Abstract
In this study, a 4D dosimetry concept was developed. This concept included a method for calculation of 3D reference absorbed dose matrices at every control point of the delivery using a clinical treatment planning system (TPS). Further, the gamma evaluation method was extended to incorporate the 4th dimension of the TPS calculated dose distributions. The applications of the 4D dosimetry concept on pre-treatment quality control and real-time in vivo dosimetry were investigated.
Export citation and abstract BibTeX RIS
Content from this work may be used under the terms of the Creative Commons Attribution 3.0 licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI.