Low magnetic fields for flow propagators in permeable rocks

https://doi.org/10.1016/j.jmr.2006.08.006Get rights and content

Abstract

Pulsed field gradient NMR flow propagators for water flow in Bentheimer sandstone are measured at low fields (1H resonance 2 MHz), using both unipolar and bipolar variants of the pulsed gradient method. We compare with propagators measured at high fields (1H resonance 85 MHz). We show that (i) measured flow propagators appear to be equivalent, in this rock, and (ii) the lower signal to noise ratio at low fields is not a serious limitation. By comparing different pulse sequences, we study the effects of the internal gradients on the propagator measurement at 2 MHz, which for certain rocks may persist even at low fields.

Introduction

Increasingly, laboratory nuclear magnetic resonance (NMR) measurements are being made at ever-higher magnetic fields to enhance resolution and signal to noise ratio (S/N or SNR). In a different context, NMR at low magnetic fields (1H resonance frequency ω0/2π  2 MHz), and with low-resolution magnets, is now widely used as a borehole measurement in petroleum and other geophysical exploration [1], [2]. With the growing demand for hydrocarbons there is an immense and imminent need for developing new NMR laboratory techniques, for application to sedimentary rocks and other porous media, at these lower frequencies.

A laboratory technique receiving much current attention is the measurement of the NMR flow propagator [3], [4], [5], [6], [7], [8], [9], [10], [11], [12], [13], [14], [15] and the NMR time-of-flight technique [16], [17], [18]. This is mainly because the largest length scales that can be probed by diffusion and relaxation measurements [19], [20], [21] are ∼100 μm, which can be less than the size of the largest pores in some rocks, especially carbonates (limestones and dolomites). Furthermore the actual physics of flow in complex porous media is of interest in a wide range of processes in chemical, geological, and biological systems. Dispersion, the transport of molecules or tracers due to combined effects of diffusion and fluid flow at low Reynolds number, is an important problem both in the fundamentals of hydrodynamics [22], [23], [24] and in its application in diverse fields including biological perfusion, chemical reactors, soil remediation and oil recovery. These flow processes are controlled by the nature of the interconnections, and the topology of the pore space over length scales equivalent to many pores. Although for highly heterogenous samples, multiple length scales can be important [7], [8], [12], [24], [25], [26], [27] and not all are accessible by NMR methods, the NMR flow propagators can nevertheless probe displacements of the order of ∼5 mm. This is almost two orders of magnitude larger than those achievable by the diffusion and relaxation methods currently used, in petroleum industry practice, as probes of the rock pore space.

Hitherto, most laboratory pulsed-field-gradient (PFG-NMR) experiments measuring flow propagators and dispersion in porous media have used high (typically 85 MHz 1H resonance) magnetic fields [3], [4], [5], [6], [7], [8], [9], [11], [12], [13], [14], [15], [16], [18]. The advantage of high fields in SNR is well-known; their draw-back is a strong increase in the deleterious effects of induced internal field gradients. Contrast in magnetic susceptibility Δχ between fluid components and solids gives rise to induced internal field gradients gint which increase with static field B0(=ω0/γ). These may be reduced by reductions in B0. The reduction is at least linear; in [28] it is shown that diffusion in the larger pores imposes a maximum effective gradient gmax, which scales super-linearly as B03/2. For the first time, we compare the propagators measured at low fields (2 MHz present work) to those determined at high fields (85 MHz [14]) using adjacent core-plugs and using similar pulse sequences. We conclude that internal gradient effects on the flow propagator can be satisfactorily ameliorated at high fields if (but only if) the correct pulse sequence is used. At low fields, internal gradients are greatly reduced, and more protocol options are available in PFG experiments [29]. In particular, we show that the unipolar PFG archetypes [30], [31] are usable when internal gradients are sufficiently small. Even when they persist, as in our study rock, the unipolar protocols remain marginally acceptable, subject to a small penalty on SNR and larger systematic errors on the moments of the propagator.

Flow propagators may also be measured at low fields with fixed-field-gradients (FFG) in the fringing field of a superconducting magnet [32]. However, unlike the PFG experiments used here, only a more limited number of FFG pulse sequences can be used in fringe fields. In particular there are no analogues of the bipolar variants available in PFG experiments. Our data implies that in the case of our study rock (a sandstone), FFG experiments would be subject to quantitative errors in the propagator measurement even at low fields, whilst in many carbonates they would not.

We consider the signal to noise ratio SNR of the low field experiments in more detail. According to [33] this scales as ω07/4 for equivalent resonator geometries. A mitigating factor is the solenoid resonator geometry used in our low-field system. This yields an advantage (relative to a saddle coil) of a factor ∼3.1 [33] additional to the frequency scaling. At 2 MHz, relative to 85 MHz, we estimate a reduction in SNR of ≃245 (see Appendix C). Nevertheless, our practical results show that this is not a serious limitation, and dispel previous misconceptions about SNR limitations for flow propagator measurements at low fields.

Section snippets

Experimental

PFG-NMR propagator experiments measure the statistical distribution of fluid displacements (ζ) for a chosen mean flow velocity (v) during a chosen flow evolution time (Δ). The (tunable) length scale in PFG-NMR experiments is given by 〈ζ0 = , where 〈ζ0 is the mean displacement during time Δ along the mean flow direction (), i.e. 〈v = [0, v, 0]. v is a volume-average “interstitial” or capillary flow velocity given by v=V˙/Aϕ, where V˙ is the imposed volumetric flow rate, and A and ϕ are the

High vs. low-field propagators

In Fig. 4 we compare the flow propagators for similar (adjacent cores, cut from the same block) Bentheimer sandstone rocks under similar experimental conditions. The 2 MHz (present study) is taken with Δ = 0.45 s and v = 1.10 mm s−1, while 85 MHz [14] used Δ = 0.45 s and v = 1.03 mm s−1. We conclude that measurements at 2 MHz present no difficulty, and that the propagators are very similar at low (2 MHz) and high (85 MHz) fields. The small differences shown in Fig. 4 could be due to (a) salinity effects on the

Unipolar vs. bipolar sequences at 2 MHz

Interpretation of the flow propagator generally assumes that there are no internal field gradient effects. The presence of internal gradients distorts the propagator and quantitative accuracy is degraded. As discussed in Section 3, the bipolar PFG sequence reduces but does not eliminate these internal gradient effects. In FFG experiments however (as in fringe fields or in a bore-hole tool), there is no analogue of the bipolar sequence. In this section, we discuss distortions of the propagator

Quantitative propagators and moments analysis

The objective of this paper is to demonstrate that propagator measurements are practical at low fields, and not a detailed discussion of these propagators. However, a brief discussion on the moments analysis is made below to show that the data is of high quality. All moments have been calculated using the “low-q” moments of S(q) [14].

Propagators for various flow evolution times Δ = {0.1, 0.25, 0.5, 0.75, 1} s, as a function of displacement are presented in Fig. 3. The axes are scaled by the nominal

Conclusions

We have demonstrated the feasibility of low field NMR measurements of fluid displacement distributions (the flow propagator) in Stokes flow in porous media, in the pre-asymptotic dispersion regime. The data quality shows that the well-known loss of SNR at low magnetic fields is not in practice a serious limitation.

The observed similarity between high field (85 MHz, [14]) and low field (2 MHz, present study) flow propagators implies that the bipolar pulse sequence is sufficient in reducing

Acknowledgments

We gratefully acknowledge Uli M. Scheven for insightful comments and discussions, and Douglas D. Griffin for his technical expertise.

References (46)

  • J.D. Seymour et al.

    J. Mag Res. Ser. A

    (1996)
  • J.A. Seeley et al.

    J. Magn. Reson.

    (2004)
  • M.D. Hürlimann

    J. Magn. Reson.

    (1998)
  • G. Leu et al.

    Magn. Reson. Imaging

    (2005)
  • U.M. Scheven

    J. Magn. Reson.

    (2005)
  • D.I. Hoult et al.

    J. Magn. Reson.

    (1976)
  • R.M. Cotts et al.

    J. Magn. Reson.

    (1989)
  • E.J. Fordham et al.

    Magn. Reson. Imaging

    (1994)
  • M.D. Hürlimann et al.

    J. Magn. Reson. A

    (1994)
  • P.Z. Sun et al.

    J. Magn. Reson.

    (2003)
  • U.M. Scheven et al.

    Magn. Reson. Imaging

    (2005)
  • R.L. Kleinberg et al.

    J. Colloid Interface Sci.

    (1993)
  • M.D. Hürlimann et al.

    J. Magn. Reson.

    (2002)
  • R. Akkurt, H.J. Vinegar, P.N. Tutunjian, A.J. Guillory, Transaction of SPWLA 36th Annual Logging Symposium paper N...
  • M.D. Hürlimann, L. Venkataramanan, C. Flaum, P. Speier, C. Karmonik, R. Freedman, N. Heaton, Transaction of SPWLA 43rd...
  • K.J. Packer et al.

    Mol. Phys.

    (1996)
  • Y.E. Kutsovsky et al.

    Phys. Fluids

    (1996)
  • A. Ding et al.

    Phys. Rev. E

    (1996)
  • L. Lebon et al.

    Phys. Fluids

    (1996)
  • L. Lebon et al.

    Phys. Fluids

    (1997)
  • J.D. Seymour et al.

    AIChE J.

    (1997)
  • J.J. Tessier et al.

    Phys. Fluids

    (1998)
  • B. Manz et al.

    Phys. Fluids

    (1999)
  • Cited by (23)

    • Can sodium NMR provide more than a tracer for brine in petrophysics?

      2016, Journal of Petroleum Science and Engineering
    • Real-time oil-saturation monitoring in rock cores with low-field NMR

      2015, Journal of Magnetic Resonance
      Citation Excerpt :

      In the latter case, the use of Earth’s field NMR may circumvent the requirement for polarizing magnets [39]. In the present work, we demonstrate that a low-cost, low-field magnet (no spectroscopic resolution) can be used for rapid process monitoring, despite the challenge of poor signal-to-noise ratio (SNR) associated with such systems [40]. In our application to oil-saturation measurements in a rock plug, real-time data allow intervention by the experimentalist, such as modification of the flood protocol or generation of a particular saturation state for additional analyses (e.g., log-calibration measurements).

    • Low-field permanent magnets for industrial process and quality control

      2014, Progress in Nuclear Magnetic Resonance Spectroscopy
      Citation Excerpt :

      The only modification required to the aforementioned pulse sequences is that the gradient amplitude is ramped between ±g(max) in successive scans. Flow propagators have been used to assess pre-asymptotic Stokes’ flow in bead packs [106–117], catalysts and chemical reactors [98,118,119], biofilms [120–122], foams [123,124], and rocks [86,115–117,125–136]. However, to date the implementation of the propagator experiment on low-field hardware has been limited [86].

    View all citing articles on Scopus
    1

    Present address: Schlumberger Cambridge Research, Cambridge, CB3 0EL, UK.

    View full text