Introduction

Correlative microscopy, i.e., the combined utilization of a range of microscopy techniques on a single specimen, is increasingly deployed to understand fundamental aspects in material science. The combination of transmission electron microscopy (TEM) and atom probe tomography (APT) enables researchers to relate the atomic structure and composition of nanoscale features of interest and has been gaining influence over the past decades. Typically, correlative TEM/APT analyses have been carried out from distinct specimens from the same alloy,1,2,3,4,5,6,7 which fails in cases where the microstructure is inhomogeneous and/or when targeting rare microstructural features for both TEM and APT investigations. As recently reviewed by Herbig,8 full electron microscopy characterization of specific features can be carried out on needle-shaped specimens prior to APT analysis by utilizing specially designed holders. This approach has, for example, led to a better understanding of diffusional mechanisms resulting in local solute segregation at crystalline imperfections9,10,11,12,13,14 that critically impact the macroscopic material behavior. APT had previously revealed details of the composition of structural imperfections,15 the presence of which were confirmed by field ion microscopy16 or TEM.11,17,18 Here, we introduce methodologies to aid target features of interest for correlative TEM/APT investigation by exploiting advanced scanning electron microscopy (SEM) techniques in deformed Ni- and CoNi-based superalloys, one of the most important classes of engineering materials for temperatures above 1000°C. Their high temperature stability is attributed to the uniform distribution of L12-ordered γ′ precipitates coherently embedded in an fcc solid solution γ matrix. Three protocols for site-specific correlative investigation are presented.

Materials and Methods

Materials

The polycrystalline Ni-based superalloy IN792, commonly used in land-based gas turbines, was studied. Its chemical composition is Ni–13.9Cr–8.8Co–1.1Mo–1.3W–7.6Al–4.9Ti–1.3Ta–0.4C–0.1B–0.012Zr (at.%). It was provided as 20-mm-diameter bars by Howmet. A hot isostatic press (HIP) process at 1195°C and 150 MPa for 2 h was followed by a solution heat treatment stage at 1121°C for 2 h and a final stage of aging at 850°C for 24 h. After heat treatment, specimens were ground and polished with abrasive media to a 1-μm finish and were isothermally exposed at 750°C for 50 h.

A single-crystal CoNi-based superalloy with the composition Co–32Ni–8Al–5W–6Cr–2.5Ti–1.5Ta–0.1Hf–0.4Si (at.%) was prepared by the Bridgman process. The heat treatment steps were 1280°C/8 h + 1050°C/5 h + 900°C/16 h to obtain a uniform γ/γ′ microstructure. Tensile specimens were cut and crept at 850°C with an applied stress of 400 MPa along the [001]-direction up to 4.6% (380 h) and 0.3% (40 h) plastic strain. Cross sections close to the {100} and {110} habit planes were cut and mechanically polished for further microstructural characterization.

Controlled Electron Channeling Contrast Imaging (cECCI)

A Zeiss Merlin scanning electron microscope (Carl Zeiss SMT AG, Germany) with a Gemini-type field emission gun electron column and a Bruker e-Flash HR EBSD detector (Bruker Corp., USA) was used. ECCI (electron channeling contrast imaging) under controlled diffraction conditions was employed to rotate and tilt the crystal into two-beam diffraction conditions using the TOCA (Tools for Orientation Determination and Crystallographic Analysis computer program; for details, see Zaefferer et al.19). The microscope was operated at 30 kV accelerating voltage and 3–4 nA probe current.

Focused Ion Beam (FIB) Preparation

Atom probe specimens were fabricated using a dual-beam SEM/focused-ion-beam (FIB) instrument (FEI Helios Nanolab 600 and 660 equipped with a STEM detector). The specimens were sharpened by FIB milling followed by a final cleaning procedure at 2 kV and 16-pA current to remove severely damaged regions induced by the previously employed high-energy (30 kV) Ga ion beam.

Transmission Electron Microscopy (TEM)

Transmission electron microscopy (TEM) on the APT specimen was carried out using a Phillips CM-20 operating at 200 kV. High-resolution imaging was carried out in STEM mode using an aberration corrected Titan FEI microscope operated at 300 kV.

Atom Probe Tomography (APT)

APT measurements were conducted using instruments from Cameca Instruments Inc.: a LEAP 5000XR, LEAP 5000XS and LEAP 4000 HR. The laser pulsing mode was applied at a pulse repetition rate between 100 and 250 kHz and a pulse energy between 30 and 40 pJ. The specimen’s base temperature was kept at 40 K, and the target detection rate was set to be 5 ions detected every 1000 pulses on average. Data analysis was performed using the IVAS 3.6.14 software package.

Results and Discussion

Method (1) cECCI-Guided Conventional Target Preparation

First, ECC imaging is used to locate a region of interest (ROI) with high dislocation density in the IN792 alloy, generated during the volume expansion of oxidized carbide in the γ/γ′ microstructure as shown in Fig. 1a.20,21 A series of APT specimens were prepared from locations highlighted in blue following the conventional procedure outlined by Thompson et al.22 and using a commercially available flat-top Si coupon and conventional clip holder. A representative tomographic reconstruction is shown in Fig. 1b: two γ′ precipitates and their interfaces with the γ matrix are imaged. Within γ′, a Cr isoconcentration surface (3 at.%) reveals γ particles that form during the 750°C heat treatment,23,24 while the tubular features correspond to solute segregation to dislocations. A linear compositional profile plotted from a cylindrical ROI perpendicular to the dislocation, along the arrow labeled #1, plotted in Fig. 1c, reveals clear segregation of Cr and Co to the dislocations, which contrasts with a depletion in Ni, Al and Ti.

Fig. 1
figure 1

Method (1): (a) cECCI micrograph from the cross section of an oxidized intragranular MC carbide in the IN792 alloy after isothermal exposure at 750°C for 50 h. Part of the oxidized MC carbide is denoted by the red dashed line, and the recrystallized zone is denoted by the green dashed line. (b) Atom probe reconstruction from a region with high dislocation density next to an oxidized MC carbide showing γ/γ′ interfaces, dislocations, and γ particles within the γ′ precipitate. (c) 1D concentration profile perpendicular to the denoted dislocation in (b) (arrow #1) (Color figure online)

Method (2) cECCI-Guided In-plane Target Preparation for TEM and APT

Here again, ECC imaging is performed to reveal and locate lattice defects within the crept CoNi-based superalloy as summarized in Fig. 2a–b. Figure 2a is obtained by controlled rotation/tilting of the sample such that its surface normal is close to the [110] direction of the fcc-crystalline lattice and with the \( g = \left( {00\bar{2}} \right) \) two-beam condition, which shows a clear atomic density contrast between the dark fcc-γ matrix and the bright L12-γ′ precipitates rafted along the creep-loading direction. High-number density of stacking faults (SFs) appears as bright features with oscillating intensity. These are visible because of shearing of γ′, which is the primary active deformation mechanism operating during high temperature creep in this type of alloy.25 From the stereographic projection in Fig. 2b, aligned according to the sample orientation, four variants of SFs are observed, marked as A, B, C and D according to their traces on the {111} planes. SF variants A and D are associated with the \( \left( {11\bar{1}} \right) \) and \( \left( {111} \right) \) planes forming angles of 38° and 36° with the sample surface and exhibit fading oscillating contrast. This fading contrast arises because of the loss of backscattered signal from the inclined stacking faults with increasing depth below the sample surface. The B and C variants appear near edge-on belonging to the \( \left( {\bar{1}11} \right) \) and \( \left( {\bar{1}1\bar{1}} \right) \) planes forming angles of 80° and 81° with the sample surface and thus appear as sharp bright lines without fading intensity of the oscillating contrast.

Fig. 2
figure 2

Method (2): (a) cECC imaging of the creep-deformed CoNi-based superalloy with the sample surface normal close to the [110] zone axis revealing numerous SFs and (b) corresponding stereographic projection. (c) Locating the defect feature in the bulk sample by ECC imaging. (d, e) Marking the feature with Pt deposition (white color) in FIB/SEM. (f–k) Steps involving the in-plane lift-out procedure to prepare the defect of interest into the apex of an APT tip to conduct correlative microscopy

Once an ROI, highlighted in yellow, is identified in the bulk sample as shown in Fig. 2c and located in the FIB as indicated by the black box in Fig. 2d, it is marked and protected by in situ Pt deposition; see Fig. 2d–e. Figure 2f–k shows the step-by-step in-plane lift-out 26,27 procedure schematically, where cut parts of the lift-out are positioned/attached on the electro-polished tips of a halved TEM Mo grid, which is held in a dedicated holder8,28 and subsequently shaped into conical APT specimens by annular milling.29

Figure 3a shows a bright-field TEM image of the APT specimen containing a C-type SF close to the [110] zone axis and in a \( g = \left( {00\bar{2}} \right) \) two-beam condition. A straight line appears with a contrast resulting from the intersection of the staking fault with the surface of the APT specimen. Figure 3b shows a HAADF-STEM image of the APT tip close to the [110] zone axis. The bright contrast, with respect to the surrounding lattice, along the SF indicates segregation of heavy elements.30,31 Figure 3c shows a high-resolution STEM image revealing the D019 ordered structure of the fault plane surrounded by the L12 γ′ precipitate.31,32 A missing A-type plane as visible in Fig. 3d reveals that this is an intrinsic SF. Figure 3e–g shows the corresponding APT reconstructions, which has the overall composition of γ′. Figure 3e–f depicts the distribution of Al atoms (brown) and W atoms (red), respectively. Towards the bottom of the tomogram, a planar depletion of Al contrasting with an enrichment in W becomes apparent. Figure 3g displays the Al compositional map plotted and projected along the y direction onto the xz plane of a cuboidal ROI, which highlights the local Al deficiency at the SF. Figure 3h shows the concentration profiles across the fault plane indicating Co/W enrichment and Ni/Al depletion. Similar compositional segregation effects in the stacking faults have been reported even in crept Co-based superalloys.32

Fig. 3
figure 3

Method (2): (a) Bright field (BF) TEM image of a γ′ precipitate taken near to the [110] zone axis. An SF is present, appearing as a dark contrast line. (b) HAADF STEM image showing bright Z contrast along the fault plane, indicating segregation of higher Z solutes. (c, d) High-resolution STEM image of the atomic structure of the SF viewed edge-on along the [110] direction. (e–h) APT reconstructed tomogram showing Al deficiency and W enrichment along the fault plane

Method (3): TSEM-Guided Target Preparation from TEM Foil for APT

A third method offers the possibility to prepare APT specimens directly from pre-thinned TEM foils. Method (3) allows a detailed TEM characterization on a larger area and the selection and target preparation of a sparsely distributed defect type from this lamella into the APT specimen tip. Figure 4 shows the schematic step-wise preparation procedure for the case of an SF ribbon in the 0.3% crept CoNi-based superalloy.

Fig. 4
figure 4

Method (3): (a) Marking of the ROI with Pt deposition or C contamination in a TEM foil. (b–d) Successive steps of the specimen preparation from a TEM foil (pink color); (e) corresponding series of electron micrographs obtained in the SEM and TEM for the final tip to carry out APT analysis (Color figure online)

Here, the ROI does not show any variation in Z-contrast, and the SFs are not detectable with SE (secondary electron) or BSE (back-scattered electron) detectors. The ROI is identified in the STEM mode in the SEM and marked with Pt or C X-shaped patterns, as shown in Fig. 4a.

In cases where TSEM imaging is not available, a convergent beam focused on the sample’s surface can be used to mark the ROI by carbon contamination in the TEM. A 3 µm × 6 µm lamella is cut out at 15 kV and attached to the micro-manipulator using Pt deposition; see Fig. 4b. Figure 4c shows the lamella transferred and welded on both sides to a pre-sharpened Mo or W grid. The lamella must be parallel to the supporting needle, and the dimensions of the stump should not exceed 1–2 µm. Finally, as depicted in Fig. 4d, annular milling is performed, with a last milling step at 2 kV with an inner pattern diameter of approximately 50 nm. The X-shaped fiducials enable estimation of the milling depth and position of the ROI. Figure 4e exemplifies the preparation procedure from defect identification and marking to a final APT specimen containing an SF parallel to the specimen’s main axis. The preparation is carried out continuously without transfer or breaking of the vacuum and is hence suitable for materials prone to oxidation. In this approach, the ROI is not milled out of bulk material but merely cut out of an already thinned area. This allows for short times under ion radiation when using 15–30 kV for cutting. Beam-sensitive materials can also be processed with voltages down to 5 kV in reasonable times. Shorter radiation times and/or lower acceleration voltages both result in lower overall ion beam damage in the ROI. In addition, the presented method only requires a small sample volume from the TEM foil and thus the foil can be reused for further TEM structural characterization.

This method is suitable for foil thicknesses in the range of 50–250 nm. Foils thinner than typical APT specimen dimensions will result in asymmetric tips and an enhanced risk for severe ion beam damage. TSEM and/or TEM investigations beforehand allow the identification of features of interest in thicker regions of the specimen and an assessment of their representativeness. Advantages of low-energy TSEM have been discussed in detail by Callahan et al.33 In the case of the example in Fig. 4, the fault ribbon was traced from the hole in the TEM foil back to a thicker region to avoid the aforementioned issues. No specimens with a thickness < 80 nm were prepared but we would recommend ion beam energies below 5 kV and “blind” cutting without the acquisition of ion images during the process. Please note that APT specimens from thin TEM foils will exhibit smaller dimensions, and the potential for ion damage in the specimen is increased.

Summary and Conclusion

Figure 5 summarizes the three experimental protocols introduced herein to characterize atomic-scale segregation at dislocations in a crept polycrystalline Ni-based superalloy and at SFs in a crept CoNi-based single-crystal alloy. These nanoscale defects critically impact the mechanical performance of high-temperature engineering alloys. These three protocols are also applicable to any other material or defect features. The strength of these approaches is to enable both structural and compositional information on the same object, which is necessary to unveil the mechanisms controlling the mechanical response of the material under investigation.

Fig. 5
figure 5

Sequence of experimental steps in the three methods adopted for the correlative microscopy