Concentric wear of the Delrin disc in a Bjork-Shiley heart valve prosthesis: Report of two cases

  • Case Reports
  • Published:
Surgery Today Aims and scope Submit manuscript

Abstract

We report herein the cases of two patients who received replacement of aortic Bjork-Shiley Delrin (BSD) valves that had been implanted for over 20 years following the development of aortic regurgitation (AR) resulting from wear of a Delrin disc. Case 1 was a 61-year-old woman in whom echocardiography revealed marked left ventricular dilatation and moderate AR 23 years after an aortic valve replacement (AVR) with a 21-mm BSD valve. Case 2 was a 51-year-old woman in whom echocardiography revealed marked dilation of the right atrium and moderate AR 23 years and 8 months after an AVR with a 21-mm BSD valve, as well as a mitral valve replacement with a 3M Starr-Edwards (SE) ball valve and tricuspid annuloplasty. In both patients, the BSD valves were replaced with other mechanical valves at reoperation. Examination of the explanted BSD valves showed that the Delrin discs contained increases in the radial gaps and strut indentation grooves on the inflow and outflow surfaces. The type and magnitude of the wear on the Delrin discs in these valves were consistent with data reported in the literature for this valve design after similar implant duration.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Bjork VO (1969) A new tilting disc valve prosthesis. Scand J Thorac Cardiovasc Surg 3:1–10

    PubMed  CAS  Google Scholar 

  2. Bjork VO (1972) Delrin as an implant material for valve occluders. Scand J Thorac Cardiovasc Surg 6:103–107

    PubMed  CAS  Google Scholar 

  3. Teoh SH, Martin RL, Lim SC, Lee KH, Mok CK, Kwok WC (1990) Delrin as an occluder material. ASAIO Trans 36:M417-M421

    PubMed  CAS  Google Scholar 

  4. Bjork VO (1971) Disc wear in the Bjork-Shiley tilting disc valve. Experimental and clinical observations. Scand J Thorac Cardiovasc Surg 5:87–91

    PubMed  CAS  Google Scholar 

  5. Takamoto T, Niwa A, Taniguchi K, Sakamoto T, Yamada T, Asano K (1981) M-mode echocardiographic findings of artificial tricuspid valve (Bjork-Shiley valve) dysfunction. Nippon Choonnpa Igakkai Zasshi (Jpn J Med Ultrason) 16 [Suppl II]:343–344

    Google Scholar 

  6. Misumi H, Matsuo K, Ishihara K, Nakae S, Hashimoto A, Koyanagi H (1989) Malfunction of the Bjork-Shiley valve prostheses due to disc defacement: report of two cases with successful reoperation (in Japanese with English abstract). Nippon Kyobu Geka Gakkai Zasshi (J Jpn Assoc Thorac Surg) 37:132–137

    Google Scholar 

  7. Katsumata T, Shimakura T, Nakano H, Shimamura Y, Hoshino K, Harada K, Maeshima H, Kawano H, Asakawa K, Yabuki M, Matsuda N (1993) Late malfunction of the Bjork-Shiley valve prosthesis due to Delrin disc defacement (in Japanese with English abstract). Kyobu Geka (Jpn J Thorac Surg) 46:1059–1062

    PubMed  CAS  Google Scholar 

  8. Mazzucco A, Morea P, Milano A, Bortolotti U (1994). Concentric wear of the Delrin disc of a Bjork-Shiley prosthesis: an uncommon cause of prosthetic incompetence (letter to Editor). J Thorac Cardiovasc Surg 107:318–319

    PubMed  CAS  Google Scholar 

  9. Abe M, Sakai A, Lim ZB, Hanayama N, Osawa M (1996) Report of a case with re-mitral valve replacedment for Bjork-Shiley Delrin disc valve prosthesis: disc wear of the Delrin disc 20 years after implantation (in Japanese with English abstract). Nippon Kyobu Geka Gakkai Zasshi (J Jpn Assoc Thorac Surg) 44:545–547

    PubMed  CAS  Google Scholar 

  10. Wieting DW (1996) The Bjork-Shiley Delrin tilting disc heart valve: Historical perspective, design and need for scientific analyses after 25 years. J Heart Valve Dis 5:S157-S168

    PubMed  Google Scholar 

  11. Dellsperger KC, Grunkmeier GL, Stein PD (1996) Clinical experience with the Bjork-Shiley Delrin tilting disc heart valve. J Heart Valve Dis 5:S169-S177

    PubMed  Google Scholar 

  12. Engel PA, Clark ES, Park SH, McKellop H, Thyagarajan K, Zhao Z, Campbell P, Wieting DW (1996) Surface and edge wear of Bjork-Shiley Delrin heart valve discs. J Heart Valve Dis 5:S190-S205

    PubMed  Google Scholar 

  13. Thyagarajan K, Conlin C, Milligan HL, Wieting DW (1996) A comparison of wear patterns in vivo and in vitro for Bjork-Shiley Delrin heart valve discs. J Heart Valve Dis 5:S206-S215

    PubMed  Google Scholar 

  14. Taams MA, Gussenhoven EJ, Cahalan MK, Roelandt JRTC, van Herwerden LA, The HK, Bon N, de Jong N (1989) Transesophageal Doppler color flow mapping in the detection of native and Bjork-Shiley mitral valve regurgitation. J Am Coll Cardiol 13:95–99

    Article  PubMed  CAS  Google Scholar 

  15. Aoyagi S, Nishi Y, Kawara T, Oryoji A, Kosuga K, Oishi K (1994) Doppler echocardiographic diagnosis of malfunction of a St. Jude Medical mitral valve. Artif Organs Today 3:299–307

    Google Scholar 

  16. Lindower PD, Dellsperger KC, Johnson B, Chandran KB, Vandenberg BF (1996) Variability of regurgitation in Bjork-Shiley mitral valves and relationship to disc occluder design: an in vitro two-dimensional color-Doppler flow mapping study. J Heart Valve Dis 5:S178-S183

    PubMed  Google Scholar 

  17. Stein PD, Dellsperger KC (1996) Summary and recommendations. J Heart Valve Dis 5:S246-S248

    PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Aoyagi, S., Tayama, K., Tayama, E. et al. Concentric wear of the Delrin disc in a Bjork-Shiley heart valve prosthesis: Report of two cases. Surg Today 29, 1115–1119 (1999). https://doi.org/10.1007/s005950050656

Download citation

  • Received:

  • Accepted:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s005950050656

Key Words

Navigation