Abstract
Objectives
To develop a deep learning–based method for contrast-enhanced breast lesion detection in ultrafast screening MRI.
Materials and methods
A total of 837 breast MRI exams of 488 consecutive patients were included. Lesion’s location was independently annotated in the maximum intensity projection (MIP) image of the last time-resolved angiography with stochastic trajectories (TWIST) sequence for each individual breast, resulting in 265 lesions (190 benign, 75 malignant) in 163 breasts (133 women). YOLOv5 models were fine-tuned using training sets containing the same number of MIP images with and without lesions. A long short-term memory (LSTM) network was employed to help reduce false positive predictions. The integrated system was then evaluated on test sets containing enriched uninvolved breasts during cross-validation to mimic the performance in a screening scenario.
Results
In five-fold cross-validation, the YOLOv5x model showed a sensitivity of 0.95, 0.97, 0.98, and 0.99, with 0.125, 0.25, 0.5, and 1 false positive per breast, respectively. The LSTM network reduced 15.5% of the false positive prediction from the YOLO model, and the positive predictive value was increased from 0.22 to 0.25.
Conclusions
A fine-tuned YOLOv5x model can detect breast lesions on ultrafast MRI with high sensitivity in a screening population, and the output of the model could be further refined by an LSTM network to reduce the amount of false positive predictions.
Clinical relevance statement
The proposed integrated system would make the ultrafast MRI screening process more effective by assisting radiologists in prioritizing suspicious examinations and supporting the diagnostic workup.
Key Points
• Deep convolutional neural networks could be utilized to automatically pinpoint breast lesions in screening MRI with high sensitivity.
• False positive predictions significantly increased when the detection models were tested on highly unbalanced test sets with more normal scans.
• Dynamic enhancement patterns of breast lesions during contrast inflow learned by the long short-term memory networks helped to reduce false positive predictions.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
As the most sensitive breast imaging modality [1], MRI has been used as a supplement tool for high-risk population-based screening where mammography alone is not sufficient [2, 3]. However, the high cost, low availability, and lack of dedicated radiologists restrict the application of MRI for an intermediate- or even low-risk population [4].
Shortening the protocol has been attempted to increase the cost-effectiveness of breast MRI screening [5]. Recently, several innovative abbreviated MRI protocols have been proposed and evaluated [6,7,8,9]. Mann et al investigated the feasibility of using ultrafast breast MRI as a standalone technique for breast MRI screening [10]. A multi-reader study showed that this time-resolved angiography with stochastic trajectories (TWIST) sequence-based protocol performed similarly to the full diagnosed protocol and significantly higher screening specificity [11]. Compared with the protocol proposed by Kuhl et al [9], the ultrafast MRI-based protocol produces twenty high spatial acquisitions within 102 s, which allows not only morphologic analysis, but also kinetic analysis during contrast agent inflow. Recent researches also showed the advantages of the early-stage dynamic information in the ultrafast MRI [12, 13].
Automatic detection of breast lesions has the potential to boost the efficiency of screening given that the majority of screening MRIs are lesion free. However, this requires the detection model being highly accurate and dependable. Lesion detection algorithms were developed and tested in prior studies using highly enriched datasets [14,15,16], where the proportion of scans with suspicious lesions (58.3 to 100%) and cancers (5.7 to 63.8%) is much higher than previously reported. According to a prospective observational study conducted by Kuhl et al [2], a total of 3861 screening MRIs resulted in 61 cancers, 175 BI-RADS 3 and 171 BI-RADS 4 or 5 diagnoses. This yielded a cancer rate of 1.6% and a lesion rate of 9.0%, and both were significantly lower than the rates aforementioned. To ensure the reliability of detection models, it is crucial to test them in a real-world screening setting. The difference in the proportion of scans with and without suspicious lesions between the model development dataset and actual clinical practice can potentially result in misleading performance in lesion detection, thereby diminishing the model’s dependability within the screening population.
In this study, a deep learning–based detection system was developed to identify enhanced lesions in ultrafast MRI. To simulate the screening situation, the detection system was evaluated with test cohorts that had mostly normal examinations. The proposed detection systems aimed to accelerate the screening process by prioritizing MRI scans and reducing the radiologist’s workload.
Materials and methods
Patients
The breast MRI scans were retrospectively collected at the University Medical Center Groningen. The institute’s local ethics committee approved this retrospective study (METc 2018/652) and the need for informed consent was waived. The same dataset was used in our earlier work, where a classification system was invented and tested [17]. Lesions in those MRI scans were delicately annotated with bounding boxes to develop a detection system in this study.
To be specific, out of 809 consecutive women who underwent breast MRI examinations at our institute between April 2016 and October 2019, 1447 examinations were first acquired. Details of the acquisition protocols are provided in the Electronic Supplementary Material. Examinations were then included if the following conditions are satisfied: (1) a complete scan contains TWIST sequences; (2) the indication for MRI should be either screening or preoperative assessment; and (3) the identified lesions should have been biopsied or had at least a 2-year follow-up to serve as a gold standard for benignancy. Breast MRIs without TWIST sequences or performed for other reasons (chemotherapy response evaluation, post-surgery follow-up, and implant check) were excluded. Similar with previous research [14], in this study, the left and right breasts of each woman are considered different data points and involved in the training and validation independently.
Data preprocessing and annotation
An overview of the preprocessing procedure of the breast MRI is shown in Fig. 1. A 3D-Unet was used to segment the breast area on the T1-weighted acquisition acquired prior to contrast agent injection to remove the redundant area behind the chest wall and the artifacts surrounding the breasts [18]. Scaling and FOV alignment were used to deploy the 3D masks to TWIST volumes, resulting in segmented volumes. Subtracted volumes were then created by subtracting the segmented pre-contrast volume from the segmented post-contrast volumes. Maximum intensity projection (MIP) images were then generated by applying the MIP operation to the segmented subtracted volumes. The MIP images were then split in two, to produce separate left and right breast images. On the MIP images, lesions were then annotated with bounding boxes (X.J., 3 years of experiences). LabelMe was used for the annotation process [19]. The location of the lesions was derived from the radiology clinical reports and confirmed by an experienced breast radiologist (M.D., 10 years of experience) when in doubt.
Breast lesion detection with YOLOv5
The proposed integrated system for detecting breast lesions in ultrafast MRI was composed of two parts: a YOLO model for lesion detection and a long short-term memory (LSTM) network for false positive reduction. The pipeline of the proposed detection method is illustrated in Fig. 2.
YOLO is a one-stage, global context informed architecture that could achieve faster execution speed without compromising the prediction accuracy [20]. Moreover, the YOLO architecture, which allows the use of pure negative images (images devoid of any target object) as input for training, matches the purpose of training models with normal breast images to reduce false positive predictions in a screening setting. MIP images generated in the data preprocessing were used to finetune YOLOv5 model that had been pretrained using the COCO dataset. Three YOLOv5 models (v5n with only 1.9 million parameters, v5m with 21.1 million parameters, and v5x with 86.7 million parameters) were selected and compared in this study to investigate the impact of model size on detection performance. To further investigate the impact of magnetic field strength on models’ detection performance, a subgroup analysis was also conducted in which 1.5-T and 3-T scans were used to train and test YOLOv5 models separately. The detailed information about model finetuning is described in Electonic Supplementary Material.
Benign lesions usually take longer than malignant lesions to be enhanced in the TWIST sequences [12, 21]. To ensure the detectability of benign lesions, MIP images of the early acquisitions in the TWIST were not involved in the development of the detection model. For each ultrafast MRI examination, only the MIP images of the last acquisition were used for the training and validation of the YOLO models.
False positive reduction
Rather than deducing the results from the YOLO models directly, the locations of the positive predictions with a probability above th0 from the output of YOLOv5 models were utilized to extract clips across the TWIST sequences. During inference, the LSTM network took each predicted bounding box’s clip (the area spanning the 14 yield MIP images of the ultrafast DCE sequences) as input. The output was a likelihood score of a breast containing a lesion. The predict scores of the YOLOv5 and LSTM networks were then merged to reach the final judgment (Fig. 2). The architecture and training process of the LSTM models are provided in the Electronic Supplementary Material.
Experiments
Five-fold cross-validation was performed at the breast level to train and evaluate each model and the integrated system. Especially, to avoid data overlap between training and test data from the same breast in patients who had multiple examinations, a group-based shuffle split method was used to ensure data from different dates of the same breast are bound together. Moreover, the YOLO and LSTM networks were trained with the same data splits and then integrated together to prevent data leakage throughout the pipeline.
The number of samples with and without lesions in this analysis was inevitably imbalanced. To overcome this imbalance, all positive samples (breasts with lesions) were merged with an equivalent amount of randomly selected negative samples (breasts without lesions) to establish a balanced development dataset. The rest of the negative samples were subsequently grouped as an isolated negative set and were not used for the model training. However, in addition to assessing the model trained with the balanced dataset, the isolated negative set was also merged to the test data during validation to evaluate the proposed system’s performance with a normal screening prevalence. The performance with and without the isolated negative set was compared. A diagram of the data split and validation set formation is shown in Electronic Supplementary Material Fig. S3.
Data analysis
Free-response operating characteristic (FROC) analysis was adopted to assess the performance of the evaluated YOLOv5 models and the integrated system. In this study, a false positive prediction is defined as a non-lesion area that is predicted to have a lesion, while a false negative prediction is defined as a lesion that deep learning models failed to detect. To illustrate the effectiveness of the LSTM network for false positive reduction, the sensitivity of the YOLO models alone and the integrated system (YOLO + LSTM) at 0.125, 0.25, 0.5, 1, and 2 false positives per breast were also calculated. Meanwhile, the detection ability of each model to detect malignant lesions was analyzed in the same way. The data analysis was performed with Scikit-learn 0.22.1 and COCO-FROC-analysis 0.2.0 packages in Python programming language.
Results
Included lesions
A total of 488 women were included in this study. The mean age of the included women (n = 488) was 48.5 years (range, 27–83 years), for women with breast lesions (n = 133) was 52.0 years (range, 27–83 years), and for women with malignant lesions (n = 58) was 57.6 years (range, 34–83 years). Five women were undergoing MRI examination with an indication of preoperative assessment, and all had malignant lesions, which account for 8.6% of the total number of women with malignant lesion and 3.8% of women with lesion. A flowchart of this process is illustrated in Fig. S4; further details of these patients have previously been reported [17].
In total, 962 single breasts were derived from the included patients; 14 breasts were excluded due to mastectomy. For included breasts, 83.0% (n = 799) were reported as lesion free, and 315 were derived from 1.5-T scans and 484 were derived from 3.0-T scans. Seventeen percent (n = 163) had at least one lesion, in which 7.3% (n = 70) contained only one lesion, and 9.7% (n = 93) contained multiple lesions. In total, 265 lesions were annotated, 71 lesions were derived from 56 1.5-T scans and 194 were derived from 112 3.0-T scans. The median size of all reported lesions was 13.0 mm (range, 5.0–110.0 mm), 9.0 mm (range, 5.0–81.0 mm) for benign lesions (n = 190) and 22.0 mm (range, 6.0–110.0 mm) for malignant lesions (n = 75). The detailed information of the lesions is illustrated in Table 1.
YOLOv5 model performance
During cross-validation, the fine-tuned YOLOv5x model showed a sensitivity of 0.95 (0.85–1.0), 0.97 (0.91–1.0), 0.98 (0.93–1.0), and 0.99 (0.96–1.0) with 0.125, 0.25, 0.5, and 1 false positive per breast, respectively, compared to 0.94 (0.91–0.98), 0.97 (0.95–1.0), 0.98 (0.96–1.0), and 0.99 (0.96–1.0) for the YOLOv5m model and 0.77 (0.69–0.86), 0.87 (0.78–0.94), 0.93 (0.89–0.98), and 0.98 (0.94–1.0) for the YOLOv5n model. With two false positives per breast, all models obtained a sensitivity of 1.0 for lesion identification. For different magnetic field strength subgroup, the YOLOv5x model achieved an overall higher sensitivity on the 3.0 T subgroup than the 1.5 T subgroup, with a sensitivity of 0.76 (0.65–0.84) and 0.69 (0.59–0.83) with 0.125 false positives per breast, respectively. Detailed results of the model on each subgroup are provided in the Electronic Supplementary Material.
For malignant lesion detection, the YOLOv5x model had a sensitivity of 0.96 (0.80–1.0), 0.97 (0.86–1.0), 0.98 (0.90–1.0), and 1.0 with 0.125, 0.25, 0.5, and 1 false positive per breast, respectively, compared to 0.97 (0.90–1.0), 1.0, 1.0, and 1.0 for the YOLOv5m model and 0.94 (0.90–1.0), 0.97 (0.92–1.0), 0.98 (0.92–1.0), and 0.98 (0.92–1.0) for the YOLOv5n model. All models had a sensitivity of 1.0 for malignant lesion detection, with two false positives per breast. The FROC curve of the YOLOv5x models is shown in Fig. 3a.
False positive reduction with LSTM
To reduce false positive predictions, an LSTM network was used to further identify candidates from the output of the YOLOv5 models. The performance of the integrated system was evaluated using the same data split as the single YOLO model during cross-validation. The FROC curve of the integrated system is shown in Fig. 3b. The integrated system retained approximately equivalent sensitivity for both malignant and all lesions detection, after integrating the LSTM network (Table 2). To demonstrate the efficacy of the false positive reduction, Table 3 shows the number of false positive predictions of the YOLOv5x model in each fold of the cross-validation with and without the LSTM network. The findings were obtained using a 0.5 th0 and 0.5 th1 threshold setting.
Example of final detection results of the integrated system is shown in Fig. 4. Only detections with probability above the threshold and with correct location estimates deemed as true positives (Fig. 4a), while normal breast tissues without any lesion predictions deemed as true negatives (Fig. 4b). Other tissues, such as nipples, lymph nodes, vessels, and enhanced parenchyma, which were incorrectly predicted as lesions were considered false positives. Any missed lesions and lesions with incorrect location estimations were considered as false negatives.
Discussion
In this study, we developed a deep learning–based method for automated detection of breast lesions for the purpose of breast cancer screening. Unlike previous research on breast MRI lesion detection, we focused on the performance of models with a screening prevalence. The model was evaluated using test sets that included mostly normal scans rather than only scans with lesions. The proposed integrated system had a sensitivity that was comparable to previous studies while producing fewer false positives.
To address the high false positive rate for breast lesion detection, an LSTM network aimed at identifying different dynamic intensity patterns during contrast inflow was employed. Using the YOLOv5x model as an example, the LSTM network could help eliminate 15% of the false positive predictions. This LSTM network operates on the output of the YOLOv5 models and has the potential to misclassify correctly predicted lesions, resulting in a decreased sensitivity. However, despite a few misclassifications induced by the LSTM network, the integrated system retained its high sensitivity in general during cross-validation, and the feared significant decline in sensitivity was not observed (Fig. 3b and Table 3).
This is a follow-up study of our previous work in which a classification system was invented to identify lesion-free scans that use only TWIST sequences [17]. However, the previous study mainly focused on the primary tumors in each breast and take no account of minor findings. Instead of generating a categorical prediction, we elaborately annotated all lesions with bounding box in the cohort and trained detection methods to localize lesions in the breasts, for the propose of providing more precise and visible results. Compared with the classification system in our previous work, which focused on excluding normal scans to minimize the reading list, the detection model developed in this study gives more immediate visual assistance, allowing radiologists to focus on the suspicious breast lesions directly.
The size of YOLOv5 models has impact on the detection performance. Three different YOLOv5 models were evaluated in this study. Compared with the 0.77 sensitivity achieved by the YOLOv5n model, both YOLOv5m and YOLOv5x achieved superior sensitivity (0.94 and 0.95) with 0.125 false positives per breast. This advantage gradually vanished with a higher false positive rate. We also investigate the impact of magnetic field strength on detection models’ performance. The YOLOv5x model achieved an overall higher sensitivity on the 3.0 T subgroup than the 1.5 T subgroup. However, despite the model’s poor performance, which likely resulted by the insufficient amount of positive training samples in each subgroup, it is difficult to draw the conclusion that a stronger magnetic field benefits the model’s performance. Collecting more data would enable us to conduct a comprehensive subgroup analysis.
The integrated system is developed to detect all enhanced breast masses (> 5 mm), not just malignant ones. Several previous studies developed and tested models for cancer detection in DCE-MRI and were successful [22, 23]. However, the diagnosis of malignancy is heavily dependent on additional MRI sequences and even biopsies. Using a single model to locate and identify just malignancies while ignoring other suspicious findings is risky and unreasonable, especially for screening purposes, dedicated AI models may be more useful for malignancy identification of candidate lesions [24].
One should keep in mind is that ultrafast MRI has not yet been evaluated in a real screening cohort [25]. Studies that compared the effectiveness of ultrafast and conventional MRI were also mainly retrospective studies involving patients with lesions in MRI. Meanwhile, the ultrafast MRI techniques are not standardized yet; parallel imaging, viewing sharing, and compressed sensing are all used to obtain a higher temporal resolution but are all referred to ultrafast MRI [26, 27].
This study has several limitations. One of the limitations was that there was no independent test set, instead, we used cross-validation to illustrate the effectiveness of the integrated system. Meanwhile, due to the rarity of the number of cancers in a screening population and the large amount of data required to train deep learning models, preoperative MRI scans were also included in the dataset to increase the number of cancers. Hence, the system should be further evaluated with additional dedicated multicenter multivendor databases. Another limitation is that only part of benign lesions was histologically examined; this reflected the radiologists’ confidence in their ability to determine the need for biopsy based on imaging outcomes.
This study demonstrates the ability of a deep learning–based method to detect candidate findings in ultrafast breast MRI. This proposed fully automated method could be helpful in detecting breast lesions in the setting of breast cancer screening, thereby potentially reducing radiologists’ workload. This, in turn, will allow breast MRI screening to apply to a larger population, resulting in better preventive health care delivery.
Abbreviations
- FROC:
-
Free-response operating characteristic
- LSTM:
-
Long short-term memory
- MIP:
-
Maximum intensity projections
- PPV:
-
Positive predictive value
- TWIST:
-
Time-resolved angiography with stochastic trajectories
References
Mann RM, Hooley R, Barr RG, Moy L (2020) Novel approaches to screening for breast cancer. Radiology 297:266–285. https://doi.org/10.1148/RADIOL.2020200172
Kuhl CK, Strobel K, Bieling H et al (2017) Supplemental breast MR imaging screening of women with average risk of breast cancer. Radiology 283:361–370. https://doi.org/10.1148/radiol.2016161444
Monticciolo DL, Newell MS, Moy L et al (2018) Breast cancer screening in women at higher-than-average risk: recommendations from the ACR. J Am Coll Radiol 15:408–414. https://doi.org/10.1016/j.jacr.2017.11.034
Mann RM, Kuhl CK, Moy L (2019) Contrast-enhanced MRI for breast cancer screening. J Magn Reson Imaging 50:1–14. https://doi.org/10.1002/jmri.26654
Tollens F, Baltzer PAT, Dietzel M et al (2022) Economic potential of abbreviated breast MRI for screening women with dense breast tissue for breast cancer. Eur Radiol. https://doi.org/10.1007/s00330-022-08777-5
Grimm LJ, Soo MS, Yoon S et al (2015) Abbreviated screening protocol for breast MRI. a feasibility study. Acad Radiol 22:1157–1162. https://doi.org/10.1016/j.acra.2015.06.004
Ohlmeyer S, Laun FB, Bickelhaupt S et al (2021) Ultra-high b-value diffusion-weighted imaging-based abbreviated protocols for breast cancer detection. Invest Radiol 56:629–636. https://doi.org/10.1097/RLI.0000000000000784
Comstock CE, Gatsonis C, Newstead GM et al (2020) Comparison of abbreviated breast MRI vs digital breast tomosynthesis for breast cancer detection among women with dense breasts undergoing screening. JAMA 323:746–756. https://doi.org/10.1001/jama.2020.0572
Kuhl CK, Schrading S, Strobel K et al (2014) Abbreviated breast magnetic resonance imaging (MRI): first postcontrast subtracted images and maximum-intensity projection—a novel approach to breast cancer screening with MRI. J Clin Oncol 32:2304–2310. https://doi.org/10.1200/JCO.2013.52.5386
Mann RM, Mus RD, Van Zelst J et al (2014) A novel approach to contrast-enhanced breast magnetic resonance imaging for screening: high-resolution ultrafast dynamic imaging. Invest Radiol 49:579–585. https://doi.org/10.1097/RLI.0000000000000057
Van Zelst JCM, Vreemann S, Witt HJ et al (2018) Multireader study on the diagnostic accuracy of ultrafast breast magnetic resonance imaging for breast cancer screening. Invest Radiol 53:579–586. https://doi.org/10.1097/RLI.0000000000000494
Pelissier M, Ambarki K, Salleron J, Henrot P (2021) Maximum slope using ultrafast breast DCE-MRI at 1.5 Tesla: a potential tool for predicting breast lesion aggressiveness. Eur Radiol 31:9556–9566. https://doi.org/10.1007/s00330-021-08089-0p
Milenkovic J, Dalmıs MU, Žgajnar J, Platel B (2017) Textural analysis of early-phase spatiotemporal changes in contrast enhancement of breast lesions imaged with an ultrafast DCE-MRI protocol. Med Phys 44:4652–4664. https://doi.org/10.1002/mp.12408
Ayatollahi F, Shokouhi SB, Mann RM, Teuwen J (2021) Automatic breast lesion detection in ultrafast DCE-MRI using deep learning. Med Phys 48:5897–5907. https://doi.org/10.1002/mp.15156
Rodríguez-Ruiz A, Krupinski E, Mordang J-J et al (2019) Detection of breast cancer with mammography: effect of an artificial intelligence support system. Radiology 290:305–314. https://doi.org/10.1148/radiol.2018181371
Watanabe AT, Lim V, Vu HX et al (2019) Improved cancer detection using artificial intelligence: a retrospective evaluation of missed cancers on mammography. J Digit Imaging 32:625–637. https://doi.org/10.1007/s10278-019-00192-5
Jing X, Wielema M, Cornelissen LJ et al (2022) Using deep learning to safely exclude lesions with only ultrafast breast MRI to shorten acquisition and reading time. Eur Radiol. https://doi.org/10.1007/s00330-022-08863-8
Zhang J, Saha A, Zhu Z, Mazurowski MA (2018) Hierarchical convolutional neural networks for segmentation of breast tumors in MRI with application to radiogenomics. IEEE Trans Med Imaging 38:435–447. https://doi.org/10.1109/TMI.2018.2865671
Russell BC, Torralba A, Murphy KP, Freeman WT (2008) LabelMe: a database and web-based tool for image annotation. Int J Comput Vis 77:157–173. https://doi.org/10.1007/s11263-007-0090-8
Fang W, Wang L, Ren P (2020) Tinier-YOLO: a real-time object detection method for constrained environments. IEEE Access 8:1935–1944. https://doi.org/10.1109/ACCESS.2019.2961959
Abe H, Mori N, Tsuchiya K et al (2016) Kinetic analysis of benign and malignant breast lesions with ultrafast dynamic contrast-enhanced MRI: comparison with standard kinetic assessment. AJR Am J Roentgenol 207:1159–1166. https://doi.org/10.2214/AJR.15.15957
Gubern-Mérida A, Martí R, Melendez J et al (2015) Automated localization of breast cancer in DCE-MRI. Med Image Anal 20:265–274. https://doi.org/10.1016/j.media.2014.12.001
Dalmış MU, Vreemann S, Kooi T et al (2018) Fully automated detection of breast cancer in screening MRI using convolutional neural networks. J Med Imaging 5:1. https://doi.org/10.1117/1.jmi.5.1.014502
Dalmış MU, Gubern-Mérida A, Vreemann S et al (2019) Artificial intelligence based classification of breast lesions imaged with a multi-parametric breast MRI protocol with ultrafast DCE-MRI, T2 and DWI. Invest Radiol 00:325–332. https://doi.org/10.1097/RLI.0000000000000544
Mann RM, van Zelst JCM, Vreemann S, Mus RDM (2019) Is ultrafast or abbreviated breast MRI ready for prime time? Curr Breast Cancer Rep 11:9–16. https://doi.org/10.1007/s12609-019-0300-8
Gao Y, Heller SL (2020) Abbreviated and ultrafast breast MRI in clinical practice. Radiographics 40:1507–1527. https://doi.org/10.1148/rg.2020200006
Tsao J (2010) Ultrafast imaging: principles, pitfalls, solutions, and applications. J Magn Reson Imaging 32:252–266. https://doi.org/10.1002/jmri.22239
Funding
The authors state that this work has not received any funding.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Guarantor
The scientific guarantor of this publication is Peter M. A. van Ooijen.
Conflict of interest
The authors of this manuscript declare no relationships with any companies, whose products or services may be related to the subject matter of the article.
Statistics and biometry
No complex statistical methods were necessary for this paper.
Informed consent
Written informed consent was waived by the institutional review board.
Ethical approval
Institutional review board approval was obtained.
Study subjects or cohorts overlap
Some study subjects or cohorts have been previously reported in Jing X, Wielema M, Cornelissen LJ, et al (2022) Using deep learning to safely exclude lesions with only ultrafast breast MRI to shorten acquisition and reading time. Eur Radiol. https://doi.org/10.1007/s00330-022-08863-8.
Methodology
• Retrospective
• Diagnostic or prognostic study
• performed at one institution
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Below is the link to the electronic supplementary material.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Jing, X., Dorrius, M.D., Zheng, S. et al. Localization of contrast-enhanced breast lesions in ultrafast screening MRI using deep convolutional neural networks. Eur Radiol 34, 2084–2092 (2024). https://doi.org/10.1007/s00330-023-10184-3
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00330-023-10184-3