioned in Ref. 9, where the authors specifically cited the CBSI method as a way to derive functional activation based on a specific characteristic of activation, such as the anticorrelation used in the CBSI or the antiphase correlation used by the HPC. Furthermore, both the HPC and the CBSI satisfy the crucial requirement to use both HbO2 and HHb as stated by the fNIRS best practices10 as it is encoded with information from both HbO2 and HHb. The HHb signal performed third best in all amplitudes, better than the HbO2 signal, possibly because it is less susceptible to the systemic noise component as suggested by previous studies.3,8

We next assessed the signals using fNIRS data acquired from a finger-tapping task. Data were acquired from 128 adult participants with 134 channels covering the entire head. Signals’ performance was evaluated on their ability to localize activation with respect to the target anatomical location (ground truth). Here our target region was the left motor cortex for a right finger tapping task. As shown in Fig. 10, we found that the HHb signal provided the best localization with a single channel localized over the Precentral Gyrus, the site of the primary motor cortex. However, the t-value was lower than the corresponding channel for HbO2, suggesting HHb has reduced statistical power compared with HbO2. This region has been implicated in fMRI and PET studies8791 as well as in fNIRS studies.9295 More specifically,93 also deduced that the HHb signal provided more localized activation than the HbO2 signal in a motor task, in agreement with our findings. Studies have shown the HHb signal is more highly correlated to the BOLD signal than HbO2,19 therefore, it was expected for the HHb signal to display results in agreement with the fMRI BOLD signal. However, as we have shown with the synthetic data a single chromophore should not be used to make interpretations due to the possibility of false positives and negatives, and results obtained from one signal should be looked at with caution. A further analysis (shown in Fig. S15 in the Supplemental Material) of this dataset showed the HHb signal was positively increasing alongside HbO2 for a widespread number of channels. HHb increasing with HbO2 is a typical characteristic of a systemic change.2,7 This manifested in a larger HbTotal increase than expected, represented in the results as a widespread HbTotal false-positive activation. These results show how the possibility of false-positives in a single chromophore can affect the statistical inference of activation. The HPC signal was able to provide more specific localization than the other signals formed from HbO2 and HHb (i.e., CBSI, HbDiff, and HbTotal). The two channels that displayed significant activation correspond to the Precentral Gyrus, the same as the HHb signal. The HPC offers increased confidence in the results because it is based on the fundamental antiphase relationship intrinsic to the hemodynamic response. The CBSI and HbDiff signals displayed 38 channels in common showing activation, with 10 different. CBSI showed one more channel located in the left motor cortex that was not present in the HbDiff signal. The similarity between these two signals is due to the formula used to form the CBSI, (see Ref. 56). They differ in that the CBSI signal has a parameter (α), formed by the ratio of the standard deviations of the measured HbO2 and HHb signals across the whole time series for that channel. This parameter reflects the ratio of the standard deviation of HbO2 and HHb, representative of the ratio of the noise amplitude in the signal. The similarity between the CBSI and HbDiff was also shown in the synthetic data SNR test, where the HbDiff was one position lower than the CBSI in amplitudes 2 and 3 and within a t-value of 1 for amplitude 1.

In comparison to the HbDiff, HbTotal, and the CBSI signals, since the HPC does not consider the amplitude of HbO2 and HHb in its formation, it is not susceptible to large increases due to false positives or negatives in its constituent signals or to unexpected noise amplitudes. We believe that its sensitivity to the fundamental antiphase of HbO2 and HHb has prevented it from detecting the channels where HbO2 and HHb were increasing or decreasing simultaneously.

Taken together, our results on synthetic and experimental data suggest that the HPC signal has the potential for the identification of functional brain activity from fNIRS data and may contribute to increased robustness against false positives and negatives. Considering the individual signals only, the results suggest the HHb signal is better able to localize brain activation compared with the HbO2 signal.

This paper did not evaluate the performance of any signals after other signal processing methods were applied. It has been noted that the acquisition of fNIRS data from short channels during data acquisition and subsequent usage as a regressor in the GLM can significantly improve the HbO2 results by essentially removing the systemic components from the analysis.96 Software-based methods such as the Spatial filter97 also provide methods of improving the individual signals acquired from the fNIRS system. Therefore, it is possible to obtain good fNIRS data while using the HbO2 and HHb signals separately. The HPC signal offers the advantage of providing adequate localization without the added complexity of physically incorporating short channels, the reduced spatial resolution of dedicating optodes to short channels, or the increased analytic complexity of using short-channel data or additional software-based signal processing methods.

In our future studies, we plan to validate the HPC analysis further using more diverse task-protocols to examine how more severe systemic influences affect the HPC, as well as develop the method to be used with varying task-frequencies and conditions occurring at the same frequency. The threshold values specified in this paper can be used as a default value for initial use, however other users can tweak the value as desired, similarly for the “n” parameter in Eq. (4), Fig. 3 we suggest a default value of 1. Further work will also be conducted to optimize the detection of reversed activation response patterns and to minimize the impact of confounding factors on the HPC-based localization of functional brain activity. In our future experiments, we will investigate and refine these values further. Furthermore, we believe the HPC can be used to determine channel quality by numerically presenting the relationship between HbO2 and HHb. For example, if there is a large systemic influence, we will see a concurrent increase in HbO2 and HHb, which the HPC will reflect. Finally, version 1.0 of the toolbox is under development which will provide all the functionality of the HPC signal in an easy-to-use graphical user interface, allowing integration with popular analysis toolboxes.

5.

Conclusion

We introduced a new combined signal and evaluates the performance of inferring activation against other signals. Among these, the HHb signal resulted to be the single individual chromophore with the highest accuracy in the identification of brain activation. However, our results further suggested that using one chromophore as a means to make interpretations of functional activation might not be adequate to account for false positives and/or false negatives. Our initial investigation on the HPC signal provided promising results, showing comparable performance to HHb but going beyond that by incorporating information from HbO2, hence being less prone to false positives and/or negatives.

The evidence provided in this work can have a significant impact in the fNIRS field. The HPC signal is able to localize activation and encodes information from HbO2 and HHb. The comparison of signals will help to reduce the heterogeneity and uncertainty around the analysis and interpretation of HbO2 and HHb.

Disclosures

The authors declare no conflicts of interest.

Acknowledgments

This work was supported by the Engineering and Physical Sciences Research Council (EPSRC, EP/R512400/1), the EPSRC-funded UCL Centre for Doctoral Training in Intelligent, Integrated Imaging in Healthcase (i4Health, EP/S021930/1), the Wellcome Trust (104580/Z/14/Z; 212979/Z/18/Z), the MRC (MR/S003134/1), the BBSRC (BB/J014567/1), the Birkbeck Wellcome Trust Institutional Strategic Support Fund (ISSF), the ESRC (ES/V012436/1). Additionally, the NIMH (R01MH107513; R01MH111629; R01MH119430), NICHD (R37HD090153), NIDCD (R21DC017821).

Code, Data, and Materials Availability

The code for the HPC v1.0 and sample data are available at https://github.com/multimodalspectroscopy/HPC.

References

1. 

S. Tak and J. C. Ye, “Statistical analysis of fNIRS data: a comprehensive review,” NeuroImage, 85 72 –91 (2014). https://doi.org/10.1016/j.neuroimage.2013.06.016 NEIMEF 1053-8119 Google Scholar

2. 

I. Tachtsidis and F. Scholkmann, “False positives and false negatives in functional near-infrared spectroscopy: issues, challenges, and the way forward,” Neurophotonics, 3 031405 (2016). https://doi.org/10.1117/1.NPh.3.3.031405 Google Scholar

3. 

E. Kirilina et al., “The physiological origin of task-evoked systemic artefacts in functional near infrared spectroscopy,” NeuroImage, 61 70 –81 (2012). https://doi.org/10.1016/j.neuroimage.2012.02.074 NEIMEF 1053-8119 Google Scholar

4. 

P. Pinti, D. Cardone and A. Merla, “Simultaneous fNIRS and thermal infrared imaging during cognitive task reveal autonomic correlates of prefrontal cortex activity,” Sci Rep, 5 17471 (2015). https://doi.org/10.1038/srep17471 Google Scholar

5. 

M. Caldwell et al., “Modelling confounding effects from extracerebral contamination and systemic factors on functional near-infrared spectroscopy,” NeuroImage, 143 91 –105 (2016). https://doi.org/10.1016/j.neuroimage.2016.08.058 NEIMEF 1053-8119 Google Scholar

6. 

F. Scholkmann et al., “End-tidal CO2: an important parameter for a correct interpretation in functional brain studies using speech tasks,” Neuroimage, 66 71 –79 (2013). https://doi.org/10.1016/j.neuroimage.2012.10.025 NEIMEF 1053-8119 Google Scholar

7. 

H. Obrig and A. Villringer, “Beyond the visible—imaging the human brain with light,” J. Cereb. Blood Flow Metab., 23 1 –18 (2003). https://doi.org/10.1097/01.WCB.0000043472.45775.29 Google Scholar

8. 

E. Kirlilna et al., “Identifying and quantifying main components of physiological noise in functional near infrared spectroscopy on the prefrontal cortex,” Front. Hum. Neurosci., 7 1 (2013). https://doi.org/10.3389/fnhum.2013.00864 Google Scholar

9. 

F. Scholkmann et al., “A review on continuous wave functional near-infrared spectroscopy and imaging instrumentation and methodology,” NeuroImage, 85 6 –27 (2014). https://doi.org/10.1016/j.neuroimage.2013.05.004 NEIMEF 1053-8119 Google Scholar

10. 

M. A. Yücel et al., “Best practices for fNIRS publications,” Neurophotonics, 8 012101 (2021). https://doi.org/10.1117/1.NPh.8.1.012101 Google Scholar

11. 

Y. Hoshi, N. Kobayashi and M. Tamura, “Interpretation of near-infrared spectroscopy signals: a study with a newly developed perfused rat brain model,” J. Appl. Physiol. Bethesda Md, 90 (5), 1657 –1662 (2001). https://doi.org/10.1152/jappl.2001.90.5.1657 Google Scholar

12. 

D. Zhang et al., “Discrimination of emotional prosodies in human neonates: a pilot fNIRS study,” Neurosci. Lett., 658 62 –66 (2017). https://doi.org/10.1016/j.neulet.2017.08.047 NELED5 0304-3940 Google Scholar

13. 

H. Sato et al., “Cerebral hemodynamics in newborn infants exposed to speech sounds: a whole-head optical topography study,” Hum. Brain Mapp., 33 2092 –2103 (2012). https://doi.org/10.1002/hbm.21350 HBRME7 1065-9471 Google Scholar

14. 

J. Hirsch et al., “Frontal temporal and parietal systems synchronize within and across brains during live eye-to-eye contact,” NeuroImage, 157 314 –330 (2017). https://doi.org/10.1016/j.neuroimage.2017.06.018 NEIMEF 1053-8119 Google Scholar

15. 

F. Gundel et al., “Meditation and the brain—neuronal correlates of mindfulness as assessed with near-infrared spectroscopy,” Psychiatr. Res. Neuroimaging, 271 24 –33 (2018). https://doi.org/10.1016/j.pscychresns.2017.04.002 Google Scholar

16. 

F. Scholkmann et al., “Effects of psilocybin on functional connectivity measured with fNIRS: Insights from a single-subject pilot study,” 11 (2019). Google Scholar

17. 

M. M. Plichta et al., “Event-related functional near-infrared spectroscopy (fNIRS) based on craniocerebral correlations: reproducibility of activation?,” Hum. Brain Mapp., 28 733 –741 (2007). https://doi.org/10.1002/hbm.20303 HBRME7 1065-9471 Google Scholar

18. 

R. B. Buxton, “Dynamic models of BOLD contrast,” NeuroImage, 62 953 –961 (2012). https://doi.org/10.1016/j.neuroimage.2012.01.012 NEIMEF 1053-8119 Google Scholar

19. 

T. J. Huppert et al., “A temporal comparison of BOLD, ASL, and NIRS hemodynamic responses to motor stimuli in adult humans,” NeuroImage, 29 368 –382 (2006). https://doi.org/10.1016/j.neuroimage.2005.08.065 NEIMEF 1053-8119 Google Scholar

20. 

A. Kleinschmidt et al., “Simultaneous recording of cerebral blood oxygenation changes during human brain activation by magnetic resonance imaging and near-infrared spectroscopy,” J. Cereb. Blood Flow Metab., 16 817 –826 (1996). https://doi.org/10.1097/00004647-199609000-00006 Google Scholar

21. 

V. Toronov et al., “The roles of changes in deoxyhemoglobin concentration and regional cerebral blood volume in the fMRI BOLD signal,” NeuroImage, 19 1521 –1531 (2003). https://doi.org/10.1016/S1053-8119(03)00152-6 NEIMEF 1053-8119 Google Scholar

22. 

N. Altvater-Mackensen and T. Grossmann, “Modality-independent recruitment of inferior frontal cortex during speech processing in human infants,” Dev. Cogn. Neurosci., 34 130 –138 (2018). https://doi.org/10.1016/j.dcn.2018.10.002 Google Scholar

23. 

M. Balconi, L. Cortesi and D. Crivelli, “Motor planning and performance in transitive and intransitive gesture execution and imagination: does EEG (RP) activity predict hemodynamic (fNIRS) response?,” Neurosci. Lett., 648 59 –65 (2017). https://doi.org/10.1016/j.neulet.2017.03.049 NELED5 0304-3940 Google Scholar

24. 

M. Balconi and G. Fronda, “The ‘gift effect’ on functional brain connectivity. Inter-brain synchronization when prosocial behavior is in action,” Sci. Rep., 10 5394 (2020). https://doi.org/10.1038/s41598-020-62421-0 SRCEC3 2045-2322 Google Scholar

25. 

D. Bandara et al., “Building predictive models of emotion with functional near-infrared spectroscopy,” Int. J. Hum.-Comput. Stud., 110 75 –85 (2018). https://doi.org/10.1016/j.ijhcs.2017.10.001 IHSTEI 1071-5819 Google Scholar

26. 

L. Bell et al., “Auditory and visual response inhibition in children with bilateral hearing aids and children with ADHD,” Brain Sci., 10 307 (2020). https://doi.org/10.3390/brainsci10050307 Google Scholar

27. 

R. Cañigueral et al., “Facial and neural mechanisms during interactive disclosure of biographical information,” NeuroImage, 226 117572 (2021). https://doi.org/10.1016/j.neuroimage.2020.117572 NEIMEF 1053-8119 Google Scholar

28. 

A. Curtin et al., “Enhancing neural efficiency of cognitive processing speed via training and neurostimulation: an fNIRS and TMS study,” NeuroImage, 198 73 –82 (2019). https://doi.org/10.1016/j.neuroimage.2019.05.020 NEIMEF 1053-8119 Google Scholar

29. 

C. C. J. M. de Klerk, A. F. de C. Hamilton and V. Southgate, “Eye contact modulates facial mimicry in 4-month-old infants: an EMG and fNIRS study,” Cortex, 106 93 –103 (2018). https://doi.org/10.1016/j.cortex.2018.05.002 Google Scholar

30. 

S. R. de Oliveira et al., “Changes of functional response in sensorimotor cortex of preterm and full-term infants during the first year: an fNIRS study,” Early Hum. Dev., 133 23 –28 (2019). https://doi.org/10.1016/j.earlhumdev.2019.04.007 EHDEDN 0378-3782 Google Scholar

31. 

J. Defenderfer et al., “Investigating the role of temporal lobe activation in speech perception accuracy with normal hearing adults: an event-related fNIRS study,” Neuropsychologia, 106 31 –41 (2017). https://doi.org/10.1016/j.neuropsychologia.2017.09.004 NUPSA6 0028-3932 Google Scholar

32. 

T. Gruber et al., “Human discrimination and categorization of emotions in voices: a functional near-infrared spectroscopy (fNIRS) study,” Front. Neurosci., 14 1 (2020). https://doi.org/10.3389/fnins.2020.00570 1662-453X Google Scholar

33. 

M. J. Herrmann et al., “Relevance of dorsolateral and frontotemporal cortex on the phonemic verbal fluency—a fNIRS-study,” Neuroscience, 367 169 –177 (2017). https://doi.org/10.1016/j.neuroscience.2017.10.028 Google Scholar

34. 

E. S. Jackson et al., “A fNIRS investigation of Speech Planning and Execution in Adults Who Stutter,” Neuroscience, 406 73 –85 (2019). https://doi.org/10.1016/j.neuroscience.2019.02.032 Google Scholar

35. 

L. Jin et al., “Differences in brain signal complexity between experts and novices when solving conceptual science problem: a functional near-infrared spectroscopy study,” Neurosci. Lett., 699 172 –176 (2019). https://doi.org/10.1016/j.neulet.2019.02.015 NELED5 0304-3940 Google Scholar

36. 

J. A. Kable and C. D. Coles, “Prefrontal cortical responses in children with prenatal alcohol-related neurodevelopmental impairment: a functional near-infrared spectroscopy study,” Clin. Neurophysiol., 128 2099 –2109 (2017). https://doi.org/10.1016/j.clinph.2017.08.009 CNEUFU 1388-2457 Google Scholar

37. 

A. Katzorke et al., “Decreased hemodynamc response in inferior frontotemporal regions in elderly with mild cognitive impairment,” Psychiatry Res. Neuroimaging, 274 11 –18 (2018). https://doi.org/10.1016/j.pscychresns.2018.02.003 Google Scholar

38. 

S. Kujach et al., “A transferable high-intensity intermittent exercise improves executive performance in association with dorsolateral prefrontal activation in young adults,” NeuroImage, 169 117 –125 (2018). https://doi.org/10.1016/j.neuroimage.2017.12.003 NEIMEF 1053-8119 Google Scholar

39. 

M. Mauri et al., “Hemodynamic and behavioral peculiarities in response to emotional stimuli in children with attention deficit hyperactivity disorder: an fNIRS study,” J. Affect. Disord., 277 671 –680 (2020). https://doi.org/10.1016/j.jad.2020.08.064 JADID7 Google Scholar

40. 

H. Obrig et al., “Impact of associative word learning on phonotactic processing in 6-month-old infants: a combined EEG and fNIRS study,” Dev. Cogn. Neurosci., Sensitive periods across development, 25 185 –197 (2017). https://doi.org/10.1016/j.dcn.2016.09.001 Google Scholar

41. 

P. H. S. Pelicioni et al., “People with Parkinson’s disease exhibit reduced cognitive and motor cortical activity when undertaking complex stepping tasks requiring inhibitory control,” Neurorehabil. Neural Repair, 34 1088 –1098 (2020). https://doi.org/10.1177/1545968320969943 Google Scholar

42. 

L. Pirazzoli et al., “Hand or spoon? Exploring the neural basis of affective touch in 5-month-old infants,” Dev. Cogn. Neurosci., 35 28 –35 (2019). https://doi.org/10.1016/j.dcn.2018.06.002 Google Scholar

43. 

S. Steber et al., “Uncovering electrophysiological and vascular signatures of implicit emotional prosody,” Sci. Rep., 10 5807 (2020). https://doi.org/10.1038/s41598-020-62761-x SRCEC3 2045-2322 Google Scholar

44. 

S. Steber and S. Rossi, “So young, yet so mature? Electrophysiological and vascular correlates of phonotactic processing in 18-month-olds,” Dev. Cogn. Neurosci., 43 100784 (2020). https://doi.org/10.1016/j.dcn.2020.100784 Google Scholar

45. 

F. Stephan, H. Saalbach and S. Rossi, “Inner versus overt speech production: does this make a difference in the developing brain?,” Brain Sci., 10 939 (2020). https://doi.org/10.3390/brainsci10120939 Google Scholar

46. 

F. Stephan, H. Saalbach and S. Rossi, “The brain differentially prepares inner and overt speech production: electrophysiological and vascular evidence,” Brain Sci., 10 148 (2020). https://doi.org/10.3390/brainsci10030148 Google Scholar

47. 

A. van der Kant et al., “Negative affect is related to reduced differential neural responses to social and non-social stimuli in 5-to-8-month-old infants: a functional near-infrared spectroscopy-study,” Dev. Cogn. Neurosci., 30 23 –30 (2018). https://doi.org/10.1016/j.dcn.2017.12.003 Google Scholar

48. 

A. van der Kant et al., “Linguistic and non-linguistic non-adjacent dependency learning in early development,” Dev. Cogn. Neurosci., 45 100819 (2020). https://doi.org/10.1016/j.dcn.2020.100819 Google Scholar

49. 

Y. Wei et al., “Functional near-infrared spectroscopy (fNIRS) as a tool to assist the diagnosis of major psychiatric disorders in a Chinese population,” Eur. Arch. Psychiatry Clin. Neurosci., 271 (4), 745 –757 (2021). https://doi.org/10.1007/s00406-020-01125-y Google Scholar

50. 

S. C. Wriessnegger et al., “Imagine squeezing a cactus: cortical activation during affective motor imagery measured by functional near-infrared spectroscopy,” Brain Cogn., 126 13 –22 (2018). https://doi.org/10.1016/j.bandc.2018.07.006 Google Scholar

51. 

S. C. Wriessnegger et al., “Force related hemodynamic responses during execution and imagery of a hand grip task: a functional near infrared spectroscopy study,” Brain Cognit., 117 108 –116 (2017). https://doi.org/10.1016/j.bandc.2017.06.010 Google Scholar

52. 

Z. Zhang et al., “The effects of CACNA1C gene polymorphism on prefrontal cortex in both schizophrenia patients and healthy controls,” Schizophr. Res., 204 193 –200 (2019). https://doi.org/10.1016/j.schres.2018.09.007 SCRSEH 0920-9964 Google Scholar

53. 

F. Scholkmann et al., “Effect of short-term colored-light exposure on cerebral hemodynamics and oxygenation, and systemic physiological activity,” Neurophotonics, 4 045005 (2017). https://doi.org/10.1117/1.NPh.4.4.045005 Google Scholar

54. 

T. Karen et al., “Cerebral hemodynamic responses in preterm-born neonates to visual stimulation: classification according to subgroups and analysis of frontotemporal-occipital functional connectivity,” Neurophotonics, 6 045005 (2019). https://doi.org/10.1117/1.NPh.6.4.045005 Google Scholar

55. 

H. Zohdi et al., “Long-term blue light exposure changes frontal and occipital cerebral hemodynamics: not all subjects react the same,” Oxygen Transport to Tissue XLII, 217 –222 Springer International Publishing, Cham (2021). Google Scholar

56. 

T. Sukal-Moulton et al., “Relationship between sensorimotor cortical activation as assessed by functional near infrared spectroscopy and lower extremity motor coordination in bilateral cerebral palsy,” NeuroImage Clin., 20 275 –285 (2018). https://doi.org/10.1016/j.nicl.2018.07.023 Google Scholar

57. 

K. N. de Winkel et al., “Neural correlates of decision making on whole body yaw rotation: An fNIRS study,” Neurosci. Lett., 654 56 –62 (2017). https://doi.org/10.1016/j.neulet.2017.04.053 NELED5 0304-3940 Google Scholar

58. 

D. T. Corp et al., “Reduced motor cortex inhibition and a ‘cognitive-first’ prioritisation strategy for older adults during dual-tasking,” Exp. Gerontol., 113 95 –105 (2018). https://doi.org/10.1016/j.exger.2018.09.018 EXGEAB 0531-5565 Google Scholar

59. 

X. Cui, S. Bray and A. L. Reiss, “Functional near infrared spectroscopy (NIRS) signal improvement based on negative correlation between oxygenated and deoxygenated hemoglobin dynamics,” NeuroImage, 49 3039 –3046 (2010). https://doi.org/10.1016/j.neuroimage.2009.11.050 NEIMEF 1053-8119 Google Scholar

60. 

A. Anderson et al., “Evaluation of neurocognitive function of prefrontal cortex in ornithine transcarbamylase deficiency,” Mol. Genet. Metab., 129 207 –212 (2020). https://doi.org/10.1016/j.ymgme.2019.12.014 Google Scholar

61. 

C. Artemenko et al., “Functional lateralization of arithmetic processing in the intraparietal sulcus is associated with handedness,” Sci. Rep., 10 1 –775 (2020). https://doi.org/10.1038/s41598-020-58477-7 SRCEC3 2045-2322 Google Scholar

62. 

J. Liu et al., “Interplay between prior knowledge and communication mode on teaching effectiveness: Interpersonal neural synchronization as a neural marker,” NeuroImage, 193 93 –102 (2019). https://doi.org/10.1016/j.neuroimage.2019.03.004 NEIMEF 1053-8119 Google Scholar

63. 

M. J. Maier et al., “Forgiveness and cognitive control—provoking revenge via theta-burst-stimulation of the DLPFC,” NeuroImage, 183 769 –775 (2018). https://doi.org/10.1016/j.neuroimage.2018.08.065 NEIMEF 1053-8119 Google Scholar

64. 

F. G. Metzger et al., “Functional brain imaging of walking while talking—an fNIRS study,” Neuroscience, 343 85 –93 (2017). https://doi.org/10.1016/j.neuroscience.2016.11.032 Google Scholar

65. 

A. Mirelman et al., “Effects of aging on prefrontal brain activation during challenging walking conditions,” Brain Cognit., 115 41 –46 (2017). https://doi.org/10.1016/j.bandc.2017.04.002 Google Scholar

66. 

Y. Pan et al., “Instructor-learner brain coupling discriminates between instructional approaches and predicts learning,” NeuroImage, 211 116657 (2020). https://doi.org/10.1016/j.neuroimage.2020.116657 NEIMEF 1053-8119 Google Scholar

67. 

Y. Pan et al., “Interpersonal brain synchronization with instructor compensates for learner’s sleep deprivation in interactive learning,” Biochem. Pharmacol., 191 114111 (2021). https://doi.org/10.1016/j.bcp.2020.114111 Google Scholar

68. 

Y. Pan et al., “Interpersonal synchronization of inferior frontal cortices tracks social interactive learning of a song,” NeuroImage, 183 280 –290 (2018). https://doi.org/10.1016/j.neuroimage.2018.08.005 NEIMEF 1053-8119 Google Scholar

69. 

D. Rosenbaum et al., “Neural correlates of mindful emotion regulation in high and low ruminators,” Sci. Rep., 10 15617 (2020). https://doi.org/10.1038/s41598-020-71952-5 SRCEC3 2045-2322 Google Scholar

70. 

D. Rosenbaum et al., “Stress-related dysfunction of the right inferior frontal cortex in high ruminators: an fNIRS study,” NeuroImage Clin., 18 510 –517 (2018). https://doi.org/10.1016/j.nicl.2018.02.022 Google Scholar

71. 

D. Rosenbaum et al., “Cortical hemodynamic changes during the trier social stress test: an fNIRS study,” NeuroImage, 171 107 –115 (2018). https://doi.org/10.1016/j.neuroimage.2017.12.061 NEIMEF 1053-8119 Google Scholar

72. 

T. Sharon et al., “Which obstacle attributes place additional demands on higher-level cognitive function in patients with Parkinson’s disease?,” Parkinsonism Relat Disord., 78 178 –183 (2020). https://doi.org/10.1016/j.parkreldis.2020.08.002 Google Scholar

73. 

L. Gagnon et al., “Short separation channel location impacts the performance of short channel regression in NIRS,” NeuroImage, 59 2518 –2528 (2012). https://doi.org/10.1016/j.neuroimage.2011.08.095 NEIMEF 1053-8119 Google Scholar

74. 

D. T. Delpy and M. Cope, “Quantification in tissue near-infrared spectroscopy,” Philos. Trans. R. Soc. Lond. B, 352 649 –659 (1997). https://doi.org/10.1098/rstb.1997.0046 Google Scholar

75. 

M. A. Yücel et al., “Targeted principle component analysis: a new motion artifact correction approach for near-infrared spectroscopy,” J. Innov. Opt. Health Sci., 7 1350066 (2014). https://doi.org/10.1142/S1793545813500661 Google Scholar

76. 

P. Pinti et al., “Current status and issues regarding pre-processing of fNIRS neuroimaging data: an investigation of diverse signal filtering methods within a general linear model framework,” Front. Hum. Neurosci., 12 505 (2019). https://doi.org/10.3389/fnhum.2018.00505 Google Scholar

78. 

T. J. Huppert et al., “HomER: a review of time-series analysis methods for near-infrared spectroscopy of the brain,” Appl. Opt., 48 (10), D280 –298 (2009). https://doi.org/10.1364/AO.48.00D280 APOPAI 0003-6935 Google Scholar

79. 

B. Molavi and G. A. Dumont, “Wavelet based motion artifact removal for functional near infrared spectroscopy,” in Annu. Int. Conf. IEEE Eng. Med. Biol. Soc., 5 –8 (2010). https://doi.org/10.1109/IEMBS.2010.5626589 Google Scholar

80. 

G. R. J. Cooper and D. R. Cowan, “Comparing time series using wavelet-based semblance analysis,” Comput. Geosci., 34 95 –102 (2008). https://doi.org/10.1016/j.cageo.2007.03.009 CGEODT 0098-3004 Google Scholar

81. 

L. Duan et al., “Wavelet-based method for removing global physiological noise in functional near-infrared spectroscopy,” Biomed. Opt. Express, 9 3805 (2018). https://doi.org/10.1364/BOE.9.003805 BOEICL 2156-7085 Google Scholar

82. 

X. Zhang et al., “Activation detection in functional near-infrared spectroscopy by wavelet coherence,” J. Biomed. Opt., 20 016004 (2015). https://doi.org/10.1117/1.JBO.20.1.016004 JBOPFO 1083-3668 Google Scholar

83. 

C. Torrence and G. P. Compo, “A practical guide to wavelet analysis,” Bull. Am. Meteorol. Soc., 79 61 –78 (1998). https://doi.org/10.1175/1520-0477(1998)079<0061:APGTWA>2.0.CO;2 BAMIAT 0003-0007 Google Scholar

84. 

P. Abry, “Ondelettes et turbulences: multirésolutions, algorithmes de décomposition, invariance d’échelle et signaux de pression,” Paris, New York (1997). Google Scholar

85. 

K. J. Friston et al., “Statistical parametric maps in functional imaging: a general linear approach,” Hum. Brain Mapp., 2 189 –210 (1994). https://doi.org/10.1002/hbm.460020402 HBRME7 1065-9471 Google Scholar

86. 

Y. Benjamini and Y. Hochberg, “Controlling the false discovery rate: a practical and powerful approach to multiple testing,” J. R. Stat. Soc. Ser. B Methodol., 57 289 –300 (1995). https://doi.org/10.1111/j.2517-6161.1995.tb02031.x Google Scholar

87. 

L. M. Carey et al., “The functional neuroanatomy and long-term reproducibility of brain activation associated with a simple finger tapping task in older healthy volunteers: a serial PET study,” NeuroImage, 11 124 –144 (2000). https://doi.org/10.1006/nimg.1999.0522 NEIMEF 1053-8119 Google Scholar

88. 

S. C. Cramer et al., “Activation of distinct motor cortex regions during ipsilateral and contralateral finger movements,” J. Neurophysiol., 81 383 –387 (1999). https://doi.org/10.1152/jn.1999.81.1.383 JONEA4 0022-3077 Google Scholar

89. 

M. Guye et al., “Combined functional MRI and tractography to demonstrate the connectivity of the human primary motor cortex in vivo,” NeuroImage, 19 1349 –1360 (2003). https://doi.org/10.1016/S1053-8119(03)00165-4 NEIMEF 1053-8119 Google Scholar

90. 

P. E. Roland and R. J. Seitz, “Positron emission tomography studies of the somatosensory system in man,” Ciba Found. Symp., 163 113 –120 (1991). https://doi.org/10.1002/9780470514184.ch7 CIBSB4 0300-5208 Google Scholar

91. 

T. K. Turesky et al., “An fMRI study of finger tapping in children and adults,” Hum. Brain Mapp., 39 3203 –3215 (2018). https://doi.org/10.1002/hbm.24070 HBRME7 1065-9471 Google Scholar

92. 

D. A. Benaron et al., “Noninvasive functional imaging of human brain using light,” J. Cereb. Blood Flow Metab., 20 469 –477 (2000). https://doi.org/10.1097/00004647-200003000-00005 Google Scholar

93. 

W. N. Colier et al., “Human motor-cortex oxygenation changes induced by cyclic coupled movements of hand and foot,” Exp. Brain Res., 129 457 –461 (1999). https://doi.org/10.1007/s002210050913 EXBRAP 0014-4819 Google Scholar

94. 

C. Drenckhahn et al., “A validation study of the use of near-infrared spectroscopy imaging in primary and secondary motor areas of the human brain,” Epilepsy Behav., 49 118 –125 (2015). https://doi.org/10.1016/j.yebeh.2015.04.006 Google Scholar

95. 

C. Hirth et al., “Non-invasive functional mapping of the human motor cortex using near-infrared spectroscopy,” NeuroReport, 7 1977 –1981 (1996). https://doi.org/10.1097/00001756-199608120-00024 NERPEZ 0959-4965 Google Scholar

96. 

D. Wyser et al., “Short-channel regression in functional near-infrared spectroscopy is more effective when considering heterogeneous scalp haemodynamics,” Neurophotonics, 7 (3), 035011 (2020). https://doi.org/10.1117/1.NPh.7.3.035011 Google Scholar

97. 

X. Zhang, J. A. Noah and J. Hirsch, “Separation of the global and local components in functional near-infrared spectroscopy signals using principal component spatial filtering,” Neurophotonics, 3 015004 (2016). https://doi.org/10.1117/1.NPh.3.1.015004 Google Scholar

Biographies of the authors are not available.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Uzair Hakim, Paola Pinti, J. Adam Noah, Xian Z. Zhang, Paul Burgess, Antonia Hamilton, Joy Hirsch, and Ilias Tachtsidis "Investigation of functional near-infrared spectroscopy signal quality and development of the hemodynamic phase correlation signal," Neurophotonics 9(2), 025001 (18 May 2022). https://doi.org/10.1117/1.NPh.9.2.025001
Received: 16 September 2021; Accepted: 13 April 2022; Published: 18 May 2022