CC BY-NC-ND 4.0 · Endosc Int Open 2018; 06(12): E1417-E1422
DOI: 10.1055/a-0754-2290
Original article
Owner and Copyright © Georg Thieme Verlag KG 2018

Usefulness of the “newly designed” short-type single-balloon enteroscope for ERCP in patients with Roux-en-Y gastrectomy: a pilot study

Yuki Tanisaka
1   Department of Gastroenterology, Saitama Medical University International Medical Center, Saitama, Japan
,
Shomei Ryozawa
1   Department of Gastroenterology, Saitama Medical University International Medical Center, Saitama, Japan
,
Masafumi Mizuide
1   Department of Gastroenterology, Saitama Medical University International Medical Center, Saitama, Japan
,
Masanori Kobayashi
1   Department of Gastroenterology, Saitama Medical University International Medical Center, Saitama, Japan
,
Akashi Fujita
1   Department of Gastroenterology, Saitama Medical University International Medical Center, Saitama, Japan
,
Kazuhiro Minami
1   Department of Gastroenterology, Saitama Medical University International Medical Center, Saitama, Japan
,
Tsutomu Kobatake
1   Department of Gastroenterology, Saitama Medical University International Medical Center, Saitama, Japan
,
Kumiko Omiya
1   Department of Gastroenterology, Saitama Medical University International Medical Center, Saitama, Japan
,
Hirotoshi Iwano
1   Department of Gastroenterology, Saitama Medical University International Medical Center, Saitama, Japan
,
Ryuichiro Araki
2   Community Health Science Center, Saitama Medical University, Saitama, Japan
› Author Affiliations
Further Information

Publication History

submitted07 May 2018

accepted after revision22 August 2018

Publication Date:
29 November 2018 (online)

Abstract

Background and aims In recent years, the short single-balloon enteroscope (SBE) has been used during endoscopic retrograde cholangiopancreatography (ERCP) for patients with surgically altered anatomy and has been reported to be useful. However, difficulties remain, and the procedures and devices need improvements. We assessed the usefulness and superiority of a new short SBE equipped with passive bending and high-force transmission by comparisons with outcomes using the conventional short SBE.

Methods This study evaluated short SBE-assisted ERCP-related procedures for Roux-en-Y gastrectomy between September 2011 and October 2017. Outcomes including the procedural success rate, which was the primary outcome, were assessed to compare the conventional short SBE (SIF-Y0004 [prototype]) and the new short SBE (SIF-H290S).

Results Of 74 procedures performed in 61 patients, 51 procedures in 39 patients involved the SIF-Y0004, and 23 procedures in 22 patients involved the SIF-H290S. The procedural success rates were 70.6 % for SIF-Y0004, and 95.7 % for SIF-H290S, representing better results for the new short SBE (P = 0.02). The new short SBE also had a superior diagnostic success rate (P = 0.047) and median time to reach the blind end (P < 0.001).

Conclusions Roux-en-Y gastrectomy patients treated with the new short SBE had better outcomes than those treated with conventional short SBE. More cases need to be studied; however, the new short SBE has the potential to improve ERCP outcomes in patients with surgically altered anatomy.

 
  • References

  • 1 Elton E, Hanson BL, Qaseem T. et al. Diagnostic therapeutic ERCP using an enteroscope and pediatric colonoscope in long-limb surgical bypass patients. Gastrointest Endosc 1998; 47: 62-67
  • 2 Wright BE, Cass OW, Freeman ML. ERCP in patients with long-limb Roux-en-Y gastrojejunostomy and intact papilla. Gastrointest Endosc 2002; 56: 225-232
  • 3 Yamamoto H, Sekine Y, Sato Y. et al. Total enteroscopy with a nonsurgical steerable double-balloon method. Gastrointest Endosc 2001; 53: 216-220
  • 4 Aabakken L, Bretthauer M, Line PD. Double-balloon enteroscopy for endoscopic retrograde cholangiography in patients with a Roux-en-Y anastomosis. Endoscopy 2007; 39: 1068-1071
  • 5 Chu YC, Yang CC, Yeh YH. et al. Double-balloon enteroscopy application in biliary tract disease – its therapeutic and diagnostic functions. Gastrointest Endosc 2008; 68: 585-591
  • 6 Iwamoto S, Ryozawa S, Yamamoto H. et al. Double balloon endoscope facilitates endoscopic retrograde cholangiopancreatography in roux-en-y anastomosis patients. Dig Endosc 2010; 22: 64-68
  • 7 Shimatani M, Hatanaka H, Kogure H. et al. Diagnostic and therapeutic endoscopic retrograde cholangiography using a short-type double-balloon endoscope in patients with altered gastrointestinal anatomy: A multicenter prospective study in Japan. Am J Gastroenterol 2016; 111: 1750-1758
  • 8 Kawamura T, Yasuda K, Tanaka K. et al. Clinical evaluation of a newly developed single-balloon enteroscope. Gastrointest Endosc 2008; 68: 1112-1116
  • 9 Wang AY, Sauer BG, Behm BW. et al. Single-balloon enteroscopy effectively enables diagnostic and therapeutic retrograde cholangiography in patients with surgically altered anatomy. Gastrointest Endosc 2010; 71: 641-649
  • 10 Saleem A, Baron TH, Goustout CJ. et al. Endoscopic retrograde cholangiopancreatography using a single-balloon enteroscope in patients with altered Roux-en-Y anatomy. Endoscopy 2010; 42: 656-660
  • 11 Yamauchi H, Kida M, Okuwaki K. et al. Short-type single balloon enteroscope for endoscopic retrograde cholangiopancreatography with altered gastrointestinal anatomy. World J Gastroenterol 2013; 19: 1728-1735
  • 12 Iwai T, Kida M, Yamauchi H. et al. Short-type and conventional single-balloon enteroscopes for endoscopic retrograde cholangiopancreatography in patients with surgically altered anatomy: single-center experience. Dig Endosc 2014; 26 (Suppl. 02) 156-163
  • 13 Shimatani M, Takaoka M, Ikeura T. et al. Evaluation of endoscopic retrograde cholangiopancreatography using a newly developed short-type single-balloon endoscope in patients with altered gastrointestinal anatomy. Dig Endosc 2014; 26 (Suppl. 02) 147-155
  • 14 Kawamura T, Uno K, Suzuki A. et al. Clinical usefulness of a short-type, prototype single-balloon enteroscope for endoscopic retrograde cholangiopancreatography in patients with altered gastrointestinal anatomy: preliminary experiences. Dig Endosc 2015; 27: 82-86
  • 15 Yane K, Katanuma A, Maguchi H. et al. Short-type single-balloon enteroscope-assisted ERCP in postsurgical altered anatomy: potential factors affecting procedural failure. Endoscopy 2017; 49: 69-74
  • 16 Cotton PB, Eisen GM, Aabakken L. et al. A lexicon for endoscopic adverse events: report of an ASGE workshop. Gastrointest Endosc 2010; 71: 446-454
  • 17 Trindade AJ, Mella JM, Slattery E. et al. Use of a cap in single-balloon enteroscopy-assisted endoscopic retrograde cholangiography. Endoscopy 2015; 47: 453-456
  • 18 Sato K, Ito S, Kitagawa T. et al. A prospective randomized study on the benefits of a new small-caliber colonoscope. Endoscopy 2012; 44: 746-753
  • 19 Mizukami T, Ogata H, Hibi T. “Passive-bending colonoscope” significantly improves cecal intubation in difficult cases. World J Gastroenterol 2012; 18: 4454-4456
  • 20 Yamauchi H, Kida M, Okuwaki K. et al. Passive-bending, short-type single-balloon enteroscope for endoscopic retrograde cholangiopancreatography in Roux-en-Y anastomosis patients. World J Gastroenterol 2015; 21: 1546-1553