Effect of Gender and Device Mouthpiece Shape on Bolus Insulin Aerosol Delivery Using the AERx Pulmonary Delivery System

  • Published:
Pharmaceutical Research Aims and scope Submit manuscript

Abstract

Purpose. A study was designed to compare differences in insulin aerosol deposition profiles in healthy male and female subjects, as well as examine the effect of mouthpiece cross-sectional shape, volume, and taper on deposition profiles using a developmental AERx pulmonary delivery system.

Methods. >Six mouthpieces were screened in the laboratory, and three were selected for clinical investigation: a cylindrical mouthpiece with constant-cross-sectional area, an elliptical mouthpiece with constant-cross-sectional area, and a tapered elliptical mouthpiece with an exit cross-sectional area equal to one half the entrance cross-sectional area.

Results. There was no significant difference in the lung dose or in the deposition pattern between males and females (p > 0.05, by ANOVA). The cross-sectional shape of the mouthpiece had no significant effect on the clinical lung dose or the deposition pattern (p > 0.05, by ANOVA), although in vitrotesting showed lower emitted dose values with the tapered elliptical mouthpiece (by ANOVA and Duncan's multiple range test, α = 0.05). Using the tapered mouthpiece in the clinic resulted in significantly more deposition on the mouthpiece itself when compared to the nontapered mouthpieces.

Conclusion. Inhalation of insulin using the AERx system was insensitive to differences in male and female respiratory tract geometry across all mouthpiece designs examined.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Boyd, B., Noymer, P., Liu, K. et al. Effect of Gender and Device Mouthpiece Shape on Bolus Insulin Aerosol Delivery Using the AERx Pulmonary Delivery System. Pharm Res 21, 1776–1782 (2004). https://doi.org/10.1023/B:PHAM.0000045228.32419.ba

Download citation

  • Issue Date:

  • DOI: https://doi.org/10.1023/B:PHAM.0000045228.32419.ba

Navigation