Outcome of short- to medium-term migration analysis of a cementless short stem total hip arthroplasty using EBRA-FCA: a radiological and clinical study

  • Hip Arthroplasty
  • Published:
Archives of Orthopaedic and Trauma Surgery Aims and scope Submit manuscript

Abstract

Introduction

Stress shielding may lead to aseptic loosening which is a common reason for implant failure. An established method to identify implants with risk of implant failure caused by aseptic loosening is to measure early migration of the stem with the “Ein Bild Roentgen Analyse” femoral component analysis (EBRA-FCA). Therefore, the aim of this study was to measure the migration of a cementless short stem prosthesis via EBRA-FCA to predict the future performance.

Materials and methods

A total collective of 71 patients were treated with a cementless short hip stem prosthesis. Indications for surgery were primary coxarthrosis, dysplasia coxarthrosis or femoral head necrosis. After surgery, the patients were followed-up immediately after 3, 6, 12, 24, 36, and 48 months and X-ray images for EBRA-FCA measurements were taken. Axial caudal migration as well as the varus/valgus tilting of the prosthesis was determined. Possible influencing factors like BMI, age, diagnosis, gender or Harris Hip Score (HHS) on the migration of the stem were assessed.

Results

HHS increased significantly direct postoperatively (p < 0.001). At the second follow-up the average caudal migration was 0.42 ± 0.52 mm (range: 0.00–2.85 mm) (p < 0.001). A total of 14 patients underwent a caudal migration greater than 1.5 mm until 48 months. The initial varus and valgus tilts within the first 3 months were significant (p < 0.001). No correlations between BMI, age, diagnosis, gender or HHS and the migration as well as the tilting of the cementless short hip stem prosthesis were found.

Conclusions

Although initial axial caudal migration as well as tilting tendencies in varus or valgus position can be detected, there is no marked migration of the examined prosthesis after the first 48 months. Likewise, no aseptic early loosening was detected throughout the study period, which indicates good osseointegration of the short stem prosthesis.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3

Similar content being viewed by others

References

  1. Pivec R, Johnson AJ, Mears SC, Mont MA (2012) Hip arthroplasty. Lancet 380:1768–1777. https://doi.org/10.1016/S0140-6736(12)60607-2

    Article  PubMed  Google Scholar 

  2. Wittenberg RH, Steffen R, Windhagen H, Bücking P, Wilcke A (2013) Five-year results of a cementless short-hip-stem prosthesis. Orthop Rev 5:e4. https://doi.org/10.4081/or.2013.e4

    Article  Google Scholar 

  3. Gronewold J, Berner S, Olender G, Hurschler C, Windhagen H, von Lewinski G, Floerkemeier T (2014) Changes in strain patterns after implantation of a short stem with metaphyseal anchorage compared to a standard stem: an experimental study in synthetic bone. Orthop Rev 6:5211. https://doi.org/10.4081/or.2014.5211

    Article  Google Scholar 

  4. Jahnke A, Fonseca Ulloa CA, Seeger JB, Rickert M, Jahnke GW, Ahmed GA, Ishaque BA (2018) Analysis of the elastic bending characteristics of cementless short hip stems considering the valgus alignment of the prosthetic stem. Clin Biomech 52:49–56. https://doi.org/10.1016/j.clinbiomech.2018.01.006

    Article  Google Scholar 

  5. Roth A, Richartz G, Sander K, Sachse A, Fuhrmann R, Wagner A, Venbrocks R-A (2005) Verlauf der periprothetischen Knochendichte nach Hüfttotalendoprothesenimplantation. Abhängigkeit von Prothesentyp und knöcherner Ausgangssituation. Orthopade 34:334–344. https://doi.org/10.1007/s00132-005-0773-1

    Article  CAS  PubMed  Google Scholar 

  6. Streit MR, Haeussler D, Bruckner T, Proctor T, Innmann MM, Merle C et al (2016) Early migration predicts aseptic loosening of cementless femoral stems: a long-term study. Clin Orthop Relat Res 474:1697–1706. https://doi.org/10.1007/s11999-016-4857-5

    Article  PubMed  PubMed Central  Google Scholar 

  7. Krismer M, Biedermann R, Stöckl B, Fischer M, Bauer R, Haid C (1999) The prediction of failure of the stem in THR by measurement of early migration using EBRA-FCA. Einzel-Bild-Roentgen-Analyse-femoral component analysis. J Bone Joint Surg Br 81:273–280

    Article  CAS  Google Scholar 

  8. Kroell A, Beaulé P, Krismer M, Behensky H, Stoeckl B, Biedermann R (2009) Aseptic stem loosening in primary THA: migration analysis of cemented and cementless fixation. Int Orthop 33:1501–1505. https://doi.org/10.1007/s00264-008-0701-1

    Article  PubMed  Google Scholar 

  9. Ishaque BA, Stürz H, Basad E (2011) Fatigue fracture of a short stem hip replacement: a failure analysis with electron microscopy and review of the literature. J Arthroplasty 26:665.e17–20. https://doi.org/10.1016/j.arth.2010.05.021

    Article  PubMed  Google Scholar 

  10. Yan SG, Weber P, Steinbrück A, Hua X, Jansson V, Schmidutz F (2018) Periprosthetic bone remodelling of short-stem total hip arthroplasty: a systematic review. Int Orthop. 42:2077–2086. https://doi.org/10.1007/s00264-017-3691-z

    Article  PubMed  Google Scholar 

  11. Jahnke A, Engl S, Seeger JB, Basad E, Rickert M, Ishaque BA (2015) Influences of fit and fill following hip arthroplasty using a cementless short-stem prosthesis. Arch Orthop Trauma Surg 135:1609–1614. https://doi.org/10.1007/s00402-015-2302-y

    Article  PubMed  Google Scholar 

  12. von Lewinski G, Floerkemeier T (2015) 10-year experience with short stem total hip arthroplasty. Orthopedics 38:S51–S56. https://doi.org/10.3928/01477447-20150215-57

    Article  Google Scholar 

  13. Kutzner KP, Kovacevic MP, Freitag T, Fuchs A, Reichel H, Bieger R (2016) Influence of patient-related characteristics on early migration in calcar-guided short-stem total hip arthroplasty: a 2-year migration analysis using EBRA-FCA. J Orthop Surg Res 11:29. https://doi.org/10.1186/s13018-016-0363-4

    Article  PubMed  PubMed Central  Google Scholar 

  14. Kaipel M, Grabowiecki P, Sinz K, Farr S, Sinz G (2015) Migration characteristics and early clinical results of the NANOS® short-stem hip arthroplasty. Wien Klin Wochenschr 127:375–378. https://doi.org/10.1007/s00508-015-0756-0

    Article  PubMed  Google Scholar 

  15. van Oldenrijk J, Molleman J, Klaver M, Poolman RW, Haverkamp D (2014) Revision rate after short-stem total hip arthroplasty: a systematic review of 49 studies. Acta Orthop 85:250–258. https://doi.org/10.3109/17453674.2014.908343

    Article  PubMed  PubMed Central  Google Scholar 

  16. Loppini M, Grappiolo G (2018) Uncemented short stems in primary total hip arthroplasty: the state of the art. EFORT Open Rev 3:149–159. https://doi.org/10.1302/2058-5241.3.170052

    Article  PubMed  PubMed Central  Google Scholar 

  17. Schmidutz F, Graf T, Mazoochian F, Fottner A, Bauer-Melnyk A, Jansson V (2012) Migration analysis of a metaphyseal anchored short-stem hip prosthesis. Acta Orthop 83:360–365. https://doi.org/10.3109/17453674.2012.712891

    Article  PubMed  PubMed Central  Google Scholar 

  18. Krismer M, Tschupik JP, Bauer R, Mayrhofer P, Stöckl B, Fischer M, Biedermann R (1997) Einzel-Bild-Röntgen-Analyse (EBRA) zur Messung der Migration von Hüftendoprothesen. Orthopade 26:229–236. https://doi.org/10.1007/PL00003378

    Article  CAS  PubMed  Google Scholar 

  19. Harris WH (1969) Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. An end-result study using a new method of result evaluation. J Bone Joint Surg Am 51:737–755

    Article  CAS  Google Scholar 

  20. Biedermann R, Krismer M, Stöckl B, Mayrhofer P, Ornstein E, Franzén H (1999) Accuracy of EBRA-FCA in the measurement of migration of femoral components of total hip replacement. Einzel-Bild-Röntgen-Analyse-femoral component analysis. J Bone Joint Surg Br 81:266–272

    Article  CAS  Google Scholar 

  21. Wilkinson JM, Hamer AJ, Elson RA, Stockley I, Eastell R (2002) Precision of EBRA-Digital software for monitoring implant migration after total hip arthroplasty. J Arthroplasty 17:910–916

    Article  CAS  Google Scholar 

  22. Augustin L, Boller S, Bobach C, Jahnke A, Ahmed GA, Rickert M, Ishaque BA (2018) Development of periprosthetic bone mass density around the cementless Metha® short hip stem during three year follow up—a prospective radiological and clinical study. Int Orthop 62:579. https://doi.org/10.1007/s00264-018-4126-1

    Article  Google Scholar 

  23. Boller S, Jahnke A, Augustin L, Ahmed G, Rickert M, Ishaque BA (2018) Age-related osseointegration of a short hip stem: a clinical and radiological 24 months follow-up. Arch Orthop Trauma Surg. https://doi.org/10.1007/s00402-018-3082-y

    Article  PubMed  Google Scholar 

  24. Jahnke A, Engl S, Altmeyer C, Jakubowitz E, Seeger JB, Rickert M, Ishaque BA (2014) Changes of periprosthetic bone density after a cementless short hip stem: a clinical and radiological analysis. Int Orthop 38:2045–2050. https://doi.org/10.1007/s00264-014-2370-6

    Article  PubMed  Google Scholar 

  25. Kutzner KP, Freitag T, Donner S, Kovacevic MP, Bieger R (2017) Outcome of extensive varus and valgus stem alignment in short-stem THA: clinical and radiological analysis using EBRA-FCA. Arch Orthop Trauma Surg 137:431–439. https://doi.org/10.1007/s00402-017-2640-z

    Article  PubMed  PubMed Central  Google Scholar 

  26. Fottner A, Woiczinski M, Kistler M, Schröder C, Schmidutz TF, Jansson V, Schmidutz F (2017) Influence of undersized cementless hip stems on primary stability and strain distribution. Arch Orthop Trauma Surg. 137:1435–1441. https://doi.org/10.1007/s00402-017-2784-x

    Article  PubMed  Google Scholar 

  27. Lerch M, von der Haar-Tran A, Windhagen H, Behrens BA, Wefstaedt P, Stukenborg-Colsman CM (2012) Bone remodelling around the Metha short stem in total hip arthroplasty: a prospective dual-energy X-ray absorptiometry study. Int Orthop. 2012(36):533–538. https://doi.org/10.1007/s00264-011-1361-0

    Article  Google Scholar 

  28. Epinette J-A, Brax M, Chammaï Y (2017) A predictive radiological analysis of short stems versus both shortened and long stems in primary hip replacement: a case-control study of 100 cases of Metha versus ABG II and Omnifit HA at 2–8 years’ follow-up. Orthop Traumatol Surg Res. 103:981–986. https://doi.org/10.1016/j.otsr.2017.07.014

    Article  PubMed  Google Scholar 

  29. Schnurr C, Loucif A, Patzer T, Schellen B, Beckmann J, Eysel P (2018) Short stem survival after osteonecrosis of the femoral head. Arch Orthop Trauma Surg. 138:573–579. https://doi.org/10.1007/s00402-018-2892-2

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Bernd Alexander Ishaque.

Ethics declarations

Conflict of interest

We thank Aesculap for the financial support for the EBRA software costs.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Jahnke, A., Wiesmair, AK., Fonseca Ulloa, C.A. et al. Outcome of short- to medium-term migration analysis of a cementless short stem total hip arthroplasty using EBRA-FCA: a radiological and clinical study. Arch Orthop Trauma Surg 140, 247–253 (2020). https://doi.org/10.1007/s00402-019-03315-3

Download citation

  • Received:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00402-019-03315-3

Keywords

Navigation