Whole-body single-photon emission computed tomography using dual, large-field-of-view scintillation cameras

, , and

Published under licence by IOP Publishing Ltd
, , Citation R J Jaszczak et al 1979 Phys. Med. Biol. 24 1123 DOI 10.1088/0031-9155/24/6/003

0031-9155/24/6/1123

Abstract

A whole-body single-photon emission computed tomography system (SPECT) consisting of two large-field-of-view scintillation cameras mounted on a rotatable gantry, a minicomputer and a display station has been designed, constructed and evaluated. In its usual mode of operation, eleven contiguous transverse sections, each 12.5 or 25 mm thick, are reconstructed from projection data acquired during a single, continuous 360 degrees rotation lasting from 2 to 22 min. A generalised filtered and weighted backprojection algorithm is used to reconstruct data obtained with conventional parallel-hole collimators in the case of body scanning, or with specially designed fan beam collimators in the case of centrally positioned organs. A simple yet effective, correction is used to compensate for the effects of gamma ray attenuation within the patient.