Elsevier

Physica Medica

Volume 32, Issue 9, September 2016, Pages 1072-1081
Physica Medica

Original paper
Grid patterns, spatial inter-scan variations and scanning reading repeatability in radiochromic film dosimetry

https://doi.org/10.1016/j.ejmp.2016.08.003Get rights and content

Highlights

  • Three new sources of uncertainty in radiochromic film dosimetry are presented.

  • The variance of the scanner noise follows periodical patterns in both axes.

  • Inter-scan variability depends on each autocalibration of the CCD detectors.

  • Points in the film translate to different pixel coordinates in repeated scans.

  • A novel method to correct for spatial inter-scan variations is proposed.

Purpose

When comparing different scans of the same radiochromic film, several patterns can be observed. These patterns are caused by different sources of uncertainty, which affect the repeatability of the scanner. The purpose of this work was to study these uncertainties.

Methods

The variance of the scanner noise, as a function of the pixel position, was studied for different resolutions. The inter-scan variability of the scanner response was analyzed taking into account spatial discrepancies. Finally, the distance between the position of the same point in different scans was examined.

Results

The variance of noise follows periodical patterns in both axes, causing the grid patterns. These patterns were identified for resolutions of 50, 72 and 96 dpi, but not for 150 dpi. Specially recognizable is the sinusoidal shape with a period of 8.5 mm that is produced with 72 dpi. Inter-scan variations of the response caused systematic relative dose deviations larger than 1% in 5% of the red channel images, 9% of the green and 51% of the blue. No systematic deviation larger than 1% was found after applying response corrections. The initial positioning and the speed of the scanner lamp vary between scans.

Conclusions

Three new sources of uncertainty, which influence radiochromic film dosimetry with flatbed scanners, have been identified and analyzed in this work: grid patterns, spatial inter-scan variations and scanning reading repeatability. A novel correction method is proposed, which mitigates spatial inter-scan variations caused by deviations in the autocalibration of the individual Charge Coupled Device detectors.

Introduction

The system composed of radiochromic films and a flatbed scanner is the dosimeter of choice for many applications in radiology and radiation therapy [1]. This dosimetry system is affected by several sources of uncertainty. Some of them involve only the film: for example, the thickness variations of the active layer [2], the change in film darkening as a function of post-irradiation time [3], the influence of humidity and temperature [4], [5], the UV-induced polymerization [6], etc. Some other uncertainties are a consequence of the interaction of the characteristics of both the film and scanner: for example, the lateral artifact [7], [8], Newton rings [9], the dependency with the orientation of the film on the scanner bed [10], the cross talk effect [8], the dependency on film-to-light source distance [11], [12], etc. Finally, other uncertainties are intrinsic to the scanner: for example, noise [13], [14], the inter-scan variability of the scanner response [11], warming-up of the lamp [15], [16], differences between color channels [17], [18], [19], [20], etc.

Despite all those perturbations, GAFChromic films (Ashland Inc., Wayne, NJ) have been repeatedly found to be capable of delivering accurate dose measurements [20], [21], [22], [23]. Still, to further improve the accuracy of the dosimetry system, thorough knowledge of its uncertainties is necessary.

GAFChromic EBT3 films were used in this study, in combination with the Epson Expression 10000XL scanner (Seiko Epson Corporation, Nagano, Japan). In the literature, the Epson Expression 10000XL scanner has been selected numerous times [3], [7], [8], [11], [24], [25] for radiochromic film dosimetry. In this work, the repeatability of this scanner has been examined. As a result, three new artifacts have been identified and analyzed: grid patterns, spatial inter-scan variations and scanning reading repeatability.

Section snippets

Methods and materials

GAFChromic EBT3 films from lot 06061401 were employed. They were irradiated with a Novalis Tx accelerator (Varian, Palo Alto, CA, USA). The darkening of the films was measured with an Epson Expression 10000XL scanner. The scanner was powered on 30 min before readings and five scans were taken to warm up its lamp. The films were placed on the center of the scanner with an opaque frame. To avoid the Callier effect [12], [24], a glass sheet, with a thickness of 3 mm, was placed on top of the films.

Grid pattern

Fig. 5 plots the MAD of the differences in pixel value with respect to the mean image as a function of the column, resolution and color channel for the unexposed film, white and black backgrounds. For the sake of clarity, only 100 columns are included. Nevertheless, the same patterns with the same periodicity appear in the rows and in the rest of the columns.

To discard that the patterns found in the black background were caused by scattered light, measurements were repeated covering the scanner

Grid pattern

Measurements of the scanner are affected by noise. It is well known [16] that the variance of the noise depends on the resolution of the scanner: the larger the resolution, then the larger the variance. However, this variance is not constant throughout the entire scanner bed. For the scanner and scanning software being studied, periodical patterns in both axes have been found using resolutions of 50, 72 and 96 dpi. These patterns are independent of films: they even appear in the absence of

Conclusions

For the scanner and scanning software under study, three new sources of uncertainty in radiochromic film dosimetry have been identified and analyzed: the grid pattern, spatial inter-scan variations and scanning reading repeatability.

The grid patterns appear because the variance of noise is not constant throughout the entire scanner bed: it follows periodical patterns in both axes. These patterns have been identified using resolutions of 50, 72 and 96 dpi. The mean dose uncertainty due to noise

Acknowledgements

The authors would like to thank Primož Peterlin and Juan José Rovira for their contributions to this work.

One of the authors (I.M.) is co-founder of Radiochromic.com.

References (28)

  • S. Devic

    Radiochromic film dosimetry: past, present, and future

    Physica Med

    (2011)
  • R. Dreindl et al.

    Radiochromic film dosimetry: considerations on precision and accuracy for EBT2 and EBT3 type films

    Zeitschrift für Medizinische Physik

    (2014)
  • S. Devic et al.

    Reference radiochromic film dosimetry: review of technical aspects

    Phys Med

    (2016)
  • B. Hartmann et al.

    Technical note: homogeneity of Gafchromic EBT2 film

    Med Phys

    (2010)
  • C. Andrés et al.

    A comprehensive study of the Gafchromic EBT2 radiochromic film. A comparison with EBT

    Med Phys

    (2010)
  • A. Rink et al.

    Temperature and hydration effects on absorbance spectra and radiation sensitivity of a radiochromic medium

    Med Phys

    (2008)
  • F. Girard et al.

    Reference dosimetry using radiochromic film

    J Appl Clin Med Phys

    (2012)
  • A. Niroomand-Rad et al.

    Radiochromic film dosimetry: recommendations of AAPM radiation therapy committee task group 55

    Med Phys

    (1998)
  • A.A. Schoenfeld et al.

    The artefacts of radiochromic film dosimetry with flatbed scanners and their causation by light scattering from radiation-induced polymers

    Phys Med Biol

    (2014)
  • L. van Battum et al.

    How flatbed scanners upset accurate film dosimetry

    Phys Med Biol

    (2015)
  • M.J. Butson et al.

    Evaluation of the magnitude of EBT Gafchromic film polarization effects

    Australas Phys Eng Sci Med

    (2009)
  • D. Lewis et al.

    Correcting scan-to-scan response variability for a radiochromic film-based reference dosimetry system

    Med Phys

    (2015)
  • A.L. Palmer et al.

    Evaluation and mitigation of potential errors in radiochromic film dosimetry due to film curvature at scanning

    J Appl Clin Med Phys

    (2015)
  • H. Bouchard et al.

    On the characterization and uncertainty analysis of radiochromic film dosimetry

    Med Phys

    (2009)
  • Cited by (0)

    View full text