Bioelectromagnetic localization of a pacing catheter in the heart

, , , , , , , , , and

Published under licence by IOP Publishing Ltd
, , Citation K Pesola et al 1999 Phys. Med. Biol. 44 2565 DOI 10.1088/0031-9155/44/10/314

0031-9155/44/10/2565

Abstract

The accuracy of localizing source currents within the human heart by non-invasive magneto- and electrocardiographic methods was investigated in 10 patients. A non-magnetic stimulation catheter inside the heart served as a reference current source. Biplane fluoroscopic imaging with lead ball markers was used to record the catheter position. Simultaneous multichannel magnetocardiographic (MCG) and body surface potential mapping (BSPM) recordings were performed during catheter pacing. Equivalent current dipole localizations were computed from MCG and BSPM data, employing standard and patient-specific boundary element torso models. Using individual models with the lungs included, the average MCG localization error was 7±3 mm, whereas the average BSPM localization error was 25±4 mm. In the simplified case of a single homogeneous standard torso model, an average error of 9±3 mm was obtained from MCG recordings. The MCG localization accuracies obtained in this study imply that the capability of multichannel MCG to locate dipolar sources is sufficient for clinical purposes, even without constructing individual torso models from x-ray or from magnetic resonance images.