Skip to main content
Log in

A digital esophageal stethoscope system for monitoring anesthetized patients’ cardiovascular system

  • Original Article
  • Published:
Biomedical Engineering Letters Aims and scope Submit manuscript

Abstract

Purpose

Esophageal stethoscope is a device to measure heart sound or respiratory sound by inserting a catheter into the patients’ esophagus during operation. It is easy to use, noninvasive, and not influenced by ambient noise, but does not provide the quantitative information about cardiovascular system. Thus, in this study, we tried to develop a digital esophageal stethoscope system (DESS) that can be used to monitor the heart sounds of anesthetized patients during operation.

Methods

It consists of the wireless sensor module and the base station. The wireless sensor module acquires and amplifies heart sound data with microphone connected to esophageal catheter and transmits the data to the base station through Bluetooth. The base station is composed of a receiver and a PC. The receiver processes and forwards the heart sound data to a PC, where MatLab based software (Cardiac Sound Analyzer) records and displays the heart sound as phonocardiogram and spectrogram in time and frequency domain in real time. Besides, it can quantify and objectify heart sound information by calculating its intervals, heart rates, and intensity values in dB.

Results

We performed a pilot experiment with patients under general anesthetic surgery and evaluated the performance of the developed system.

Conclusions

This system is so simple and useful for anesthesiologist to evaluate the included information on cardiac condition without being disturbed by ambient noise and is helpful for comparing a patient’s states before and after surgery.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Reference

  1. The Korean society of anesthesiologists. Anesthesiology & Pain Medicine. Seoul: Ryo Moon Gak.P.Co; 2010.

    Google Scholar 

  2. The Association of anaesthetists of Great Britain and Ireland. Recommendatons for standards of monitoring during aneaesthesia and recovery. 4th ed. London: Association of Anaesthetists of Great Britain and Northern Ireland; 1994.

    Google Scholar 

  3. Ara G. Tilkian. Understanding heart sounds and murmurs with an introduction to lung sounds. 4th ed. Philadelphia: W.B. Saunders; 2004.

    Google Scholar 

  4. Luisa A, Singhal A, Knighten V. New index of cardiac contractility during stress testing with treadmill. Acta Cardiol. 1986; 41(1):31–39.

    Google Scholar 

  5. Stein P, Sabbah H, Khaha F, Anbe D. Exploration of the cause of the low intensity aortic component of the second heart sound in nonhypotensive patients with poor ventricular performance. Circulation. 1978; 57(3):590–593.

    Article  Google Scholar 

  6. Chen D, Pibarot P, Honos G, Durano L. Estimation of pulmonary artery pressure by spectral analysis of the second heart sound. Am J Cardiol. 1996; 78(7): 785–789.

    Article  Google Scholar 

  7. The Korean society of anesthesiologists. Anesthesiology. Seoul: Yeomungak; 1996.

    Google Scholar 

  8. Smith C. An endo-esophageal stethoscope. Anesthesiology. 1951; 15(5):556.

    Article  Google Scholar 

  9. Murray L Rice, D John Doyle. Comparison of Phonocardiographic Monitoring Locations. Conf Proc IEEE Eng Med Biol Soc. 1995; 1:685–686.

    Google Scholar 

  10. Ahlstrom C. Nonlinear phonocardiographic signal processing. Dissertation. Linköping Univ. Sweden. 2008.

  11. Topal T, Polat H, Guler I. Software Development for the analysis of heartbeat sounds with labVIEW in diagnosis of cardiovascular disease. J Med Cyst. 2008; 32(5):409–421.

    Article  Google Scholar 

  12. Prielipp R, Kelly J, Roy R. Use of esophageal or precordial stethoscopes by anesthesia providers; are we listening of our patients?. J Clin Anesth. 1995; 7(5):367–372.

    Article  Google Scholar 

  13. Watson A, Visram A. Survey of the use of oesophageal and precordial stethoscope in current paediatric anaesthetic practice. Paediatr Anaesth. 2001; 11(4):437–442.

    Article  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Tae Soo Lee.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Shin, J.Y., Kim, Y.C., Lim, S.W. et al. A digital esophageal stethoscope system for monitoring anesthetized patients’ cardiovascular system. Biomed. Eng. Lett. 2, 265–270 (2012). https://doi.org/10.1007/s13534-012-0080-9

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s13534-012-0080-9

Keywords

Navigation