Abstract
Purpose
Esophageal stethoscope is a device to measure heart sound or respiratory sound by inserting a catheter into the patients’ esophagus during operation. It is easy to use, noninvasive, and not influenced by ambient noise, but does not provide the quantitative information about cardiovascular system. Thus, in this study, we tried to develop a digital esophageal stethoscope system (DESS) that can be used to monitor the heart sounds of anesthetized patients during operation.
Methods
It consists of the wireless sensor module and the base station. The wireless sensor module acquires and amplifies heart sound data with microphone connected to esophageal catheter and transmits the data to the base station through Bluetooth. The base station is composed of a receiver and a PC. The receiver processes and forwards the heart sound data to a PC, where MatLab based software (Cardiac Sound Analyzer) records and displays the heart sound as phonocardiogram and spectrogram in time and frequency domain in real time. Besides, it can quantify and objectify heart sound information by calculating its intervals, heart rates, and intensity values in dB.
Results
We performed a pilot experiment with patients under general anesthetic surgery and evaluated the performance of the developed system.
Conclusions
This system is so simple and useful for anesthesiologist to evaluate the included information on cardiac condition without being disturbed by ambient noise and is helpful for comparing a patient’s states before and after surgery.
Similar content being viewed by others
Reference
The Korean society of anesthesiologists. Anesthesiology & Pain Medicine. Seoul: Ryo Moon Gak.P.Co; 2010.
The Association of anaesthetists of Great Britain and Ireland. Recommendatons for standards of monitoring during aneaesthesia and recovery. 4th ed. London: Association of Anaesthetists of Great Britain and Northern Ireland; 1994.
Ara G. Tilkian. Understanding heart sounds and murmurs with an introduction to lung sounds. 4th ed. Philadelphia: W.B. Saunders; 2004.
Luisa A, Singhal A, Knighten V. New index of cardiac contractility during stress testing with treadmill. Acta Cardiol. 1986; 41(1):31–39.
Stein P, Sabbah H, Khaha F, Anbe D. Exploration of the cause of the low intensity aortic component of the second heart sound in nonhypotensive patients with poor ventricular performance. Circulation. 1978; 57(3):590–593.
Chen D, Pibarot P, Honos G, Durano L. Estimation of pulmonary artery pressure by spectral analysis of the second heart sound. Am J Cardiol. 1996; 78(7): 785–789.
The Korean society of anesthesiologists. Anesthesiology. Seoul: Yeomungak; 1996.
Smith C. An endo-esophageal stethoscope. Anesthesiology. 1951; 15(5):556.
Murray L Rice, D John Doyle. Comparison of Phonocardiographic Monitoring Locations. Conf Proc IEEE Eng Med Biol Soc. 1995; 1:685–686.
Ahlstrom C. Nonlinear phonocardiographic signal processing. Dissertation. Linköping Univ. Sweden. 2008.
Topal T, Polat H, Guler I. Software Development for the analysis of heartbeat sounds with labVIEW in diagnosis of cardiovascular disease. J Med Cyst. 2008; 32(5):409–421.
Prielipp R, Kelly J, Roy R. Use of esophageal or precordial stethoscopes by anesthesia providers; are we listening of our patients?. J Clin Anesth. 1995; 7(5):367–372.
Watson A, Visram A. Survey of the use of oesophageal and precordial stethoscope in current paediatric anaesthetic practice. Paediatr Anaesth. 2001; 11(4):437–442.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Shin, J.Y., Kim, Y.C., Lim, S.W. et al. A digital esophageal stethoscope system for monitoring anesthetized patients’ cardiovascular system. Biomed. Eng. Lett. 2, 265–270 (2012). https://doi.org/10.1007/s13534-012-0080-9
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s13534-012-0080-9