Elsevier

Radiation Measurements

Volume 45, Issue 1, January 2010, Pages 92-97
Radiation Measurements

Study on energy dependence of PAGAT polymer gel dosimeter evaluated using X-Ray CT

https://doi.org/10.1016/j.radmeas.2009.11.003Get rights and content

Abstract

The normoxic polymer gel dosimeter evaluated with X-Ray computed tomography has emerged as a promising tool for measuring the dose delivered during radiotherapy in three dimensions. This study presents the dependence of PAGAT normoxic polymer gel sensitivity to different photon and electron energies. PAGAT polymer gel was prepared under normal atmospheric condition and irradiated with different photon energies of 1.25 MeV from Co-60 and 6 MV and 15 MV from linear accelerator and electron energies of 6, 9, 12, 15, 18 and 21 MeV from linear accelerator. Evaluation of dosimeter was performed with an X-Ray CT scanner. Images were acquired with optimum scanning protocols to reduce the signal-to-noise ratio. The averaged image was subtracted from the unirradiated polymer gel image for background. Central axis depth dose (PDD) curves obtained for each energy and polymer gel dosimeter measurements were in good agreement with diode and film measurements. Hounsfield (HU) – dose response curve for each photon and electron energy were derived from the PDD curve obtained from the gel dosimeter measurements. From the study it was clear that the HU-dose response curve was linear in the region 1–10 Gy. The dosimeter sensitivity was defined as a slope of these linear HU-dose response curves and found that the sensitivity of polymer gel decreases with increase in both photon and electron energies. This trend in dependence of PAGAT gel dosimeter sensitivity to different photon and electron energies was not dosimetrically significant. However, to evaluate the test phantom exposed with one energy using the calibration curve derived at another energy can produce clinically significant error.

Introduction

Modern radiotherapy centers realize the highly localized conformal techniques such as Intensity Modulated Radiation Therapy (IMRT), Streotactic Radiosurgery (SRS) and Stereotactic Radiotherapy (SRT) are highly desirable to minimize the dose to nearby critical structures while maximizing the tumor control probability with high uniformity and precision. Since these techniques provide complex three dimensional conformal dose distributions, the precise measurement of the dose delivered to the patient is one of the main goals in clinical dosimetry for treatment plan validation. So the verification of treatment plans is of substantial interest and these techniques have created an urgent need for three dimensional dosimetry with high resolution and accuracy. Dosimetry has been performed conventionally with small volume ion chamber, diode detector, diamond detector and radiographic/radiochromic films etc. However, most current standard dosimetry methods do not meet modern requirements because they allow for dose measurement in one or two dimensions only.

In recent years certain gel dosimeters like Fricke gel and polymer gel are promising dosimetry tools because of their high sensitivity, superior tissue equivalence and the ability to simulate the response of real tissue to radiation exposure (Gore et al., 1984, Maryanski et al., 1993, Maryanski et al., 1994, Olsson et al., 1989, Olsson et al., 1990). Gel dosimetry offers advantages over the conventional planar and point dosimeters in that the dose distributions can be measured three dimensionally with high spatial resolution and accuracy (Ibbott et al., 1997, De Deene et al., 1998, De Deene et al., 2000). It has been proved that new PAGAT (Polyacrylamide, Gelatin and Tetrakis) polymer gel dosimeters fulfill the requirements of conformal radiotherapy accurately and are suitable for the verification of complex 3D dose distribution (Venning et al., 2005a). The PAGAT gel utilizes the principle of radiation induced polymerization and cross-linking of monomers dispersed in a water based matrix like gelatin (Maryanski et al., 1993, Maryanski et al., 1994, Maryanski et al., 1996a). It has been shown that the rate of polymerization is directly proportional to the amount of dose deposited to the polymer gel and causes the change in relaxation rate, optical properties and attenuation of X-Rays etc. So, one can calculate the dose deposition of the gel by measuring change in the relaxation time with magnetic resonance imaging (MRI) (Maryanski et al., 1993, Maryanski et al., 1994, Maryanski et al., 1996a), by measuring optical properties with optical computed tomography (CT) scanner (Gore et al., 1996, Oldham et al., 2001, Maryanski et al., 1996b), by measuring changes in dosimeter X-Ray linear attenuation coefficient with X-Ray CT scanner (Hilts et al., 2000, Trapp et al., 2001, Audet et al., 2002), by measuring ultrasonic speed of propagation and attenuation (Mather et al., 2002, Mather and Baldock, 2003) and also with vibrational spectroscopy (Baldock et al., 1998, Jirasek et al., 2001).

Recently it has been reported that PAGAT polymer gels evaluated with an X-Ray CT scanner provides a sensitive and stable dosimetry system for the verification of dose distributions in radiotherapy (Jirasek et al., 2006). The CT number or Hounsfield units (HU) obtained from the CT images related to the X-Ray linear attenuation coefficient is proportional to change in physical density of polymer gel upon radiation (Trapp et al., 2001, Trapp et al., 2002, Brindha et al., 2006).

The most important dosimetry properties of polymer gel are a linear dose response and a constant sensitivity irrespective of energy and dose rate. For clinical use, one need to evaluate the response of the polymer gel dosimeter to monomer concentration, energy of radiation, beam modality, dose rate, temperature during evaluation, time between irradiation and evaluation, and the imaging parameters like tube voltage, tube current, and exposure time. Since, these factors affect the response of the dosimeter, it is necessary to study the influence of each factor independently (Hill et al., 2005, Hilts et al., 2000, Hilts and Duzenli, 2004a, Hilts and Duzenli, 2004b).

The response of the PAG (Polyacrylamide and Gelatin) gel has been reported to be independent of energy for a wide range of electron and photon beams by several authors (Baldock et al., 1996, Maryanski et al., 1996a, Maryanski et al., 1996b). Based on these results Ibbott et al. (1997) determined the dose response curve with a 6 MV photon beam and applied it to the gel irradiated with Co-60 beam from Gamma knife unit without any corrections. But Novotny et al., 2001a, Novotny et al., 2001b has observed a trend in polymer gel sensitivity dependence on both photon and electron energy. Similarly Farajollahi et al. (1999) and Baldock et al. (1996) irradiated the polymer gel with different photon energies of 300 kV X-Rays, 660 keV, 1.25 MeV, 6 MV, 8 MV and 16 MV and with electron energies of 5, 7, 9, 12, 15, 17 and 20 MeV and they concluded that polymer gels dose response is independent of photon and electron energies. Wuu et al. (2003) has studied the dose response of BANG (Bis Acrylic acid Nitrogen Gelatin) polymer gel for 6 MV photon, 6 MeV electron and 662 keV photon beams and found to be same and linear from 0 to 30 Gy and used the same calibration curve to evaluate the dose distribution around a Re-188 brachytherapy source. Radiological transport properties studied by Pantelis et al. (2004) and Venning et al. (2005b) have reported that normoxic polymer gel dosimeters are independent of energy over the wide range of 100 keV–10 MeV. Based on these results Hurley et al. (2005) have used the calibration curve determined with 6 MV photons for the evaluation of MAGIC (Methacrylic acid, Ascorbic acid in Gelatin Initiated by Copper) gel irradiated with Ir-192 source. De Deene et al., 2001, De Deene et al., 2006 have reported that no change was observed in the dose response curve for both PAG and nPAG (normoxic PolyAcrylamid) gel dosimeters and a small increase with photon energy was observed in the nMAG (normoxic methacrylic based gel) gel dosimeter for the photon beams of 6 MV and 25 MV. Several other workers have reported a decrease in gel dosimeter response when irradiated with high linear energy transfer (LET) particles (Ramm et al., 2000, Jirasek and Duzenli, 2002, Heufelder et al., 2003) and shown that there is a energy dependence in the polymer gel dosimeter response.

The aim of this study was to evaluate the dependence of the PAGAT polymer gel dosimeter response for different photon and electron energies. The response of the polymer gel dosimeter was determined by calculating the slope of the linear portion of dose response curve for each beam energy. The dose response curve was obtained from the percentage depth dose (PDD) curve of each beam energy.

Section snippets

Gel preparation

The polymer gel dosimeter used in this study was the normoxic “PAGAT” gel described elsewhere (Venning et al., 2005a, Brindha et al., 2006) and was prepared under normal atmospheric conditions. The PAGAT polymer gel consisted of 3.5% (w/w) N, N′-methylene-bis-acrylamide (BIS), 3.5% (w/w) acrylamide, 5% (w/w) gelatin, 89% triple distilled water and 10 mM tetrakis hydroxymethyl phosphonium chloride (THP) as antioxidant. Details of elemental composition of PAGAT gel and water are summarized in

Results and discussion

For each photon and electron energy central axis depth dose (PDD) curves were obtained by measuring the Hounsfield (HU) number as a function of depth and values were normalized at a depth of maximum Hounsfield (HU) number. All gel dosimeters data were compared with diode and film measurements. From Fig. 1, Fig. 2, Fig. 3, Fig. 4, Fig. 5, Fig. 6, Fig. 7, Fig. 8, Fig. 9 it is clear that the gel dosimeter measurements were found to be good agreement with that of diode and film measurements with

Conclusion

This study shows the feasibility of preparing the PAGAT polymer gel under normal atmospheric conditions and the evaluation of dose distribution using the X-Ray CT scanner. Our results shows PAGAT polymer gel has a linear dose response up to 10 Gy with satisfactory correlation between the polymer gel data, diode and film measurements. From the study it was clear that dependence of PAGAT gel dosimeter sensitivity to different modality of the beam used was insignificant and comparable with results

References (44)

  • C. Audet et al.

    CT gel dosimetry technique: comparison of planned and measured 3D stereotactic dose volume

    J. Appl. Clin. Med. Phys.

    (2002)
  • C. Baldock et al.

    Energy response and tissue equivalence of polymer gels for radiation dosimetry by MRI

    Proc. Eur. Soc. Magn. Reson. Med. Biol.

    (1996)
  • C. Baldock et al.

    Fourier transform Raman spectroscopy of polyacrylamide gels (PAGs) for radiation dosimetry

    Phys. Med. Biol.

    (1998)
  • B.E. Bjarngard et al.

    Attenuation in high energy X-ray beams

    Med. Phys.

    (1994)
  • S. Brindha et al.

    Optimization of the imaging protocol of an X-ray CT scanner for evaluation of normoxic polymer gel dosimeters

    J. Med. Phys.

    (2006)
  • J.R. Cunningham et al.

    The dependence of mass energy absorption coefficient ratios on beam size and depth in a phantom

    Med. Phys.

    (1986)
  • Y. De Deene et al.

    Three-dimensional dosimetry using polymer gel and magnetic resonance imaging applied to the verification of conformal radiation therapy in head and- neck cancer

    Radiat. Oncol.

    (1998)
  • Y. De Deene et al.

    Validation of MR based polymer gel dosimetry as a preclinical three-dimensional verification

    Magn. Reson. Med.

    (2000)
  • Y. De Deene et al.

    On the accuracy of monomer/polymer gel dosimetry in the proximity of a high dose rate 192-Ir source

    Phys. Med. Biol.

    (2001)
  • Y. De Deene et al.

    The fundamental radiation properties of normoxic polymer gel dosimeters: a comparison between a methacrylic acid based gel and acrylamide based gels

    Phys. Med. Biol.

    (2006)
  • A.R. Farajollahi et al.

    An investigation into the use of polymer gel dosimetry in low dose rate brachytherapy

    Br. J. Radiol.

    (1999)
  • J.C. Gore et al.

    Measurement of radiation dose distribution by nuclear magnetic resonance (NMR) imaging

    Phys. Med. Biol.

    (1984)
  • J.C. Gore et al.

    Radiation dose distributions in three dimensions from tomographic optical density scanning of polymer gels: I. Development of optical scanner

    Phys. Med. Biol.

    (1996)
  • J. Heufelder et al.

    Use of BANG polymer gel for dose measurements in a 68 MeV proton beam

    Med. Phys.

    (2003)
  • B. Hill et al.

    The dose response of normoxic polymer gel dosimeters measured using X-ray CT

    Br. J. Radiol.

    (2005)
  • M. Hilts et al.

    Polymer gel dosimetry using X-ray computed tomography: a feasibility study

    Phys. Med. Biol.

    (2000)
  • M. Hilts et al.

    Image filtering for improved dose resolution in CT polymer gel dosimetry

    Med. Phys.

    (2004)
  • M. Hilts et al.

    Image noise in X-ray CT polymer gel dosimetry

    J. Phys.: Conf. Ser.

    (2004)
  • C. Hurley et al.

    A study of normoxic polymer gel dosimeter comprising methacrylic acid, gelatin and tetrakis (hydroxymethyl) phosphonium chloride (MAGAT)

    Appl. Radiat. Isot.

    (2005)
  • G.S. Ibbott et al.

    Three-dimensional visualization and measurement of conformal dose distributions using magnetic resonance imaging of BANG polymer gel dosimeters

    Int. J. Radiat. Oncol. Biol. Phys.

    (1997)
  • Absorbed Dose Determination in External Beam Radiotherapy: an International Code of Practice for Dosimetry Based on Standards of Absorbed Dose to Water

    (2000)
  • A. Jirasek et al.

    Relative effectiveness of polyacrylamide gel dosimeters applied to proton beams: Fourier transform Raman observations and track structure calculations

    Med. Phys.

    (2002)
  • Cited by (0)

    View full text