Spatial-resolved electrochemiluminescence ratiometry based on bipolar electrode for bioanalysis
Introduction
Electrochemiluminescence (ECL) is a luminescence excitation process triggered electrochemically (Richter, 2004; Shan et al., 2009). Combining the advantages of zero optical background and easy reaction control by applying electrode potential, ECL has been proved to be a highly sensitive method in the areas of biological (Dai et al., 2015, Wu et al., 2015), environmental (McCall et al., 1999; McCall and Richter, 2000) and food analysis (Cheng and Stanker, 2013). Through monitoring the change in ECL intensity, species that directly take part in ECL reaction (Lei et al., 2015, Wang et al., 2013) or indirectly influence the reaction (Zhang et al., 2015b) could be quantified. However, other factors such as environmental conditions can interfere with the signal output, especially during trace analysis, which may cause false positive or negative errors.
Ratiometric detection is an ideal strategy to limit the interference factors via normalizing environmental variation by self-calibration, which has been widely developed in fluorescence (Wu et al., 2016). In the past 5 years, some works based on the methodology of ratiometric ECL have been reported in biological and environmental analysis (Wu et al., 2016, Zhang et al., 2014). The ratiometric ECL should include both dual-potential and dual-wavelength signal ratiometric assays. Since the first dual-potential ECL ratiometry for DNA sensing was reported by Xu and co-workers in 2013 (Zhang et al., 2013), a series of researches have been published for the detections of cancer cell, miRNA and Mg2+ using similar strategy (Cheng et al., 2014, Hao et al., 2014, Wang et al., 2016). Recently, applying graphite-like carbon nitride nanosheet (g-C3N4 NS), an effective, strong and stable ECL emitter, Xu et al. reported the first dual-wavelength ratiometric ECL approach for the highly sensitive detection of miRNA (Feng et al., 2016). Despite the advantage of the ratiometric ECL, it usually makes the detection system more complicated, since two ECL substances with special characters and typical co-reactants should exist in the same solution. How to choose the emitters and balance their ECL excitation potentials, intensities, wavelengths and co-reactants remains a challenge. In addition, optical filters or spectrometer have to be applied for dual-wavelength ratiometry, which decreased the ECL intensity to a great extent (Doeven et al., 2012). Therefore, pursuing a simple strategy with self-calibration function but easy to handle is valuable for further study.
Aiming at this issue, we developed a closed bipolar electrode (BPE)-ECL device in the present work for ratiometric detection of prostate specific antigen (PSA), which belongs to the human kallikrein family and has been recognized as the premier tumor marker for the detection of early stage prostate cancer and to monitor the recurrence of the disease after treatment. BPE is an electrically conductive material that promotes electrochemical reactions at its extremities (Chow et al., 2008, Wu et al., 2012, Wu et al., 2011). The electroneutrality across the BPE makes the reactions on each pole of BPE electrically coupled (Fosdick et al., 2013, Shi et al., 2014, Wu et al., 2013, Zhang et al., 2016). Thus one electrochemical sensing reaction at one pole of BPE could be quantified at both ends. The closed BPE is an excellent choice for ratiometric assay since the solutions contacting the BPE anode and cathode are physically separated from one another. Therefore, two ECL emitter/co-reactant systems could be separated in space which limits mutual interference. The sensing principle is shown in Scheme 1. The cathode modified with Au@g-C3N4 as the ECL substance served as sensing pole. The anode of the BPE served for calibration via another ECL substance, Ru(bpy)32+ adding in the reservoir. As a sensing probe, Pt-PAMAM-DNAzyme was prepared and assembled at the cathode through the combination of probe DNA and PSA aptamer. The composite acted as the quencher of cathode ECL emission, but the promoter of anode ECL ascribed to the electric equilibrium of the BPE. In the presence of target PSA, its recognition by aptamer led to the release of Pt-PAMAM-DNAzyme from the cathode and caused partial recover of the ECL emission. Accordingly, the anode ECL emission decreased because of the decreased faradiac current flowing through BPE. Calculating the ratio of ECL cathode to ECL anode resulted in the quantitative analysis of PSA. Compared with conventional dual-potential and dual-wavelength ECL ratiometries, the spatial-resolved ECL ratiometry based on closed BPE is more simple and sensitive since the two ECL reactions could be spatially separated and no optical filters were required for the data acquisition.
Section snippets
Materials and reagents
Indium-tinoxide (ITO)-coated (thickness: 100 nm; resistance: 10 Ω/square) aluminosilicate glass slides were obtained from China Southern Glass (Shenzhen, China). Sylgard 184 (including poly (dimethylsiloxane) (PDMS) monomer and curing agent) was from Dow Corning (Midland, MI). Melamine was obtained from Fuchen Chemical Reagent Co. (Tianjin, China). H2SO4 was supplied by Shanghai Chemical Reagent Co. Ltd., China. HAuCl4, H2PtCl6, NaBH4, trisodium citrate, PAMAM dendrimer (ethylenediamine core,
Characterization of Pt and Pt-PAMAM Nanoparticles
As shown in Fig. 1A, the Pt NPs were well dispersed with the average diameter of about 5 nm. Partially quarternized PAMAM dendrimers can be used as a polymeric template reservoir to prepare stable platinum-dendrimer nanocomposites due to its quaternary ammonium groups and primary amines. It was found that the resulted nanocomposites remained stable for several months. As shown in Fig. 1B, TEM image showed that well-defined nanosized platinum nanoparticles were almost monodispersed. The spectrum
Conclusions
In conclusion, we reported a spatial-resolved ratiometric BPE-ECL biosensor for the detection of PSA. Two ECL emission substrates, g-C3N4 and Ru(bpy)32+ were introduced for sensing and self-calibration in separated reservoirs. Common concern for the mutual interference between the ECL reactions in ratiometric system was limited and no optical filter was applied in the detection, which increased the sensitivity as 3 times higher. The integration of Pt-PAMAM and G-4 DNAzyme at BPE cathode showed
Acknowledgments
We gratefully acknowledge the Ministry of Science and Technology Program of China (2016YFA0201200), the National Natural Science Foundation of China (Grants 21327902, 21475058) and the Priority Academic Program Development of Jiangsu Higher Education Institutions.
References (30)
- et al.
Biosens. Bioelectron.
(2014) - et al.
Biosens. Bioelectron.
(2013) - et al.
Biosens. Bioelectron.
(2016) - et al.
Biosens. Bioelectron.
(2015) - et al.
J. Phys. Chem. Lett.
(2012) - et al.
Nanoscale
(2013) - et al.
Anal. Chem.
(2014) - et al.
J. Agric. Food Chem.
(2013) - et al.
Anal. Chem.
(2014) - et al.
J. Am. Chem. Soc.
(2008)
Anal. Chem.
Angew. Chem. Int. Ed.
Anal. Chem.
Angew. Chem. Int. Ed.
Chem. Commun.
Cited by (57)
A simple and sensitive electrochemiluminescence spectrum measurement platform and spectrum-resolved ratiometric sensor for miroRNA-141 determination
2022, Electrochimica ActaCitation Excerpt :Reproducibility is one of the important parameters in an analytical method. Ratiometric strategies are widely used to improve the reproducibility thanking to their self-correction ability [47–52]. Ratiometric ECL sensors include spatial-resolved [50,51], potential-resolved [52–56], and spectrum-resolved [17–21] approaches.
Shared-cathode closed bipolar electrochemiluminescence cloth-based chip for multiplex detection
2022, Analytica Chimica ActaProximity binding induced nucleic acid cascade amplification strategy for ultrasensitive homogeneous detection of PSA
2021, Analytica Chimica ActaCitation Excerpt :Therefore, the development of novel ultrasensitive and accurate methods for the early detection of PSA is crucial. Up to dates, various techniques have been well developed for PSA detection, such as surface-enhanced Raman scattering [2,3], surface plasmon resonance [4], chemiluminescence [5,6], fluorescence immunoassay [7–9], electrochemiluminescence [10,11], electrochemical [12,13] and photoelectrochemical immunoassays [14,15]. Among which, the fluorescence homogeneous immunoassay has been widely applied for protein analysis due to its easy operation.
Bipolar electrochemiluminescence sensors: From signal amplification strategies to sensing formats
2021, Coordination Chemistry Reviews