Patient dosimetry in abdominal arteriography

, , , and

Published under licence by IOP Publishing Ltd
, , Citation G J Kemerink et al 1999 Phys. Med. Biol. 44 1133 DOI 10.1088/0031-9155/44/5/002

0031-9155/44/5/1133

Abstract

This study aims at accurate quantification of x-ray exposure and effective dose to the patient in abdominal arteriography. Using an automatic monitoring system, all relevant exposure parameters were determined during 172 abdominal arteriographies. Common projections were extracted for a `normal' reference group of procedures and used in Monte Carlo calculations of dose-area product to organ dose conversion coefficients. Dose-area product, organ doses and effective dose were quantified for intravenous and intra-arterial procedures. The large data sets describing exposure could be condensed to a set of 28 common views. New coefficients to convert dose-area product to organ equivalent dose and effective dose were calculated for nine views contributing approximately 80% to the total dose-area product. The average dose-area product was in intravenous procedures and in intra-arterial procedures. The corresponding average effective doses to the patient were 4 mSv and 6 mSv respectively (range 2-12 mSv, actual value depending on procedure type and gender). It is concluded that automatic monitoring of x-ray exposure parameters, complemented by the calculation of Monte Carlo organ dose conversion coefficients, is a feasible and promising approach to accurate dosimetry of complex arteriographic procedures.