Abstract
Objectives
To assess the learning curve for performing reliable liver stiffness measurements using a new hybrid machine composed of transient elastography (TE) interfaced with an ultrasound device for radiographers and radiologists with different levels of expertise in ultrasound imaging.
Methods
Ten novice operators who had never performed TE measurements were prospectively evaluated from April to October 2018: senior radiologists, young radiologists, fellows, radiographers, and residents, with different levels of experience in abdominal ultrasound imaging. All operators had a short theoretical training followed by a training session under supervision in three patients. Then, each operator had to perform TE in 50 consecutive patients with chronic liver disease, using beforehand ultrasound examination to select measurement area in the right liver lobe, and if needed, the XL probe. Percentages of failures and reliable measurements were compared.
Results
The rates of failures of measurements, poorly reliable, reliable, and very reliable results, were of 4.2% (21/500), 2.4% (12/500), 47.6% (238/500), and 45.8% (229/500), respectively. The rates of reliable plus very reliable results were excellent, ranging from 91 to 96% among all the subgroups. The rates of very reliable, reliable, and unreliable results did not differ between operator subgroups and especially between junior radiologists, senior radiologists, and radiographers. No breaking point was observed in the interquartile range/median values over time.
Conclusion
TE interfaced with ultrasound in this hybrid machine presents no learning curve effect. After a short initial training session, a novice observer is able to perform high rates of reliable and very reliable TE measurements.
Key Points
• When performing liver stiffness measurements using a new hybrid machine composed of transient elastography interfaced with ultrasound, the rate of failures of measurements is very low, below 5%.
• After a short training session and using ultrasound planning, a novice operator, whatever its expertise in ultrasound imaging, is capable of performing high rates of reliable and very reliable measurements.
• No learning curve is needed for performing reliable liver stiffness measurements using this new hybrid machine.
Similar content being viewed by others
Abbreviations
- BMI:
-
Body mass index
- ICC:
-
Intra-class correlation coefficient
- IQR/M:
-
Interquartile range-to-liver stiffness median ratio
- LSM:
-
Liver stiffness measurement
- TE:
-
Transient elastography
References
Foucher J, Chanteloup E, Vergniol J et al (2006) Diagnosis of cirrhosis by transient elastography (FibroScan): a prospective study. Gut 55:403–408
Friedrich-Rust M, Ong MF, Martens S et al (2008) Performance of transient elastography for the staging of liver fibrosis: a meta-analysis. Gastroenterology 134:960–974
Vergniol J, Foucher J, Terrebonne E et al (2011) Noninvasive tests for fibrosis and liver stiffness predict 5-year outcomes of patients with chronic hepatitis C. Gastroenterology 140:1970–9, 9 e1-3
de Lédinghen V, Vergniol J, Barthe C et al (2013) Non-invasive tests for fibrosis and liver stiffness predict 5-year survival of patients chronically infected with hepatitis B virus. Aliment Pharmacol Ther 37:979–988
Vergniol J, Boursier J, Coutzac C et al (2014) Evolution of noninvasive tests of liver fibrosis is associated with prognosis in patients with chronic hepatitis C. Hepatology 60:65–76
Sandrin L, Catheline S, Tanter M, Hennequin X, Fink M (1999) Time-resolved pulsed elastography with ultrafast ultrasonic imaging. Ultrason Imaging 21:259–272
European Association for the Study of the L (2011) EASL Clinical Practice Guidelines: management of hepatitis C virus infection. J Hepatol 55:245–264
European Association For The Study Of The L (2012) EASL clinical practice guidelines: management of chronic hepatitis B virus infection. J Hepatol 57:167–185
Piscaglia F, Salvatore V, Mulazzani L, Cantisani V, Schiavone C (2016) Ultrasound shear wave elastography for liver disease. A critical appraisal of the many actors on the stage. Ultraschall Med 37:1–5
Cassinotto C, Lapuyade B, Mouries A et al (2014) Non-invasive assessment of liver fibrosis with impulse elastography: comparison of Supersonic Shear Imaging with ARFI and FibroScan. J Hepatol 61:550–557
Sandrin L, Fourquet B, Hasquenoph JM et al (2003) Transient elastography: a new noninvasive method for assessment of hepatic fibrosis. Ultrasound Med Biol 29:1705–1713
Degos F, Perez P, Roche B et al (2010) Diagnostic accuracy of FibroScan and comparison to liver fibrosis biomarkers in chronic viral hepatitis: a multicenter prospective study (the FIBROSTIC study). J Hepatol 53:1013–1021
Nobili V, Vizzutti F, Arena U et al (2008) Accuracy and reproducibility of transient elastography for the diagnosis of fibrosis in pediatric nonalcoholic steatohepatitis. Hepatology 48:442–448
Zarski JP, Sturm N, Guechot J et al (2012) Comparison of nine blood tests and transient elastography for liver fibrosis in chronic hepatitis C: the ANRS HCEP-23 study. J Hepatol 56:55–62
Boursier J, Zarski JP, de Lédinghen V et al (2013) Determination of reliability criteria for liver stiffness evaluation by transient elastography. Hepatology 57:1182–1191
Molinari N, Morena M, Cristol JP, Daurè JP (2002) Free knot splines for biochemical data. Comput Methods Prog Biomed 67:163–167
Kettaneh A, Marcellin P, Douvin C et al (2007) Features associated with success rate and performance of FibroScan measurements for the diagnosis of cirrhosis in HCV patients: a prospective study of 935 patients. J Hepatol 46:628–634
Boursier J, Konate A, Guilluy M et al (2008) Learning curve and interobserver reproducibility evaluation of liver stiffness measurement by transient elastography. Eur J Gastroenterol Hepatol 20:693–701
Castera L, Foucher J, Bernard PH et al (2010) Pitfalls of liver stiffness measurement: a 5-year prospective study of 13,369 examinations. Hepatology 51:828–835
Pang JX, Pradhan F, Zimmer S et al (2014) The feasibility and reliability of transient elastography using Fibroscan(R): a practice audit of 2335 examinations. Can J Gastroenterol Hepatol 28:143–149
de Lédinghen V, Vergniol J, Foucher J, El-Hajbi F, Merrouche W, Rigalleau V (2010) Feasibility of liver transient elastography with FibroScan using a new probe for obese patients. Liver Int 30:1043–1048
Myers RP, Pomier-Layrargues G, Kirsch R et al (2012) Feasibility and diagnostic performance of the FibroScan XL probe for liver stiffness measurement in overweight and obese patients. Hepatology 55:199–208
Armstrong MJ, Corbett C, Hodson J et al (2013) Operator training requirements and diagnostic accuracy of Fibroscan in routine clinical practice. Postgrad Med J 89:685–692
Funding
The authors state that this work has not received any funding.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Guarantor
The scientific guarantor of this publication is Dr. Christophe Cassinotto.
Conflict of interest
The authors of this manuscript declare no relationships with any companies whose products or services may be related to the subject matter of the article.
Statistics and biometry
One of the authors has significant statistical expertise (Dr Molinari Nicolas).
Informed consent
Written informed consent was obtained from all subjects (patients) in this study.
Ethical approval
Institutional Review Board approval was obtained.
Methodology
• Prospective
• Observational
• Performed at one institution
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Electronic supplementary material
Supplementary Table 1
(DOCX 26 kb)
Rights and permissions
About this article
Cite this article
Cassinotto, C., Pierredon-Foulongne, MA., Belgour, A. et al. Learning curve of liver stiffness measurement using a new hybrid machine composed of transient elastography interfaced with ultrasound. Eur Radiol 30, 1088–1095 (2020). https://doi.org/10.1007/s00330-019-06388-1
Received:
Revised:
Accepted:
Published:
Issue Date: