Brought to you by:
Paper The following article is Open access

Improving contrast between gray and white matter of Logan graphical analysis' parametric images in positron emission tomography through least-squares cubic regression and principal component analysis

, , , , and

Published 15 March 2021 © 2021 The Author(s). Published by IOP Publishing Ltd
, , Citation Paulus Kapundja Shigwedha et al 2021 Biomed. Phys. Eng. Express 7 035003 DOI 10.1088/2057-1976/abec18

2057-1976/7/3/035003

Abstract

Logan graphical analysis (LGA) is a method for in vivo quantification of tracer kinetics in positron emission tomography (PET). The shortcoming of LGA is the presence of a negative bias in the estimated parameters for noisy data. Various approaches have been proposed to address this issue. We recently applied an alternative regression method called least-squares cubic (LSC), which considers the errors in both the predictor and response variables to estimate the LGA slope. LSC reduced the bias in non-displaceable binding potential estimates while causing slight increases in the variance. In this study, we combined LSC with a principal component analysis (PCA) denoising technique to counteract the effects of variance on parametric image quality, which was assessed in terms of the contrast between gray and white matter. Tissue time–activity curves were denoised through PCA, prior to estimating the regression parameters using LSC. We refer to this approach as LSC–PCA. LSC–PCA was assessed against OLS–PCA (PCA with ordinary least-squares (OLS)), LSC, and conventional OLS-based LGA. Comparisons were made for simulated 11 C-carfentanil and 11C Pittsburgh compound B (11C-PiB) data, and clinical 11 C-PiB PET images. PCA-based methods were compared over a range of principal components, varied by the percentage variance they account for in the data. The results showed reduced variances in distribution volume ratio estimates in the simulations for LSC–PCA compared to LSC, and lower bias compared to OLS–PCA and OLS. Contrasts were not significantly improved in clinical data, but they showed a significant improvement in simulation data —indicating a potential advantage of LSC–PCA over OLS–PCA. The effects of bias reintroduction when many principal components are used were also observed in OLS–PCA clinical images. We therefore encourage the use of LSC–PCA. LSC–PCA can allow the use of many principal components with minimal risk of bias, thereby strengthening the interpretation of PET parametric images.

Export citation and abstract BibTeX RIS

Original content from this work may be used under the terms of the Creative Commons Attribution 4.0 licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI.

1. Introduction

Logan graphical analysis (LGA) is a well-accepted method used for quantitative analysis of positron emission tomography (PET) data from reversible binding radiotracer-receptor systems [13]. Generally, LGA [2] can be interpreted as a linearization technique that transforms PET tissue time-activity curves (tTACs) into linear relationships between two variables, the slope of which is a physiologically interpretable distribution volume ratio (DVR). DVR is a measure of the ratio of distribution volume in a receptor-containing to non-receptor region, and is used for assessing receptor availability [2].

LGA is highly efficient in terms of computation time, making it widely accepted; however, it introduces noise-induced bias in the DVR estimates [3, 4]. The effect of noise on the LGA parameter estimates has been well investigated [5], with the bias found to increase with increase in magnitude of both noise and the LGA slope [35].

Different techniques have been proposed to address the LGA bias problem [69]. Each method has its limitations, the most notable of which is the reduction in bias at the expense of precision [3, 4]. We recently applied an alternative regression method called least-squares cubic (LSC) [10] and demonstrated that it performs better than the conventional ordinary least-squares (OLS)-based LGA and multilinear reference tissue model 2 [8]. The LSC approach is relatively simple as it uses a different linear regression method than OLS to estimate the LGA slope. LSC-based LGA performs well as it can account for errors in both the predictor and response variables. This is suitable for the LGA variables that are contaminated with correlated noise components. The LSC approach significantly reduces the bias in the non-displaceable binding potential (BPND ) estimates [10]. BPND denotes the density of specific binding sites such as neuroreceptors, and is related to DVR by, BPND = DVR − 1. However, the reduction of bias by LSC is accompanied by a slight increase in estimate variance [10]. The slight increase influences contrast of resultant images, and accordingly, a countermeasure is required to obtain an image with higher contrast. Since the increase in estimate variance is caused by noise, we expect a combination of LSC and a denoising technique to solve the problem. Principal component analysis (PCA) is a common tool in PET parametric imaging and was previously used to denoise tTACs, where it was noted for its minimal variance [11]. In this study, we therefore applied PCA denoising to tTACs prior to parameter estimation through LSC regression, expecting bias reduction of LSC and variance reduction of PCA. We then assessed its effect on the contrast of the parametric images. Images with minimal bias and variance would make for improved contrast, since contrast is in general a form of the ratio between average estimates to the standard deviations.

PCA denoising comes at the cost of information loss, of which it increases with the number of discarded principal components (PCs) [12]. To denoise the data, only a few PCs should be retained. However, if more than necessary PCs are discarded, they may take with them a significant proportion of the data information, leading to data degradation. Furthermore, in the context of PET parametric imaging, if more than necessary PCs are retained to denoise the data, the noise is reintroduced in the data, leading to biased estimates [11]. This, therefore calls for due deligence to balance between denoising and information loss. Since the LSC method is less biased, we also expect the combination of LSC and PCA to address this issue. This is because, in combining LSC and PCA, the emphasis is not necessarily about denoising for bias reduction, but rather to just remove a few PCs to help reduce the variance in the estimates. We denote the method combining LSC and PCA as LSC–PCA, and the standard PCA that uses OLS to estimate the LGA slope as OLS–PCA.

This study was conducted on both simulated and real data. Of the two sets of simulation data, the first set mimicked a common radiotracer 11 C-carfentanil (CFN), which binds to μ-opioid in the brain, and the second set mimicked 11C Pittsburgh compound B (11C-PiB). The real data consisted of the 11 C-PiB PET brain images, the radiotracer mimicked by the second set of simulation data. 11 C-PiB binds to beta-amyloid (Aβ); therefore, 11 C-PiB PET imaging is used for in vivo quantification of Aβ deposits, mainly in patients of Alzheimer's disease.

2. Methods

2.1. Logan graphical analysis

LGA is derived from a set of differential equations describing the in vivo behavior of an administered radiotracer. LGA [2] is described by the following equation:

Equation (1)

where C(t) and CR(t) denote the radioactivity in the target and reference tissue, respectively. ${k}_{2}^{R}$ denotes the dissociation rate constant in the reference region. The target tissue, with the target receptors, is the region of interest, whereas the reference region is chosen to have a negligible concentration of target receptors. For 11 C-PiB PET data used in this study, the cerebellum gray matter region was used as the reference region [13].

The y-intercept term (int) in (1) becomes constant after a time t*. Then, a linear relationship between $\tfrac{{\int }_{0}^{t}C(u){du}}{C(t)}$ and $\tfrac{{\int }_{0}^{t}{C}^{R}(u){du}+\tfrac{1}{{k}_{2}^{R}}{C}^{R}(t)}{C(t)}$ can be obtained from (1). Applying linear regression to (1), the DVR can be estimated for time T > t*.

2.2. Least squares cubic (LSC) linear regression

The limitation of OLS-based LGA estimates is that OLS only accounts for noise in the response variable, but both the predictor and response variables of the LGA in (1) contain noise due to the noisy term C(t) in their denominators. As a linear regression method, LSC was developed to account for noise in both the predictor and response variables [14]. Therefore, if applied to LGA, LSC can account for noise in both the predictor and response variables by minimizing the weighted squared residuals in both the variables. In addition, LSC also accounts for noise correlation in the variables. Accounting for noise correlation is also important as the noise in the LGA predictor and response variables are highly correlated [5].

Mathematical details of LSC as a regression method can be found in [14]. For specific details of LSC in LGA applications, see [10].

2.3. Principal component analysis (PCA)

PCA is a dimensionality reduction technique used mostly for feature extraction and noise filtering. PCA transforms variables into new sets of variables that are linear functions of the original variables [15]. The new sets of variables are uncorrelated and are defined by sets of orthogonal basis vectors and principal components that optimally describe the variance in the data. The principal components and corresponding variance are obtained by solving for the eigenvalues and eigenvectors of the covariance matrix of the data. The eigenvectors are the principal components and eigenvalues the corresponding variances.

The original variables of the noise filtering applications used in this study are recovered using lower-dimensional projections, resulting in denoised estimates of the original variables.

The number of principal components retained to estimate the denoised data in this study were set such that the specified percentage variance could be realized. pecifically, we evaluated the estimates of LSC–PCA and OLS–PCA methods obtained from the data denoised at 9 percentage variance levels of 91% to 99%. This is because in real practical situations, the variance level at which the best trade-off between denoising and information loss is not known. It is therefore necessary to investigate the two methods over a range of percentage variance levels.

2.4. Simulation studies

A set of kinetic parameters—the exchange rate constants K1, k2−−4 and non-displaceable distribution volume (VND )—obtained from other studies, such as [11] and [16], was used to make the first simulation data set, for the CFN radiotracer.

  • •  
    11 values of DVR were set in the range of [0.0, 4.0], covering a range of [0.0, 0.35] for k3.
  • •  
    $[{K}_{1}\,\,{k}_{4}]=[0.1835\,{\text{mL cm}}^{-1}\,{{\rm{\min }}}^{-1}\,\,\,\,0.115\,{{\rm{\min }}}^{-1}]$.
  • •  
    [k2 k3] = [K1/VND (DVR − 1) · k4].
  • •  
    VND = 1.59 mLcm−1.

The data was simulated for 60 min, in 27-frame scans (6 × 10 s, 3 × 30 s, $5\,\times \,1\,{\rm{\min }}$, $5\,\times \,2.5\,{\rm{\min }}$, and $8\,\times \,5\,{\rm{\min }}$). Two-tissue compartment PET data were simulated using a clinically measured plasma time-activity curve. The reference region was formed with a minimal noise level, and DVR = 1. As per the physiological assumptions of the reference region, the delivery (K1) and clearance (k2) rates were set identical to those of the target tissues.

For 11 C-PiB simulations, the kinetic parameters, K1, k2−−4 and VND , obtained from [17], were used to simulate the two-tissue compartmental model PET data.

  • •  
    11 values of DVR were set in the range of [0.0, 4.0], covering a range of [0.0, 0.045] for k3.
  • •  
    $[{K}_{1}\,\,{k}_{2}]=[0.262\,{\text{mL cm}}^{-1}\,{{\rm{\min }}}^{-1}\,\,\,\,0.121\,{{\rm{\min }}}^{-1}]$.
  • •  
    [k3 k4] = [(DVR − 1) · k4 0.015].
  • •  
    VND = K1/k2.

For contrast comparison in simulation data, a second set of hypothetical white matter regions were simulated. For 11 C-PiB Aβ-negative images, white matter regions can have high retention of the radiotracer due to nonspecific retention [18, 19], and slow accumulation and slower clearance [18]. Therefore, the white matter regions were simulated with the values of K1 (0.190 mLcm−1) and k2 (0.100 ${{\rm{\min }}}^{-1}$) lower than those of the gray matters, and DVR values high than those of the gray matters. Thus, to simulate white matter with high DVR values than the garay matters, we added the 11 unit vector, [0.4: 0.055: 0.95], to the 11 unit vector of the gray matter DVRs. This gave the lowest white matter DVR a value of 1.6727 (the lowest for gray matter is 1.2727), and the highest of 4.95 (the highest for gray matter is 4). The ratio of these gray and white matter DVR values are within the ranges reported in a previous study [20].

The scan period and number of frames for the 11 C-PiB simulation were set to be similar to those of the clinical 11 C-PiB PET images used in this study. The scan period was set at 70 min with 25-frame scans (6 × 10 s, 3 × 20 s, $2\,\times \,1\,{\rm{\min }}$, $2\,\times \,3\,{\rm{\min }}$, and $12\,\times \,5\,{\rm{\min }}$). The reference region was formed as for CFN; with minimal noise level; DVR = 1; K1 and k2 identical to those of the target tissues.

For the two radiotracers, a noise-free tTAC was formed for each DVR value. Statistical noise was added to the noise-free tTACs to form noisy tTACs. For each noise-free tTAC, 1024 noisy tTACs were simulated as a slice of 32 × 32 voxels. Four methods, LSC–PCA, OLS–PCA, LSC, and conventional OLS-based LGA, were then used to estimate the DVR values from the noisy tTACs, and the results were compared. The true value for the dissociation rate of the reference region (${k}_{2}^{R}$) used to construct the simulation data was used for all methods.

Following the common approach [3, 7, 11], the noise model was based on counting statistics via the half-life of the radionuclide and scanning time, such that the noise increases in the later time frames, which have lower count rates. The noise was assumed to have a zero-mean Gaussian distribution with variance proportional to the true curves. This approach is well-described in [7]. The noise was scaled to within the magnitude range observed in the actual voxel-based PET data.

The simulated data of the two radiotracers, CFN and 11 C-PiB, were denoised individually by PCA. The gray and white matter of the simulation data for 11 C-PiB were denoised together. For each data set (1024 simulated tTACS for each of the 11 true DVRs), DVR estimation from the noisy tTACs were performed thrice at each percentage variance level (i.e., 27 times) for the PCA based methods. The three estimates at each variance level were differed by a slight change of <5% in the magnitude of noise level. The magnitude of the noise level refers to the percentage ratio of the means of "standard deviation in the noisy tTACs to the mean of the noise-free tTAC".

To have equal set of data for all methods, the LSC and OLS estimates were performed 27 times at noise levels corresponding to those of the PCA methods. The mean and standard deviations were then calculated for 27 estimates for each 1024 simulated tTACs for each true DVR value.

For 11 C-PiB data, in addition to the averages and standard deviations, the DVR parametric estimates were compared numerically in terms of the contrast between the simulated gray and white matter regions. The contrast index was calculated as [21]:

Equation (2)

where μG and μW denote the mean DVR of the gray and white matter regions, respectively, and σG and σW denote the standard deviations of DVR in the gray and white matter, respectively.

2.5. Human data studies

A cohort of 12 (11 Aβ-negative and 1 Aβ-positive) subjects was included in this study. The subjects underwent 70 min 11 C-PiB PET scans of 25 frames (6 × 10 s, 3 × 20 s, $2\,\times \,1\,{\rm{\min }}$, $2\,\times \,3\,{\rm{\min }}$, and $12\,\times \,5\,{\rm{\min }}$). The scan procedure was carried out according to the Alzheimer's Disease Neuroimaging Initiative (ADNI) 11 C-PiB standards. The study was approved by the Ethics Committee of Kindai University Hospital, and written informed consent was obtained from all participants.

The obtained PET data were in the form of voxels of size 2.1 × 2.1 × 3.4 mm3 in matrices of 128 × 128 × 47 slices. DVR parametric images were then generated from the PET data using the LSC–PCA and OLS–PCA methods.

As with the simulation data, DVR estimates were performed at the 9 different percentage variance levels. However, the estimates were only performed once at each percentage variance since there is no noise adjustment. Contrasts were then calculated and compared for LSC–PCA and OLS–PCA for these estimates. Contrasts were calculated between the four main gray matter cortices (frontal, temporal, occipital, and parietal) and white matter (corona radiata).

All data analyses were performed in MATLAB (The MathWorks, Inc., Natick, MA, USA).

3. Results

3.1. Simulation studies

figure 1 presents the simulated tTACs, mimicking the two radiotracers, 1(a) CFN and 1(b) 11 C-PiB. Noise-free tTACs are represented by the thick curves, whereas the thin curves represent the noisy tTACs. For each noise-free curve, out of the 1024 simulated noisy curves, only one is shown in each of the subfigures in figure 1. The lowermost thick curve denotes the reference region, and the upper 11 curves denote the 11 DVR values.

Figure 1.

Figure 1. Simulated tTACs for the radiotracers, (a) CFN and (b) 11 C-PiB. In each figure, the thick and thin curves represent the noise-free and noisy tTACs, respectively. The upper 11 thick curves correspond to the 11 DVR values, whereas the 12th and lowest curve correspond to the time-activity curve in the reference region.

Standard image High-resolution image

figure 2 shows examples of PCA-denoised tTACs for the two radiotracers, 2(a) CFN and 2(b) 11 C-PiB. The percentage labels of the tTACs, PCA (91%)—PCA (99%), in the legends denote the tTACs denoised with the number of PCs that respectively explain 91%—99% variance in the data. The corresponding noise-free and noisy (unprocessed by PCA) tTACs are also included. The average number of PCs that denoised the data at the specified percentage variance levels are shown in table 1. Table 1 shows that the CFN data required more PCs to attain the specified percentage variance value, compared with 11 C-PiB. The total number of PCs for the CFN and 11 C-PiB data are 27 and 25, respectively, corresponding to their number of time points. For both radiotracers, the tTACs denoised at 98% variance level are showing to be still noisy. At 95% and 91%, noise reduction could be seen in the tTACs. One can consider lowering the number of PCs, but that would equate to lowering the amount of explained variance, and hence information loss.

Figure 2.

Figure 2. Simulated noisy tTACs and their corresponding PCA-denoised tTACs for the two radiotracers, (a) CFN and (b) 11 C-PiB. Nine PCA-denoised tTACs are shown for each radiotracer. The 9 denoised tTACs (as labelled) correspond to PCA with a number of PCs that explain 91%—99% variance in the data.

Standard image High-resolution image

Table 1. The average (and standard deviations in brackets) number of PCs that reached the specified percentage variance for denoising the data by PCA. (see full table in "Additional file" (table A 1)).

Number of PCs
  91% 93% 95% 96% 98% 99%
CFN 9.59 (2.27)12.07 (2.09)14.85 (1.72)16.33 (1.54)19.85 (1.06)21.89 (0.42)
11 C-PiB 5.33 (3.03)6.33 (4.04)9.52 (3.09)10.96 (2.83)15.48 (2.49)18.67 (2.04)

figure 3 compares the percentage recovered DVR in the averages of the values estimated from the simulated noisy tTACs using four methods, LSC–PCA, OLS–PCA, LSC, and OLS. In other words, the difference from 100% denotes the percentage bias. For LSC–PCA and OLS–PCA, the averages are for estimates from tTACs denoised at 9 different percentage variance levels, estimated thrice at each level (27 estimates). LSC and OLS averages are from tTACs at 27 different noise levels. As expected, LSC–PCA estimates have smaller error bars than the LSC estimates. This is especially noticeable for 11 C-PiB estimates in figure 3(b), in which some larger variances seen in LSC estimates are well stabilized in LSC–PCA. This demonstrates PCA's effectiveness in reducing the variance in the estimates —the purpose for which it is needed in combination with LSC. In figure 3(a), we see minimal effects in the bias and standard deviations in the estimates for both LSC–PCA compared to LSC, and OLS–PCA compared to OLS. In this case, LSC–PCA maintains its advantage of smaller biases, whereas the bias in OLS–PCA estimates approach those for OLS. The difference in the observations for the two radiotracer data could be due to that CFN data required more number of PCs, thus reintroducing the noise in the data. Radiotracer-focused studies may help determine the ranges of percentage variance that should be retained for various radiotracers. Tables 2 and 3 show the average values of the estimated DVR for the respective CFN and 11 C-PiB radiotracer, and respectively corresponding to figures, 3(a) and figure 3(b).

Figure 3.

Figure 3. Percentage recovered DVR of the average (N = 1024*10) DVR values estimated from noisy tTACs using the four methods, LSC–PCA, OLS–PCA, LSC, and OLS, for the two radiotracers, CFN in (a), and 11 C-PiB in (b). The error bars denote standard deviations. For LSC–PCA and OLS–PCA, the estimates were performed thrice at 9 different percentage variances, 91—99%. For LSC and OLS, the estimates were performed 27 times (27 different noise levels).

Standard image High-resolution image

Table 2. Average DVR values (and standard deviations in brackets) estimated from the noisy tTACs by the four methods, LSC–PCA, LSC, OLS–PCA, and OLS, for simulated CFN radiotracer. These data correspond to figure 3(a).

T rue DVR 1.250 1.800 2.350 2.900 3.450 4.00
Estimated DVR
LSC-PCA 1.249 (0.069)1.803 (0.123)2.356 (0.191)2.904 (0.268)3.454 (0.359)3.987 (0.462)
OLS-PCA 1.207 (0.066)1.719 (0.114)2.217 (0.173)2.698 (0.236)3.165 (0.304)3.605 (0.372)
LSC 1.250 (0.074)1.795 (0.133)2.341 (0.208)2.892 (0.299)3.449 (0.414)4.002 (0.537)
OLS 1.206 (0.069)1.707 (0.118)2.193 (0.176)2.666 (0.224)3.123 (0.317)3.556 (0.396)

Table 3. Average DVR values (and standard deviations in brackets) estimated from the noisy tTACs by the four methods, LSC–PCA, LSC, OLS–PCA, and OLS, for simulated 11 C-PiB radiotracer. These data correspond to figure 3(b).

True DVR 1.250 1.800 2.350 2.900 3.450 4.00
Estimated DVR
LSC-PCA 1.167 (0.059)1.559 (0.121)1.994 (0.222)2.465 (0.322)2.950 (0.440)3.455 (0.574)
OLS-PCA 1.128 (0.053)1.459 (0.105)1.812 (0.172)2.187 (0.254)2.551 (0.341)2.926 (0.438)
LSC 1.167 (0.072)1.549 (0.152)1.971 (0.264)2.430 (0.419)2.896 (0.586)3.400 (0.142)
OLS 1.119 (0.058)1.426 (0.112)1.742 (0.174)2.070 (0.254)2.379 (0.323)2.694 (0.407)

figure 4 compares the contrast between the simulated gray and white matter regions. The comparisons include all calculated DVR values (three times at 9 percentage variance levels for LSC–PCA and OLS–PCA, and at 27 different noise levels for LSC and OLS). Statistical comparisons of these contrast are shown in table 4, using the Friedman test [22] and Post-Hoc multiple comparison. A statistical significant difference is observed between LSC–PCA and all other methods. On the other hand, no statistical significant difference is observed between OLS–PCA and LSC.

Figure 4.

Figure 4. Box plots of the DVR contrast between the hypothetical simulated gray and white matter regions, calculated for all DVR values estimated at different percentage variance levels for LSC–PCA and OLS–PCA, and at 27 different noise levels for LSC and OLS. The circles in the box plots denote the means, whereas the horizontal lines denote the medians.

Standard image High-resolution image

Table 4. Friedman test and Post-Hoc multiple comparisons of the p-values of the contrast values of simulated 11 C-PiB data plotted in figure 4. Tabulated numerical values correspond to the p-values between the two methods in the corresponding rows and columns.

Friedman Test (p value < 0.001)
Post-Hoc Multiple Comparisons p-values
 OLS-PCALSCOLS
LSC–PCA0.0410.022<0.001
OLS–PCA 0.804<0.009
LSC  0.018

3.2. Human data studies

figure 5 compares the contrast between gray matter cortices and white matter for clinical 11 C-PiB DVR images obtained by LSC–PCA and OLS–PCA, calculated for the 11 Aβ-negative subjects, at 9 different percentage variance levels. The Friedman test achieved an overall p value of 0.008, and a Pot-Hoc p value of 0.187 between LSC-PCA and OLS-PCA.

Figure 5.

Figure 5. Box plots of the DVR contrast between gray matter regions; the frontal, temporal, occipital and parietal cortices, and white matter region (corona radiata), calculated for the 11 Aβ-negative subjects. The calculations included all DVR estimates obtained by denoising the data at 9 different percentage variance levels. Only LSC–PCA and OLS–PCA are included since LSC and OLS could only be estimated once. The circles in the box plots denote the means, whereas the horizontal lines denote the medians.

Standard image High-resolution image

figure 6 shows axial slices of the DVR parametric images obtained using four methods, LSC–PCA, OLS–PCA, LSC, and OLS. PCA-based methods were denoised with the number of PCs that explained 95% variance for these images. The two images in the upper row are from two Aβ-negative subjects, while the bottom row is from the Aβ-positive subject. The means and standard deviations of the number of PCs that denoised the data at the specified percentage variance levels for the 12 clinical 11 C-PiB brain images are shown in table 5. These data shows the expected trend of the number of PCs increasing with the amount explained percentage variance in the data.

Figure 6.

Figure 6.  DVR parametric images as obtained by the four methods, LSC–PCA, OLS–PCA, LSC and OLS. The two upper rows are for Aβ-negative subjects, whereas the bottom row is for the Aβ-positive subject. The tTACs were denoised with PCs to explain 95% variance for these images.

Standard image High-resolution image

Table 5. Average (and standard deviations in brackets) number of PCs required to reach the specified percentage variance for denoising the data by PCA. (see full table in "Additional file" (table A 1).

Number of PCs
91% 93% 95% 96% 98% 99%
3.83 (1.19)4.92 (1.16)6.33 (1.15)7.33 (1.15)10.75 (1.96)15.08 (2.31)

In figure 6, noise components in the images estimated by the LSC method can be seen appearing as random noisy pixels/regions with high DVR values in the brain region (see the black arrows in figure 6). Images obtained by the LSC–PCA method exhibit a rather fair distribution of the DVR values with no sharp increases in the pixels/regions in the main brain region. However, some bright pixels could be seen on the outskirts of the LSC–PCA images in the middle and bottom row. These could be attributed to the LSC, as this is part of the variance concept which is being addressed by combining LSC and PCA. Further analysis should be carried out to investigate if these bright pixels could be observed within the brain region in LSC–PCA images. The images obtained by OLS–PCA appear similar to those for LSC–PCA. The LSC images thus appear slightly chunky (compared to the LSC–PCA and OLS–PCA images), for these specific settings, which can hinder the contrast between the brain structures. The lowest DVR estimates are observed in OLS-based images, as expected.

figure 7 shows DVR parametric images in which the difference between LSC–PCA and OLS–PCA could be seen at some levels of percentage variance for PCA denoising. In this figure we can see that as the percentage variance to be explained is increased to 97% from 95%, the OLS–PCA estimates get biased. Calculated contrast values for this subject for DVR estimates from the data denoised at 97% variance are shown in table 6, showing lower values for OLS–PCA. Another case for an image showing biased estimates for OLS–PCA is shown in "Additional file" (figure 2A). These results reveals the effect of noise reintroduction in the data, when many PCs are used, on the OLS–PCA estimates.

Figure 7.

Figure 7.  DVR parametric images as obtained by the four methods, LSC–PCA, OLS–PCA, LSC and OLS for a different Aβ-negative subject from those in figure 6 . Two images are displayed for LSC–PCA (95%) and OLS–PCA (97%), denoting that the image is estimated from the data denoised at 95% and 97%, respectively. The total number of PCs adding up to these amounts of explained variances for this subject are 7 and 14, respectively.

Standard image High-resolution image

Table 6. Contrast values for the single 11 C-PiB subject plotted in figure 7, for 97% variance for PCA-based methods.

Contrast
 FrontalTemporalOccipitalParietal
LSC-PCA 1.1621.0301.1101.049
OLS-PCA 1.0300.9390.9620.847
LSC 1.0070.9150.9190.789
OLS 0.9050.8310.8130.624

4. Discussion

In this study we observed that OLS–PCA estimates can be affected by noise reintroduction when many number of PCs are used. For example, the data of the subject of the slices shown in figure 7 required the most number of PCs (14) to denoise whilst maintaining 98% of variance in the data. The effect of noise reintroduction can be seen in this figure in terms lower estimates for OLS–PCA method. LSC–PCA images remained less biased for this subject. LSC reduces the bias, and PCA fends off variation effects in the estimates, which could help maintain the contrast over a wide range of percentage variance denoising levels.

The significant p value observed between LSC–PCA and OLS–PCA in simulation results could be an indication that the same can be obtained for clinical data. This can be investigated with a larger cohort of clinical 11 C-PiB brain data.

For CFN simulations, we saw that the OLS–PCA estimates remained biased almost similar to the OLS estimates. In this case, the LSC–PCA estimates remained unbiased, denoting the importance of employing LSC for bias reduction.

For 11 C-PiB simulations, we saw the effectiveness of PCA in reducing the variance in the estimates of LSC–PCA leading to estimates with both low bias and variance. The bias in the OLS–PCA estimates could still be seen in the DVR estimates for 11 C-PiB simulated data.

The results in this study highlights the importance of adopting a less-biased regression method, such as LSC, even with the PCA tTACs denoising method. This is important, especially when the number of PCs to be used is uncertain, which is usually the case in practical settings.

PCA has been shown to reduce bias when using only a few PCs [11]. However, using a minimal number of PCs might lead to information loss, leading to lower contrasts (see additional file, figure 1A). This means that information is lost when PCs are discarded, i.e., the fewer the number of PCs used, the more the information loss. The number of PCs to retain depends on the purpose of the study. For PET imaging, to avoid information loss, it is necessary to retain sufficient variance to reconstruct the denoised tTACs.

For LSC, in addition to the two input variables (predictor and response), estimation of the weights and correlation of the errors in the two variables are required. The appropriateness of the weight functions used, specifically for LGA, is demonstrated in [10]. However, they are only estimates, and there is room for further improvements.

In this study we focused on PCA, which is most common, and assess the effects of combining it with a less biased regression method —instead of OLS— for DVR parametric estimation. However, the study of PET data denoising is ongoing, and several other approaches have been proposed. Some recent advanced methods include the use of Deep Convolutional Neural Networks [23], and HighlY constrained backPRojection (HYPR) is an image-processing technique [24], which involves the creation of a composite image from the time series, then denoise the individual time frames using the high SNR of the composite image. The scope of this study is however about the effect of employing LSC with a tTAC denoising method, as opposed to using that tTAC denoising method with OLS. PCA is a common tool, and thus it was considered for the denoising method in this study.

5. Conclusion

In this study, we focused on the fusion of LSC and PCA methods, LSC–PCA, to improve the contrast of PET parametric images. LSC–PCA was compared to OLS–PCA, LSC and OLS for simulation data. Comparisons of the contrast values for simulation data of 11 C-PiB showed a significant difference between LSC–PCA and the other methods. The comparison of contrast in clinical images showed a non significant difference from that of OLS–PCA, but the significant difference observed in simulation data is indicative of a significant difference in clinical data, which should be investigated with a larger cohort of data. Several cases were also revealed in clinical images where OLS–PCA estimates get biased when the number of PCs used to denoise the data is increased. This study demonstrates the importance of using LSC as a regression method even when the data are denoised with PCA. The complementary capabilities of LSC and PCA assure that both bias and variances will be reduced, unlike in OLS–PCA with bias risk. These findings should be useful in PET parametric imaging for appropriate interpretation of the extracellular binding in the brain. Specifically for 11 C-PiB imaging, improved contrast will help in the assessment of Alzheimer's Disease and associated cognitive impairments.

The study conducted for the results reported in this study is limited to only two radiotracers, CFN and 11 C-PiB for simulation studies, and one radiotracer (11C-PiB) for clinical data. Furthermore, the clinical data cohort was limited and consisted mostly of Aβ-negative subjects. Future studies could seek to expand on these matters. Other future studies could also focus on self-improving LSC for weight estimation, and providing a tracer-specific analysis of PCA denoising. Using other tTACs denoising methods such as neural networks, and evaluating clinical images with professional physicians are other important aspects to be considered in further studies.

Acknowledgments

We would like to thank Editage (www.editage.com) for English language editing.

Author's contribution

Concepts and designPKS, YF, YK. Literature researchPKS, YK, YF. Clinical data collectionKI, KH. Clinical data handling and preprocessingYK, TY. Data analysisPKS, YF. Manuscript preparationPKS. Manuscript editingYK, YF, PKS. All authors have read and approved the final manuscript.

Funding

None.

Data availability

The data that support the findings of this study are available upon reasonable request from the authors.

Competing interests

The authors declare that they have no competing interests.

Please wait… references are loading.
10.1088/2057-1976/abec18