Paper The following article is Open access

Operation of NIM5 fountain with 1.5x10-15 uncertainty and design of new NIM6 in NIM

, , , , and

Published under licence by IOP Publishing Ltd
, , Citation F Fang et al 2016 J. Phys.: Conf. Ser. 723 012009 DOI 10.1088/1742-6596/723/1/012009

1742-6596/723/1/012009

Abstract

The cesium fountain primary frequency standard NIM5 started to operate since 2008 and started to report to BIPM since 2014. The major constrains of NIM5 is a relatively large background signal at the detection and microwave leakages due to the Ramsey cavity. A new fountain clock NIM6 is under construction. Besides some improvements on the vacuum system, a new Ramsey cavity and a microwave synthesizer are made to reduce the Type B uncertainty. Another feature of NIM6 is collecting atoms from a MOT loading optical molasses to get more atoms with a more uniform density distribution. With a new frequency synthesizer based on cryogenic sapphire oscillator (CSO), NIM6 is aiming to reach the quantum projection noise, thus leading to a reduced Type A uncertainty compared with NIM5.

Export citation and abstract BibTeX RIS

Content from this work may be used under the terms of the Creative Commons Attribution 3.0 licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI.

Please wait… references are loading.
10.1088/1742-6596/723/1/012009