Investigation of the imaging properties of an X-ray film scanner

, , and

Published under licence by IOP Publishing Ltd
, , Citation W Doler et al 1994 Phys. Med. Biol. 39 917 DOI 10.1088/0031-9155/39/5/010

0031-9155/39/5/917

Abstract

The imaging properties of a CCD X-ray film scanner were investigated. The relationship between optical density and the logarithm of output pixel value provides information on the linearity of the system. The resolution properties of the scanner can be characterized by the presampling modulation transfer function (MTF) which was calculated from a set of line spread functions with various alignments relative to the sampling grid. The authors' results show that the scanner is linear up to 2.0 optical density units, which leads to restrictions in digitizing X-ray films with higher optical density. The measured MTF shows the good spatial resolution of the CCD scanner which is demonstrated on an example.

Export citation and abstract BibTeX RIS