Original paperGrid patterns, spatial inter-scan variations and scanning reading repeatability in radiochromic film dosimetry
Introduction
The system composed of radiochromic films and a flatbed scanner is the dosimeter of choice for many applications in radiology and radiation therapy [1]. This dosimetry system is affected by several sources of uncertainty. Some of them involve only the film: for example, the thickness variations of the active layer [2], the change in film darkening as a function of post-irradiation time [3], the influence of humidity and temperature [4], [5], the UV-induced polymerization [6], etc. Some other uncertainties are a consequence of the interaction of the characteristics of both the film and scanner: for example, the lateral artifact [7], [8], Newton rings [9], the dependency with the orientation of the film on the scanner bed [10], the cross talk effect [8], the dependency on film-to-light source distance [11], [12], etc. Finally, other uncertainties are intrinsic to the scanner: for example, noise [13], [14], the inter-scan variability of the scanner response [11], warming-up of the lamp [15], [16], differences between color channels [17], [18], [19], [20], etc.
Despite all those perturbations, GAFChromic films (Ashland Inc., Wayne, NJ) have been repeatedly found to be capable of delivering accurate dose measurements [20], [21], [22], [23]. Still, to further improve the accuracy of the dosimetry system, thorough knowledge of its uncertainties is necessary.
GAFChromic EBT3 films were used in this study, in combination with the Epson Expression 10000XL scanner (Seiko Epson Corporation, Nagano, Japan). In the literature, the Epson Expression 10000XL scanner has been selected numerous times [3], [7], [8], [11], [24], [25] for radiochromic film dosimetry. In this work, the repeatability of this scanner has been examined. As a result, three new artifacts have been identified and analyzed: grid patterns, spatial inter-scan variations and scanning reading repeatability.
Section snippets
Methods and materials
GAFChromic EBT3 films from lot 06061401 were employed. They were irradiated with a Novalis Tx accelerator (Varian, Palo Alto, CA, USA). The darkening of the films was measured with an Epson Expression 10000XL scanner. The scanner was powered on 30 min before readings and five scans were taken to warm up its lamp. The films were placed on the center of the scanner with an opaque frame. To avoid the Callier effect [12], [24], a glass sheet, with a thickness of 3 mm, was placed on top of the films.
Grid pattern
Fig. 5 plots the MAD of the differences in pixel value with respect to the mean image as a function of the column, resolution and color channel for the unexposed film, white and black backgrounds. For the sake of clarity, only 100 columns are included. Nevertheless, the same patterns with the same periodicity appear in the rows and in the rest of the columns.
To discard that the patterns found in the black background were caused by scattered light, measurements were repeated covering the scanner
Grid pattern
Measurements of the scanner are affected by noise. It is well known [16] that the variance of the noise depends on the resolution of the scanner: the larger the resolution, then the larger the variance. However, this variance is not constant throughout the entire scanner bed. For the scanner and scanning software being studied, periodical patterns in both axes have been found using resolutions of 50, 72 and 96 dpi. These patterns are independent of films: they even appear in the absence of
Conclusions
For the scanner and scanning software under study, three new sources of uncertainty in radiochromic film dosimetry have been identified and analyzed: the grid pattern, spatial inter-scan variations and scanning reading repeatability.
The grid patterns appear because the variance of noise is not constant throughout the entire scanner bed: it follows periodical patterns in both axes. These patterns have been identified using resolutions of 50, 72 and 96 dpi. The mean dose uncertainty due to noise
Acknowledgements
The authors would like to thank Primož Peterlin and Juan José Rovira for their contributions to this work.
One of the authors (I.M.) is co-founder of Radiochromic.com.
References (28)
Radiochromic film dosimetry: past, present, and future
Physica Med
(2011)- et al.
Radiochromic film dosimetry: considerations on precision and accuracy for EBT2 and EBT3 type films
Zeitschrift für Medizinische Physik
(2014) - et al.
Reference radiochromic film dosimetry: review of technical aspects
Phys Med
(2016) - et al.
Technical note: homogeneity of Gafchromic EBT2 film
Med Phys
(2010) - et al.
A comprehensive study of the Gafchromic EBT2 radiochromic film. A comparison with EBT
Med Phys
(2010) - et al.
Temperature and hydration effects on absorbance spectra and radiation sensitivity of a radiochromic medium
Med Phys
(2008) - et al.
Reference dosimetry using radiochromic film
J Appl Clin Med Phys
(2012) - et al.
Radiochromic film dosimetry: recommendations of AAPM radiation therapy committee task group 55
Med Phys
(1998) - et al.
The artefacts of radiochromic film dosimetry with flatbed scanners and their causation by light scattering from radiation-induced polymers
Phys Med Biol
(2014) - et al.
How flatbed scanners upset accurate film dosimetry
Phys Med Biol
(2015)