Predictive value of rapid-rate non-sustained ventricular tachycardia in the occurrence of appropriate implantable cardioverter-defibrillator therapy

  • Published:
Journal of Interventional Cardiac Electrophysiology Aims and scope Submit manuscript

Abstract

Purpose

Rapid-rate non-sustained ventricular tachycardia (RR-NSVT) that meets detection criteria but terminates itself before the delivery of implantable cardioverter-defibrillator (ICD) therapy is not rare in routine ICD interrogation. Whether sustained ventricular tachycardia/fibrillation will occur in a short time after RR-NSVT has not been fully elucidated.

Methods

Clinical features and follow-up data of 828 ICD patients with home monitoring were retrospectively collected. RR-NSVT characteristics and time interval between the first episode of RR-NSVT and subsequent appropriate ICD therapy were analyzed.

Results

During a mean follow-up of 44.75 ± 20.87 months, 335 episodes of RR-NSVT were documented in 145 patients. A total of 119 patients had both RR-NSVT and appropriate ICD therapy. In multivariate COX regression models, RR-NSVT was an independent predictor of appropriate ICD therapy (HR 7.599, 95%CI 5.926–9.745, P < 0.001), appropriate shock (HR 6.222, 95%CI 4.667–8.294, P < 0.001), and all-cause mortality (HR 2.156, 95%CI 1.499–3.099, P < 0.001). Appropriate ICD therapy was administered after the first RR-NSVT episode in 101 patients, with a median interval of 21 days. Compared to RR-NSVT with appropriate ICD therapy occurring beyond 21 days, RR-NSVT within 21 days prior to appropriate ICD therapy had a longer median duration time (14 s vs. 12 s, P = 0.013), but without significant difference in mean RR interval at initial detection and mean RR interval after episode termination.

Conclusions

Rapid-rate non-sustained VT was an independent predictor of appropriate ICD therapy and all-cause mortality. The presence of RR-NSVT should be considered a possible herald of more serious cardiac events in ICD patients.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1

Similar content being viewed by others

References

  1. Myerburg RJ, Junttila MJ. Sudden cardiac death caused by coronary heart disease. Circulation. 2012;125:1043–52.

    Article  Google Scholar 

  2. Singh SN, Fisher SG, Carson PE, et al. Prevalence and significance of nonsustained ventricular tachycardia in patients with premature ventricular contractions and heart failure treated with vasodilator therapy. Department of Veterans Affairs CHF STAT Investigators. J Am Coll Cardiol. 1998;32:942–7.

    Article  CAS  Google Scholar 

  3. Teerlink JR, Jalaluddin M, Anderson S, Kukin ML, Eichhorn EJ, Francis G, et al. Ambulatory ventricular arrhythmias in patients with heart failure do not specifically predict an increased risk of sudden death. PROMISE (Prospective Randomized Milrinone Survival Evaluation) Investigators. Circulation. 2000;101:40–6.

    Article  CAS  Google Scholar 

  4. Yokoshiki H, Shimizu A, Mitsuhashi T, Furushima H, Sekiguchi Y, Manaka T, et al. Prognostic significance of nonsustained ventricular tachycardia in patients receiving cardiac resynchronization therapy for primary prevention: analysis of the Japan cardiac device treatment registry database. J Arrhythm. 2018;34:139–47.

    Article  Google Scholar 

  5. Doval HC, Nul DR, Grancelli HO, Varini SD, Soifer S, Corrado G, et al. Nonsustained ventricular tachycardia in severe heart failure. Independent marker of increased mortality due to sudden death. GESICA-GEMA investigators. Circulation. 1996;94:3198–203.

    Article  CAS  Google Scholar 

  6. Huikuri HV, Tapanainen JM, Lindgren K, Raatikainen P, Mäkikallio TH, Juhani Airaksinen KE, et al. Prediction of sudden cardiac death after myocardial infarction in the beta-blocking era. J Am Coll Cardiol. 2003;42:652–8.

    Article  Google Scholar 

  7. Mittal S, Aktas MK, Moss AJ, et al. The impact of nonsustained ventricular tachycardia on reverse remodeling, heart failure, and treated ventricular tachyarrhythmias in MADIT-CRT. J Cardiovasc Electrophysiol. 2014;25:1082–7.

    Article  Google Scholar 

  8. Clementy N, Bisson A, Challal F, Andre C, Pierre B, Fauchier L, et al. Nonsustained ventricular tachycardia at the time of implantation predicts appropriate therapies on rapid ventricular arrhythmia in primary prevention patients with nonischemic cardiomyopathy: Results From the Very-High-Rate Registry. JACC Clin Electrophysiol. 2017;3:1338–9.

    Article  Google Scholar 

  9. Pedersen CT, Kay GN, Kalman J, Borggrefe M, Della-Bella P, Dickfeld T, et al. EHRA/HRS/APHRS expert consensus on ventricular arrhythmias. Europace. 2014;16:1257–83.

    Article  Google Scholar 

  10. Epstein AE, Dimarco JP, Ellenbogen KA, et al. ACC/AHA/HRS 2008 guidelines for device-based therapy of cardiac rhythm abnormalities. Heart Rhythm. 2008;5:e1–62.

    Article  Google Scholar 

  11. Chen J, Johnson G, Hellkamp AS, Anderson J, Mark DB, Lee KL, et al. Rapid-rate nonsustained ventricular tachycardia found on implantable cardioverter-defibrillator interrogation: relationship to outcomes in the SCD-HeFT (Sudden Cardiac Death in Heart Failure Trial). J Am Coll Cardiol. 2013;61:2161–8.

    Article  Google Scholar 

  12. Jiménez-Candil J, Hernández J, Perdiguero P, Martín A, Moríñigo J, Bravo L, et al. Prognostic significance of nonsustained ventricular tachycardia episodes occurring early after implantable cardioverter-defibrillator implantation among patients with left ventricular dysfunction. Am J Cardiol. 2016;118:1503–10.

    Article  Google Scholar 

  13. Wang W, Lian Z, Rowin EJ, et al. Prognostic implications of nonsustained ventricular tachycardia in high-risk patients with hypertrophic cardiomyopathy. Circ Arrhythm Electrophysiol. 2017;10:e004604.

    PubMed  Google Scholar 

  14. Wilkoff BL, Fauchier L, Stiles MK, Morillo CA, al-Khatib SM, Almendral J, Aguinaga L, Berger RD, Cuesta A, Daubert JP, Dubner S, Ellenbogen KA, Mark Estes III NA, Fenelon G, Garcia FC, Gasparini M, Haines DE, Healey JS, Hurtwitz JL, Keegan R, Kolb C, Kuck KH, Marinskis G, Martinelli M, McGuire M, Molina LG, Okumura K, Proclemer A, Russo AM, Singh JP, Swerdlow CD, Teo WS, Uribe W, Viskin S, Wang CC, Zhang S 2015 HRS/EHRA/APHRS/SOLAECE expert consensus statement on optimal implantable cardioverter-defibrillator programming and testing. Heart Rhythm 2016;13: e50–e86.

  15. Wilkoff BL, Williamson BD, Stern RS, Moore SL, Lu F, Lee SW, et al. Strategic programming of detection and therapy parameters in implantable cardioverter-defibrillators reduces shocks in primary prevention patients: results from the PREPARE (Primary Prevention Parameters Evaluation) study. J Am Coll Cardiol. 2008;52:541–50.

    Article  Google Scholar 

  16. Gasparini M, Menozzi C, Proclemer A, Landolina M, Iacopino S, Carboni A, et al. A simplified biventricular defibrillator with fixed long detection intervals reduces implantable cardioverter defibrillator (ICD) interventions and heart failure hospitalizations in patients with non-ischaemic cardiomyopathy implanted for primary prevention: the RELEVANT [Role of long dEtection window programming in patients with LEft VentriculAr dysfunction, non-ischemic eTiology in primary prevention treated with a biventricular ICD] study. Eur Heart J. 2009;30:2758–67.

    Article  Google Scholar 

  17. Gasparini M, Proclemer A, Klersy C, Kloppe A, Lunati M, Ferrer JBM, et al. Effect of long-detection interval vs standard-detection interval for implantable cardioverter-defibrillators on antitachycardia pacing and shock delivery: the ADVANCE III randomized clinical trial. JAMA. 2013;309:1903–11.

    Article  CAS  Google Scholar 

  18. Moss AJ, Schuger C, Beck CA, Brown MW, Cannom DS, Daubert JP, et al. Reduction in inappropriate therapy and mortality through ICD programming. N Engl J Med. 2012;367:2275–83.

    Article  CAS  Google Scholar 

  19. Hayes DL, Boehmer JP, Day JD, Gilliam FR III, Heidenreich PA, Seth M, et al. Cardiac resynchronization therapy and the relationship of percent biventricular pacing to symptoms and survival. Heart Rhythm. 2011;8:1469–75.

    Article  Google Scholar 

  20. Zhou X, Gunderson BD, Olson WH, et al. Incidence of nonsustained and sustained ventricular tachyarrhythmias in patients with an implantable cardioverter defibrillator. J Cardiovasc Electrophysiol. 2004;15:14–20.

    Article  Google Scholar 

  21. Nemec J, Hammill SC, Shen WK. Increase in heart rate precedes episodes of ventricular tachycardia and ventricular fibrillation in patients with implantable cardioverter defibrillators: analysis of spontaneous ventricular tachycardia database. Pacing Clin Electrophysiol. 1999;22:1729–38.

    Article  CAS  Google Scholar 

  22. Kuck KH, Schaumann A, Eckardt L, Willems S, Ventura R, Delacrétaz E, et al. Catheter ablation of stable ventricular tachycardia before defibrillator implantation in patients with coronary heart disease (VTACH): a multicentre randomised controlled trial. Lancet. 2010;375:31–40.

    Article  Google Scholar 

  23. Qian Z, Guo J, Zhang Z, et al. Optimal programming management of ventricular tachycardia storm in ICD patients. J Biomed Res. 2015;29:35–43.

    PubMed  PubMed Central  Google Scholar 

Download references

Funding

This work was supported by the National Science & Technology Pillar Program during the 12th Five-Year Plan Period (2011BAI11B02).

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Shu Zhang.

Ethics declarations

Conflict of interest

The authors declare that they have no conflict of interest.

Additional information

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Zhou, Y., Zhao, S., Chen, K. et al. Predictive value of rapid-rate non-sustained ventricular tachycardia in the occurrence of appropriate implantable cardioverter-defibrillator therapy. J Interv Card Electrophysiol 57, 473–480 (2020). https://doi.org/10.1007/s10840-019-00557-4

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10840-019-00557-4

Keywords

Navigation