SHORT COMMUNICATION

Comparison of the BIPM cryogenic radiometer with a mechanically cooled cryogenic radiometer from the NPL

, , , and

Published under licence by IOP Publishing Ltd
, , Citation R Goebel et al 1996 Metrologia 33 177 DOI 10.1088/0026-1394/33/2/8

0026-1394/33/2/177

Abstract

The direct comparison at the Bureau International des Poids et Mesures (BIPM) of the BIPM cryogenic radiometer (Radiox) with a mechanically cooled cryogenic radiometer (MCR) from the National Physical Laboratory (NPL) is described. The radiometers were compared using a laser beam at λ = 647 nm with power levels of 400 μW and 650 μW. They were found to be in good agreement, as the measured difference in responsivity was found to be 1,5 parts in 105, with a combined standard uncertainty of 8 parts in 105.

Export citation and abstract BibTeX RIS

10.1088/0026-1394/33/2/8