Design and calibration of electric field probes in the range 10 - 120 MHz

, and

Published under licence by IOP Publishing Ltd
, , Citation Helen C Taylor et al 1997 Phys. Med. Biol. 42 1387 DOI 10.1088/0031-9155/42/7/011

0031-9155/42/7/1387

Abstract

In view of potential thermal hazards, there is a need to determine the specific absorption rate (SAR) distributions associated with radiofrequency coils used in magnetic resonance imaging (MRI) (typically 10 - 120 MHz). Electric field (E-field) distributions in tissue-equivalent phantoms may be determined using a probe comprising a dipole antenna and a detector. The geometry of the dipole dictates the sensitivity of the device, thus two designs are discussed in this paper. Both probes are compact, have a spatial resolution of , operate at MR frequencies and have a response independent of the dielectric characteristics of the phantom material. Calibration of these probes requires a system capable of producing a known E-field both in air and in a tissue-like medium at frequencies between 10 and 120 MHz. Transverse electromagnetic wave (TEM) cells answering these specifications are described and the calibration procedure outlined. Accurately calibrated E-field probes can make field measurements in phantoms which can be used to verify predictions from numerical models. These numerical techniques may then be used to predict E-fields, and hence SAR, in patients.

Export citation and abstract BibTeX RIS

10.1088/0031-9155/42/7/011