Abstract
Recent investigation using the Sysmex XE-2100 hematology analyzer (Sysmex Corporation, Japan) has demonstrated erroneously high eosinophil counts and abnormal white blood cell (WBC) scattergrams in malaria cases. This study was conducted to assess the diagnostic efficiency of the Sysmex XE-2100 analyzer for malaria. One hundred forty-four patients initially diagnosed with Plasmodium vivax infection, 319 patients with febrile illness, and 24 patients who underwent malaria treatment were analyzed. Atypical features on Sysmex XE-2100 analyzer were categorized as pseudoeosinophilia (a gap of more than 5% in eosinophil counts between the Sysmex XE-2100 analyzer and microscopic examination) and abnormal WBC scattergram. Pseudoeosinophilia or abnormal WBC scattergram were detected in 100 of 144 malaria-positive samples (sensitivity 69.4%, specificity 100%). The samples with pseudoeosinophilia or abnormal WBC scattergrams showed significantly higher parasite counts than the samples without pseudoeosinophilia or an abnormal WBC scattergram (P < 0.05). All 24 samples from patients for whom the malaria smear was repeated after malaria treatment was initiated showed a normalized eosinophil count and a normal WBC histogram. In conclusion, attention to differential count and WBC scattergrams provided by the Sysmex XE-2100 would be a valuable tool in malaria detection.
Similar content being viewed by others
References
Hanscheid T (1999) Diagnosis of malaria: a review of alternatives to conventional microscopy. Clin Lab Haematol 21:235–245
Zalis MG, Ferreira-da-Cruz MF, Balthazar-Guedes HC, Banic DM, Alecrim W, Souza JM, Druilhe P, Daniel-Ribeiro CT (1996) Malaria diagnosis: standardization of a polymerase chain reaction for the detection of Plasmodium falciparum parasites in individuals with low-grade parasitemia. Parasitol Res 82:612–616
Hanscheid T, Pinto BG, Cristino JM, Grobusch MP (2000) Malaria diagnosis with the haematology analyser Cell-Dyn 3500: what does the instrument detect. Clin Lab Haematol 22:259–261
Hanscheid T, Melo-Cristino J, Pinto BG (2001) Automated detection of malaria pigment in white blood cells for the diagnosis of malaria in Portugal. Am J Trop Med Hyg 64:290–292
Hanscheid T, Pinto BG, Pereira I, Cristino JM, Valadas E (1999) Avoiding misdiagnosis of malaria: a novel automated method allows specific diagnosis, even in the absence of clinical suspicion. Emerg Infect Dis 5:836–838
Huh J, Jung J, Yoon H, Chung W (2005) Pseudoeosinophilia associated with malaria infection determined in the Sysmex XE-2100 hematology analyzer. Ann Hematol 84:400–402
Boisier P, Jambou R, Raharimalala L, Roux J (2002) Relationship between parasite density and fever risk in a community exposed to a low level of malaria transmission in Madagascar highlands. Am J Trop Med Hyg 67:137–140
Mendelow BV, Lyons C, Nhlangothi P, Tana M, Munster M, Wypkema E, Liebowitz L, Marshall L, Scott S, Coetzer TL (1999) Automated malaria detection by depolarization of laser light. Br J Haematol 104:499–503
Hanscheid T, Valadas E, Grobusch MP (2000) Automated malaria diagnosis using pigment detection. Parasitol Today 16:549–551
Park GB, Cha YJ (2006) Three cases of pseudoeosinophilia associated with malaria determined in the Sysmex XE-2100 Automated Hematology Analyzer. Korean J Lab Med 26:77–80
Nguyen PH, Day N, Pram TD, Ferguson DJ, White NJ (1995) Intraleucocytic malaria pigment and prognosis in severe malaria. Trans R Soc Trop Med Hyg 89:200–204
Hanscheid T, Egan TJ, Grobusch MP (2007) Haemozoin: from melatonin pigment to drug target, diagnostic tool, and immune modulator. Lancet Infect Dis 7:675–685
Grobusch MP, Hanscheid T, Kramer B, Neukammer J, May J, Seybold J, Kun JF, Suttorp N (2003) Sensitivity of hemozoin detection by automated flow cytometry in non- and semi-immune malaria patients. Cytometry B Clin Cytom 55:46–51
Suh IB, Kim HJ, Kim JY, Lee SW, An SS, Kim WJ, Lim CS (2003) Evaluation of the Abbott Cell-Dyn 4000 hematology analyzer for detection and therapeutic monitoring of Plasmodium vivax in the Republic of Korea. Trop Med Int Health 8:1074–1081
Padial MM, Subirats M, Puente S, Lago M, Crespo S, Palacios G, Baquero M (2005) Sensitivity of laser light depolarization analysis for detection of malaria in blood samples. J Med Microbiol 54:449–452
Chai JY (1999) Re-emerging Plasmodium vivax malaria in the Republic of Korea. Korean J Parasitol 37:129–143
Day NP, Pham TD, Phan TL, Dinh XS, Pham PL, Ly VC, Tran TH, Nguyen TH, Bethell DB, Nguyan HP, White NJ (1996) Clearance kinetics of parasites and pigment-containing leukocytes in severe malaria. Blood 88:4694–4700
Metzger WG, Mordmuller BG, Kremsner PG (1995) Malaria pigment in leucocytes. Trans R Soc Trop Med Hyg 89:637–638
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Huh, H.J., Oh, G.Y., Huh, J.W. et al. Malaria detection with the Sysmex XE-2100 hematology analyzer using pseudoeosinophilia and abnormal WBC scattergram. Ann Hematol 87, 755–759 (2008). https://doi.org/10.1007/s00277-008-0486-8
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00277-008-0486-8