Indicator for hydration balance during haemodialysis based on anisotropic FEM

, , , , and

Published 11 June 2008 2008 Institute of Physics and Engineering in Medicine
, , Citation M Freiberger et al 2008 Physiol. Meas. 29 S479 DOI 10.1088/0967-3334/29/6/S40

0967-3334/29/6/S479

Abstract

The current paper proposes a new indicator for well-balanced dehydration of patients undergoing a haemodialysis. It is based on an estimator for the extra-cellular volume and the ultrafiltration rate. The extra-cellular fluid was computed from continuous tetrapolar bio-impedance measurements taken on the lower leg in a frequency range of several kilohertz up to 500 kHz. Finite element simulations on different leg models with anisotropic conductivities calculated with Cole models were carried out in order to incorporate the significant anisotropy of human tissue into the estimation process. The indicator was tested on measurement data gathered from 25 persons during 150 haemodialysis sessions. Its performance was determined by computing ROC curves. Results of the data analysis are reported.