Evaluation of Mycobacteria Growth Indicator Tube (MGIT), an Automated Culture System for Detection of Mycobacteria from Clinical Specimens
Intetsu KOBAYASHIHaruyo TODAEtsuko KOYAMAMiyuki HASEGAWANorishige HASHIGUCHIHideo ARAIToshie FUKAMIAkira WATANABE
Author information
JOURNAL FREE ACCESS

1999 Volume 73 Issue 2 Pages 172-178

Details
Abstract

We compared the Mycobacteria Growth Indicator Tube 960 (MGIT 960) and Ogawa medium (OM) for the detection of mycobacteria (acid-fast bacteria: AFB) using 882 sputum specimens. Overall, 120 strains of AFB were isolated by the MGIT 960 system and 99 strains of AFB were isolated by using OM. As far as Mycobacterium tuberculosis is concerned, 88 and 71 isolates were achieved by the MGIT 960 and OM respectively. A total of 28 isolates (18 isolates of M. tuberculosis and 10 isolates of nontuberculous mycobacteria: NTM) were detected by the MGIT 960 only whereas only 2 isolates (1 M. tuberculosis and 1 NTM) were detected by OM only. Of these sputum specimens, 72 were smear positive for AFB. The rates of smear negative but culture positive specimens were 8.0% (65 out of 809) for the MGIT 960 system and 6.2% (50 out of 809) for OM. The contamination rate for MGIT 960 was only 1.2%. The average time required for detection of M. tuberculosis was 14.1 days by the MGIT 960 system and 24.6 days by OM. For the NTM, the average detection time were 8.3 days for the MGIT 960 system and 22.8 days for OM. These results indicate that the MGIT 960 system allows detection of mycobacteria significantly faster than OM.

Content from these authors
© The Japansese Association for Infectious Diseases
Previous article Next article
feedback
Top