Metal cell for use in vapour pressure measurements by the effusion method

, and

Published under licence by IOP Publishing Ltd
, , Citation S Blairs et al 1961 J. Sci. Instrum. 38 469 DOI 10.1088/0950-7671/38/12/305

0950-7671/38/12/469

Abstract

Full constructional details are given for an effusion cell in Armco iron, which in conjunction with a calibrated glass spring and cathetometer has been used to measure the vapour pressure over solid ferric chloride in the temperature range 100-200° C. The chief advantage claimed for the cell is that the effusion orifices, in interchangeable push fit end-caps, are of the form of true circular knife edged holes with easily determined dimensions.

Export citation and abstract BibTeX RIS

10.1088/0950-7671/38/12/305