28 July 2016 Microvascular contrast enhancement in optical coherence tomography using microbubbles
Author Affiliations +
Abstract
Gas microbubbles (MBs) are investigated as intravascular optical coherence tomography (OCT) contrast agents. Agar + intralipid scattering tissue phantoms with two embedded microtubes were fabricated to model vascular blood flow. One was filled with human blood, and the other with a mixture of human blood + MB. Swept-source structural and speckle variance (sv) OCT images, as well as speckle decorrelation times, were evaluated under both no-flow and varying flow conditions. Faster decorrelation times and higher structural and svOCT image contrasts were detected in the presence of MB in all experiments. The effects were largest in the svOCT imaging mode, and uniformly diminished with increasing flow velocity. These findings suggest the feasibility of utilizing MB for tissue hemodynamic investigations and for microvasculature contrast enhancement in OCT angiography.

1.

Introduction

Various contrast agents have been suggested to enhance tissue microvasculature in optical coherence tomography (OCT),1 including gold nanoparticles,2 nanorods,3 microspheres,4 and nanoshells.5 Gold nanoparticles enhanced the microvascular contrast in vivo in an oral dysplasia hamster cheek pouch model by as much as 150%.6 Microspheres with liquid-filled cores and scattering particles within their shell have also shown significant contrast improvement, and deployment of these contrast media has been demonstrated for in-vivo OCT.7 However, while initially in the blood stream, these agents leak from the vessels to the surrounding tissues which decreases the contrast, and their accumulation (often in liver) may cause a number of side effects.8 In light of these drawbacks, microbubbles [MBs; currently Food and Drug Administration (FDA) approved intravascular contrast agents in diagnostic ultrasound9] may have a role to play as OCT contrast agents.10,11

MBs are micrometer sized shell-encapsulated gas-core spheres. They are considered intravascular agents, i.e., following intravenous injection, they remain within the blood compartment and do not extravasate. Their pharmacokinetics suggest a lifetime of 10 to 40 min following injection, depending on their shell composition and gas type.12 In addition to diagnostic/imaging applications, MBs (in combination with therapeutic ultrasound) are being investigated to enhance chemotherapy,13,14 radiotherapy,15 and thermal therapy with gold nanoparticles.16

Speckle variance (sv) OCT has been actively pursued as an attractive alternative for three-dimensional (3-D) visualization of the microvascular network, utilizing temporal variation of the scattering (speckle) pattern between frames as the contrast mechanism.17,18 In addition to the high resolution svOCT imaging of microvasculature, it has potential to provide new insight into the dynamics of vascular remodeling.1921

In our previous phantom study,11 we demonstrated that MBs can be used as an optical contrast agent, yielding increased scattering (1.4× at 300 nm and 6× at 850 nm). Similarly, Barton et al.10 have shown that MBs enhanced the backscattered OCT signal in model scattering media. Initial in-vivo work also yielded enhanced visualization of blood vessels with lumen diameters less than a few hundred micrometers.10 However, a quantitative assessment of MB contrast enhancement under no-flow and flowing blood conditions in a well-controlled model system with known parameters is lacking, and is addressed in the current study.

2.

Materials and Methods

2.1.

Tissue Phantoms

To model biological tissue, 10 phantoms were made by heating 1% TCBS agar (Sigma-Aldrich, Missouri) and 2% intralipid diluted from 20% stock (Sigma-Aldrich, St. Louis, Missouri) in water. Intralipid mimics the scattering properties of biological tissue,22,23 and agar solidifies the resultant suspension upon cooling. The heated mixture was cooled in chambers containing two microtubes to simulate blood vessels. Five prepared phantom chambers contained microtubes with internal diameters of 600  μm, and five more with internal diameters of 300  μm. A double-syringe high-precision pump (Model NE-4002X, New Era Pump Systems, Farmingdale, New York) was used to fill the two microtubes simultaneously; OCT measurements under both no-flow and flow conditions were then possible. The average blood flow velocities in both microtubes were 0, 0.5, 1, 2, 3, 5, 8, and 12  mm/s corresponding to venules, capillaries, and arterioles.2426 One syringe was filled with blood, the other with the mixture of blood and MBs. Rotating magnets inside both syringes were used to avoid red blood cells (RBC) settling and MBs floating during all experiments.

2.2.

Human Blood

Ethics approval was granted by the Research Ethics Boards of Ryerson University and the Canadian Blood Services. All experiments were conducted using whole blood donated by healthy volunteers recruited by the Canadian Blood Services’ Network Center for Applied Development (Vancouver, BC, Canada). The whole blood was 1-week old and contained 14% citrate phosphate dextrose anticoagulant.

2.3.

Microbubbles

Commercially available Definity® microbubbles (perflutren lipid microspheres, Lantheus Medical Imaging, Billerica, Massachusetts), FDA approved for diagnostic ultrasound contrast, were used in this study. The mean diameter of MBs was 1.2±0.5  μm. A 2.5% volume fraction of a stock suspension (average 1.2×1010  MBs/mL) was injected into the blood microtube. The resultant MB concentration was comparable to that used in ex-vivo ultrasound tests.27

2.4.

Optical Coherence Tomography System

A home-built and previously described swept source OCT system28 was used in this study. Briefly, a 36 kHz short cavity laser source with a polygon-based tuneable filter and wavelength centered at 1310 nm had a sweep range of 110 nm and an average output power of 40 mW. The system contained a 50/50 Mach-Zehnder interferometer and an imaging probe mounted on three-axis micropositioning stage in the sample arm. A dual-balanced photo detector was used in the output arm of the interferometer for DC signal component removal. A fiber Bragg grating provided the A-scan trigger at the beginning of each laser sweep. An OCT signal recalibration was performed with a standard Mach–Zehnder recalibration interferometer to generate linear k-space data from the raw data sampled with a 125  MS/s data acquisition card. After the data were equally spaced in k-space, it was inversely Fourier transformed for further analysis. The axial resolution (in air) of the system was 8  μm. The experimental setup is shown in Fig. 1.

Fig. 1

Experimental setup.

JBO_21_7_076014_f001.png

2.5.

Data Acquisition and Processing

OCT imaging was performed for both no-flow and flow conditions for blood (left microtube on subsequent images) and blood + MB mixture (right microtube). Data were recorded using Thorlabs software (Thorlabs Inc., Newton, New Jersey) in one-dimensional (1-D), two-dimensional (2-D), and 3-D modes.

In 1-D, A-scans were acquired from both microtubes to quantify MB-induced differences in OCT signal intensities. The autocorrelation function (ACF) was calculated from 18 pixels along the chosen A-line (shown in Fig. 2 as a vertical white dashed line) within the region of interest (ROI) of the corresponding microtube. The average of speckle decorrelation times (1/e level) of these 18 pixels is displayed in Fig. 3. Autocorrelation was considered as the correlation between intensity values Ii of the same pixel at times ti and ti+τ with increasing time delay τ:

Eq. (1)

ACFi=i=1Nτ(IiI¯)(Ii+τI¯)i=1N(IiI¯)2.
For each phantom, A-scans were recorded continuously for 8 s with the optical beam parked at the center of the chosen microtube (with blood or blood + MB) for all flow conditions. Each resultant record of 288,000 A-scans was divided into 240 intervals of N=1200 A-scans (33 ms acquisition time). ACFs were computed for each interval and then averaged.

Fig. 2

Typical structural B-scan of the phantom. The left microtube was filled with blood, and the right with blood + MBs [2.5% volume fraction of a stock suspension (average 1.2×1010  MBs/mL) were injected into the blood microtube]. ROIs of blood and blood + MB mixture are labeled with half-circles at the top circular section of microtube lumens. Background ROI was chosen at the same depth as shown. The vertical dashed line indicates the location of the acquired A-scans for ACF calculation (Fig. 3).

JBO_21_7_076014_f002.png

Fig. 3

Decorrelation time plots for blood and blood + MB mixture under eight flow conditions in 300-μm tube.

JBO_21_7_076014_f003.png

2-D images (B-scans) of each phantom were acquired with 4 mm of lateral scanning. MATLAB software (MATLAB, Mathworks, Natick, Massachusetts) was used for data processing. Images of both blood and blood + MB microtubes were used to quantify the difference in resultant phantom contrasts. Since the majority of the signal from the bottom of the microtube was low due to high scattering of blood, the ROI was chosen as the top-half circular section of the microtube lumen (see Fig. 2). The averaged ROI intensity was used to quantify the contrast via

Eq. (2)

CW=II0I0,
where I is the average intensity within ROI and I0 is the average intensity of the background ROI0. The background ROI0 was chosen to be in the middle of the phantom between the microtubes, at the same depth (Fig. 2).

3-D OCT images were acquired in sv mode in order to obtain sv average intensity projections for 3×1×1  mm3 data volume as previously described,28 with the number of sequential same location B-scans equal to 8, and 25 ms step size in slow-axis (Y) direction. svOCT calculates the interframe intensity variance of structural images, where contrast is based on the difference in time-varying properties. Specifically,

Eq. (3)

SVjk=1Ni=1N(IijkIjk¯)2,
where N=8 is the number of B-scans acquired from the same i-location, Iijk is the intensity of the pixel of the i’th B-scan, j is the axial pixel, k is the lateral pixel, and Ijk¯ is the mean intensity of that pixel over eight sequential B-scans. Statistical analysis (two-tailed test) was performed using SPSS statistics software, using results from the five phantoms for any given experimental point. Results with p<0.05 were considered statistically significant.

3.

Results and Discussion

3.1.

One-Dimensional Decorrelation Analysis

The ACFs of OCT signals with and without MBs at no-flow and varying flow conditions are plotted in Fig. 3. The results from the 300-μm-diameter microtube are shown; those from the larger microtube show similar trends with slightly higher values at low flows. As expected, the decorrelation time decreases with blood flow velocity. The decorrelation times were typically 15  ms at no-flow condition, decreasing to 1  ms for velocities >5  mm/s. As seen, the presence of MB causes faster decorrelation, and this differential effect diminishes with increasing blood velocities. Thus the potential utility of these MB contrast agents, at least as suggested by this 1-D analysis, appears enhanced in the slower flow regimes. This may be relevant for improved visualization of smaller vessels with slower blood velocities with svOCT imaging.

The reasons for the observed decrease in decorrelation times may be twofold: (i) in the no-flow regime, larger particles (RBC) undergo slower Brownian motion than the scatterers in the blood + MB mixture, with the latter’s smaller average diameter of 1.2  μm increasing the rapidity of temporal variation and signal intensity fluctuations; (ii) in the flow regime, a given MB ensemble is continuously being replaced by another, resulting in a higher temporal variation of the scattered OCT signal and higher intensity fluctuations for a given applied velocity (flow rate). As seen from Fig. 3, the former effect is considerably stronger than the latter.

3.2.

Contrast Enhancement of Two-Dimensional Structural Images

Structural OCT images, corresponding image contrasts, and ratios of blood + MB to blood contrasts acquired for different flow conditions are shown in Fig. 4. Although structural snapshot images show little visual evidence of MB contrast enhancement in Figs. 4(a) and 4(d), their averaged ROI intensities used for CW calculation via Eq. (2) clearly demonstrate the contrast difference for both microtube diameters in Figs. 4(b) and 4(e). Dynamic OCT signal changes better reflect the contrast difference as shown in standard deviation image [Fig. 4(c)], calculated from eight subsequent structural images at 0.5  mm/s blood flow velocity.

Fig. 4

Structural images, corresponding contrast of blood and blood + MB mixture and ratios of blood + MB to blood contrasts under eight flow conditions for 600- and 300-μm microtubes. (a) Structural image of the phantom with 600-μm microtubes at 0.5  mm/s blood flow velocity; (b) structural image contrast for 600-μm microtubes; (c) standard deviation image calculated from eight sequential structural images at 0.5  mm/s blood flow velocity; (d) structural image of the phantom with 300-μm microtubes at 0.5  mm/s blood flow velocity; (e) structural image contrast for 300-μm microtubes. Error bars in (b) and (e) were evaluated from data at 200 different locations (B-scans) within each phantom. Note the different vertical scales in (b) and (e), see text for explanation; (f) ratio of blood + MB/blood structural contrasts for both microtube diameters.

JBO_21_7_076014_f004.png

Similar to the temporal 1-D analysis (Fig. 3), the level of enhancement in Figs. 4(b) and 4(e) diminishes with increasing flow, but does not disappear (asymptote to unity) even for fast flows; as seen, there are still differences in the blue versus red histogram bars. Interestingly, despite the more than twofold difference in contrast values between 300- and 600-μm diameter microtubes [note the different vertical scales in Figs. 4(b) and 4(e)], ratios of blood + MB to blood contrasts (height of red histogram bars divided by height of blue ones) for both of them are similar [Fig. 4(f)]. This likely stems from the fact that contrast in OCT structural images is based on the refractive index mismatches in the interrogated material. By introducing MB to the blood, the refractive index mismatches increase which increases scattering. This phenomenon is likely independent of the flow rate and yields additional structural image contrast, compared to blood-only conditions.

The two fold difference in contrast values between 300- and 600-μm microtubes in Figs. 4(b) and 4(e) can be explained by the chosen contrast evaluation method. The OCT signal is lower at the bottom of the chosen half-circles of microtube lumens [see Figs. 2, 4(a), and 4(d)] for 600-μm diameter microtubes in comparison to 300-μm diameter ones. As contrast evaluation is based on averaged ROI intensity, more low intensity pixels in the larger microtube’s center will result in lower contrast value. The reason for the observed difference (especially at low flows) between contrast ratio curves in Fig. 4(f) may be in the faster RBC settling at low flows due to less intensive mixing/laminarity within the microtube of bigger diameter. We consider these effects not significant in the context of real in-vivo blood flow velocities reported in the literature for comparable vessel diameters (8  mm/s or faster). What is important here is a higher structural image contrast in the presence of MB in all experiments, more pronounced at low flows. Findings in Secs. 3.1 and 3.2 now set the stage for dynamic OCT image contrast investigation, as per sv studies discussed as follows.

3.3.

Contrast Enhancement of Two-Dimensional and Three-Dimensional Speckle Variance Optical Coherence Tomography Images

As pointed out in Sec. 2, in the svOCT mode, the B-scan acquisition time was set to 25 ms between the N=8 frames. This was (1) fast enough that signals from stationary agar did not decorrelate between frames for svOCT signal to approach zero, and (2) sufficiently slow to ensure complete interframe decorrelation (high svOCT signal) for pixels representing blood (see Fig. 3 for the decorrelation time plots).

Figures 5(a) and 5(c) show the svOCT images corresponding to the structural images in Figs. 4(a) and 4(d) at 0.5  mm/s blood flow velocity. As seen, there are more bright pixels in the lumen area of 300-μm diameter microtubes in comparison to 600  μm ones, likely due to OCT signal wash-out in bigger microtubes. Similar to structural contrast case, this also caused ∼ twofold difference between corresponding sv contrasts in the two microtubes [Figs. 5(b) and 5(d)]; note the different vertical scales.

Fig. 5

(a) svOCT image of the phantom with 600-μm microtubes calculated from eight subsequent structural images at 0.5  mm/s blood flow velocity; (b) svOCT image contrast for 600-μm microtubes; (c) svOCT image of the phantom with 300-μm microtubes at 0.5  mm/s blood flow velocity; (d) svOCT image contrast for 300-μm microtubes. Error bars in (b) and (d) were evaluated from data at 200 different locations within each phantom. Note the different vertical scales in (b) and (d), see text for explanation; (e) ratio of blood + MB/blood sv contrasts for both microtube diameters.

JBO_21_7_076014_f005.png

Assessment of the ratio between blood + MB and blood contrasts [Fig. 5(e)] revealed a slightly higher contrast for smaller diameter microtube at low flows (similar trends to the structural contrast ratio metric above). Despite this similarity, it is important to emphasize that svOCT contrast values were found to be (i) 2 orders of magnitude higher than structural contrast values [see Figs. 4(b), 4(e), 5(b), and 5(d) for comparison], and (ii) 1.3 to 2 times higher for blood + MB mixture than for blood for flows up to 3  mm/s [see Fig. 5(e)].

Moving on potential practical implementation of svOCT vascular contrast enhancement with MB added to the blood stream, we present en-face (xy lateral dimensions) average intensity projections of volumetric phantom scans in Fig. 6. There are eight projections for different flow conditions of a phantom with embedded 600-μm microtubes. The average velocities are indicated in each panel, and the flow direction is labeled with black arrows.

Fig. 6

En-face (xy lateral dimensions) svOCT average intensity projections of volumetric phantom scans. Eight projections corresponding to eight flow conditions in 600-μm diameter microtubes with blood (left microtube) and blood + MB mixture (right microtube). Image intensities are normalized to the maximum value. The flow direction is labeled with black arrows. The OCT slow scanning axis is antiparallel to the flow direction.

JBO_21_7_076014_f006.png

The average contrast values for these panels with deviations from phantom to phantom (error bars) can be found in Fig. 5(b). It is noticeable that at 0 to 1  mm/s flows there is a slight gradient of svOCT signal in Y-direction along microtubes. That is because svOCT imaging takes time (tens of seconds for each panel) and images reflect RBC gradual settlement at low flows. As other parameters (blood flow, MB concentration, hematocrit) remain unchanged, from these low flow images, where brighter svOCT signal is seen at right microtubes (filled with blood + MB mixture), we can assume that gradual RBC settlement occurs slower than in left microtubes (filled with blood). The possible reason for that may be MB particles exhibiting faster Brownian motion relative to RBC at low flows due to their smaller size (as mentioned in Sec. 3.1). One of the recent reports29 on OCT study of diffusive versus flow contributions to backscattered OCT signal at low flows for small particle sizes supports this hypothesis. As a result, faster motion of surrounding smaller particles contributes to slower RBC settlement, leading to more rapid intensity fluctuations and higher svOCT contrast.

However, at higher velocities the flow contribution to OCT scattering dominated over diffusive (Brownian motion) effects.29 In our experiments, this leads to increased role of RBC, dominating over MB contribution and therefore diminishing the difference between blood and blood + MB mixture contrasts in both snapshot structural OCT and dynamic svOCT images. Indeed, the common trends observed here from different analysis methods [Figs. 3, 4(b), 4(e), 5(b), and 5(d), as well as faster flow panels in Fig. 6] support these suggested mechanisms.

For potential in-vivo applications of MB-enhanced OCT, assessment of microvascular response to treatment (ultrasound ablation, photodynamic, chemical and radiation therapies of tissues) appears promising. For example, in the last scenario,30 smallest microvessels with low blood flow are known to be most sensitive to irradiation; however, their detection with svOCT is difficult due to low SNR especially at lower depths. Adding MB to the blood stream before OCT imaging may increase SNR for these vessels and improve their detection and thus help visualize their response to irradiation. Further, vascular-targeted treatments (e.g., angiogenic or antiangiogenic therapies) generally affect the vascular functionality. Preimaging the vasculature, then adding MB and imaging vasculature again may reveal “compromised” stagnant/blocked vessels with unchanged contrast (no MB) versus “functioning” increased-contrast vessels (with MB).

4.

Conclusions

We have demonstrated that the addition of MBs to blood improves the OCT structural and sv contrast at different flow conditions in two differently sized model vessels. Various OCT measurement modes were employed to demonstrate these effects, which decreased overall with increasing blood flow rates. This is potentially significant for in vivo deployment, perhaps leading to MB-enabled detection of the finest microvascular features in tissues (capillaries). These smallest microvessels are difficult to detect otherwise, yet are important of early pathology detection and treatment response monitoring.31

Further, as MB-based OCT contrast appears to decrease with flow velocity, this methodology may potentially help visualize stationary (no blood flow) conditions as blocked/stagnant microvessels amongst the flowing microvascular network, thus isolating dysfunctional perfusion foci that are otherwise difficult to detect. These findings warrant further investigations in utilizing MB for tissue hemodynamic investigations and for microvasculature contrast enhancement in OCT angiography.

Acknowledgments

This study was supported by the Canadian Institutes of Health Research, Ryerson University Health Research Grant. V.D. was also supported by Alexander Graham Bell Canada Graduate Scholarship from the Natural Sciences and Engineering Research Council of Canada.

References

1. 

A. K. Dunn et al., “Introduction: feature issue on in vivo microcirculation imaging,” Biomed. Opt. Express, 2 (7), 1861 –1863 (2011). http://dx.doi.org/10.1364/BOE.2.001861 BOEICL 2156-7085 Google Scholar

2. 

E. V. Zagaynova et al., “Contrasting properties of gold nanoparticles for optical coherence tomography: phantom, in vivo studies and Monte Carlo simulation,” Phys. Med. Biol., 53 (18), 4995 –5009 (2008). http://dx.doi.org/10.1088/0031-9155/53/18/010 PHMBA7 0031-9155 Google Scholar

3. 

T. S. Troutman, J. K. Barton and M. Romanowski, “Optical coherence tomography with plasmon resonant nanorods of gold,” Opt. Lett., 32 (11), 1438 –1440 (2007). http://dx.doi.org/10.1364/OL.32.001438 OPLEDP 0146-9592 Google Scholar

4. 

T. M. Lee et al., “Engineered microsphere contrast agents for optical coherence tomography,” Opt. Lett., 28 (17), 1546 –1548 (2003). http://dx.doi.org/10.1364/OL.28.001546 OPLEDP 0146-9592 Google Scholar

5. 

J. C. Y. Kah et al., “Control of optical contrast using gold nanoshells for optical coherence tomography imaging of mouse xenograft tumor model in vivo,” J. Biomed. Opt., 14 (5), 054015 (2009). http://dx.doi.org/10.1117/1.3233946 JBOPFO 1083-3668 Google Scholar

6. 

C. S. Kim et al., “Enhanced detection of early-stage oral cancer in vivo by optical coherence tomography using multimodal delivery of gold nanoparticles,” J. Biomed. Opt., 14 (3), 034008 (2009). http://dx.doi.org/10.1117/1.3130323 JBOPFO 1083-3668 Google Scholar

7. 

C. Yang, “Molecular contrast optical coherence tomography: a review,” Photochem. Photobiol., 81 (2), 215 –237 (2005). http://dx.doi.org/10.1562/2004-08-06-IR-266.1 PHCBAP 0031-8655 Google Scholar

8. 

K. M. Au et al., “Anti-biofouling conducting polymer nanoparticles as a label-free optical contrast agent for high resolution subsurface biomedical imaging,” Biomaterials, 34 (35), 8925 –8940 (2013). http://dx.doi.org/10.1016/j.biomaterials.2013.07.094 BIMADU 0142-9612 Google Scholar

9. 

M. J. Hsu et al., “Characterization of individual ultrasound microbubble dynamics with a light-scattering system,” J. Biomed. Opt., 16 (6), 067002 (2011). http://dx.doi.org/10.1117/1.3583575 JBOPFO 1083-3668 Google Scholar

10. 

J. K. Barton, J. B. Hoying and C. J. Sullivan, “Use of microbubbles as an optical coherence tomography contrast agent,” Acad. Radiol., 9 (1), S52 –S55 (2002). http://dx.doi.org/10.1016/S1076-6332(03)80395-1 Google Scholar

11. 

H. Assadi, R. Karshafian and A. Douplik, “Optical scattering properties of intralipid phantom in presence of encapsulated microbubbles,” Int. J. Photoenergy, 2014 1 –9 (2014). http://dx.doi.org/10.1155/2014/471764 Google Scholar

12. 

M. A. Borden et al., “Ultrasound radiation force modulates ligand availability on targeted contrast agents,” Mol. Imaging, 5 (3), 139 –147 (2006). http://dx.doi.org/10.2310/7290.2006.00016 Google Scholar

13. 

M. Todorova et al., “Antitumor effects of combining metronomic chemotherapy with the antivascular action of ultrasound stimulated microbubbles,” Int. J. Cancer, 132 (12), 2956 –2966 (2013). http://dx.doi.org/10.1002/ijc.27977 IJCNAW 1097-0215 Google Scholar

14. 

D. E. Goertz et al., “Antitumor effects of combining docetaxel (taxotere) with the antivascular action of ultrasound stimulated microbubbles,” PLoS One, 7 (12), e52307 (2012). http://dx.doi.org/10.1371/journal.pone.0052307 POLNCL 1932-6203 Google Scholar

15. 

G. J. Czarnota et al., “Tumor radiation response enhancement by acoustical stimulation of the vasculature,” Proc. Natl. Acad. Sci. U. S. A., 109 (30), E2033 –E2041 (2012). http://dx.doi.org/10.1073/pnas.1200053109 Google Scholar

16. 

C. Tarapacki and R. Karshafian, “Enhancing laser therapy using PEGylated gold nanoparticles combined with ultrasound and microbubbles,” Ultrasonics, 57 36 –43 (2015). http://dx.doi.org/10.1016/j.ultras.2014.10.015 ULTRA3 0041-624X Google Scholar

17. 

A. Mariampillai et al., “Speckle variance detection of microvasculature using swept-source optical coherence tomography,” Opt. Lett., 33 (13), 1530 –1532 (2008). http://dx.doi.org/10.1364/OL.33.001530 OPLEDP 0146-9592 Google Scholar

18. 

M. S. Mahmud et al., “Review of speckle and phase variance optical coherence tomography to visualize microvascular networks,” J. Biomed. Opt., 18 (5), 050901 (2013). http://dx.doi.org/10.1117/1.JBO.18.5.050901 JBOPFO 1083-3668 Google Scholar

19. 

I. Gorczynska et al., “Comparison of amplitude-decorrelation, speckle-variance and phase-variance OCT angiography methods for imaging the human retina and choroid,” Biomed. Opt. Express, 7 (3), 911 (2016). http://dx.doi.org/10.1364/BOE.7.000911 BOEICL 2156-7085 Google Scholar

20. 

L. A. Matveev et al., “Hybrid M-mode-like OCT imaging of three-dimensional microvasculature in vivo using reference-free processing of complex valued B-scans,” Opt. Lett., 40 (7), 1472 –1475 (2015). http://dx.doi.org/10.1364/OL.40.001472 OPLEDP 0146-9592 Google Scholar

21. 

B. J. Vakoc et al., “Cancer imaging by optical coherence tomography: preclinical progress and clinical potential,” Nat. Rev. Cancer, 12 (5), 363 –368 (2012). http://dx.doi.org/10.1038/nrc3235 NRCAC4 1474-175X Google Scholar

22. 

H. J. van Staveren et al., “Light scattering in intralipid-10% in the wavelength range of 400–1100 nm,” Appl. Opt., 30 (31), 4507 –4514 (1991). http://dx.doi.org/10.1364/AO.30.004507 APOPAI 0003-6935 Google Scholar

23. 

P. Di Ninni, F. Martelli and G. Zaccanti, “Intralipid: towards a diffusive reference standard for optical tissue phantoms,” Phys. Med. Biol., 56 (2), N21 –N28 (2011). http://dx.doi.org/10.1088/0031-9155/56/2/N01 PHMBA7 0031-9155 Google Scholar

24. 

B. Fagrell, A. Fronek and M. Intaglietta, “A microscope-television system for studying flow velocity in human skin capillaries,” Am. J. Physiol. Heart Circ. Physiol., 233 (2), H318 –H321 (1977). Google Scholar

25. 

P. Bagchi, “Mesoscale simulation of blood flow in small vessels,” Biophys. J., 92 (6), 1858 –1877 (2007). http://dx.doi.org/10.1529/biophysj.106.095042 BIOJAU 0006-3495 Google Scholar

26. 

H. N. Mayrovitz, D. Larnard and G. Duda, “Blood velocity measurement in human conjunctival vessels,” Cardiovasc. Dis., 8 (4), 509 –526 (1981). CADIDW 0093-3546 Google Scholar

27. 

T. A. Fritz and E. C. Unger, “Phase I clinical trials of MRX-115. A new ultrasound contrast agent,” Invest. Radiol., 32 (12), 735 –740 (1998). http://dx.doi.org/10.1097/00004424-199712000-00003 INVRAV 0020-9996 Google Scholar

28. 

A. Mariampillai et al., “Optimized speckle variance OCT imaging of microvasculature,” Opt. Lett., 35 (8), 1257 –1259 (2010). http://dx.doi.org/10.1364/OL.35.001257 OPLEDP 0146-9592 Google Scholar

29. 

N. Weiss, T. G. van Leeuwen and J. Kalkman, “Simultaneous and localized measurement of diffusion and flow using optical coherence tomography,” Opt. Express, 23 (3), 3448 –3459 (2015). http://dx.doi.org/10.1364/OE.23.003448 OPEXFF 1094-4087 Google Scholar

30. 

A. Maeda et al., “In-vivo optical imaging of tumor and microvascular response to ionizing radiation,” PLoS One, 7 (8), e42133 (2012). http://dx.doi.org/10.1371/journal.pone.0042133 POLNCL 1932-6203 Google Scholar

31. 

M.-T. Tsai et al., “Investigation of temporal vascular effects induced by focused ultrasound treatment with speckle-variance optical coherence tomography,” Biomed. Opt. Express, 5 (7), 2009 –2022 (2014). http://dx.doi.org/10.1364/BOE.5.002009 BOEICL 2156-7085 Google Scholar

Biography

Homa Assadi is a biomedical physics PhD candidate at the Ryerson University, Canada. She received her BS and MS degrees in physics from the University of Isfahan and Azad University, respectively. Her current research interests in optical devices such as spectroscopy, diffuse reflectance imaging, OCT imaging, LS imaging, and microbubbles contrast agents. She is a member of OSA and SPIE.

Valentin Demidov is a medical biophysics PhD candidate at University of Toronto, Canada. His research is focused on development and application of new diagnostic techniques based on the use of optical coherence tomography (OCT). Current initiatives include volumetric optical microangiography, quantification of microvascular changes in tissues with biologically relevant metrics, and alternative contrast mechanisms in OCT (image texture analysis, speckle temporal synchronization effects, interstitial fluids mapping in healthy and diseased tissues).

Raffi Karshafian: Biography is not available.

Alexandre Douplik is an associate professor of the Department of Physics at Ryerson University, Toronto, since August 2011, and also residential scientist of Keenan Research Centre at St. Michael Hospital, Toronto (since 2015). In 2007 to 2011, he was a guest professor of Nuremberg-Erlangen University, Germany. Before 2007, he was a scientific associate at Biophysics/Bioimaging Division in Ontario Cancer Institute in Toronto (2006 to 2007) and senior researcher of Xillix Technologies Corporation, Vancouver (2002 to 2006).

I. Alex Vitkin is a professor of Medical Biophysics and Radiation Oncology at the University of Toronto, a senior scientist at the University Health Network, and a clinical physicist at Princess Margaret Cancer Centre (all in Toronto, Canada). He has published ~170 papers + book chapters on biophotonics, primarily on tissue polarimetry and optical coherence tomography. He is also a fellow of OSA and SPIE, and a professor at the Nizhny Novgorod State Medical Academy (Russia).

© 2016 Society of Photo-Optical Instrumentation Engineers (SPIE) 1083-3668/2016/$25.00 © 2016 SPIE
Homa Assadi, Valentin Demidov, Raffi Karshafian, Alexandre Douplik, and I. Alex Vitkin "Microvascular contrast enhancement in optical coherence tomography using microbubbles," Journal of Biomedical Optics 21(7), 076014 (28 July 2016). https://doi.org/10.1117/1.JBO.21.7.076014
Published: 28 July 2016
Lens.org Logo
CITATIONS
Cited by 15 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Blood

Optical coherence tomography

Blood circulation

Tissues

Scattering

Speckle

In vivo imaging

Back to Top